Published in: Archives of Orthopaedic and Trauma Surgery 2/2008

01-02-2008 | Trauma Surgery

Proximal humeral fractures: what is semi-rigid? Biomechanical properties of semi-rigid implants, a biomechanical cadaver based evaluation

Authors: F. Kralinger, M. Gschwentner, M. Wambacher, V. Smekal, C. Haid

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 2/2008

Login to get access

Abstract

Introduction

Proximal humerus fractures remain challenging especially in the elderly. Biomechanical data put semi-rigid implants in favour of osteopenic or osteoporotic situation. Little surgical side damage is associated with a minimal invasive approach of these implants. The aim of this study was to evaluate the mechanical properties of three such implants.

Material and methods

Fresh frozen cadaver specimens were mounted as proposed by the distributors. Three different implants were used: LCP-PH (locking compression plate proximal humerus, Synthes, Austria), HB (humerus block, Synthes, Austria), and IMC (intramedullary claw, ITS, Austria). Subcapital fracture was simulated by resecting a 5 mm gap. All specimens were comparable in “B” (one), “M” (ineral) and “D” (ensity). Four load cases were tested: varus bending, medial shearing and axial torque. A cyclic test (1,000 cycles) was performed in the first load case (varus stress) for all three implants.

Results

The LCP-PH was the most rigid in all three load cases, always followed by the HB. The IMC was the most elastic device with almost immeasurable values in axial torque. In the cyclic setting, the load reduction of the HB followed by the LCP-PH was significantly better than that for the IMC.

Conclusion

The differences in stiffness are varying tremendously. The IMC is the implant with the lowest stiffness in all load cases and the highest load reduction. New “semi-rigids” claim good clinical performance, yet prospective clinical studies have to prove this. It is unlikely that the IMC can maintain fracture reduction in fracture situations of complex nature (no ligamentotaxis).
Literature
1.
go back to reference Anjum SN, Butt MS (2005) Treatment of comminuted proximal humerus fractures with shoulder hemiarthroplasty in elderly patients. Acta Orthop Belg 71:388–395PubMed Anjum SN, Butt MS (2005) Treatment of comminuted proximal humerus fractures with shoulder hemiarthroplasty in elderly patients. Acta Orthop Belg 71:388–395PubMed
2.
go back to reference Ballmer FT, Hertel R (1998) Indications and results of shoulder prosthetics in complex proximal humerus fractures. Ther Umsch 55:197–202PubMed Ballmer FT, Hertel R (1998) Indications and results of shoulder prosthetics in complex proximal humerus fractures. Ther Umsch 55:197–202PubMed
3.
go back to reference Boileau P, Krishnan SG, Tinsi L, Walch G, Coste JS, Mole D (2002) Tuberosity malposition and migration: reasons for poor outcomes after hemiarthroplasty for displaced fractures of the proximal humerus. J Shoulder Elbow Surg 11:401–412PubMedCrossRef Boileau P, Krishnan SG, Tinsi L, Walch G, Coste JS, Mole D (2002) Tuberosity malposition and migration: reasons for poor outcomes after hemiarthroplasty for displaced fractures of the proximal humerus. J Shoulder Elbow Surg 11:401–412PubMedCrossRef
4.
go back to reference Coleman SH, Craig EV (2002) Hemiarthroplasty for complex fractures of the proximal humerus: surgical technique and results with the Atlas trimodular prosthesis. Am J Orthop 31:11–17PubMedCrossRef Coleman SH, Craig EV (2002) Hemiarthroplasty for complex fractures of the proximal humerus: surgical technique and results with the Atlas trimodular prosthesis. Am J Orthop 31:11–17PubMedCrossRef
5.
go back to reference Duda GN, Epari DR, Babst R, Lambert SM, Matthys R, Sudkamp NP (2007) Mechanical evaluation of a new minimally invasive device for stabilization of proximal humeral fractures in elderly patients. A cadaver study. Acta Orthop 78:430–435PubMedCrossRef Duda GN, Epari DR, Babst R, Lambert SM, Matthys R, Sudkamp NP (2007) Mechanical evaluation of a new minimally invasive device for stabilization of proximal humeral fractures in elderly patients. A cadaver study. Acta Orthop 78:430–435PubMedCrossRef
6.
go back to reference Hertel R (2005) Fractures of the proximal humerus in osteoporotic bone. Osteoporos Int 16(Suppl 2):S65–S72PubMedCrossRef Hertel R (2005) Fractures of the proximal humerus in osteoporotic bone. Osteoporos Int 16(Suppl 2):S65–S72PubMedCrossRef
7.
go back to reference Kralinger F, Schwaiger R, Wambacher M, Farrell E, Menth-Chiari W, Lajtai G, Hubner C, Resch H (2004) Outcome after primary hemiarthroplasty for fracture of the head of the humerus. A retrospective multicentre study of 167 patients. J Bone Joint Surg Br 86:217–219PubMedCrossRef Kralinger F, Schwaiger R, Wambacher M, Farrell E, Menth-Chiari W, Lajtai G, Hubner C, Resch H (2004) Outcome after primary hemiarthroplasty for fracture of the head of the humerus. A retrospective multicentre study of 167 patients. J Bone Joint Surg Br 86:217–219PubMedCrossRef
8.
go back to reference Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN (2003) Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg 123:74–81PubMed Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN (2003) Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg 123:74–81PubMed
9.
go back to reference Lill H, Hepp P, Rose T, Konig K, Josten C (2004) The angle stable locking-proximal-humerus-plate (LPHP) for proximal humeral fractures using a small anterior-lateral-deltoid-splitting-approach—technique and first results. Zentralbl Chir 129:43–48PubMedCrossRef Lill H, Hepp P, Rose T, Konig K, Josten C (2004) The angle stable locking-proximal-humerus-plate (LPHP) for proximal humeral fractures using a small anterior-lateral-deltoid-splitting-approach—technique and first results. Zentralbl Chir 129:43–48PubMedCrossRef
10.
go back to reference Povacz P, Resch H (1998) Osteosynthesis of proximal humerus fractures. Ther Umsch 55:192–196PubMed Povacz P, Resch H (1998) Osteosynthesis of proximal humerus fractures. Ther Umsch 55:192–196PubMed
12.
go back to reference Resch H, Hubner C, Schwaiger R (2001) Minimally invasive reduction and osteosynthesis of articular fractures of the humeral head. Injury 32(Suppl 1):SA25–SA32PubMed Resch H, Hubner C, Schwaiger R (2001) Minimally invasive reduction and osteosynthesis of articular fractures of the humeral head. Injury 32(Suppl 1):SA25–SA32PubMed
13.
go back to reference Ruegsegger P, Kalender WA (1993) A phantom for standardization and quality-control in peripheral bone measurements by Pqct and Dxa. Phys Med Biol 38:1963–1970CrossRef Ruegsegger P, Kalender WA (1993) A phantom for standardization and quality-control in peripheral bone measurements by Pqct and Dxa. Phys Med Biol 38:1963–1970CrossRef
14.
go back to reference Siffri PC, Peindl RD, Coley ER, Norton J, Connor PM, Kellam JE (2006) Biomechanical analysis of blade plate versus locking plate fixation for a proximal humerus fracture: Comparison using cadaveric and synthetic humeri. J Ortho Trauma 20:547–554CrossRef Siffri PC, Peindl RD, Coley ER, Norton J, Connor PM, Kellam JE (2006) Biomechanical analysis of blade plate versus locking plate fixation for a proximal humerus fracture: Comparison using cadaveric and synthetic humeri. J Ortho Trauma 20:547–554CrossRef
Metadata
Title
Proximal humeral fractures: what is semi-rigid? Biomechanical properties of semi-rigid implants, a biomechanical cadaver based evaluation
Authors
F. Kralinger
M. Gschwentner
M. Wambacher
V. Smekal
C. Haid
Publication date
01-02-2008
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 2/2008
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-007-0512-7

Other articles of this Issue 2/2008

Archives of Orthopaedic and Trauma Surgery 2/2008 Go to the issue