Skip to main content
Top
Published in: Archives of Orthopaedic and Trauma Surgery 6/2008

01-06-2008 | Basic Science

Reference markers in computer aided orthopaedic surgery: rotational stability testings and clinical implications

Authors: Daniel Kendoff, Michael J. Gardner, Christian Krettek, Tobias Hüfner, Musa Citak

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 6/2008

Login to get access

Abstract

Navigation procedures rely on the stability of the reference arrays (RA) fixed to the bony anatomy. The risk of inadvertent collision and unnoticed movements of the RA may occur, and limb movements might also provocate collisions. Consequently, relevant measurements failures during the navigated procedure might occur and reduce the overall precision of the system. The magnitude of torque to destabilize an RA from its bony-fixation is unknown. The purpose of this study was to determine the ability of standard RA’s to resist applied torque. A digital torque application device was developed to allow for precise torque application to the RA system at four cadavers. Clockwise, gradually increasing rotational force was applied to the RA in 1, 2, or 3 Nm, held for 1 s and released, repeated in 10 cycles. One pin fixation systems with 4.0 Schanz pins were used. A second RA was fixed 20 cm proximal to the tested RA. A navigation system was used to measure the relative positions of both RA’s during torque application. The rotational differences at torque application were calculated and statistically evaluated. Results revealed averaged 1.0° rotation [range (r), 1.0–1.1°] with first cycle of 1 Nm torque, the last cycle averaged 1.0° (r, 1.0–1.1°) with no significant differences in rotation between any of the cycles (p > 0.5). Torque of 2 Nm resulted in 1.8° rotation (r, 1.5–1.8°) with no significant increases between first and second cycle (p = 0.56), second and third trials (p = 0.35) while fourth cycle averaged 2.0°, with significant increase (p = 0.011). All subsequent cycles resulted in significant increases. Torque of 3 Nm produced 2.9° of rotation for initial cylce (r,2.5–3.3°), with significant increases with each cycle (p < 005). Torques of ≥2 Nm may cause loosening of the RA, thus may lose its original position relative to the bone. Surgeons using computer aided surgery systems should be aware of possible unrecognized movements of the RA, however, gentle collisions do not appear to cause significant motion or destabilization.
Literature
1.
go back to reference Ensini A, Catani F, Leardini A, Romagnoli M, Giannini S (2007) Alignments and clinical results in conventional and navigated total knee arthroplasty. Clin Orthop Relat Res 457:156–162PubMed Ensini A, Catani F, Leardini A, Romagnoli M, Giannini S (2007) Alignments and clinical results in conventional and navigated total knee arthroplasty. Clin Orthop Relat Res 457:156–162PubMed
2.
go back to reference Hüfner T, Kendoff D, Citak M, Geerling J, Krettek C (2006) Precision in orthopaedic computer navigation. Orthopade 35:1043–1055PubMedCrossRef Hüfner T, Kendoff D, Citak M, Geerling J, Krettek C (2006) Precision in orthopaedic computer navigation. Orthopade 35:1043–1055PubMedCrossRef
3.
go back to reference Hüfner T, Kfuri M Jr, Kendoff D, Richter M, Geerling J, Krettek C (2003) Navigated osteosynthesis of the proximal femur. An experimental study. Unfallchirurg 106:975–979PubMed Hüfner T, Kfuri M Jr, Kendoff D, Richter M, Geerling J, Krettek C (2003) Navigated osteosynthesis of the proximal femur. An experimental study. Unfallchirurg 106:975–979PubMed
4.
go back to reference Kahler DM (2004) Image guidance: fluoroscopic navigation. Clin Orthop Relat Res April(421):70–76CrossRef Kahler DM (2004) Image guidance: fluoroscopic navigation. Clin Orthop Relat Res April(421):70–76CrossRef
5.
go back to reference Kendoff D, Bogojevic A, Citak M, Citak M, Maier C, Maier G, Krettek C, Hufner T (2007) Experimental validation of noninvasive referencing in navigated procedures on long bones. J Orthop Res 25:201–207PubMedCrossRef Kendoff D, Bogojevic A, Citak M, Citak M, Maier C, Maier G, Krettek C, Hufner T (2007) Experimental validation of noninvasive referencing in navigated procedures on long bones. J Orthop Res 25:201–207PubMedCrossRef
6.
go back to reference Kendoff D, Citak M, Pearle A, Gardner MJ, Hankemeier S, Krettek C, Hufner T (2007) Influence of lower limb rotation in navigated alignment analysis: implications for high tibial osteotomies. Knee Surg Sports Traumatol Arthrosc 15(8):1003–1008PubMedCrossRef Kendoff D, Citak M, Pearle A, Gardner MJ, Hankemeier S, Krettek C, Hufner T (2007) Influence of lower limb rotation in navigated alignment analysis: implications for high tibial osteotomies. Knee Surg Sports Traumatol Arthrosc 15(8):1003–1008PubMedCrossRef
7.
go back to reference Khadem R, Yeh CC, Sadeghi-Tehrani M, Bax MR, Johnson JA, Welch JN, Wilkinson EP, Shahidi R (2000) Comparative tracking error analysis of five different optical tracking systems. Comput Aided Surg 5:98–107PubMedCrossRef Khadem R, Yeh CC, Sadeghi-Tehrani M, Bax MR, Johnson JA, Welch JN, Wilkinson EP, Shahidi R (2000) Comparative tracking error analysis of five different optical tracking systems. Comput Aided Surg 5:98–107PubMedCrossRef
8.
go back to reference Li Q, Zamorano L, Jiang Z, Gong JX, Pandya A, Perez R, Diaz F (1999) Effect of optical digitizer selection on the application accuracy of a surgical localization system-a quantitative comparison between the OPTOTRAK and flashpoint tracking systems. Comput Aided Surg 4:314–321PubMedCrossRef Li Q, Zamorano L, Jiang Z, Gong JX, Pandya A, Perez R, Diaz F (1999) Effect of optical digitizer selection on the application accuracy of a surgical localization system-a quantitative comparison between the OPTOTRAK and flashpoint tracking systems. Comput Aided Surg 4:314–321PubMedCrossRef
9.
go back to reference Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89:236–243PubMedCrossRef Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89:236–243PubMedCrossRef
10.
go back to reference Mayr E, de la Barrera JL, Eller G, Bach C, Nogler M (2006) The effect of fixation and location on the stability of the markers in navigated total hip arthroplasty: a cadaver study. J Bone Joint Surg Br 88:168–172PubMedCrossRef Mayr E, de la Barrera JL, Eller G, Bach C, Nogler M (2006) The effect of fixation and location on the stability of the markers in navigated total hip arthroplasty: a cadaver study. J Bone Joint Surg Br 88:168–172PubMedCrossRef
11.
go back to reference Mihalko WM, Duquin T, Axelrod JR, Bayers-Thering M, Krackow KA (2006) Effect of one- and two-pin reference anchoring systems on marker stability during total knee arthroplasty computer navigation. Comput Aided Surg 11:93–98PubMedCrossRef Mihalko WM, Duquin T, Axelrod JR, Bayers-Thering M, Krackow KA (2006) Effect of one- and two-pin reference anchoring systems on marker stability during total knee arthroplasty computer navigation. Comput Aided Surg 11:93–98PubMedCrossRef
12.
go back to reference Ohnsorge JA, Weisskopf M, Siebert CH (2005) epiDRB—a new minimally invasive concept for referencing in the field of computer-assisted orthopaedic surgery. Z Orthop Ihre Grenzgeb 143:316–322PubMedCrossRef Ohnsorge JA, Weisskopf M, Siebert CH (2005) epiDRB—a new minimally invasive concept for referencing in the field of computer-assisted orthopaedic surgery. Z Orthop Ihre Grenzgeb 143:316–322PubMedCrossRef
13.
go back to reference Parratte S, Argenson JN (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am 89:494–499PubMedCrossRef Parratte S, Argenson JN (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am 89:494–499PubMedCrossRef
14.
go back to reference Wagner A, Schicho K, Birkfellner W, Figl M, Seemann R, Konig F, Kainberger F, Ewers R (2002) Quantitative analysis of factors affecting intraoperative precision and stability of optoelectronic and electromagnetic tracking systems. Med Phys 29:905–912PubMedCrossRef Wagner A, Schicho K, Birkfellner W, Figl M, Seemann R, Konig F, Kainberger F, Ewers R (2002) Quantitative analysis of factors affecting intraoperative precision and stability of optoelectronic and electromagnetic tracking systems. Med Phys 29:905–912PubMedCrossRef
Metadata
Title
Reference markers in computer aided orthopaedic surgery: rotational stability testings and clinical implications
Authors
Daniel Kendoff
Michael J. Gardner
Christian Krettek
Tobias Hüfner
Musa Citak
Publication date
01-06-2008
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 6/2008
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-007-0496-3

Other articles of this Issue 6/2008

Archives of Orthopaedic and Trauma Surgery 6/2008 Go to the issue