Published in: European Archives of Oto-Rhino-Laryngology 8/2017

01-08-2017 | Otology

A multicenter study on objective and subjective benefits with a transcutaneous bone-anchored hearing aid device: first Nordic results

Authors: Dan Dupont Hougaard, Soren Kjaergaard Boldsen, Anne Marie Jensen, Soren Hansen, Per Cayé Thomassen

Published in: European Archives of Oto-Rhino-Laryngology | Issue 8/2017

Login to get access

Abstract

Examination of objective as well as subjective outcomes with a new transcutaneous bone-anchored hearing aid device. The study was designed as a prospective multicenter consecutive case-series study involving tertiary referral centers at two Danish University Hospitals. A total of 23 patients were implanted. Three were lost to follow-up. Patients had single-sided deafness, conductive or mixed hearing loss. Intervention: Rehabilitative. Aided and unaided sound field hearing was evaluated objectively using (1) pure warble tone thresholds, (2) pure-tone average (PTA4), (3) speech discrimination score (SDS) in quiet, and (4) speech reception threshold 50% at 70 dB SPL noise level (SRT50%). Subjective benefit was evaluated by three validated questionnaires: (1) the IOI-HA, (2) the SSQ-12, and (3) a questionnaire evaluating both the frequency and the duration of hearing aid usage. The mean aided PTA4 was lowered by 14.7 dB. SDS was increased by 37.5% at 50 dB SPL, SRT50% in noise improved 1.4 dB. Aided thresholds improved insignificantly at frequencies above 2 kHz. 52.9% of the patients used their device every day, and 76.5% used the device at least 5 days a week. Mean IOI-HA score was 3.4, corresponding to a good benefit. In SSQ-12, “quality of hearing” scored especially high. Patients with a conductive and/or mixed hearing loss benefitted the most. This device demonstrates a significant subjective hearing benefit 8 month post surgery. In patients with conductive and/or mixed hearing losses, patient satisfaction and frequency of use were high. Objective gain measures showed less promising results especially in patients with single-sided deafness (SSD) compared to other bone conduction devices.
Literature
1.
go back to reference Tjellström A, Lindström J, Hallén O et al (1981) Osseointegrated titanium implants in the temporal bone. A clinical study on bone-anchored hearing aids. Am J Otol 2:304–310PubMed Tjellström A, Lindström J, Hallén O et al (1981) Osseointegrated titanium implants in the temporal bone. A clinical study on bone-anchored hearing aids. Am J Otol 2:304–310PubMed
2.
go back to reference Snik AF, Mylanus EA, Proops DW et al (2005) Consensus statements on the BAHA system: where do we stand at present? Ann Otol Rhinol Laryngol Suppl 195:2–12CrossRefPubMed Snik AF, Mylanus EA, Proops DW et al (2005) Consensus statements on the BAHA system: where do we stand at present? Ann Otol Rhinol Laryngol Suppl 195:2–12CrossRefPubMed
3.
go back to reference Dun CA, Faber HT, de Wolf MJ, Cremers CW, Hol MK (2011) An overview of different systems: the bone-anchored hearing aid. Adv Otorhinolaryngol 71:22–31PubMed Dun CA, Faber HT, de Wolf MJ, Cremers CW, Hol MK (2011) An overview of different systems: the bone-anchored hearing aid. Adv Otorhinolaryngol 71:22–31PubMed
4.
go back to reference Hakansson B, Tjellström A, Rosenhall U, Carlsson P (1985) The bone-anchored hearing aid. Principal design and a psychoacoustical evaluation. Acta Otolaryngol 100(3–4):229–239CrossRefPubMed Hakansson B, Tjellström A, Rosenhall U, Carlsson P (1985) The bone-anchored hearing aid. Principal design and a psychoacoustical evaluation. Acta Otolaryngol 100(3–4):229–239CrossRefPubMed
5.
go back to reference Kiringoda R, Lustig LR (2013) A meta-analysis of the complications associated with osseointegrated hearing aids. Otol Neurotol 34:790–794CrossRefPubMed Kiringoda R, Lustig LR (2013) A meta-analysis of the complications associated with osseointegrated hearing aids. Otol Neurotol 34:790–794CrossRefPubMed
6.
go back to reference Holgers KM, Tjellström A, Bjursten LM, Erlandsson BE et al (1988) Soft tissue reactions around percutaneous implants: a clinical study of soft tissue conditions around skin-penetrating titanium implants for bone anchored hearing aids. Am J Otol 9:56–59PubMed Holgers KM, Tjellström A, Bjursten LM, Erlandsson BE et al (1988) Soft tissue reactions around percutaneous implants: a clinical study of soft tissue conditions around skin-penetrating titanium implants for bone anchored hearing aids. Am J Otol 9:56–59PubMed
7.
go back to reference Reyes RA, Tjellstöm A, Granström G (2000) Evaluation of implant losses and skin reactions around extraoral bone-anchored implants: a 0-to 8-year follow-up. Otolaryngol Head Neck Surg 122:272–276CrossRefPubMed Reyes RA, Tjellstöm A, Granström G (2000) Evaluation of implant losses and skin reactions around extraoral bone-anchored implants: a 0-to 8-year follow-up. Otolaryngol Head Neck Surg 122:272–276CrossRefPubMed
8.
go back to reference Hobson JC, Roper AJ, Andrew R, Rothera MP, Hill P, Green KM (2010) Complications of bone-anchored hearing aid implantation. J Laryngol Otol 124:132–136CrossRefPubMed Hobson JC, Roper AJ, Andrew R, Rothera MP, Hill P, Green KM (2010) Complications of bone-anchored hearing aid implantation. J Laryngol Otol 124:132–136CrossRefPubMed
9.
go back to reference Calvo Bodnia N, Foghsgaard S, Nue Moller M, Cayé-Thomasen P (2014) Long-term results of 185 consecutive osseointegrated hearing device implantations: a comparison among children, adults, and elderly. Otol Neurotol 35:e301–e306CrossRefPubMed Calvo Bodnia N, Foghsgaard S, Nue Moller M, Cayé-Thomasen P (2014) Long-term results of 185 consecutive osseointegrated hearing device implantations: a comparison among children, adults, and elderly. Otol Neurotol 35:e301–e306CrossRefPubMed
10.
go back to reference Kurz A, Flynn M, Caversaccio M, Kompis M (2014) Speech understanding with a new implant technology: a comparative study with a new Nonskin penetrating BAHA system. Biomed Res Int 2014:416205CrossRefPubMedPubMedCentral Kurz A, Flynn M, Caversaccio M, Kompis M (2014) Speech understanding with a new implant technology: a comparative study with a new Nonskin penetrating BAHA system. Biomed Res Int 2014:416205CrossRefPubMedPubMedCentral
11.
go back to reference Elberling C, Ludvigsen C, Lyregaard PE (1989) DANTALE: a new Danish speech material. Scand Audiol 18(3):169–175CrossRefPubMed Elberling C, Ludvigsen C, Lyregaard PE (1989) DANTALE: a new Danish speech material. Scand Audiol 18(3):169–175CrossRefPubMed
12.
go back to reference Cox RM, Alexander GC (2002) The international outcome inventory for hearing aids (IOI-HA): psychometric properties of the English version. Int J Audiol 41(1):30–35CrossRefPubMed Cox RM, Alexander GC (2002) The international outcome inventory for hearing aids (IOI-HA): psychometric properties of the English version. Int J Audiol 41(1):30–35CrossRefPubMed
13.
go back to reference Thunberg Jespersen C, Bille M, Legarth JV (2014) Psychometric properties of a revised Danish translation of the international outcome inventory for hearing aids (IOI-HA). Int J Audiol 53(5):302–308CrossRefPubMedPubMedCentral Thunberg Jespersen C, Bille M, Legarth JV (2014) Psychometric properties of a revised Danish translation of the international outcome inventory for hearing aids (IOI-HA). Int J Audiol 53(5):302–308CrossRefPubMedPubMedCentral
14.
go back to reference Noble W, Jensen NS, Naylor G, Bhullar N, Akeroyd MA (2013) A short form of the speech, spatial and qualities of hearing scale suitable for clinical use: the SSQ12. Int J Audiol 52(6):409–412CrossRefPubMed Noble W, Jensen NS, Naylor G, Bhullar N, Akeroyd MA (2013) A short form of the speech, spatial and qualities of hearing scale suitable for clinical use: the SSQ12. Int J Audiol 52(6):409–412CrossRefPubMed
15.
go back to reference Rasmussen J, Olsen SO, Nielsen LH (2012) Evaluation of long-term patient satisfaction and experience with the Baha(R) bone conduction implant. Int J Audiol 51(3):194–199CrossRefPubMed Rasmussen J, Olsen SO, Nielsen LH (2012) Evaluation of long-term patient satisfaction and experience with the Baha(R) bone conduction implant. Int J Audiol 51(3):194–199CrossRefPubMed
16.
go back to reference Briggs R, Van Hasselt A, Luntz Michal et al (2015) Clinical performance of a new magnetic bone conduction hearing implant system: results from a prospective, multicenter, clinical investigation. Otol Neurotol 36:834–841CrossRefPubMed Briggs R, Van Hasselt A, Luntz Michal et al (2015) Clinical performance of a new magnetic bone conduction hearing implant system: results from a prospective, multicenter, clinical investigation. Otol Neurotol 36:834–841CrossRefPubMed
17.
go back to reference Denoyelle F, Leboulanger N, Coudert C, Mazzaschi O et al (2013) New closed skin bone-anchored implant: preliminary results in 6 children with ear Atresia. Otol Neurotol 34:275–281CrossRefPubMed Denoyelle F, Leboulanger N, Coudert C, Mazzaschi O et al (2013) New closed skin bone-anchored implant: preliminary results in 6 children with ear Atresia. Otol Neurotol 34:275–281CrossRefPubMed
18.
go back to reference Eberhard KE, Olsen SØ, Miyazaki H, Bille M, Caye-Thomasen P (2016) Objective and subjective outcome of a new transcutaneous bone conduction hearing device: half-year follow-up of the first 12 nordic implantations. Otol Neurotol 37(3):267–275. doi:10.1097/MAO.0000000000000969 PubMed Eberhard KE, Olsen SØ, Miyazaki H, Bille M, Caye-Thomasen P (2016) Objective and subjective outcome of a new transcutaneous bone conduction hearing device: half-year follow-up of the first 12 nordic implantations. Otol Neurotol 37(3):267–275. doi:10.​1097/​MAO.​0000000000000969​ PubMed
19.
go back to reference Cox RM, Stephens D, Kramer SE (2002) Translations of the international outcome inventory for hearing aids (IOI-HA). Int J Audiol 41(1):3–26CrossRefPubMed Cox RM, Stephens D, Kramer SE (2002) Translations of the international outcome inventory for hearing aids (IOI-HA). Int J Audiol 41(1):3–26CrossRefPubMed
20.
go back to reference Tjellstrom A, Hakansson B, Granstrom G (2001) Bone-anchored hearing aids: current status in adults and children. Otolaryngol Clin N Am 34(2):337–364CrossRef Tjellstrom A, Hakansson B, Granstrom G (2001) Bone-anchored hearing aids: current status in adults and children. Otolaryngol Clin N Am 34(2):337–364CrossRef
21.
go back to reference Verstraeten N, Zarowski AJ, Somers T, Riff D, Offeciers EF (2009) Comparison of the audiologic results obtained with the bone anchored hearing aid attached to the headband, the test band, and to the ‘‘snap’’ abutment. Otol Neurotol 30:70–75CrossRefPubMed Verstraeten N, Zarowski AJ, Somers T, Riff D, Offeciers EF (2009) Comparison of the audiologic results obtained with the bone anchored hearing aid attached to the headband, the test band, and to the ‘‘snap’’ abutment. Otol Neurotol 30:70–75CrossRefPubMed
22.
go back to reference Gerdes T, Salcher RB, Schwab B, Lenarz T et al (2016) Comparison of audiological results between a transcutaneous and a percutaneous bone conduction instrument in conductive hearing loss. Otol Neurotol 27:685–691CrossRef Gerdes T, Salcher RB, Schwab B, Lenarz T et al (2016) Comparison of audiological results between a transcutaneous and a percutaneous bone conduction instrument in conductive hearing loss. Otol Neurotol 27:685–691CrossRef
23.
go back to reference Powell HRF, Rolfe AM, Birman CS (2015) A comparative study of audiologic outcomes for two transcutaneous bone-anchored hearing devices. Otol Neurotol 36:1525–1531CrossRefPubMed Powell HRF, Rolfe AM, Birman CS (2015) A comparative study of audiologic outcomes for two transcutaneous bone-anchored hearing devices. Otol Neurotol 36:1525–1531CrossRefPubMed
24.
go back to reference Riss D, Arnoldner C, Baumgartner WD, Blineder M et al (2014) Indication criteria and outcomes with the bonebridge transcutaneous bone-conduction implant. Laryngoscope 124:2802–2806CrossRefPubMed Riss D, Arnoldner C, Baumgartner WD, Blineder M et al (2014) Indication criteria and outcomes with the bonebridge transcutaneous bone-conduction implant. Laryngoscope 124:2802–2806CrossRefPubMed
25.
go back to reference Nadaraja GS, Gurgel RK, KiM J, Chang KW (2013) Hearing outcomes of atresia surgery versus osseointegrated bone conduction device in patients with congenital aural atresia: a systematic review. Otol Neurotol 34(8):1394–1399CrossRefPubMed Nadaraja GS, Gurgel RK, KiM J, Chang KW (2013) Hearing outcomes of atresia surgery versus osseointegrated bone conduction device in patients with congenital aural atresia: a systematic review. Otol Neurotol 34(8):1394–1399CrossRefPubMed
26.
go back to reference Gawecki W, Stieler OM, Balcerowiak A et al. (2016) Surgical, functional and audiological evaluation of new BAHA attract system implantations. Eur Arch Otorhinolaryngol 273(10):3123–3130CrossRefPubMedPubMedCentral Gawecki W, Stieler OM, Balcerowiak A et al. (2016) Surgical, functional and audiological evaluation of new BAHA attract system implantations. Eur Arch Otorhinolaryngol 273(10):3123–3130CrossRefPubMedPubMedCentral
27.
go back to reference Carr D, Moraleda J, Procter V, Wright K, Ray J (2015) Initial UK experience with a novel transcutaneous bone conduction device. Otol Neurotol 36:1399–1402CrossRefPubMed Carr D, Moraleda J, Procter V, Wright K, Ray J (2015) Initial UK experience with a novel transcutaneous bone conduction device. Otol Neurotol 36:1399–1402CrossRefPubMed
28.
go back to reference Dimitriadis PA, Farr MR, Allam A, Ray J (2016) Three year experience with the cochlear BAHA attract implant: a systematic review of the literature. BMC Ear Nose Throat Disord 16:12CrossRefPubMedPubMedCentral Dimitriadis PA, Farr MR, Allam A, Ray J (2016) Three year experience with the cochlear BAHA attract implant: a systematic review of the literature. BMC Ear Nose Throat Disord 16:12CrossRefPubMedPubMedCentral
29.
go back to reference Hakansson BE, Carlsson PU, Tjellstrom A, Liden G (1994) The bone-anchored hearing aid: principal design and audiometric results. Ear Nose Throat J 73(9):670–675PubMed Hakansson BE, Carlsson PU, Tjellstrom A, Liden G (1994) The bone-anchored hearing aid: principal design and audiometric results. Ear Nose Throat J 73(9):670–675PubMed
30.
go back to reference Gurgel RK, Jackler RK, Dobie RA, Popelka GR (2012) A new standardized format for reporting hearing outcome in clinical trials. Otolaryngol Head Neck Surg 147(5):803–807CrossRefPubMed Gurgel RK, Jackler RK, Dobie RA, Popelka GR (2012) A new standardized format for reporting hearing outcome in clinical trials. Otolaryngol Head Neck Surg 147(5):803–807CrossRefPubMed
Metadata
Title
A multicenter study on objective and subjective benefits with a transcutaneous bone-anchored hearing aid device: first Nordic results
Authors
Dan Dupont Hougaard
Soren Kjaergaard Boldsen
Anne Marie Jensen
Soren Hansen
Per Cayé Thomassen
Publication date
01-08-2017
Publisher
Springer Berlin Heidelberg
Published in
European Archives of Oto-Rhino-Laryngology / Issue 8/2017
Print ISSN: 0937-4477
Electronic ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-017-4614-8

Other articles of this Issue 8/2017

European Archives of Oto-Rhino-Laryngology 8/2017 Go to the issue