Published in: Archives of Orthopaedic and Trauma Surgery 7/2006

01-09-2006 | Orthopaedic Surgery

Fracture of ceramic heads in total hip replacement

Authors: B. Habermann, W. Ewald, M. Rauschmann, L. Zichner, A. A. Kurth

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 7/2006

Login to get access

Abstract

After introduction of ceramics in total hip replacement, there have been several studies on wear and fracture of the femoral head component. Though reports on fractures are few, we saw four fractures within 2 months. In all patients, a cementless hip prosthesis by four different surgeons was implanted between 3/2001 and 2/2004. In three patients, a ceramic-on-polyethylene pair and in one, a ceramic-on-ceramic pair was used. Only one patient suffered an adequate trauma. The mean survival of the ceramic head was 27 months (11–42). In two patients with polyethylene inlays, the inlay showed signs of wear out due to the fractured head. All four revision surgeries had a good outcome with satisfying results and no complications. Though we observe the postoperative development after implantation of ceramic components closely, we still believe that ceramics in total hip replacement in young and active patients are indicated with good long term results.
Literature
1.
go back to reference Allain J, Le Mouel S, Goutallier D, Voisin MC (1999) Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br 81(5):835–842PubMedCrossRef Allain J, Le Mouel S, Goutallier D, Voisin MC (1999) Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br 81(5):835–842PubMedCrossRef
2.
3.
go back to reference Bohler M, Mochida Y, Bauer TW, Plenk Jr H, Salzer M (2000) Wear debris from two different alumina-on-alumina total hip arthroplasties. J Bone Joint Surg 82B:901–909CrossRef Bohler M, Mochida Y, Bauer TW, Plenk Jr H, Salzer M (2000) Wear debris from two different alumina-on-alumina total hip arthroplasties. J Bone Joint Surg 82B:901–909CrossRef
4.
go back to reference Borssen B, Kerrholm J, Snorrason F (1991) Osteolysis after ceramic-on-ceramic hip arthorplasty. A case report. Acta Orthop Scand 62(1):73–75 Borssen B, Kerrholm J, Snorrason F (1991) Osteolysis after ceramic-on-ceramic hip arthorplasty. A case report. Acta Orthop Scand 62(1):73–75
5.
go back to reference Boutin P (1972) Arthroplastie totale de hanche par prothèse en alumine frittée. Rev Chir Orthop 58:229–246PubMed Boutin P (1972) Arthroplastie totale de hanche par prothèse en alumine frittée. Rev Chir Orthop 58:229–246PubMed
6.
go back to reference Boutin P, Blanquaert D (1981) Le frottement Al/Al en chirurgie de la hanche–1205 arthroplasties totales. Rev Chir Orthop; 279–287 Boutin P, Blanquaert D (1981) Le frottement Al/Al en chirurgie de la hanche–1205 arthroplasties totales. Rev Chir Orthop; 279–287
7.
go back to reference Dawihl W, Dorre E, Altmeyer G (1980) Determination of the service life of a friction locking fixation element for use in ceramic total hips. Biomed Tech 25(12):311–315CrossRef Dawihl W, Dorre E, Altmeyer G (1980) Determination of the service life of a friction locking fixation element for use in ceramic total hips. Biomed Tech 25(12):311–315CrossRef
8.
go back to reference Dorre E, Dawihl W, Altmeyer G (1977) Fatigue strength of ceramic hip endoprostheses. Biomed Tech 22(1–2):3–7 Dorre E, Dawihl W, Altmeyer G (1977) Fatigue strength of ceramic hip endoprostheses. Biomed Tech 22(1–2):3–7
9.
go back to reference Dorre E, Dawihl W, Krohn U, Altmeyer G, Semlithsc M (1982) Do ceramic components of artificial hip joints keep their strength in the human body? Biomed Tech 27(12): 303–308 Dorre E, Dawihl W, Krohn U, Altmeyer G, Semlithsc M (1982) Do ceramic components of artificial hip joints keep their strength in the human body? Biomed Tech 27(12): 303–308
10.
go back to reference Fritsch EW, Gleitz M (1996) Ceramic femoral head fractures in total hip arthoplasty. Clin Orthop 328:129–136PubMedCrossRef Fritsch EW, Gleitz M (1996) Ceramic femoral head fractures in total hip arthoplasty. Clin Orthop 328:129–136PubMedCrossRef
11.
go back to reference Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L (2005) Ceramics in total hip replacement. Clin Orhop Relat Res 430:62–71CrossRef Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L (2005) Ceramics in total hip replacement. Clin Orhop Relat Res 430:62–71CrossRef
12.
go back to reference Harms J, Mausle E (1979) Tissue reaction to ceramic implant material. J Biomed Mater Res 13:67–87PubMedCrossRef Harms J, Mausle E (1979) Tissue reaction to ceramic implant material. J Biomed Mater Res 13:67–87PubMedCrossRef
13.
go back to reference Heck DA, Partridge CM, Reuben JD et al. (1995) Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty 10:575–580PubMed Heck DA, Partridge CM, Reuben JD et al. (1995) Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty 10:575–580PubMed
14.
go back to reference Heck DA, Partridge CM, Reuben Jd, Lanzer WL, Lewis CG, Keating EM (1995) Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty 10:575–580PubMed Heck DA, Partridge CM, Reuben Jd, Lanzer WL, Lewis CG, Keating EM (1995) Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty 10:575–580PubMed
15.
16.
go back to reference Hummer CD, Rothmann RH, Hozack WJ (1995) Catastrophic failure of modular zirconia-ceramic femoral head components after total hip arthroplasty. J Arthroplasty 10(6):848–850PubMedCrossRef Hummer CD, Rothmann RH, Hozack WJ (1995) Catastrophic failure of modular zirconia-ceramic femoral head components after total hip arthroplasty. J Arthroplasty 10(6):848–850PubMedCrossRef
17.
go back to reference Kern S, Schreiber A, Hilfiker B (1990) Ceramic head fracture—a rare complication in hip endoprosthesis. Case report and literature survey. Z Orthop Ihre Grenzgeb 128(5):543–548PubMedCrossRef Kern S, Schreiber A, Hilfiker B (1990) Ceramic head fracture—a rare complication in hip endoprosthesis. Case report and literature survey. Z Orthop Ihre Grenzgeb 128(5):543–548PubMedCrossRef
18.
go back to reference Kontinnen YT, Zhao D, Beklen A, Ma G et al (2005) The Microenvironment around total hip replacement prostheses. Clin Orthop Relat Res 430:28–38CrossRef Kontinnen YT, Zhao D, Beklen A, Ma G et al (2005) The Microenvironment around total hip replacement prostheses. Clin Orthop Relat Res 430:28–38CrossRef
19.
go back to reference Maccauro G, Piconi C, Burger W, Pilloni L, De Santis E, Muratori F, Learmonth ID (2004) Fracture of a Y-TZP ceramic femoral head. J Bone Joint Surg Br 86-B:1192–1196CrossRef Maccauro G, Piconi C, Burger W, Pilloni L, De Santis E, Muratori F, Learmonth ID (2004) Fracture of a Y-TZP ceramic femoral head. J Bone Joint Surg Br 86-B:1192–1196CrossRef
20.
go back to reference Michaud RJ, Rasha SY (1995) Sontaneaous fracture of the ceramic ball in a ceramic-polyethylene total hip arthroplasty. J Arthroplasty 10(6):863–867PubMedCrossRef Michaud RJ, Rasha SY (1995) Sontaneaous fracture of the ceramic ball in a ceramic-polyethylene total hip arthroplasty. J Arthroplasty 10(6):863–867PubMedCrossRef
21.
go back to reference Mittelmeier TH, Walter A (1987) The influence of prosthesis design on wear and loosening phenomena. CRC Crit Rev Biocompatibility 3:19 Mittelmeier TH, Walter A (1987) The influence of prosthesis design on wear and loosening phenomena. CRC Crit Rev Biocompatibility 3:19
22.
go back to reference Norton MR, Yarlagadda R, Anderson GH (2002) Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head. J Bone Joint Surg Br 84(1):631–635PubMedCrossRef Norton MR, Yarlagadda R, Anderson GH (2002) Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head. J Bone Joint Surg Br 84(1):631–635PubMedCrossRef
23.
go back to reference Schwägerl W, Zenz P, Pospisil C, Gertschak W. (1995) Aspekte zur Zukunft der Hüftendoprothese an Hand von 10-Jahresergebnissen mit dem Zweymueller-Schaft und dem McKee-System. Die Metallpaarung „Metasul” in der Hüftendoprothetik. Huber, Bern Schwägerl W, Zenz P, Pospisil C, Gertschak W. (1995) Aspekte zur Zukunft der Hüftendoprothese an Hand von 10-Jahresergebnissen mit dem Zweymueller-Schaft und dem McKee-System. Die Metallpaarung „Metasul” in der Hüftendoprothetik. Huber, Bern
24.
go back to reference Sedel L, Bizot P, Nizar R, Meunier A (1998) Perspective on a 25 years experience with ceramic on ceramic articulation in total hip replacement. Semin Arthroplasty 9:123–124 Sedel L, Bizot P, Nizar R, Meunier A (1998) Perspective on a 25 years experience with ceramic on ceramic articulation in total hip replacement. Semin Arthroplasty 9:123–124
25.
26.
go back to reference Simon JA, Dayan AJ, Ergas E et al (1998) Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty 13:108–113PubMedCrossRef Simon JA, Dayan AJ, Ergas E et al (1998) Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty 13:108–113PubMedCrossRef
28.
go back to reference Walter A (1997) Ceramic-to-ceramic combination. Relic or renaissance? Orthopade 26:110–116PubMed Walter A (1997) Ceramic-to-ceramic combination. Relic or renaissance? Orthopade 26:110–116PubMed
29.
go back to reference Willman G (1998) Ceramics for total joint replacement—what a surgeon should know. Orthopedics 21:173–177 Willman G (1998) Ceramics for total joint replacement—what a surgeon should know. Orthopedics 21:173–177
30.
go back to reference Willman G (2000) The evolution of ceramics in total hip replacement. Hip Int 10:193–203 Willman G (2000) The evolution of ceramics in total hip replacement. Hip Int 10:193–203
31.
32.
go back to reference Wirganowicz PZ, Thomas BJ (1997) Massive osteolysis after ceramic on ceramic total hip arthroplasty. A case report. Clin Orthop 338:100–104CrossRef Wirganowicz PZ, Thomas BJ (1997) Massive osteolysis after ceramic on ceramic total hip arthroplasty. A case report. Clin Orthop 338:100–104CrossRef
33.
go back to reference Wirganowicz PZ, Thomas BJ (1998) Massive osteolysis after ceramic on ceramic total hip arthroplasty. A case report. Clin Orthop Relat Res 349:273–274CrossRef Wirganowicz PZ, Thomas BJ (1998) Massive osteolysis after ceramic on ceramic total hip arthroplasty. A case report. Clin Orthop Relat Res 349:273–274CrossRef
34.
go back to reference Wroblewski BM, Siney PD, Fleming PA (2003) Wear of enhanced ultra-high molecular-weight polyethylene (Hylamer) in comination with a 22.225 mm diameter zirconia fermoral head. J bone Joint Surg 85B:376–379CrossRef Wroblewski BM, Siney PD, Fleming PA (2003) Wear of enhanced ultra-high molecular-weight polyethylene (Hylamer) in comination with a 22.225 mm diameter zirconia fermoral head. J bone Joint Surg 85B:376–379CrossRef
35.
go back to reference Yoon TR, Rowe SM, Jung ST, Seon KJ, Maloney WJ (1998) Osteolysis in association with a total hip arthroplasty with ceramic bearing surfaces. J Bone Joint Surg Am 80(10):1459–1468PubMed Yoon TR, Rowe SM, Jung ST, Seon KJ, Maloney WJ (1998) Osteolysis in association with a total hip arthroplasty with ceramic bearing surfaces. J Bone Joint Surg Am 80(10):1459–1468PubMed
36.
go back to reference Zichner L, Willert HG (1992) Comparison of Alumina-Polyethylene and Metal-Polyethylene in Clinical Trials. Clin Orthop Relat Res 282:86–94PubMed Zichner L, Willert HG (1992) Comparison of Alumina-Polyethylene and Metal-Polyethylene in Clinical Trials. Clin Orthop Relat Res 282:86–94PubMed
37.
go back to reference Zichner L, Lindenfeld T (1997) In-vivo wear of the slide combinations ceramics-polyethylene as opposed to metal-polyethylene. Orthopade 26:129–134PubMed Zichner L, Lindenfeld T (1997) In-vivo wear of the slide combinations ceramics-polyethylene as opposed to metal-polyethylene. Orthopade 26:129–134PubMed
Metadata
Title
Fracture of ceramic heads in total hip replacement
Authors
B. Habermann
W. Ewald
M. Rauschmann
L. Zichner
A. A. Kurth
Publication date
01-09-2006
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 7/2006
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-006-0173-y

Other articles of this Issue 7/2006

Archives of Orthopaedic and Trauma Surgery 7/2006 Go to the issue