Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 2/2012

01-02-2012 | Glaucoma

Comparison of measurement error of Cirrus HD-OCT and Heidelberg Retina Tomograph 3 in patients with early glaucomatous visual field defect

Authors: Alexander A. Shpak, Maria K. Sevostyanova, Svetlana N. Ogorodnikova, Irina N. Shormaz

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 2/2012

Login to get access

Abstract

Background

To compare the measurement error of Cirrus HD-OCT and Heidelberg Retina Tomograph 3 (HRT3) in patients with early glaucomatous visual field defect.

Methods

Thirty-nine consecutive patients (39 eyes) with early glaucomatous visual field defect were included. One eye of a patient was examined with Cirrus HD-OCT and HRT3 in one session. Each instrument was used by two operators, each taking two measurements in turn. We performed measurements of average retinal nerve fiber layer (RNFL) thickness and RNFL thickness in four quadrants with Cirrus HD-OCT and 13 stereometric parameters of the optic nerve head and RNFL with HRT3.

Results

Cirrus HD-OCT parameters performed much better than HRT3 stereometric parameters. Mean (for both operators) intraoperator within-subject coefficient of variation of the best Cirrus HD-OCT parameter (average RNFL thickness) was 2.6–2.7 times lower than the best HRT3 parameters [mean cup depth and rim area (P < 0.001)]. Mean intraoperator variability of RNFL thickness in quadrants (except nasal quadrant) was also significantly lower with OCT than with HRT. The interoperator within-subject coefficients of variation for both average RNFL thickness and RNFL thickness in all quadrants were significantly lower than the interoperator variability of best HRT3 parameter [mean cup depth (P < 0.001)]. The within-subject coefficient of variation of the average/mean RNFL thickness assessed by both instruments was 5.4–7.3 times lower for Cirrus HD-OCT. Among HRT3 parameters, mean cup depth, rim area and linear cup/disk ratio were the least variable, while cup volume, cup area and cup/disc area ratio were the most variable parameters.

Conclusions

Cirrus HD-OCT provides excellent intrasession intra- and interoperator repeatability of the RNFL measurements, especially of the average RNFL thickness in primary open-angle glaucoma (POAG) patients and POAG suspects with early glaucomatous visual field defect. The measurement error (expressed as within-subject coefficient of variation) of RNFL measurements by Cirrus HD-OCT is much lower than the error of HRT3 measurements of stereometric parameters of the optic nerve head and RNFL.
Literature
1.
go back to reference Miglior S, Albé E, Guareschi M, Rossetti L, Orzalesi N (2002) Intraobserver and interobserver reproducibility in the evaluation of optic disc stereometric parameters by Heidelberg Retina Tomograph. Ophthalmology 109:1072–1077PubMedCrossRef Miglior S, Albé E, Guareschi M, Rossetti L, Orzalesi N (2002) Intraobserver and interobserver reproducibility in the evaluation of optic disc stereometric parameters by Heidelberg Retina Tomograph. Ophthalmology 109:1072–1077PubMedCrossRef
2.
go back to reference Verdonck N, Zeyen T, Van Malderen L, Spileers W (2002) Short-term intra-individual variability in Heidelberg Retina Tomograph II. Bull Soc Belge Ophtalmol 286:51–57PubMed Verdonck N, Zeyen T, Van Malderen L, Spileers W (2002) Short-term intra-individual variability in Heidelberg Retina Tomograph II. Bull Soc Belge Ophtalmol 286:51–57PubMed
3.
go back to reference Tan JC, Garway-Heath DF, Hitchings RA (2003) Variability across the optic nerve head in scanning laser tomography. Br J Ophthalmol 87:557–559PubMedCrossRef Tan JC, Garway-Heath DF, Hitchings RA (2003) Variability across the optic nerve head in scanning laser tomography. Br J Ophthalmol 87:557–559PubMedCrossRef
4.
go back to reference Funk J, Mueller H (2003) Comparison of long-term fluctuations: laser scanning tomography versus automated perimetry. Graefes Arch Clin Exp Ophthalmol 241:721–724PubMedCrossRef Funk J, Mueller H (2003) Comparison of long-term fluctuations: laser scanning tomography versus automated perimetry. Graefes Arch Clin Exp Ophthalmol 241:721–724PubMedCrossRef
5.
go back to reference Watkins RJ, Broadway DC (2005) Intraobserver and interobserver reliability indices for drawing scanning laser ophthalmoscope optic disc contour lines with and without the aid of optic disc photographs. J Glaucoma 14:351–357PubMedCrossRef Watkins RJ, Broadway DC (2005) Intraobserver and interobserver reliability indices for drawing scanning laser ophthalmoscope optic disc contour lines with and without the aid of optic disc photographs. J Glaucoma 14:351–357PubMedCrossRef
6.
go back to reference Strouthidis NG, White ET, Owen VM, Ho TA, Hammond CJ, Garway-Heath DF (2005) Factors affecting the test-retest variability of Heidelberg retina tomograph and Heidelberg retina tomograph II measurements. Br J Ophthalmol 89:1427–1432PubMedCrossRef Strouthidis NG, White ET, Owen VM, Ho TA, Hammond CJ, Garway-Heath DF (2005) Factors affecting the test-retest variability of Heidelberg retina tomograph and Heidelberg retina tomograph II measurements. Br J Ophthalmol 89:1427–1432PubMedCrossRef
7.
go back to reference Prata TS, Meira-Freitas D, Lima VC, Guedes LM, Magalhães FP, Paranhos Junior A (2010) Factors affecting the variability of the Heidelberg Retina Tomograph III measurements in newly diagnosed glaucoma patients. Arq Bras Oftalmol 73:354–357PubMedCrossRef Prata TS, Meira-Freitas D, Lima VC, Guedes LM, Magalhães FP, Paranhos Junior A (2010) Factors affecting the variability of the Heidelberg Retina Tomograph III measurements in newly diagnosed glaucoma patients. Arq Bras Oftalmol 73:354–357PubMedCrossRef
8.
go back to reference Leung CK, Cheung CY, Weinreb RN, Qiu Q, Liu S, Li H, Xu G, Fan N, Huang L, Pang CP, Lam DS (2009) Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: a variability and diagnostic performance study. Ophthalmology 116:1257–1263PubMedCrossRef Leung CK, Cheung CY, Weinreb RN, Qiu Q, Liu S, Li H, Xu G, Fan N, Huang L, Pang CP, Lam DS (2009) Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: a variability and diagnostic performance study. Ophthalmology 116:1257–1263PubMedCrossRef
9.
go back to reference Vizzeri G, Weinreb RN, Gonzalez-Garcia AO, Bowd C, Medeiros FA, Sample PA, Zangwill LM (2009) Agreement between spectral-domain and time-domain OCT for measuring RNFL thickness. Br J Ophthalmol 93:775–781PubMedCrossRef Vizzeri G, Weinreb RN, Gonzalez-Garcia AO, Bowd C, Medeiros FA, Sample PA, Zangwill LM (2009) Agreement between spectral-domain and time-domain OCT for measuring RNFL thickness. Br J Ophthalmol 93:775–781PubMedCrossRef
10.
go back to reference Mwanza JC, Chang RT, Budenz DL, Durbin MK, Gendy MG, Shi W, Feuer WJ (2010) Reproducibility of peripapillary retinal nerve fiber layer thickness and optic nerve head parameters measured with Cirrus HD-OCT in glaucomatous eyes. Invest Ophthalmol Vis Sci 51:5724–5730PubMedCrossRef Mwanza JC, Chang RT, Budenz DL, Durbin MK, Gendy MG, Shi W, Feuer WJ (2010) Reproducibility of peripapillary retinal nerve fiber layer thickness and optic nerve head parameters measured with Cirrus HD-OCT in glaucomatous eyes. Invest Ophthalmol Vis Sci 51:5724–5730PubMedCrossRef
11.
go back to reference Savini G, Carbonelli M, Parisi V, Barboni P (2010) Effect of pupil dilation on retinal nerve fibre layer thickness measurements and their repeatability with Cirrus HD-OCT. Eye (Lond) 24:1503–1508CrossRef Savini G, Carbonelli M, Parisi V, Barboni P (2010) Effect of pupil dilation on retinal nerve fibre layer thickness measurements and their repeatability with Cirrus HD-OCT. Eye (Lond) 24:1503–1508CrossRef
12.
go back to reference Cremasco F, Massa G, Gonçalves Vidotti V, de Carvalho P, Lupinacci A, Costa VP (2011) Intrasession, intersession, and interexaminer variabilities of retinal nerve fiber layer measurements with spectral-domain OCT. Eur J Ophthalmol 21:264–270PubMedCrossRef Cremasco F, Massa G, Gonçalves Vidotti V, de Carvalho P, Lupinacci A, Costa VP (2011) Intrasession, intersession, and interexaminer variabilities of retinal nerve fiber layer measurements with spectral-domain OCT. Eur J Ophthalmol 21:264–270PubMedCrossRef
13.
go back to reference Labiris G, Giarmoukakis A, Katsanos A, Gkika MG, Fanariotis M, Pavlidou E, Kozobolis VP (2010) Reproducibility of OCT/SLO measurements in healthy eyes. Eur J Ophthalmol 20:552–558PubMed Labiris G, Giarmoukakis A, Katsanos A, Gkika MG, Fanariotis M, Pavlidou E, Kozobolis VP (2010) Reproducibility of OCT/SLO measurements in healthy eyes. Eur J Ophthalmol 20:552–558PubMed
14.
go back to reference González-García AO, Vizzeri G, Bowd C, Medeiros FA, Zangwill LM, Weinreb RN (2009) Reproducibility of RTVue retinal nerve fiber layer thickness and optic disc measurements and agreement with Stratus optical coherence tomography measurements. Am J Ophthalmol 147:1067–1074PubMedCrossRef González-García AO, Vizzeri G, Bowd C, Medeiros FA, Zangwill LM, Weinreb RN (2009) Reproducibility of RTVue retinal nerve fiber layer thickness and optic disc measurements and agreement with Stratus optical coherence tomography measurements. Am J Ophthalmol 147:1067–1074PubMedCrossRef
15.
go back to reference Mills RP, Budenz DL, Lee PP, Noecker RJ, Walt JG, Siegartel LR, Evans SJ, Doyle JJ (2006) Categorizing the stage of glaucoma from pre-diagnosis to end-stage disease. Am J Ophthalmol 141:24–30PubMedCrossRef Mills RP, Budenz DL, Lee PP, Noecker RJ, Walt JG, Siegartel LR, Evans SJ, Doyle JJ (2006) Categorizing the stage of glaucoma from pre-diagnosis to end-stage disease. Am J Ophthalmol 141:24–30PubMedCrossRef
20.
go back to reference Zar JH (2010) Biostatistical analysis, 5th edn. Pearson Prentice-Hall, Upper Saddle River, NJ, pp 160–161 Zar JH (2010) Biostatistical analysis, 5th edn. Pearson Prentice-Hall, Upper Saddle River, NJ, pp 160–161
21.
go back to reference Abdi H (2010) Coefficient of variation. In: Salkind N, Dougherty DM, Frey B (eds) Encyclopedia of research design. Sage, Thousand Oaks, CA, pp 169–171 Abdi H (2010) Coefficient of variation. In: Salkind N, Dougherty DM, Frey B (eds) Encyclopedia of research design. Sage, Thousand Oaks, CA, pp 169–171
Metadata
Title
Comparison of measurement error of Cirrus HD-OCT and Heidelberg Retina Tomograph 3 in patients with early glaucomatous visual field defect
Authors
Alexander A. Shpak
Maria K. Sevostyanova
Svetlana N. Ogorodnikova
Irina N. Shormaz
Publication date
01-02-2012
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 2/2012
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-011-1808-4

Other articles of this Issue 2/2012

Graefe's Archive for Clinical and Experimental Ophthalmology 2/2012 Go to the issue