Skip to main content
Top
Published in: Archives of Orthopaedic and Trauma Surgery 2/2015

01-02-2015 | Trauma Surgery

Breakage of cephalomedullary nailing in operative treatment of trochanteric and subtrochanteric femoral fractures

Authors: Christian von Rüden, Sven Hungerer, Peter Augat, Oliver Trapp, Volker Bühren, Christian Hierholzer

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 2/2015

Login to get access

Abstract

Introduction

Mechanical breakage of cephalomedullary nail osteosynthesis is a rare complication attributed to delayed fracture union or nonunion. This study presents a series of cases of breakage and secondary lag screw dislocation after cephalomedullary nailing. The aim of this study was to identify factors that contribute to cephalomedullary nail breakage.

Materials and methods

In a retrospective case series review between 02/2005 and 12/2013, we analyzed 453 patients with trochanteric and subtrochanteric fracture who had been treated by cephalomedullary nailing. Fractures were classified according to AO/OTA classification. 13 patients with cephalomedullary nail breakage were included (failure rate 2.9 %).

Results

Seven patients were women, and six men with a mean age of 72 years (range 35–94). Implant breakage occurred 6 months postoperatively (range 1–19 months). In ten cases, breakage was secondary to delayed or nonunion, which was thought to be mainly due to insufficient reduction of the fracture, and in two cases due to loss of the lag screw because of missing set screw. In one case, breakage was apparent during elective metal removal following complete fracture healing. Short-term outcome was evaluated 6 months after operative revision using Harris hip score in 11 out of 13 patients showing a mean score of 84 %. Complete radiological fracture healing has been found in 11 patients available for follow-up within 6 months after revision surgery.

Discussion

Breakage of cephalomedullary nail osteosynthesis of trochanteric fractures is a severe complication. The results of our study demonstrate that revision surgery provides good clinical and radiological short-term results. Predominately, failures of trochanteric fractures are related to lack of surgeon performance. Therefore, application of the implant requires accurate preoperative planning, advanced surgical experience to evaluate the patient and the fracture classification, and precise surgical technique including attention to detail and anatomical reduction of the fracture fragments.
Literature
1.
go back to reference Matre K, Vinje T, Havelin LI, Gjertsen JE, Furnes O, Espehaug B, Kjellevold SH, Fevang JM (2013) TRIGEN INTERTAN intramedullary nail versus sliding hip screw: a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and 1 year of follow-up. J Bone Joint Surg Am 95(3):200–208. doi:10.2106/JBJS.K.01497 PubMedCrossRef Matre K, Vinje T, Havelin LI, Gjertsen JE, Furnes O, Espehaug B, Kjellevold SH, Fevang JM (2013) TRIGEN INTERTAN intramedullary nail versus sliding hip screw: a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and 1 year of follow-up. J Bone Joint Surg Am 95(3):200–208. doi:10.​2106/​JBJS.​K.​01497 PubMedCrossRef
2.
go back to reference Kreder HJ (2013) Principles and evidence: the optimal treatment of pertrochanteric hip fractures: commentary on an article by Kjell Matre, MD, et al.: “TRIGEN INTERTAN intramedullary nail versus sliding hip screw. a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and one year of follow-up”. J Bone Joint Surg Am 95(3):e16(1–2). doi:10.2106/JBJS.L.01553 CrossRef Kreder HJ (2013) Principles and evidence: the optimal treatment of pertrochanteric hip fractures: commentary on an article by Kjell Matre, MD, et al.: “TRIGEN INTERTAN intramedullary nail versus sliding hip screw. a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and one year of follow-up”. J Bone Joint Surg Am 95(3):e16(1–2). doi:10.​2106/​JBJS.​L.​01553 CrossRef
3.
go back to reference Valverde J, Alonso M, Porro J et al (1998) Use of the gamma nail in the treatment of fractures of the proximal femur. Clin Orthop Relat Res 350:56–61PubMed Valverde J, Alonso M, Porro J et al (1998) Use of the gamma nail in the treatment of fractures of the proximal femur. Clin Orthop Relat Res 350:56–61PubMed
4.
go back to reference Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51(4):737–755PubMed Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51(4):737–755PubMed
5.
go back to reference Dimitriou R, Dahabreh Z, Katsoulis E, Matthews SJ, Branfoot T, Giannoudis PV (2005) Application of recombinant BMP-7 on persistent upper and lower limb non-unions. Injury 36(Suppl 4):51–59CrossRef Dimitriou R, Dahabreh Z, Katsoulis E, Matthews SJ, Branfoot T, Giannoudis PV (2005) Application of recombinant BMP-7 on persistent upper and lower limb non-unions. Injury 36(Suppl 4):51–59CrossRef
6.
7.
go back to reference Yoshino N, Watanabe Y, Tanekaka N et al (2006) Implant failure of long gamma nail in a patient with intertrochanteric-subtrochanteric fracture. J Orthop Sci 11:638–643PubMedCrossRef Yoshino N, Watanabe Y, Tanekaka N et al (2006) Implant failure of long gamma nail in a patient with intertrochanteric-subtrochanteric fracture. J Orthop Sci 11:638–643PubMedCrossRef
8.
go back to reference Koval KJ, Zuckerman JD (1994) Hip fractures. Evaluation and treatment of intertrochanteric fracture. J Am Acad Orthop Surg 2:150–156PubMed Koval KJ, Zuckerman JD (1994) Hip fractures. Evaluation and treatment of intertrochanteric fracture. J Am Acad Orthop Surg 2:150–156PubMed
9.
go back to reference Rodriguez Alvarez J, Casteleiro Gonzolez C, Laguna Aranda R et al (1998) Indications for use of the long gamma nail. Clin Orthop Relat Res 350:62–66PubMed Rodriguez Alvarez J, Casteleiro Gonzolez C, Laguna Aranda R et al (1998) Indications for use of the long gamma nail. Clin Orthop Relat Res 350:62–66PubMed
10.
go back to reference Bhandari M, Schemitsch E, Jonsson A et al (2009) Gamma nails revisited: gamma nails versus compression hip screw in the management of intertrochanteric fractures of the hip: a meta-analysis. J Orthop Trauma 23:460–464PubMedCrossRef Bhandari M, Schemitsch E, Jonsson A et al (2009) Gamma nails revisited: gamma nails versus compression hip screw in the management of intertrochanteric fractures of the hip: a meta-analysis. J Orthop Trauma 23:460–464PubMedCrossRef
11.
go back to reference Saarenpää I, Heikkinen T, Ristiniemi J et al (2009) Functional comparison of the dynamic hip screw and the gamma locking nail in trochanteric hip fractures: a matched-pair study of 268 patients. Int Orthop 33:255–260PubMedCentralPubMedCrossRef Saarenpää I, Heikkinen T, Ristiniemi J et al (2009) Functional comparison of the dynamic hip screw and the gamma locking nail in trochanteric hip fractures: a matched-pair study of 268 patients. Int Orthop 33:255–260PubMedCentralPubMedCrossRef
12.
go back to reference Barton T, Fleeson R, Topliss C et al (2010) A comparison of the long gamma nail with the sliding hip screw for the treatment of AO/OTA 31-A2 fractures of the proximal part of the femur. J Bone Joint Surg Am 92(4):792–798. doi:10.2106/JBJS.I.00508 PubMedCrossRef Barton T, Fleeson R, Topliss C et al (2010) A comparison of the long gamma nail with the sliding hip screw for the treatment of AO/OTA 31-A2 fractures of the proximal part of the femur. J Bone Joint Surg Am 92(4):792–798. doi:10.​2106/​JBJS.​I.​00508 PubMedCrossRef
14.
go back to reference Maes M, Deboer Y, Brabants K (2012) Failure of the titanium trochanteric gamma nail in ununited metastatic fractures. Acta Orthop Belg 78:552–557PubMed Maes M, Deboer Y, Brabants K (2012) Failure of the titanium trochanteric gamma nail in ununited metastatic fractures. Acta Orthop Belg 78:552–557PubMed
15.
go back to reference Alvarez D, Aparicio J, Fernandez E et al (2004) Implant breakage, a rare complication with the gamma nail. A review of 843 fractures of the proximal femur treated with a gamma nail. Acta Orthop Belg 70(5):435–443PubMed Alvarez D, Aparicio J, Fernandez E et al (2004) Implant breakage, a rare complication with the gamma nail. A review of 843 fractures of the proximal femur treated with a gamma nail. Acta Orthop Belg 70(5):435–443PubMed
17.
go back to reference Van den Brink WA, Janssen MC (1994) Failure of the gamma nail in a highly unstable proximal femur fracture: report of four cases encountered in the Netherlands. J Orthop Trauma 9(1):53–56CrossRef Van den Brink WA, Janssen MC (1994) Failure of the gamma nail in a highly unstable proximal femur fracture: report of four cases encountered in the Netherlands. J Orthop Trauma 9(1):53–56CrossRef
18.
go back to reference Najibi S, Lemos M, Fehnel D (2010) Mechanical failure of the long gamma nail in two proximal femur fractures. Iowa Orthop J 30:205–210PubMedCentralPubMed Najibi S, Lemos M, Fehnel D (2010) Mechanical failure of the long gamma nail in two proximal femur fractures. Iowa Orthop J 30:205–210PubMedCentralPubMed
19.
go back to reference Wozasek G, Radler C, Vecsei V (2002) Multiple gamma nail failure. Orthopedics 25(2):183–184PubMed Wozasek G, Radler C, Vecsei V (2002) Multiple gamma nail failure. Orthopedics 25(2):183–184PubMed
20.
go back to reference Gaebler C, Stanzl-Tschegg S, Tschegg E et al (1999) Implant failure of the gamma nail. Injury 30:91–99PubMedCrossRef Gaebler C, Stanzl-Tschegg S, Tschegg E et al (1999) Implant failure of the gamma nail. Injury 30:91–99PubMedCrossRef
21.
go back to reference Wang CJ, Yettram AL, Yao MS et al (1998) Finite element analysis of a gamma nail within a fractured femur. Med Eng Phys 20:677–683PubMedCrossRef Wang CJ, Yettram AL, Yao MS et al (1998) Finite element analysis of a gamma nail within a fractured femur. Med Eng Phys 20:677–683PubMedCrossRef
23.
24.
go back to reference Kasimatis GB, Lambiris E, Tyllianakis M et al (2007) Gamma nail breakage: a report of four cases. J Orthop Surg (Hong Kong) 15(3):368–372 Kasimatis GB, Lambiris E, Tyllianakis M et al (2007) Gamma nail breakage: a report of four cases. J Orthop Surg (Hong Kong) 15(3):368–372
25.
go back to reference Eberle S, Gabel J, Hungerer S, Hoffmann S, Pätzold R, Augat P, Bühren V (2012) Auxiliary locking plate improves fracture stability and healing in intertrochanteric fractures fixated by intramedullary nail. Clin Biomech 27(10):1006–1010. doi:10.1016/j.clinbiomech.2012.07.008 CrossRef Eberle S, Gabel J, Hungerer S, Hoffmann S, Pätzold R, Augat P, Bühren V (2012) Auxiliary locking plate improves fracture stability and healing in intertrochanteric fractures fixated by intramedullary nail. Clin Biomech 27(10):1006–1010. doi:10.​1016/​j.​clinbiomech.​2012.​07.​008 CrossRef
Metadata
Title
Breakage of cephalomedullary nailing in operative treatment of trochanteric and subtrochanteric femoral fractures
Authors
Christian von Rüden
Sven Hungerer
Peter Augat
Oliver Trapp
Volker Bühren
Christian Hierholzer
Publication date
01-02-2015
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 2/2015
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-014-2121-6

Other articles of this Issue 2/2015

Archives of Orthopaedic and Trauma Surgery 2/2015 Go to the issue