Treatment of midshaft clavicular delayed and non-unions with anteroinferior locking compression plating

Authors: Sjoerd A. Stufkens, P. Kloen

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 2/2010

Login to get access

Abstract

Introduction

Pain and impaired shoulder function are the predominant symptoms of midshaft clavicle non-unions. Obtaining consolidation and improvement of shoulder function is often successfully achieved with osteosynthesis and bone grafting. Most data in the literature pertain to plate osteosynthesis, placing the plate on the subcutaneous superior aspect of the clavicle. Although union rates are generally high, most patients require hardware removal as the plate is prominent under the skin causing pain and cosmetic problems.

Materials and methods

In the current retrospective study, we followed a cohort of 21 consecutive cases (20 patients) with a midshaft clavicular delayed or non-union, treated with anteroinferior plating using a 3.5 mm locking compression plate (LCP) for a mean of 30 months.

Results

We operated on 10 males and 10 females with a mean age of 48.2 years (range 16–65). There was one early plate failure that needed revision. Two patients required hardware removal because of prominence of the plate. All but two patients were satisfied with the final cosmetic result. The average DASH score at follow up was 22.8.

Discussion and conclusions

Anteroinferior plating with a 3.5 mm LCP is a reliable and reproducible treatment of midshaft clavicular delayed and non-union regarding consolidation, function, cosmesis and reduction of second surgery.
Literature
1.
go back to reference Eskola A, Vainionpaa S, Myllynen P, Patiala H, Rokkanen P (1986) Outcome of clavicular fracture in 89 patients. Arch Orthop Trauma Surg 105(6):337–338CrossRefPubMed Eskola A, Vainionpaa S, Myllynen P, Patiala H, Rokkanen P (1986) Outcome of clavicular fracture in 89 patients. Arch Orthop Trauma Surg 105(6):337–338CrossRefPubMed
2.
go back to reference Olsen BS, Vaesel MT, Sojbjerg JO (1995) Treatment of midshaft clavicular nonunion with plate fixation and autologous bone grafting. J Shoulder Elbow Surg 4(5):337–344CrossRefPubMed Olsen BS, Vaesel MT, Sojbjerg JO (1995) Treatment of midshaft clavicular nonunion with plate fixation and autologous bone grafting. J Shoulder Elbow Surg 4(5):337–344CrossRefPubMed
3.
go back to reference Robinson CM (1998) Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br 80(3):476–484CrossRefPubMed Robinson CM (1998) Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br 80(3):476–484CrossRefPubMed
4.
go back to reference Wilkins RM, Johnston RM (1983) Ununited fractures of the clavicle. J Bone Joint Surg Am 65(6):773–778PubMed Wilkins RM, Johnston RM (1983) Ununited fractures of the clavicle. J Bone Joint Surg Am 65(6):773–778PubMed
5.
go back to reference Zlowodzki M, Zelle BA, Cole PA, Jeray K, McKee MD (2005) Treatment of acute midshaft clavicle fractures: systematic review of 2144 fractures: on behalf of the Evidence-Based Orthopaedic Trauma Working Group. J Orthop Trauma 19(7):504–507CrossRefPubMed Zlowodzki M, Zelle BA, Cole PA, Jeray K, McKee MD (2005) Treatment of acute midshaft clavicle fractures: systematic review of 2144 fractures: on behalf of the Evidence-Based Orthopaedic Trauma Working Group. J Orthop Trauma 19(7):504–507CrossRefPubMed
6.
go back to reference Jupiter JB, Leffert RD (1987) Non-union of the clavicle. Associated complications and surgical management. J Bone Joint Surg Am 69(5):753–760PubMed Jupiter JB, Leffert RD (1987) Non-union of the clavicle. Associated complications and surgical management. J Bone Joint Surg Am 69(5):753–760PubMed
7.
go back to reference Collinge C, Devinney S, Herscovici D, DiPasquale T, Sanders R (2006) Anterior-inferior plate fixation of middle-third fractures and nonunions of the clavicle. J Orthop Trauma 20(10):680–686CrossRefPubMed Collinge C, Devinney S, Herscovici D, DiPasquale T, Sanders R (2006) Anterior-inferior plate fixation of middle-third fractures and nonunions of the clavicle. J Orthop Trauma 20(10):680–686CrossRefPubMed
8.
go back to reference Coupe BD, Wimhurst JA, Indar R, Calder DA, Patel AD (2005) A new approach for plate fixation of midshaft clavicular fractures. Injury 36(10):1166–1171CrossRefPubMed Coupe BD, Wimhurst JA, Indar R, Calder DA, Patel AD (2005) A new approach for plate fixation of midshaft clavicular fractures. Injury 36(10):1166–1171CrossRefPubMed
9.
go back to reference Kloen P, Sorkin AT, Rubel IF, Helfet DL (2002) Anteroinferior plating of midshaft clavicular nonunions. J Orthop Trauma 16(6):425–430CrossRefPubMed Kloen P, Sorkin AT, Rubel IF, Helfet DL (2002) Anteroinferior plating of midshaft clavicular nonunions. J Orthop Trauma 16(6):425–430CrossRefPubMed
10.
go back to reference Brouwer KM, Wright TC, Ring DC (2009) Failure of superior locking clavicle plate by axial pull-out of the lateral screws: a report of four cases. J Shoulder Elbow Surg 18(1):e22–e25CrossRefPubMed Brouwer KM, Wright TC, Ring DC (2009) Failure of superior locking clavicle plate by axial pull-out of the lateral screws: a report of four cases. J Shoulder Elbow Surg 18(1):e22–e25CrossRefPubMed
11.
go back to reference Khan SA, Shamshery P, Gupta V, Trikha V, Varshney MK, Kumar A (2008) Locking compression plate in long standing clavicular nonunions with poor bone stock. J Trauma 64(2):439–441CrossRefPubMed Khan SA, Shamshery P, Gupta V, Trikha V, Varshney MK, Kumar A (2008) Locking compression plate in long standing clavicular nonunions with poor bone stock. J Trauma 64(2):439–441CrossRefPubMed
12.
go back to reference Celestre P, Roberston C, Mahar A, Oka R, Meunier M, Schwartz A (2008) Biomechanical evaluation of clavicle fracture plating techniques: does a locking plate provide improved stability? J Orthop Trauma 22(4):241–247CrossRefPubMed Celestre P, Roberston C, Mahar A, Oka R, Meunier M, Schwartz A (2008) Biomechanical evaluation of clavicle fracture plating techniques: does a locking plate provide improved stability? J Orthop Trauma 22(4):241–247CrossRefPubMed
13.
go back to reference Iannotti MR, Crosby LA, Stafford P, Grayson G, Goulet R (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Shoulder Elbow Surg 11(5):457–462CrossRefPubMed Iannotti MR, Crosby LA, Stafford P, Grayson G, Goulet R (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Shoulder Elbow Surg 11(5):457–462CrossRefPubMed
14.
go back to reference Der TJ, Davison JN, Dias JJ (2002) Clavicular fracture non-union surgical outcome and complications. Injury 33(2):135–143CrossRef Der TJ, Davison JN, Dias JJ (2002) Clavicular fracture non-union surgical outcome and complications. Injury 33(2):135–143CrossRef
15.
go back to reference Rosenberg N, Neumann L, Wallace AW (2007) Functional outcome of surgical treatment of symptomatic nonunion and malunion of midshaft clavicle fractures. J Shoulder Elbow Surg 16(5):510–513CrossRefPubMed Rosenberg N, Neumann L, Wallace AW (2007) Functional outcome of surgical treatment of symptomatic nonunion and malunion of midshaft clavicle fractures. J Shoulder Elbow Surg 16(5):510–513CrossRefPubMed
16.
go back to reference Boehme D, Curtis RJ Jr, DeHaan JT, Kay SP, Young DC, Rockwood CA Jr (1991) Non-union of fractures of the mid-shaft of the clavicle. Treatment with a modified Hagie intramedullary pin and autogenous bone-grafting. J Bone Joint Surg Am 73(8):1219–1226PubMed Boehme D, Curtis RJ Jr, DeHaan JT, Kay SP, Young DC, Rockwood CA Jr (1991) Non-union of fractures of the mid-shaft of the clavicle. Treatment with a modified Hagie intramedullary pin and autogenous bone-grafting. J Bone Joint Surg Am 73(8):1219–1226PubMed
17.
go back to reference Mullaji AB, Jupiter JB (1994) Low-contact dynamic compression plating of the clavicle. Injury 25(1):41–45CrossRefPubMed Mullaji AB, Jupiter JB (1994) Low-contact dynamic compression plating of the clavicle. Injury 25(1):41–45CrossRefPubMed
18.
go back to reference Canadian Orthopaedic Trauma Society (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am 89(1):1–10CrossRef Canadian Orthopaedic Trauma Society (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am 89(1):1–10CrossRef
19.
go back to reference Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539CrossRefPubMed Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539CrossRefPubMed
Metadata
Title
Treatment of midshaft clavicular delayed and non-unions with anteroinferior locking compression plating
Authors
Sjoerd A. Stufkens
P. Kloen
Publication date
01-02-2010
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 2/2010
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-009-0864-2

Other articles of this Issue 2/2010

Archives of Orthopaedic and Trauma Surgery 2/2010 Go to the issue