Published in: World Journal of Urology 4/2015

01-04-2015 | Topic Paper

Impact of pulse duration on Ho:YAG laser lithotripsy: treatment aspects on the single-pulse level

Authors: Ronald Sroka, Thomas Pongratz, Gabriel Scheib, Wael Khoder, Christian G. Stief, Thomas Herrmann, Udo Nagele, Markus J. Bader

Published in: World Journal of Urology | Issue 4/2015

Login to get access

Abstract

Objectives

Holmium–YAG (Ho:YAG) laser lithotripsy is a multi-pulse treatment modality with stochastic effects on the fragmentation. In vitro investigation on the single-pulse-induced effects on fiber, repulsion as well as fragmentation was performed to identify potential impacts of different Ho:YAG laser pulse durations.

Materials and methods

A Ho:YAG laser system (Swiss LaserClast, EMS S.A., Nyon, Switzerland) with selectable long- or short-pulse mode was tested with regard to fiber burn back, the repulsion capacity using an underwater pendulum setup and single-pulse-induced fragmentation capacity using artificial (BEGO) stones. The laser parameters were chosen in accordance with clinical application modes (laser fiber: 365 and 200 µm; output power: 4, 6 and 10 W in different combinations of energy per pulse and repetition rate). Evaluation parameters were reduction in fiber length, pendulum deviation and topology of the crater.

Results

Using the long-pulse mode, the fiber burn back was nearly negligible, while in short-pulse mode, an increased burn back could be observed. The results of the pendulum test showed that the deviation induced by the momentum of short pulses was by factor 1.5–2 higher compared to longer pulses at identical energy per pulse settings. The ablation volumes induced by single pulses either in short-pulse or long-pulse mode did not differ significantly although different crater shapes appeared.

Conclusion

Reduced stone repulsion and reduced laser fiber burn back with longer laser pulses may result in a more convenient handling during clinical application and thus in an improved clinical outcome of laser lithotripsy.
Appendix
Available only for authorised users
Literature
1.
2.
go back to reference Bagley DH (2004) Ureteroscopic surgery: changing times and perspectives. Urol Clin N Am 31(1):1–4CrossRef Bagley DH (2004) Ureteroscopic surgery: changing times and perspectives. Urol Clin N Am 31(1):1–4CrossRef
3.
go back to reference Chow GK et al (2003) Ureteroscopy: effect of technology and technique on clinical practice. J Urol 170(1):99–102CrossRefPubMed Chow GK et al (2003) Ureteroscopy: effect of technology and technique on clinical practice. J Urol 170(1):99–102CrossRefPubMed
4.
go back to reference Kauer PC et al (2005) Present practice and treatment strategies in endourological stone management: results of a survey of the European Society of Uro-technology (ESUT). Eur Urol 48(2):182–188CrossRefPubMed Kauer PC et al (2005) Present practice and treatment strategies in endourological stone management: results of a survey of the European Society of Uro-technology (ESUT). Eur Urol 48(2):182–188CrossRefPubMed
5.
go back to reference Marks AJ, Teichman JM (2007) Lasers in clinical urology: state of the art and new horizons. World J Urol 25(3):227–233CrossRefPubMed Marks AJ, Teichman JM (2007) Lasers in clinical urology: state of the art and new horizons. World J Urol 25(3):227–233CrossRefPubMed
6.
go back to reference Sofer M et al (2002) Holmium:YAG laser lithotripsy for upper urinary tract calculi in 598 patients. J Urol 167(1):31–34CrossRefPubMed Sofer M et al (2002) Holmium:YAG laser lithotripsy for upper urinary tract calculi in 598 patients. J Urol 167(1):31–34CrossRefPubMed
8.
go back to reference Sroka R et al (2012) In vitro investigations of repulsion during laser lithotripsy using a pendulum set-up. Lasers Med Sci 27(3):637–643CrossRefPubMed Sroka R et al (2012) In vitro investigations of repulsion during laser lithotripsy using a pendulum set-up. Lasers Med Sci 27(3):637–643CrossRefPubMed
9.
go back to reference Eisner BH, Pengune W, Stoller ML (2009) Use of an antiretropulsion device to prevent stone retropulsion significantly increases the efficiency of pneumatic lithotripsy: an in vitro study. BJU Int 104(6):858–861CrossRefPubMed Eisner BH, Pengune W, Stoller ML (2009) Use of an antiretropulsion device to prevent stone retropulsion significantly increases the efficiency of pneumatic lithotripsy: an in vitro study. BJU Int 104(6):858–861CrossRefPubMed
10.
go back to reference Mues AC, Teichman JM, Knudsen BE (2009) Quantification of holmium:yttrium aluminum garnet optical tip degradation. J Endourol 23(9):1425–1428CrossRefPubMed Mues AC, Teichman JM, Knudsen BE (2009) Quantification of holmium:yttrium aluminum garnet optical tip degradation. J Endourol 23(9):1425–1428CrossRefPubMed
11.
go back to reference Wezel F et al (2010) Effect of pulse energy, frequency and length on holmium:yttrium-aluminum-garnet laser fragmentation efficiency in non-floating artificial urinary calculi. J Endourol 24(7):1135–1140CrossRefPubMed Wezel F et al (2010) Effect of pulse energy, frequency and length on holmium:yttrium-aluminum-garnet laser fragmentation efficiency in non-floating artificial urinary calculi. J Endourol 24(7):1135–1140CrossRefPubMed
12.
go back to reference Finley DS et al (2005) Effect of holmium:YAG laser pulse width on lithotripsy retropulsion in vitro. J Endourol 19(8):1041–1044CrossRefPubMed Finley DS et al (2005) Effect of holmium:YAG laser pulse width on lithotripsy retropulsion in vitro. J Endourol 19(8):1041–1044CrossRefPubMed
13.
go back to reference Kalra P, Le NB, Bagley D (2007) Effect of pulse width on object movement in vitro using holmium:YAG laser. J Endourol 21(2):228–231CrossRefPubMed Kalra P, Le NB, Bagley D (2007) Effect of pulse width on object movement in vitro using holmium:YAG laser. J Endourol 21(2):228–231CrossRefPubMed
15.
go back to reference Bader MJ et al Impact of pulse duration on Ho:YAG laser lithotripsy: fragmentation and dusting performance. World J Urol. 2014 Nov 4. [Epub ahead of print]. doi:10.1007/s00345-014-1429-8 Bader MJ et al Impact of pulse duration on Ho:YAG laser lithotripsy: fragmentation and dusting performance. World J Urol. 2014 Nov 4. [Epub ahead of print]. doi:10.​1007/​s00345-014-1429-8
16.
go back to reference Vogel A, Venugopalan V (2003) Mechanisms of pulsed laser ablation of biological tissues. Chem Rev 103(2):577–644CrossRefPubMed Vogel A, Venugopalan V (2003) Mechanisms of pulsed laser ablation of biological tissues. Chem Rev 103(2):577–644CrossRefPubMed
Metadata
Title
Impact of pulse duration on Ho:YAG laser lithotripsy: treatment aspects on the single-pulse level
Authors
Ronald Sroka
Thomas Pongratz
Gabriel Scheib
Wael Khoder
Christian G. Stief
Thomas Herrmann
Udo Nagele
Markus J. Bader
Publication date
01-04-2015
Publisher
Springer Berlin Heidelberg
Published in
World Journal of Urology / Issue 4/2015
Print ISSN: 0724-4983
Electronic ISSN: 1433-8726
DOI
https://doi.org/10.1007/s00345-015-1504-9

Other articles of this Issue 4/2015

World Journal of Urology 4/2015 Go to the issue