Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 10/2008

01-10-2008 | Pediatrics

Comparison of intraocular tonometry using three different non-invasive tonometers in children

Authors: Julia Lasseck, Thomas Jehle, Nicolas Feltgen, Wolf Alexander Lagrèze

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 10/2008

Login to get access

Abstract

Background

In childhood glaucoma, the correct determination of intraocular pressure (IOP) is crucial in clinical decision-making. We therefore investigated how intraocular tonometry readings correlate with readings from commonly used tonometers.

Methods

IOP was measured unilaterally in 20 children suffering from congenital (n = 7) or secondary glaucoma (n = 13), 10 minutes after the induction of general anaesthesia. The children were aged from 1 month to 17 years (mean age 4.3 years, median age 1.3 years). Non-invasive applanation (Tono-Pen XL, Perkins tonometer) and indentation tonometry (Schiötz tonometer) were performed in random order prior to intraocular tonometry with a 26-gauge needle connected to a pressure sensor. Linear regression analysis and the coefficients of variance (CV) were used to compare the data obtained from the various tonometers.

Results

Compared with intraocular pressure, the CV was 10% for the Tono-Pen XL, 17% for the Schiötz, and 19% for the Perkins tonometer. The coefficient of determination (r2) was 0.74 for Tono-Pen XL, 0.60 for Schiötz and 0.78 for Perkins tonometry. The IOP values obtained with the Tono-Pen XL scattered homogeneously around the intraocular IOP, while the Perkins and Schiötz tonometers underestimated intraocular measured IOP.

Conclusion

Of the three tonometers evaluated in this series of paediatric glaucoma patients, the Tono-Pen XL most closely reflected true IOP.
Literature
1.
go back to reference Amaral WO, Teixeira RM, Alencar LM, Cronemberger S, Calixto N (2006) Central and peripheral corneal thickness: influence on the iop measurement by Tonopen. Arq Bras Oftalmol 69:41–45PubMedCrossRef Amaral WO, Teixeira RM, Alencar LM, Cronemberger S, Calixto N (2006) Central and peripheral corneal thickness: influence on the iop measurement by Tonopen. Arq Bras Oftalmol 69:41–45PubMedCrossRef
2.
go back to reference Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS (2002) Effect of Corneal Thickness on Intraocular Pressure Measurements with the Pneumotonometer, Goldmann Applanation Tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 43:1389–1392PubMed Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS (2002) Effect of Corneal Thickness on Intraocular Pressure Measurements with the Pneumotonometer, Goldmann Applanation Tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 43:1389–1392PubMed
3.
go back to reference Blumberg D, Congdon N, Jampel H, Gilbert D, Elliott R, Rivers R, Munoz B, Quigley H (2007) The effects of sevoflurane and ketamine on intraocular pressure in children during examination under anesthesia. Am J Ophthalmol 143:494–499PubMedCrossRef Blumberg D, Congdon N, Jampel H, Gilbert D, Elliott R, Rivers R, Munoz B, Quigley H (2007) The effects of sevoflurane and ketamine on intraocular pressure in children during examination under anesthesia. Am J Ophthalmol 143:494–499PubMedCrossRef
4.
go back to reference Bordon AF, Katsumi O, Hirose T (1995) Tonometry in pediatric patients: a comparative study among Tono-pen, Perkins, and Schiötz tonometers. J Pediatr Ophthalmol Strabismus 32:373–377PubMed Bordon AF, Katsumi O, Hirose T (1995) Tonometry in pediatric patients: a comparative study among Tono-pen, Perkins, and Schiötz tonometers. J Pediatr Ophthalmol Strabismus 32:373–377PubMed
5.
go back to reference Deuter CM, Schlote T, Hahn GA, Bende T, Derse M (2002) Messung des Augeninnendrucks mit dem Tono-Pen im Vergleich zum Applanations-tonometer nach Goldmann - eine klinische Studie an 100 Augen. Klin Monatsbl Augenheilkd 219:138–142PubMedCrossRef Deuter CM, Schlote T, Hahn GA, Bende T, Derse M (2002) Messung des Augeninnendrucks mit dem Tono-Pen im Vergleich zum Applanations-tonometer nach Goldmann - eine klinische Studie an 100 Augen. Klin Monatsbl Augenheilkd 219:138–142PubMedCrossRef
6.
go back to reference Dietlein TS, Jacobi PC, Krieglstein GK (1996) Klinische Diskrepanz zwischen Schiötz- und Perkins-Tonometrie beim kindlichen Glaukom. Klin Monatsbl Augenheilkd 209:299–303PubMed Dietlein TS, Jacobi PC, Krieglstein GK (1996) Klinische Diskrepanz zwischen Schiötz- und Perkins-Tonometrie beim kindlichen Glaukom. Klin Monatsbl Augenheilkd 209:299–303PubMed
7.
go back to reference Dohadwala AA, Munger R, Damji KF (1998) Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 105:1849–1854PubMedCrossRef Dohadwala AA, Munger R, Damji KF (1998) Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 105:1849–1854PubMedCrossRef
8.
go back to reference Eisenberg DL, Sherman BG, McKeown CA, Schuman JS (1998) Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105:1173–1181PubMedCrossRef Eisenberg DL, Sherman BG, McKeown CA, Schuman JS (1998) Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105:1173–1181PubMedCrossRef
9.
go back to reference Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87PubMedCrossRef Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87PubMedCrossRef
10.
go back to reference Friedenwald JS (1937) Contribution to the theory and practice of tonometry. Am J Ophthalmol 20:985–1024 Friedenwald JS (1937) Contribution to the theory and practice of tonometry. Am J Ophthalmol 20:985–1024
11.
12.
go back to reference Horowitz GS, Byles J, Lee J, D’Este C (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Experiment Ophthalmol 32:584–589PubMedCrossRef Horowitz GS, Byles J, Lee J, D’Este C (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Experiment Ophthalmol 32:584–589PubMedCrossRef
13.
go back to reference Kniestedt C, Nee M, Stamper RL (2005) Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243:359–366PubMedCrossRef Kniestedt C, Nee M, Stamper RL (2005) Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243:359–366PubMedCrossRef
14.
go back to reference Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef
15.
go back to reference Krieglstein G, Langham ME (1975) Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica 171:132–145PubMed Krieglstein G, Langham ME (1975) Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica 171:132–145PubMed
16.
go back to reference Lam DS, Leung DY, Chiu TY, Fan DS, Cheung EY, Wong TY, Lai JS, Tham CC (2004) Pressure phosphene self-tonometry: a comparison with Goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci 45:3131–3136PubMedCrossRef Lam DS, Leung DY, Chiu TY, Fan DS, Cheung EY, Wong TY, Lai JS, Tham CC (2004) Pressure phosphene self-tonometry: a comparison with Goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci 45:3131–3136PubMedCrossRef
17.
go back to reference Lee GA, Khaw PT, Ficker LA, Shah P (2002) The corneal thickness and intraocular pressure story: where are we now? Clin Experiment Ophthalmol 30:334–337PubMedCrossRef Lee GA, Khaw PT, Ficker LA, Shah P (2002) The corneal thickness and intraocular pressure story: where are we now? Clin Experiment Ophthalmol 30:334–337PubMedCrossRef
18.
go back to reference Levy J, Lifshitz T, Rosen S, Tessler Z, Biedner BZ (2005) Is the tono-pen accurate for measuring intraocular pressure in young children with congenital glaucoma? J AAPOS 9:321–325PubMedCrossRef Levy J, Lifshitz T, Rosen S, Tessler Z, Biedner BZ (2005) Is the tono-pen accurate for measuring intraocular pressure in young children with congenital glaucoma? J AAPOS 9:321–325PubMedCrossRef
19.
go back to reference Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef
20.
go back to reference Madjlessi F, Marx W, Reinhard T, Althaus C, Sundmacher R (2000) Impressions- und Applanationstonometrie bei pathologischen Hornhäuten im Vergleich mit der intraokularen Nadeldruckmessung. Ophthalmologe 97:478–481PubMedCrossRef Madjlessi F, Marx W, Reinhard T, Althaus C, Sundmacher R (2000) Impressions- und Applanationstonometrie bei pathologischen Hornhäuten im Vergleich mit der intraokularen Nadeldruckmessung. Ophthalmologe 97:478–481PubMedCrossRef
21.
go back to reference Marx W, Madjlessi F, Reinhard T, Althaus C, Sundmacher R (1999) Mehr als vier Jahre Erfahrung mit der elektronischen, intraokularen Nadel-Druckmessung bei irregulären Hornhäuten. Ophthalmologe 96:498–502PubMedCrossRef Marx W, Madjlessi F, Reinhard T, Althaus C, Sundmacher R (1999) Mehr als vier Jahre Erfahrung mit der elektronischen, intraokularen Nadel-Druckmessung bei irregulären Hornhäuten. Ophthalmologe 96:498–502PubMedCrossRef
22.
go back to reference Mirakhur RK, Elliott P, Shepherd WF, McGalliard JN (1990) Comparison of the effects of isoflurane and halothane on intraocular pressure. Acta Anaesthesiol Scand 34:282–285PubMed Mirakhur RK, Elliott P, Shepherd WF, McGalliard JN (1990) Comparison of the effects of isoflurane and halothane on intraocular pressure. Acta Anaesthesiol Scand 34:282–285PubMed
23.
go back to reference Mirakhur RK, Shepherd WF, Darrah WC (1987) Propofol or thiopentone: effects on intraocular pressure associated with induction of anaesthesia and tracheal intubation (facilitated with suxamethonium). Br J Anaesth 59:431–436PubMedCrossRef Mirakhur RK, Shepherd WF, Darrah WC (1987) Propofol or thiopentone: effects on intraocular pressure associated with induction of anaesthesia and tracheal intubation (facilitated with suxamethonium). Br J Anaesth 59:431–436PubMedCrossRef
24.
go back to reference Nagdeve NG, Yaddanapudi S, Pandav SS (2006) The effect of different doses of ketamine on intraocular pressure in anesthetized children. J Pediatr Ophthalmol Strabismus 43:219–223PubMed Nagdeve NG, Yaddanapudi S, Pandav SS (2006) The effect of different doses of ketamine on intraocular pressure in anesthetized children. J Pediatr Ophthalmol Strabismus 43:219–223PubMed
25.
go back to reference Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR (1988) Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol 106:1697–1700PubMed Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR (1988) Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol 106:1697–1700PubMed
26.
go back to reference Schiötz H (1905) Ein neuer Tonometer. Arch Augenheilkunde 52:410–424 Schiötz H (1905) Ein neuer Tonometer. Arch Augenheilkunde 52:410–424
Metadata
Title
Comparison of intraocular tonometry using three different non-invasive tonometers in children
Authors
Julia Lasseck
Thomas Jehle
Nicolas Feltgen
Wolf Alexander Lagrèze
Publication date
01-10-2008
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 10/2008
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-008-0863-y

Other articles of this Issue 10/2008

Graefe's Archive for Clinical and Experimental Ophthalmology 10/2008 Go to the issue