Published in: European Radiology 3/2019

01-03-2019 | Musculoskeletal

Comparison of radiographs, tomosynthesis and CT with metal artifact reduction for the detection of hip prosthetic loosening

Authors: Romain Gillet, Pedro Teixeira, Chloé Bonarelli, Henry Coudane, François Sirveaux, Mathias Louis, Alain Blum

Published in: European Radiology | Issue 3/2019

Login to get access

Abstract

Objectives

To evaluate the diagnostic performance of digital tomosynthesis (DTS) for the diagnosis of hip prosthesis loosening (PL) compared with conventional radiographs and CT with metal artifact reduction (CT-MAR).

Methods

Forty-nine patients with painful hip prosthesis were prospectively included and underwent anteroposterior and lateral radiographs, anteroposterior DTS and CT-MAR of the hip. This study was approved by the local ethics committee, and all patients signed an informed consent form. Images were evaluated independently by two radiologists. Periprosthetic radiolucent lines wider than 2 mm found in two or more Gruen or De Lee and Charnley zones were considered diagnostic of PL. All cases of PL were confirmed surgically. Patients with a stable radiological follow-up for at least 1 year with an alternative cause for the symptoms or with no surgical evidence of PL were considered PL negative.

Results

There were 21 cases of PL, 9 unilateral and 12 bilateral. For both the acetabular and femoral sides, DTS had a specificity for PL detection similar to that of conventional radiographs and CT-MAR (98.5–100%, 96.9%–100% and 96.9–95.4% respectively for both readers) and a sensitivity similar to conventional radiographs (39.9–45.4% versus 33.3–51.5% for both readers) but lower than CT-MAR (84.85% for both readers). The interobserver agreement was 0.84 for CT-MAR, 0.53 for DTS and 0.39 for conventional radiographs.

Conclusion

DTS has a similar diagnostic performance to radiographs for the diagnosis of PL with a better interobserver agreement. The sensitivity however remains lower than that of CT-MAR.

Key Points

• Plain radiograph is still the first imaging step when hip prosthesis loosening is suspected.
• Interobserver agreement is better with digital tomosynthesis than radiographs.
• Sensitivity of CT with state-of-the-art metal artifact reduction is superior to that of digital tomosynthesis.
Literature
1.
go back to reference Nemes S, Gordon M, Rogmark C, Rolfson O (2014) Projections of total hip replacement in Sweden from 2013 to 2030. Acta Orthop 85:238–243CrossRef Nemes S, Gordon M, Rogmark C, Rolfson O (2014) Projections of total hip replacement in Sweden from 2013 to 2030. Acta Orthop 85:238–243CrossRef
2.
go back to reference Labek G, Thaler M, Janda W, Agreiter M, Stöckl B (2011) Revision rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Joint Surg Br 93:293–297 Labek G, Thaler M, Janda W, Agreiter M, Stöckl B (2011) Revision rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Joint Surg Br 93:293–297
3.
go back to reference Delaunay C, Hamadouche M, Girard J, Duhamel A; SoFCOT Group (2013) What are the causes for failures of primary hip arthroplasties in France? Clin Orthop Relat Res 471:3863–3869 Delaunay C, Hamadouche M, Girard J, Duhamel A; SoFCOT Group (2013) What are the causes for failures of primary hip arthroplasties in France? Clin Orthop Relat Res 471:3863–3869
4.
go back to reference Duffy PJ, Masri BA, Garbuz DS, Duncan CP (2005) Evaluation of patients with pain following total hip replacement. J Bone Joint Surg Am 87:2565–2575 Duffy PJ, Masri BA, Garbuz DS, Duncan CP (2005) Evaluation of patients with pain following total hip replacement. J Bone Joint Surg Am 87:2565–2575
5.
go back to reference Patel AR, Sweeney P, Ochenjele G, Wixson R, Stulberg SD, Puri LM (2015) Radiographically silent loosening of the acetabular component in hip arthroplasty. Am J Orthop (Belle Mead NJ) 44:406–410 Patel AR, Sweeney P, Ochenjele G, Wixson R, Stulberg SD, Puri LM (2015) Radiographically silent loosening of the acetabular component in hip arthroplasty. Am J Orthop (Belle Mead NJ) 44:406–410
6.
go back to reference Eklund K, Jonsson K, Lindblom G et al (2004) Are digital images good enough? A comparative study of conventional film-screen vs digital radiographs on printed images of total hip replacement. Eur Radiol 14:865–869CrossRef Eklund K, Jonsson K, Lindblom G et al (2004) Are digital images good enough? A comparative study of conventional film-screen vs digital radiographs on printed images of total hip replacement. Eur Radiol 14:865–869CrossRef
7.
go back to reference Claus AM, Engh CA Jr, Sychterz CJ, Xenos JS, Orishimo KF, Engh CA Sr (2003) Radiographic definition of pelvic osteolysis following total hip arthroplasty. J Bone Joint Surg Am 85:1519–1526 Claus AM, Engh CA Jr, Sychterz CJ, Xenos JS, Orishimo KF, Engh CA Sr (2003) Radiographic definition of pelvic osteolysis following total hip arthroplasty. J Bone Joint Surg Am 85:1519–1526
8.
go back to reference Blum A, Meyer JB, Raymond A et al (2016) CT of hip prosthesis: new techniques and new paradigms. Diagn Interv Imaging 97:725–733 Blum A, Meyer JB, Raymond A et al (2016) CT of hip prosthesis: new techniques and new paradigms. Diagn Interv Imaging 97:725–733
9.
go back to reference Gondim Teixeira PA, Meyer JB, Baumann C et al (2014) Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skeletal Radiol 43:1237–1246 Gondim Teixeira PA, Meyer JB, Baumann C et al (2014) Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skeletal Radiol 43:1237–1246
10.
go back to reference Roth TD, Maertz NA, Parr JA, Buckwalter KA, Choplin RH (2012) CT of the hip prosthesis: appearance of components, fixation, and complications. Radiographics 32:1089–1107 Roth TD, Maertz NA, Parr JA, Buckwalter KA, Choplin RH (2012) CT of the hip prosthesis: appearance of components, fixation, and complications. Radiographics 32:1089–1107
11.
go back to reference Chen Z, Pandit H, Taylor A, Gill H, Murray D, Ostlere S (2011) Metal-on-metal hip resurfacings—a radiological perspective. Eur Radiol 21:485–491 Chen Z, Pandit H, Taylor A, Gill H, Murray D, Ostlere S (2011) Metal-on-metal hip resurfacings—a radiological perspective. Eur Radiol 21:485–491
12.
go back to reference Laukamp KR, Lennartz S, Neuhaus VF et al (2018) CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol. https://doi.org/10.1007/s00330-018-5414-2 Laukamp KR, Lennartz S, Neuhaus VF et al (2018) CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol. https://​doi.​org/​10.​1007/​s00330-018-5414-2
13.
go back to reference Lee YH, Park KK, Song HT, Kim S, Suh JS (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340 Lee YH, Park KK, Song HT, Kim S, Suh JS (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340
14.
go back to reference West AT, Marshall TJ, Bearcroft PW (2009) CT of the musculoskeletal system: what is left is the days of MRI? Eur Radiol 19:152–164CrossRef West AT, Marshall TJ, Bearcroft PW (2009) CT of the musculoskeletal system: what is left is the days of MRI? Eur Radiol 19:152–164CrossRef
15.
go back to reference Agten CA, Sutter R, Dora C, Pfirrmann CW (2017) MR imaging of soft tissue alterations after total hip arthroplasty: comparison of classic surgical approaches. Eur Radiol 27:1312–1321CrossRef Agten CA, Sutter R, Dora C, Pfirrmann CW (2017) MR imaging of soft tissue alterations after total hip arthroplasty: comparison of classic surgical approaches. Eur Radiol 27:1312–1321CrossRef
16.
go back to reference Kretzschmar M, Nardo L, Han MM et al (2015) Metal artefact suppression at 3 T MRI: comparison of MAVRIC-SL with conventional fast spin echo sequences in patients with hip joint arthroplasty. Eur Radiol 25:2403–2411CrossRef Kretzschmar M, Nardo L, Han MM et al (2015) Metal artefact suppression at 3 T MRI: comparison of MAVRIC-SL with conventional fast spin echo sequences in patients with hip joint arthroplasty. Eur Radiol 25:2403–2411CrossRef
17.
go back to reference Fritz J, Lurie B, Miller TT, Potter HG (2014) MR imaging of hip arthroplasty implants. Radiographics 34:E106–E132CrossRef Fritz J, Lurie B, Miller TT, Potter HG (2014) MR imaging of hip arthroplasty implants. Radiographics 34:E106–E132CrossRef
18.
go back to reference Fritz J, Lurie B, Miller TT (2013) Imaging of hip arthroplasty. Semin Musculoskelet Radiol 17:316–327CrossRef Fritz J, Lurie B, Miller TT (2013) Imaging of hip arthroplasty. Semin Musculoskelet Radiol 17:316–327CrossRef
19.
go back to reference Blum A, Noël A, Regent D, Villani N, Gillet R, Gondim Teixeira P (2018) Tomosynthesis in musculoskeletal pathology. Diagn Interv Imaging 99:423–441 Blum A, Noël A, Regent D, Villani N, Gillet R, Gondim Teixeira P (2018) Tomosynthesis in musculoskeletal pathology. Diagn Interv Imaging 99:423–441
20.
go back to reference Machida H, Yuhara T, Tamura M et al (2016) Whole-body clinical applications of digital tomosynthesis. Radiographics 36:735–750CrossRef Machida H, Yuhara T, Tamura M et al (2016) Whole-body clinical applications of digital tomosynthesis. Radiographics 36:735–750CrossRef
21.
go back to reference Geijer M, Gunnlaugsson E, Götestrand S, Weber L, Geijer H (2017) Tomosynthesis of the thoracic spine: added value in diagnosing vertebral fractures in the elderly. Eur Radiol 27:491–497 Geijer M, Gunnlaugsson E, Götestrand S, Weber L, Geijer H (2017) Tomosynthesis of the thoracic spine: added value in diagnosing vertebral fractures in the elderly. Eur Radiol 27:491–497
22.
go back to reference Gomi T, Hirano H (2008) Clinical potential of digital linear tomosynthesis imaging of total joint arthroplasty. J Digit Imaging 21:312–322CrossRef Gomi T, Hirano H (2008) Clinical potential of digital linear tomosynthesis imaging of total joint arthroplasty. J Digit Imaging 21:312–322CrossRef
23.
go back to reference Gomi T, Hirano H, Umeda T (2009) Evaluation of the X-ray digital linear tomosynthesis reconstruction processing method for metal artifact reduction. Comput Med Imaging Graph 33:267–274CrossRef Gomi T, Hirano H, Umeda T (2009) Evaluation of the X-ray digital linear tomosynthesis reconstruction processing method for metal artifact reduction. Comput Med Imaging Graph 33:267–274CrossRef
24.
go back to reference Gomi T, Sakai R, Goto M, Watanabe Y, Takeda T, Umeda T (2016) Comparison of reconstruction algorithms for decreasing the exposure dose during digital tomosynthesis for arthroplasty: a phantom study. J Digit Imaging 29:488–495 Gomi T, Sakai R, Goto M, Watanabe Y, Takeda T, Umeda T (2016) Comparison of reconstruction algorithms for decreasing the exposure dose during digital tomosynthesis for arthroplasty: a phantom study. J Digit Imaging 29:488–495
26.
go back to reference Göthlin JH, Geijer M (2013) The utility of digital linear tomosynthesis imaging of total hip joint arthroplasty with suspicion of loosening: a prospective study in 40 patients. Biomed Res Int 2013:1–6CrossRef Göthlin JH, Geijer M (2013) The utility of digital linear tomosynthesis imaging of total hip joint arthroplasty with suspicion of loosening: a prospective study in 40 patients. Biomed Res Int 2013:1–6CrossRef
27.
go back to reference Tang H, Yang D, Guo S et al (2016) Digital tomosynthesis with metal artifact reduction for assessing cementless hip arthroplasty: a diagnostic cohort study of 48 patients. Skeletal Radiol 45:1523–1532CrossRef Tang H, Yang D, Guo S et al (2016) Digital tomosynthesis with metal artifact reduction for assessing cementless hip arthroplasty: a diagnostic cohort study of 48 patients. Skeletal Radiol 45:1523–1532CrossRef
28.
go back to reference Chang YB, Xu D, Zamyatin AA (2012) Metal artifact reduction algorithm for single energy and dual energy CT scans. In: 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), Anaheim, CA, 2012, pp. 3426–3429 Chang YB, Xu D, Zamyatin AA (2012) Metal artifact reduction algorithm for single energy and dual energy CT scans. In: 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), Anaheim, CA, 2012, pp. 3426–3429
29.
go back to reference Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for ct using dose–length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection Publication 103 or dual-energy scanning. AJR Am J Roentgenol 194:881–889CrossRef Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for ct using dose–length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection Publication 103 or dual-energy scanning. AJR Am J Roentgenol 194:881–889CrossRef
30.
go back to reference DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 20–32 DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 20–32
31.
go back to reference Gruen TA, McNeice G, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 141:17–2723 Gruen TA, McNeice G, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 141:17–2723
32.
go back to reference Abrahams JM, Kim YS, Callary SA et al (2017) The diagnostic performance of radiographic criteria to detect aseptic acetabular component loosening after revision total hip arthroplasty. Bone Joint J 99-B:458–464CrossRef Abrahams JM, Kim YS, Callary SA et al (2017) The diagnostic performance of radiographic criteria to detect aseptic acetabular component loosening after revision total hip arthroplasty. Bone Joint J 99-B:458–464CrossRef
33.
go back to reference Wu T, Moore RH, Rafferty EA, Kopans DB (2004) A comparison of reconstruction algorithms for breast tomosynthesis. Med Phys 31:2636–2647CrossRef Wu T, Moore RH, Rafferty EA, Kopans DB (2004) A comparison of reconstruction algorithms for breast tomosynthesis. Med Phys 31:2636–2647CrossRef
34.
go back to reference Mermuys K, De Geeter F, Bacher K et al (2010) Digital Tomosynthesis in the detection of urolithiasis: diagnostic performance and dosimetry compared with digital radiography with MDCT as the reference standard. AJR Am J Roentgenol 195:161–167CrossRef Mermuys K, De Geeter F, Bacher K et al (2010) Digital Tomosynthesis in the detection of urolithiasis: diagnostic performance and dosimetry compared with digital radiography with MDCT as the reference standard. AJR Am J Roentgenol 195:161–167CrossRef
35.
go back to reference Ha AS, Lee AY, Hippe DS, Chou SH, Chew FS (2015) Digital tomosynthesis to evaluate fracture healing: prospective comparison with radiography and CT. AJR Am J Roentgenol 205:136–141 Ha AS, Lee AY, Hippe DS, Chou SH, Chew FS (2015) Digital tomosynthesis to evaluate fracture healing: prospective comparison with radiography and CT. AJR Am J Roentgenol 205:136–141
36.
go back to reference Simoni P, Gérard L, Kaiser MJ et al (2015) Use of Tomosynthesis for detection of bone erosions of the foot in patients with established rheumatoid arthritis: comparison with radiography and CT. AJR Am J Roentgenol 205:364–370CrossRef Simoni P, Gérard L, Kaiser MJ et al (2015) Use of Tomosynthesis for detection of bone erosions of the foot in patients with established rheumatoid arthritis: comparison with radiography and CT. AJR Am J Roentgenol 205:364–370CrossRef
37.
go back to reference Canella C, Philippe P, Pansini V, Salleron J, Flipo RM, Cotten A (2011) Use of tomosynthesis for erosion evaluation in rheumatoid arthritic hands and wrists. Radiology 258:199–205 Canella C, Philippe P, Pansini V, Salleron J, Flipo RM, Cotten A (2011) Use of tomosynthesis for erosion evaluation in rheumatoid arthritic hands and wrists. Radiology 258:199–205
38.
go back to reference Ottenin MA, Jacquot A, Grospretre O et al (2012) Evaluation of the diagnostic performance of tomosynthesis in fractures of the wrist. AJR Am J Roentgenol 198:180–186 Ottenin MA, Jacquot A, Grospretre O et al (2012) Evaluation of the diagnostic performance of tomosynthesis in fractures of the wrist. AJR Am J Roentgenol 198:180–186
39.
go back to reference Becker AS, Martini K, Higashigaito K, Guggenberger R, Andreisek G, Frauenfelder T (2017) Dose reduction in tomosynthesis of the wrist. AJR Am J Roentgenol 208:159–164 Becker AS, Martini K, Higashigaito K, Guggenberger R, Andreisek G, Frauenfelder T (2017) Dose reduction in tomosynthesis of the wrist. AJR Am J Roentgenol 208:159–164
40.
go back to reference Machida H, Yuhara T, Tamura M et al (2012) Radiation dose of digital tomosynthesis for sinonasal examination: comparison with multi-detector CT. Eur J Radiol 81:1140–1145CrossRef Machida H, Yuhara T, Tamura M et al (2012) Radiation dose of digital tomosynthesis for sinonasal examination: comparison with multi-detector CT. Eur J Radiol 81:1140–1145CrossRef
41.
go back to reference Lacout A, Thariat J, Hajjam ME, Marcy PY (2013) Insight into osteo-articular digital tomosynthesis: a pictorial essay: osteo-articular digital tomosynthesis. J Med Imaging Radiat Oncol 57:45–49CrossRef Lacout A, Thariat J, Hajjam ME, Marcy PY (2013) Insight into osteo-articular digital tomosynthesis: a pictorial essay: osteo-articular digital tomosynthesis. J Med Imaging Radiat Oncol 57:45–49CrossRef
42.
go back to reference Lombard C, Gervaise A, Villani N et al (2018) The impact of dose reduction in quantitative kinematic CT of ankle joints using a full model-based iterative reconstruction algorithm: a cadaveric study. AJR Am J Roentgenol 210:396–403CrossRef Lombard C, Gervaise A, Villani N et al (2018) The impact of dose reduction in quantitative kinematic CT of ankle joints using a full model-based iterative reconstruction algorithm: a cadaveric study. AJR Am J Roentgenol 210:396–403CrossRef
43.
go back to reference Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301CrossRef Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301CrossRef
44.
go back to reference Gervaise A, Teixeira P, Villani N, Lecocq S, Louis M, Blum A (2013) CT dose optimisation and reduction in osteoarticular disease. Diagn Interv Imaging 94:371–388 Gervaise A, Teixeira P, Villani N, Lecocq S, Louis M, Blum A (2013) CT dose optimisation and reduction in osteoarticular disease. Diagn Interv Imaging 94:371–388
45.
go back to reference Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A (2014) Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging 95:47–53 Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A (2014) Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging 95:47–53
46.
go back to reference Kim YS, Abrahams JM, Callary SA et al (2017) Proximal translation of > 1 mm within the first two years of revision total hip arthroplasty correctly predicts whether or not an acetabular component is loose in 80% of cases: a case-control study with confirmed intra-operative outcomes. Bone Joint J 99-B:465–474CrossRef Kim YS, Abrahams JM, Callary SA et al (2017) Proximal translation of > 1 mm within the first two years of revision total hip arthroplasty correctly predicts whether or not an acetabular component is loose in 80% of cases: a case-control study with confirmed intra-operative outcomes. Bone Joint J 99-B:465–474CrossRef
47.
go back to reference Hayashi D, Xu L, Gusenburg J et al (2014) Reliability of semiquantitative assessment of osteophytes and subchondral cysts on tomosynthesis images by radiologists with different levels of expertise. Diagn Interv Radiol 20:353–359CrossRef Hayashi D, Xu L, Gusenburg J et al (2014) Reliability of semiquantitative assessment of osteophytes and subchondral cysts on tomosynthesis images by radiologists with different levels of expertise. Diagn Interv Radiol 20:353–359CrossRef
48.
go back to reference Tucker L, Gilbert FJ, Astley SM et al (2017) Does reader performance with digital breast tomosynthesis vary according to experience with two-dimensional mammography? Radiology 283:371–380CrossRef Tucker L, Gilbert FJ, Astley SM et al (2017) Does reader performance with digital breast tomosynthesis vary according to experience with two-dimensional mammography? Radiology 283:371–380CrossRef
Metadata
Title
Comparison of radiographs, tomosynthesis and CT with metal artifact reduction for the detection of hip prosthetic loosening
Authors
Romain Gillet
Pedro Teixeira
Chloé Bonarelli
Henry Coudane
François Sirveaux
Mathias Louis
Alain Blum
Publication date
01-03-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 3/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5717-3

Other articles of this Issue 3/2019

European Radiology 3/2019 Go to the issue