Open Access 01-08-2009 | Refractive Surgery

Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation

Authors: Muriël Doors, Lars P. J. Cruysberg, Tos T. J. M. Berendschot, John de Brabander, Frenne Verbakel, Carroll A. B. Webers, Rudy M. M. A. Nuijts

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 8/2009

Login to get access

Abstract

Background

The repeatability and interchangeability of imaging devices measuring central corneal thickness (CCT) and anterior chamber depth (ACD) are important in the assessment of patients considering refractive surgery. The purpose of this study was to investigate the agreement of CCT and ACD measurements using three imaging technologies in healthy eyes and in eyes after phakic intraocular lens implantation (pIOL).

Methods

In this comparative study, CCT and ACD were measured using anterior segment optical coherence tomography (AS-OCT), Orbscan II, and Pentacam in 33 healthy volunteers (66 eyes) and 22 patients (42 eyes) after pIOL implantation. Intraobserver repeatability was evaluated for all three devices in the healthy volunteer group.

Results

Pairwise comparison of CCT measurements showed significant differences between all devices (P < 0.001), except for the AS-OCT and Orbscan II in the healthy volunteer group (P = 0.422) and the Orbscan II and Pentacam in the pIOL group (P = 0.214). ACD measurements demonstrated significant differences between all pairwise comparisons in both groups (P ≤ 0.001). Intraobserver reliability was high for CCT and ACD measurements in the healthy volunteer group, with coefficients of variation ranging from 0.6% to 1.2% and 0.4% to 0.5% respectively.

Conclusions

CCT and ACD measurements using AS-OCT, Orbscan II, and Pentacam demonstrated high intraobserver reliability. However, these devices should not be used interchangeably for measurements of CCT and ACD in healthy subject and patients after pIOL implantation.
Literature
2.
go back to reference Cheng H, Bates AK, Wood L, McPherson K (1988) Positive correlation of corneal thickness and endothelial cell loss. Serial measurements after cataract surgery. Arch Ophthalmol 106:920–922PubMed Cheng H, Bates AK, Wood L, McPherson K (1988) Positive correlation of corneal thickness and endothelial cell loss. Serial measurements after cataract surgery. Arch Ophthalmol 106:920–922PubMed
4.
go back to reference Ho T, Cheng AC, Rao SK, Lau S, Leung CK, Lam DS (2007) Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 33:1177–1182. doi:10.1016/j.jcrs.2007.03.028 PubMedCrossRef Ho T, Cheng AC, Rao SK, Lau S, Leung CK, Lam DS (2007) Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 33:1177–1182. doi:10.​1016/​j.​jcrs.​2007.​03.​028 PubMedCrossRef
5.
go back to reference Amano S, Honda N, Amano Y, Yamagami S, Miyai T, Samejima T, Ogata M, Miyata K (2006) Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology 113:937–941. doi:10.1016/j.ophtha.2006.01.063 PubMedCrossRef Amano S, Honda N, Amano Y, Yamagami S, Miyai T, Samejima T, Ogata M, Miyata K (2006) Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology 113:937–941. doi:10.​1016/​j.​ophtha.​2006.​01.​063 PubMedCrossRef
6.
go back to reference Kim SW, Byun YJ, Kim EK, Kim TI (2007) Central corneal thickness measurements in unoperated eyes and eyes after PRK for myopia using Pentacam, Orbscan II, and ultrasonic pachymetry. J Refract Surg 23:888–894PubMed Kim SW, Byun YJ, Kim EK, Kim TI (2007) Central corneal thickness measurements in unoperated eyes and eyes after PRK for myopia using Pentacam, Orbscan II, and ultrasonic pachymetry. J Refract Surg 23:888–894PubMed
7.
go back to reference de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144:712–718. doi:10.1016/j.ajo.2007.07.021 PubMedCrossRef de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144:712–718. doi:10.​1016/​j.​ajo.​2007.​07.​021 PubMedCrossRef
8.
go back to reference Hashemi H, Mehravaran S (2007) Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia. J Cataract Refract Surg 33:1701–1707PubMedCrossRef Hashemi H, Mehravaran S (2007) Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia. J Cataract Refract Surg 33:1701–1707PubMedCrossRef
9.
go back to reference Lavanya R, Teo L, Friedman DS, Aung HT, Baskaran M, Gao H, Alfred T, Seah SK, Kashiwagi K, Foster PJ, Aung T (2007) Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography. Br J Ophthalmol 91:1023–1026. doi:10.1136/bjo.2006.113761 PubMedCrossRef Lavanya R, Teo L, Friedman DS, Aung HT, Baskaran M, Gao H, Alfred T, Seah SK, Kashiwagi K, Foster PJ, Aung T (2007) Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography. Br J Ophthalmol 91:1023–1026. doi:10.​1136/​bjo.​2006.​113761 PubMedCrossRef
10.
go back to reference Haque S, Simpson T, Jones L (2006) Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry, Orbscan II, and optical coherence tomography. J Refract Surg 22:486–493PubMed Haque S, Simpson T, Jones L (2006) Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry, Orbscan II, and optical coherence tomography. J Refract Surg 22:486–493PubMed
12.
go back to reference Yaylali V, Kaufman SC, Thompson HW (1997) Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry. J Cataract Refract Surg 23:1345–1350PubMed Yaylali V, Kaufman SC, Thompson HW (1997) Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry. J Cataract Refract Surg 23:1345–1350PubMed
14.
go back to reference Chakrabarti HS, Craig JP, Brahma A, Malik TY, McGhee CN (2001) Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-lasik eyes. J Cataract Refract Surg 27:1823–1828PubMedCrossRef Chakrabarti HS, Craig JP, Brahma A, Malik TY, McGhee CN (2001) Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-lasik eyes. J Cataract Refract Surg 27:1823–1828PubMedCrossRef
15.
go back to reference Hashemi H, Roshani M, Mehravaran S, Parsafar H, Yazdani K (2007) Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry. J Cataract Refract Surg 33:1694–1700PubMedCrossRef Hashemi H, Roshani M, Mehravaran S, Parsafar H, Yazdani K (2007) Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry. J Cataract Refract Surg 33:1694–1700PubMedCrossRef
16.
go back to reference Li H, Leung CK, Wong L, Cheung CY, Pang CP, Weinreb RN, Lam DS (2008) Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and Visante optical coherence tomography. Ophthalmology 115:796–801. doi:10.1016/j.ophtha.2007.07.006 PubMedCrossRef Li H, Leung CK, Wong L, Cheung CY, Pang CP, Weinreb RN, Lam DS (2008) Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and Visante optical coherence tomography. Ophthalmology 115:796–801. doi:10.​1016/​j.​ophtha.​2007.​07.​006 PubMedCrossRef
18.
go back to reference Goldsmith JA, Li Y, Chalita MR, Westphal V, Patil CA, Rollins AM, Izatt JA, Huang D (2005) Anterior chamber width measurement by high-speed optical coherence tomography. Ophthalmology 112:238–244. doi:10.1016/j.ophtha.2004.09.019 Goldsmith JA, Li Y, Chalita MR, Westphal V, Patil CA, Rollins AM, Izatt JA, Huang D (2005) Anterior chamber width measurement by high-speed optical coherence tomography. Ophthalmology 112:238–244. doi:10.​1016/​j.​ophtha.​2004.​09.​019
19.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
20.
go back to reference Li EY, Mohamed S, Leung CK, Rao SK, Cheng AC, Cheung CY, Lam DS (2007) Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology 114:1842–1847PubMedCrossRef Li EY, Mohamed S, Leung CK, Rao SK, Cheng AC, Cheung CY, Lam DS (2007) Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology 114:1842–1847PubMedCrossRef
21.
go back to reference Rosa N, Lanza M, Borrelli M, Polito B, Filosa ML, De Bernardo M (2007) Comparison of central corneal thickness measured with Orbscan and Pentacam. J Refract Surg 23:895–899PubMed Rosa N, Lanza M, Borrelli M, Polito B, Filosa ML, De Bernardo M (2007) Comparison of central corneal thickness measured with Orbscan and Pentacam. J Refract Surg 23:895–899PubMed
23.
go back to reference Kim HY, Budenz DL, Lee PS, Feuer WJ, Barton K (2008) Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry. Am J Ophthalmol 145:228–232. doi:10.1016/j.ajo.2007.09.030 Kim HY, Budenz DL, Lee PS, Feuer WJ, Barton K (2008) Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry. Am J Ophthalmol 145:228–232. doi:10.​1016/​j.​ajo.​2007.​09.​030
24.
go back to reference Wong AC, Wong CC, Yuen NS, Hui SP (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, orbscan and ultrasound pachymetry. Eye 16:715–721. doi:10.1038/sj.eye.6700211 Wong AC, Wong CC, Yuen NS, Hui SP (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, orbscan and ultrasound pachymetry. Eye 16:715–721. doi:10.​1038/​sj.​eye.​6700211
25.
go back to reference Suzuki S, Oshika T, Oki K, Sakabe I, Iwase A, Amano S, Araie M (2003) Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg 29:1313–1318. doi:10.1016/S0886-3350(03)00123-8 Suzuki S, Oshika T, Oki K, Sakabe I, Iwase A, Amano S, Araie M (2003) Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg 29:1313–1318. doi:10.​1016/​S0886-3350(03)00123-8
26.
27.
go back to reference Fujioka M, Nakamura M, Tatsumi Y, Kusuhara A, Maeda H, Negi A (2007) Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res 32:89–94. doi:10.1080/02713680601115010 Fujioka M, Nakamura M, Tatsumi Y, Kusuhara A, Maeda H, Negi A (2007) Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res 32:89–94. doi:10.​1080/​0271368060111501​0
Metadata
Title
Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation
Authors
Muriël Doors
Lars P. J. Cruysberg
Tos T. J. M. Berendschot
John de Brabander
Frenne Verbakel
Carroll A. B. Webers
Rudy M. M. A. Nuijts
Publication date
01-08-2009
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 8/2009
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-009-1086-6

Other articles of this Issue 8/2009

Graefe's Archive for Clinical and Experimental Ophthalmology 8/2009 Go to the issue