01-11-2020 | Mammography | Breast

Clip migration after stereotactic vacuum-assisted breast biopsy with the patient in the decubitus position

Authors: Jane Wang, Ning Chien, Hsiao-Tung Lee

Published in: European Radiology | Issue 11/2020

Login to get access

Abstract

Objective

To investigate the factors contributing to clip migration in stereotactic vacuum-assisted biopsy (VAB) of the breast using an upright unit with the patient in the decubitus position.

Methods

This retrospective study included 176 women with microcalcifications on mammograms undergoing stereotactic VAB with clip marking for analysis. The prebiopsy mammograms, stereotactic VAB images, immediate postbiopsy mammograms, and follow-up mammograms were reviewed. All VAB procedures were performed using an add-on upright unit with the patient in the decubitus position. The clip-to-lesion distance on the orthogonal view (craniocaudal view) on immediate postbiopsy mammography was estimated for each biopsy. Two cutoff points of clip-to-lesion distance of > 1 cm or > 2 cm were set for clip migration. The possible factors for clip migration based on clinical and imaging findings were then analyzed by Fisher’s exact test.

Results

When the cutoff for clip migration was set at > 1 cm, thin breast (p = 0.013) and more superficial lesion (lesion closer to the skin along the line perpendicular to the posterior nipple line, p = 0.004) were associated with clip migration. When the cutoff was set at > 2 cm, thin breast (p = 0.019), high specimen number (p = 0.030), and posterior depth (p = 0.021) were associated with clip migration.

Conclusions

Thin breasts, superficial lesion location, posterior lesion depth, and high specimen number were the factors associated with clip migration.

Key Points

We reported clip migration after ST VAB using an upright unit with the patient in the decubitus position.
The occurrence of clinically significant clip migration (> 2 cm) in our study was 8.5%, which was within the range of the other reports with the patient undergoing ST VAB in the prone position.
Thin breasts, high specimen number, and more posterior depth were the factors associated with significant clip migration.
Literature
1.
go back to reference Zuiani C, Mazzarella F, Londero V, Linda A, Puglisi F, Bazzocchi M (2007) Stereotactic vacuum-assisted breast biopsy: results, follow-up and correlation with radiological suspicion. Radiol Med 112:304–317PubMed Zuiani C, Mazzarella F, Londero V, Linda A, Puglisi F, Bazzocchi M (2007) Stereotactic vacuum-assisted breast biopsy: results, follow-up and correlation with radiological suspicion. Radiol Med 112:304–317PubMed
2.
go back to reference Uematsu T, Kasami M, Takahashi K et al (2012) Clip placement after an 11-gauge vacuum-assisted stereotactic breast biopsy: correlation between breast thickness and clip movement. Breast Cancer 19:30–36PubMed Uematsu T, Kasami M, Takahashi K et al (2012) Clip placement after an 11-gauge vacuum-assisted stereotactic breast biopsy: correlation between breast thickness and clip movement. Breast Cancer 19:30–36PubMed
3.
go back to reference Sohn V, Arthurs Z, Herbert G et al (2007) Atypical ductal hyperplasia: improved accuracy with the 11-gauge vacuum-assisted versus the 14-gauge core biopsy needle. Ann Surg Oncol 14:2497–2501PubMed Sohn V, Arthurs Z, Herbert G et al (2007) Atypical ductal hyperplasia: improved accuracy with the 11-gauge vacuum-assisted versus the 14-gauge core biopsy needle. Ann Surg Oncol 14:2497–2501PubMed
4.
go back to reference Rosen EL, Vo TT (2001) Metallic clip deployment during stereotactic breast biopsy: retrospective analysis. Radiology 218:510–516PubMed Rosen EL, Vo TT (2001) Metallic clip deployment during stereotactic breast biopsy: retrospective analysis. Radiology 218:510–516PubMed
5.
go back to reference Rosen EL, Baker JA, Soo MS (2003) Accuracy of a collagen-plug biopsy site marking device deployed after stereotactic core needle breast biopsy. AJR Am J Roentgenol 181:1295–1299PubMed Rosen EL, Baker JA, Soo MS (2003) Accuracy of a collagen-plug biopsy site marking device deployed after stereotactic core needle breast biopsy. AJR Am J Roentgenol 181:1295–1299PubMed
6.
go back to reference Reynolds HE (1999) Marker clip placement following directional, vacuum-assisted breast biopsy. Am Surg 65:59–60PubMed Reynolds HE (1999) Marker clip placement following directional, vacuum-assisted breast biopsy. Am Surg 65:59–60PubMed
7.
go back to reference Pfarl G, Helbich TH, Riedl CC et al (2002) Stereotactic 11-gauge vacuum-assisted breast biopsy: a validation study. AJR Am J Roentgenol 179:1503–1507PubMed Pfarl G, Helbich TH, Riedl CC et al (2002) Stereotactic 11-gauge vacuum-assisted breast biopsy: a validation study. AJR Am J Roentgenol 179:1503–1507PubMed
8.
go back to reference Pandelidis S, Heiland D, Jones D, Stough K, Trapeni J, Suliman Y (2003) Accuracy of 11-gauge vacuum-assisted core biopsy of mammographic breast lesions. Ann Surg Oncol 10:43–47PubMed Pandelidis S, Heiland D, Jones D, Stough K, Trapeni J, Suliman Y (2003) Accuracy of 11-gauge vacuum-assisted core biopsy of mammographic breast lesions. Ann Surg Oncol 10:43–47PubMed
9.
go back to reference Nisbet AP, Borthwick-Clarke A, Scott N (2000) 11-gauge vacuum assisted directional biopsy of breast calcifications, using upright stereotactic guidance. Eur J Radiol 36:144–146PubMed Nisbet AP, Borthwick-Clarke A, Scott N (2000) 11-gauge vacuum assisted directional biopsy of breast calcifications, using upright stereotactic guidance. Eur J Radiol 36:144–146PubMed
10.
go back to reference Lomoschitz FM, Helbich TH, Rudas M et al (2004) Stereotactic 11-gauge vacuum-assisted breast biopsy: influence of number of specimens on diagnostic accuracy. Radiology 232:897–903PubMed Lomoschitz FM, Helbich TH, Rudas M et al (2004) Stereotactic 11-gauge vacuum-assisted breast biopsy: influence of number of specimens on diagnostic accuracy. Radiology 232:897–903PubMed
11.
go back to reference Liberman L, Smolkin JH, Dershaw DD, Morris EA, Abramson AF, Rosen PP (1998) Calcification retrieval at stereotactic, 11-gauge, directional, vacuum-assisted breast biopsy. Radiology 208:251–260PubMed Liberman L, Smolkin JH, Dershaw DD, Morris EA, Abramson AF, Rosen PP (1998) Calcification retrieval at stereotactic, 11-gauge, directional, vacuum-assisted breast biopsy. Radiology 208:251–260PubMed
12.
go back to reference Agacayak F, Ozturk A, Bozdogan A et al (2014) Stereotactic vacuum-assisted core biopsy results for non-palpable breast lesions. Asian Pac J Cancer Prev 15:5171–5174PubMed Agacayak F, Ozturk A, Bozdogan A et al (2014) Stereotactic vacuum-assisted core biopsy results for non-palpable breast lesions. Asian Pac J Cancer Prev 15:5171–5174PubMed
13.
go back to reference Burbank F, Forcier N (1997) Tissue marking clip for stereotactic breast biopsy: initial placement accuracy, long-term stability, and usefulness as a guide for wire localization. Radiology 205:407–415PubMed Burbank F, Forcier N (1997) Tissue marking clip for stereotactic breast biopsy: initial placement accuracy, long-term stability, and usefulness as a guide for wire localization. Radiology 205:407–415PubMed
14.
go back to reference Carr JJ, Hemler PF, Halford PW, Freimanis RI, Choplin RH, Chen MY (2001) Stereotactic localization of breast lesions: how it works and methods to improve accuracy. Radiographics 21:463–473PubMed Carr JJ, Hemler PF, Halford PW, Freimanis RI, Choplin RH, Chen MY (2001) Stereotactic localization of breast lesions: how it works and methods to improve accuracy. Radiographics 21:463–473PubMed
15.
go back to reference Esserman LE, Cura MA, DaCosta D (2004) Recognizing pitfalls in early and late migration of clip markers after imaging-guided directional vacuum-assisted biopsy. Radiographics 24:147–156PubMed Esserman LE, Cura MA, DaCosta D (2004) Recognizing pitfalls in early and late migration of clip markers after imaging-guided directional vacuum-assisted biopsy. Radiographics 24:147–156PubMed
16.
go back to reference Georgian-Smith D, D'Orsi C, Morris E, Clark CF Jr, Liberty E, Lehman CD (2002) Stereotactic biopsy of the breast using an upright unit, a vacuum-suction needle, and a lateral arm-support system. AJR Am J Roentgenol 178:1017–1024PubMed Georgian-Smith D, D'Orsi C, Morris E, Clark CF Jr, Liberty E, Lehman CD (2002) Stereotactic biopsy of the breast using an upright unit, a vacuum-suction needle, and a lateral arm-support system. AJR Am J Roentgenol 178:1017–1024PubMed
17.
go back to reference Kruger BM, Burrowes P, MacGregor JH (2002) Accuracy of marker clip placement after mammotome breast biopsy. Can Assoc Radiol J 53:137–140PubMed Kruger BM, Burrowes P, MacGregor JH (2002) Accuracy of marker clip placement after mammotome breast biopsy. Can Assoc Radiol J 53:137–140PubMed
18.
go back to reference Sim LS, Kei PL (2008) Upright stereotactic vacuum-assisted needle biopsy of suspicious breast microcalcifications. J Med Imaging Radiat Oncol 52:358–364PubMed Sim LS, Kei PL (2008) Upright stereotactic vacuum-assisted needle biopsy of suspicious breast microcalcifications. J Med Imaging Radiat Oncol 52:358–364PubMed
19.
go back to reference Wang J, Chang CJ (2010) Use of a hemostasis introducer sheath to guide clip delivery during stereotactic directional vacuum-assisted breast biopsy when the biopsy system malfunctions. Clin Imaging 34:385–387PubMed Wang J, Chang CJ (2010) Use of a hemostasis introducer sheath to guide clip delivery during stereotactic directional vacuum-assisted breast biopsy when the biopsy system malfunctions. Clin Imaging 34:385–387PubMed
20.
go back to reference Birdwell RL, Jackman RJ (2003) Clip or marker migration 5-10 weeks after stereotactic 11-gauge vacuum-assisted breast biopsy: report of two cases. Radiology 229:541–544PubMed Birdwell RL, Jackman RJ (2003) Clip or marker migration 5-10 weeks after stereotactic 11-gauge vacuum-assisted breast biopsy: report of two cases. Radiology 229:541–544PubMed
21.
go back to reference Burnside ES, Sohlich RE, Sickles EA (2001) Movement of a biopsy-site marker clip after completion of stereotactic directional vacuum-assisted breast biopsy: case report. Radiology 221:504–507PubMed Burnside ES, Sohlich RE, Sickles EA (2001) Movement of a biopsy-site marker clip after completion of stereotactic directional vacuum-assisted breast biopsy: case report. Radiology 221:504–507PubMed
22.
go back to reference Kass R, Kumar G, Klimberg VS et al (2002) Clip migration in stereotactic biopsy. Am J Surg 184:325–331PubMed Kass R, Kumar G, Klimberg VS et al (2002) Clip migration in stereotactic biopsy. Am J Surg 184:325–331PubMed
23.
go back to reference Lehman CD, Shook JE (2003) Position of clip placement after vacuum-assisted breast biopsy: is a unilateral two-view postbiopsy mammogram necessary? Breast J 9:272–276PubMed Lehman CD, Shook JE (2003) Position of clip placement after vacuum-assisted breast biopsy: is a unilateral two-view postbiopsy mammogram necessary? Breast J 9:272–276PubMed
24.
go back to reference Parikh J (2005) Clip migration within 15 days of 11-gauge vacuum-assisted stereotactic breast biopsy. AJR Am J Roentgenol 184:S43–S46PubMed Parikh J (2005) Clip migration within 15 days of 11-gauge vacuum-assisted stereotactic breast biopsy. AJR Am J Roentgenol 184:S43–S46PubMed
25.
go back to reference Bernaerts A, De Schepper A Jr, Van Dam P, Pouillon M (2007) Clip migration after vacuum-assisted stereotactic breast biopsy: a pitfall in preoperative wire localization. JBR-BTR 90:172–175PubMed Bernaerts A, De Schepper A Jr, Van Dam P, Pouillon M (2007) Clip migration after vacuum-assisted stereotactic breast biopsy: a pitfall in preoperative wire localization. JBR-BTR 90:172–175PubMed
26.
go back to reference Chaveron C, Bachelle F, Fauquet I, Rocourt N, Faivre-Pierret M, Ceugnart L (2009) Clip migration after stereotactic macrobiopsy and presurgical localization: technical considerations and tricks. J Radiol 90:31–36PubMed Chaveron C, Bachelle F, Fauquet I, Rocourt N, Faivre-Pierret M, Ceugnart L (2009) Clip migration after stereotactic macrobiopsy and presurgical localization: technical considerations and tricks. J Radiol 90:31–36PubMed
27.
go back to reference Harris AT (2003) Clip migration within 8 days of 11-gauge vacuum-assisted stereotactic breast biopsy: case report. Radiology 228:552–554PubMed Harris AT (2003) Clip migration within 8 days of 11-gauge vacuum-assisted stereotactic breast biopsy: case report. Radiology 228:552–554PubMed
28.
go back to reference Liberman L, Dershaw DD, Morris EA, Abramson AF, Thornton CM, Rosen PP (1997) Clip placement after stereotactic vacuum-assisted breast biopsy. Radiology 205:417–422PubMed Liberman L, Dershaw DD, Morris EA, Abramson AF, Thornton CM, Rosen PP (1997) Clip placement after stereotactic vacuum-assisted breast biopsy. Radiology 205:417–422PubMed
29.
go back to reference Margolin FR, Kaufman L, Denny SR, Jacobs RP, Schrumpf JD (2003) Metallic marker placement after stereotactic core biopsy of breast calcifications: comparison of two clips and deployment techniques. AJR Am J Roentgenol 181:1685–1690PubMed Margolin FR, Kaufman L, Denny SR, Jacobs RP, Schrumpf JD (2003) Metallic marker placement after stereotactic core biopsy of breast calcifications: comparison of two clips and deployment techniques. AJR Am J Roentgenol 181:1685–1690PubMed
30.
go back to reference Parikh JR (2005) Delayed migration of Gel Mark Ultra clip within 15 days of 11-gauge vacuum-assisted stereotactic breast biopsy. AJR Am J Roentgenol 185:203–206PubMed Parikh JR (2005) Delayed migration of Gel Mark Ultra clip within 15 days of 11-gauge vacuum-assisted stereotactic breast biopsy. AJR Am J Roentgenol 185:203–206PubMed
31.
go back to reference Philpotts LE, Lee CH (2002) Clip migration after 11-gauge vacuum-assisted stereotactic biopsy: case report. Radiology 222:794–796PubMed Philpotts LE, Lee CH (2002) Clip migration after 11-gauge vacuum-assisted stereotactic biopsy: case report. Radiology 222:794–796PubMed
32.
go back to reference Jung YJ, Bae YT, Lee JY, Seo HI, Kim JY, Choo KS (2011) Lateral decubitus positioning stereotactic vacuum-assisted breast biopsy with true lateral mammography. J Breast Cancer 14:64–68PubMedPubMedCentral Jung YJ, Bae YT, Lee JY, Seo HI, Kim JY, Choo KS (2011) Lateral decubitus positioning stereotactic vacuum-assisted breast biopsy with true lateral mammography. J Breast Cancer 14:64–68PubMedPubMedCentral
33.
go back to reference Lee KE, Kim HH, Shin HJ, Cha JH (2013) Stereotactic biopsy of the breast using a decubitus table: comparison of histologic underestimation rates between 11- and 8-gauge vacuum-assisted breast biopsy. Springerplus 2:551PubMedPubMedCentral Lee KE, Kim HH, Shin HJ, Cha JH (2013) Stereotactic biopsy of the breast using a decubitus table: comparison of histologic underestimation rates between 11- and 8-gauge vacuum-assisted breast biopsy. Springerplus 2:551PubMedPubMedCentral
34.
go back to reference Lee SH, Jung YJ, Jung HJ et al (2016) Stereotactic vacuum-assisted breast biopsy under lateral decubitus position. Ann Surg Treat Res 90:16–20PubMed Lee SH, Jung YJ, Jung HJ et al (2016) Stereotactic vacuum-assisted breast biopsy under lateral decubitus position. Ann Surg Treat Res 90:16–20PubMed
35.
go back to reference Viala J, Gignier P, Perret B et al (2013) Stereotactic vacuum-assisted biopsies on a digital breast 3D-tomosynthesis system. Breast J 19:4–9PubMed Viala J, Gignier P, Perret B et al (2013) Stereotactic vacuum-assisted biopsies on a digital breast 3D-tomosynthesis system. Breast J 19:4–9PubMed
37.
go back to reference Bahl M, Maunglay M, D'Alessandro HA, Lehman CD (2019) Comparison of upright digital breast tomosynthesis-guided versus prone stereotactic vacuum-assisted breast biopsy. Radiology 290:298–304PubMed Bahl M, Maunglay M, D'Alessandro HA, Lehman CD (2019) Comparison of upright digital breast tomosynthesis-guided versus prone stereotactic vacuum-assisted breast biopsy. Radiology 290:298–304PubMed
39.
go back to reference Shih TC, Chen JH, Liu D et al (2010) Computational simulation of breast compression based on segmented breast and fibroglandular tissues on magnetic resonance images. Phys Med Biol 55:4153–4168PubMedPubMedCentral Shih TC, Chen JH, Liu D et al (2010) Computational simulation of breast compression based on segmented breast and fibroglandular tissues on magnetic resonance images. Phys Med Biol 55:4153–4168PubMedPubMedCentral
40.
go back to reference Jain A, Khalid M, Qureshi MM et al (2017) Stereotactic core needle breast biopsy marker migration: an analysis of factors contributing to immediate marker migration. Eur Radiol 27:4797–4803PubMed Jain A, Khalid M, Qureshi MM et al (2017) Stereotactic core needle breast biopsy marker migration: an analysis of factors contributing to immediate marker migration. Eur Radiol 27:4797–4803PubMed
41.
go back to reference Yen P, Dumas S, Albert A, Gordon P (2018) Post-vacuum-assisted stereotactic core biopsy clip displacement: a comparison between commercially available clips and surgical clip. Can Assoc Radiol J 69:10–15PubMed Yen P, Dumas S, Albert A, Gordon P (2018) Post-vacuum-assisted stereotactic core biopsy clip displacement: a comparison between commercially available clips and surgical clip. Can Assoc Radiol J 69:10–15PubMed
42.
go back to reference Pinkney DM, Mychajlowycz M, Shah BA (2016) A prospective comparative study to evaluate the displacement of four commercially available breast biopsy markers. Br J Radiol 89:20160149PubMedPubMedCentral Pinkney DM, Mychajlowycz M, Shah BA (2016) A prospective comparative study to evaluate the displacement of four commercially available breast biopsy markers. Br J Radiol 89:20160149PubMedPubMedCentral
Metadata
Title
Clip migration after stereotactic vacuum-assisted breast biopsy with the patient in the decubitus position
Authors
Jane Wang
Ning Chien
Hsiao-Tung Lee
Publication date
01-11-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-07015-0

Other articles of this Issue 11/2020

European Radiology 11/2020 Go to the issue