Published in: European Radiology 3/2019

Open Access 01-03-2019 | Breast

A randomized controlled trial of digital breast tomosynthesis versus digital mammography in population-based screening in Bergen: interim analysis of performance indicators from the To-Be trial

Authors: Hildegunn S. Aase, Åsne S Holen, Kristin Pedersen, Nehmat Houssami, Ingfrid S. Haldorsen, Sofie Sebuødegård, Berit Hanestad, Solveig Hofvind

Published in: European Radiology | Issue 3/2019

Login to get access

Abstract

Objectives

To describe a randomized controlled trial (RCT) of digital breast tomosynthesis including synthesized two-dimensional mammograms (DBT) versus digital mammography (DM) in a population-based screening program for breast cancer and to compare selected secondary screening outcomes for the two techniques.

Methods

This RCT, performed in Bergen as part of BreastScreen Norway, was approved by the Regional Committees for Medical Health Research Ethics. All screening attendees in Bergen were invited to participate, of which 89% (14,274/15,976) concented during the first year, and were randomized to DBT (n = 7155) or DM (n = 7119). Secondary screening outcomes were stratified by mammographic density and compared using two-sample t-tests, chi-square tests, ANOVA, negative binomial regression and tests of proportions (z tests).

Results

Mean reading time was 1 min 11 s for DBT and 41 s for DM (p < 0.01). Mean time spent at consensus was 3 min 12 s for DBT and 2 min 12 s for DM (p < 0.01), while the rate of cases discussed at consensus was 6.4% and 7.4%, respectively for DBT and DM (p = 0.03). The recall rate was 3.0% for DBT and 3.6% for DM (p = 0.03). For women with non-dense breasts, recall rate was 2.2% for DBT versus 3.4% for DM (p = 0.04). The rate did not differ for women with dense breasts (3.6% for both). Mean glandular dose per examination was 2.96 mGy for DBT and 2.95 mGy for DM (p = 0.433).

Conclusions

Interim analysis of a screening RCT showed that DBT took longer to read than DM, but had significantly lower recall rate than DM. We found no differences in radiation dose between the two techniques.

Key Points

• In this RCT, DBT was associated with longer interpretation time than DM
• Recall rates were lower for DBT than for DM
• Mean glandular radiation dose did not differ between DBT and DM
Appendix
Available only for authorised users
Literature
1.
go back to reference Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589CrossRefPubMed Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589CrossRefPubMed
2.
go back to reference Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56CrossRefPubMed Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56CrossRefPubMed
3.
go back to reference Bernardi D, Macaskill P, Pellegrini M et al (2016) Breast cancer screening with tomosynthesis (3D mammography) with acquired or synthetic 2D mammography compared with 2D mammography alone (STORM-2): a population-based prospective study. Lancet Oncol 17:1105–1113CrossRefPubMed Bernardi D, Macaskill P, Pellegrini M et al (2016) Breast cancer screening with tomosynthesis (3D mammography) with acquired or synthetic 2D mammography compared with 2D mammography alone (STORM-2): a population-based prospective study. Lancet Oncol 17:1105–1113CrossRefPubMed
4.
go back to reference Lång K, Andersson I, Rosso A, Tingberg A, Timberg P, Zackrisson S (2016) Performance of one-view breast tomosynthesis as a stand-alone breast cancer screening modality: results from the Malmo Breast Tomosynthesis Screening Trial, a population-based study. Eur Radiol 26:184–190CrossRefPubMed Lång K, Andersson I, Rosso A, Tingberg A, Timberg P, Zackrisson S (2016) Performance of one-view breast tomosynthesis as a stand-alone breast cancer screening modality: results from the Malmo Breast Tomosynthesis Screening Trial, a population-based study. Eur Radiol 26:184–190CrossRefPubMed
5.
go back to reference Skaane P, Sebuødegård S, Bandos AI et al (2018) Performance of breast cancer screening using digital breast tomosynthesis: Results from the prospective population-based Oslo Tomosynthesis Screening Trial. Breast Cancer Res Treat 169:489–496CrossRefPubMed Skaane P, Sebuødegård S, Bandos AI et al (2018) Performance of breast cancer screening using digital breast tomosynthesis: Results from the prospective population-based Oslo Tomosynthesis Screening Trial. Breast Cancer Res Treat 169:489–496CrossRefPubMed
6.
go back to reference Houssami N, Bernardi D, Caumo F et al (2018) Interval breast cancers in the 'screening with tomosynthesis or standard mammography' (STORM) population-based trial. Breast 38:150–153CrossRefPubMed Houssami N, Bernardi D, Caumo F et al (2018) Interval breast cancers in the 'screening with tomosynthesis or standard mammography' (STORM) population-based trial. Breast 38:150–153CrossRefPubMed
7.
go back to reference Bahl M, Gaffney S, McCarthy AM, Lowry KP, Dang PA, Lehman CD (2018) Breast cancer characteristics associated with 2D digital mammography versus digital breast tomosynthesis for screening-detected and interval cancers. Radiology 287:49–57CrossRefPubMed Bahl M, Gaffney S, McCarthy AM, Lowry KP, Dang PA, Lehman CD (2018) Breast cancer characteristics associated with 2D digital mammography versus digital breast tomosynthesis for screening-detected and interval cancers. Radiology 287:49–57CrossRefPubMed
8.
go back to reference Hofvind S, Hovda T, Holen ÅS et al (2018) Digital breast tomosynthesis and synthetic 2D mammography versus digital mammography: evaluation in a population-based screening program. Radiology 287:787–794CrossRefPubMed Hofvind S, Hovda T, Holen ÅS et al (2018) Digital breast tomosynthesis and synthetic 2D mammography versus digital mammography: evaluation in a population-based screening program. Radiology 287:787–794CrossRefPubMed
9.
go back to reference Caumo F, Zorzi M, Brunelli S et al (2018) Digital breast tomosynthesis with synthesized two-dimensional images versus full-field digital mammography for population screening: outcomes from the Verona screening program. Radiology 287:37–46CrossRefPubMed Caumo F, Zorzi M, Brunelli S et al (2018) Digital breast tomosynthesis with synthesized two-dimensional images versus full-field digital mammography for population screening: outcomes from the Verona screening program. Radiology 287:37–46CrossRefPubMed
10.
go back to reference Gennaro G, Bernardi D, Houssami N (2018) Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis. Eur Radiol 28:573–581CrossRefPubMed Gennaro G, Bernardi D, Houssami N (2018) Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis. Eur Radiol 28:573–581CrossRefPubMed
11.
go back to reference Svahn TM, Houssami N, Sechopoulos I, Mattsson S (2015) Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography. Breast 24:93–99CrossRefPubMed Svahn TM, Houssami N, Sechopoulos I, Mattsson S (2015) Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography. Breast 24:93–99CrossRefPubMed
12.
go back to reference Østerås BH, Skaane P, Gullien R, Martinsen ACT (2018) Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality. Phys Med Biol 63:035006CrossRefPubMed Østerås BH, Skaane P, Gullien R, Martinsen ACT (2018) Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality. Phys Med Biol 63:035006CrossRefPubMed
13.
go back to reference Skaane P, Bandos AI, Eben EB et al (2014) Two-view digital breast tomosynthesis screening with synthetically reconstructed projection images: comparison with digital breast tomosynthesis with full-field digital mammographic images. Radiology 271:655–663CrossRefPubMed Skaane P, Bandos AI, Eben EB et al (2014) Two-view digital breast tomosynthesis screening with synthetically reconstructed projection images: comparison with digital breast tomosynthesis with full-field digital mammographic images. Radiology 271:655–663CrossRefPubMed
14.
go back to reference Rafferty EA, Durand MA, Conant EF et al (2013) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 15:1784–1786 Rafferty EA, Durand MA, Conant EF et al (2013) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 15:1784–1786
15.
go back to reference Friedewald SM, Rafferty EA, Rose SL et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311:2499–2507CrossRefPubMed Friedewald SM, Rafferty EA, Rose SL et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311:2499–2507CrossRefPubMed
16.
go back to reference Gilbert FJ, Tucker L, Young KC (2016) Digital breast tomosynthesis (DBT): a review of the evidence for use as a screening tool. Clin Radiol 71:141–150CrossRefPubMed Gilbert FJ, Tucker L, Young KC (2016) Digital breast tomosynthesis (DBT): a review of the evidence for use as a screening tool. Clin Radiol 71:141–150CrossRefPubMed
18.
go back to reference Aitken Z, McCormack VA, Highnam RP et al (2010) Screen-film mammographic density and breast cancer risk: a comparison of the volumetric standard mammogram form and the interactive threshold measurement methods. Cancer Epidemiol Biomarkers Prev 19:418–428CrossRefPubMedPubMedCentral Aitken Z, McCormack VA, Highnam RP et al (2010) Screen-film mammographic density and breast cancer risk: a comparison of the volumetric standard mammogram form and the interactive threshold measurement methods. Cancer Epidemiol Biomarkers Prev 19:418–428CrossRefPubMedPubMedCentral
19.
go back to reference Highnam R, Brady M, Yaffe M, Karssemeijer N, Harvey J (2010) Robust breast composition measurement - Volpara™. In: Martí JOA, Freixenet J, Martí R (eds) Lecture notes in computer science: 10th international workshop on digital mammography. Springer-Verlag, Berlin Heidelberg, pp 342–349 Highnam R, Brady M, Yaffe M, Karssemeijer N, Harvey J (2010) Robust breast composition measurement - Volpara™. In: Martí JOA, Freixenet J, Martí R (eds) Lecture notes in computer science: 10th international workshop on digital mammography. Springer-Verlag, Berlin Heidelberg, pp 342–349
20.
go back to reference Singh T, Sharma M, Singla V, Khandelwal N (2016) Breast density estimation with fully automated volumetric method: comparison to radiologists' assessment by BI-RADS categories. Acad Radiol 23:78–83CrossRefPubMed Singh T, Sharma M, Singla V, Khandelwal N (2016) Breast density estimation with fully automated volumetric method: comparison to radiologists' assessment by BI-RADS categories. Acad Radiol 23:78–83CrossRefPubMed
21.
go back to reference Sartor H, Lång K, Rosso A, Borgquist S, Zackrisson S, Timberg P (2016) Measuring mammographic density: comparing a fully automated volumetric assessment versus European radiologists' qualitative classification. Eur Radiol 26:4354–4360CrossRefPubMedPubMedCentral Sartor H, Lång K, Rosso A, Borgquist S, Zackrisson S, Timberg P (2016) Measuring mammographic density: comparing a fully automated volumetric assessment versus European radiologists' qualitative classification. Eur Radiol 26:4354–4360CrossRefPubMedPubMedCentral
22.
go back to reference van der Waal D, den Heeten GJ, Pijnappel RM et al (2015) Comparing visually assessed BI-RADS breast density and automated volumetric breast density software: a cross-sectional study in a breast cancer screening setting. PLoS One 10:e0136667CrossRefPubMedPubMedCentral van der Waal D, den Heeten GJ, Pijnappel RM et al (2015) Comparing visually assessed BI-RADS breast density and automated volumetric breast density software: a cross-sectional study in a breast cancer screening setting. PLoS One 10:e0136667CrossRefPubMedPubMedCentral
23.
go back to reference Aujero MP, Gavenonis SC, Benjamin R, Zhang Z, Holt JS (2017) Clinical performance of synthesized two-dimensional mammography combined with tomosynthesis in a large screening population. Radiology 283:70–76CrossRefPubMed Aujero MP, Gavenonis SC, Benjamin R, Zhang Z, Holt JS (2017) Clinical performance of synthesized two-dimensional mammography combined with tomosynthesis in a large screening population. Radiology 283:70–76CrossRefPubMed
24.
go back to reference Freer PE, Riegert J, Eisenmenger L et al (2017) Clinical implementation of synthesized mammography with digital breast tomosynthesis in a routine clinical practice. Breast Cancer Res Treat 166:501–509CrossRefPubMed Freer PE, Riegert J, Eisenmenger L et al (2017) Clinical implementation of synthesized mammography with digital breast tomosynthesis in a routine clinical practice. Breast Cancer Res Treat 166:501–509CrossRefPubMed
26.
go back to reference Houssami N, Hunter K, Zackrisson S (2017) Overview of tomosynthesis (3D mammography) for breast cancer screening. Breast Cancer Manag 6:179–186 Houssami N, Hunter K, Zackrisson S (2017) Overview of tomosynthesis (3D mammography) for breast cancer screening. Breast Cancer Manag 6:179–186
27.
go back to reference Yankaskas BC, Schell MJ, Miglioretti DL (2004) Recall and detection rates in screening mammography. Cancer 101:2710–2711 author reply 1-2CrossRefPubMed Yankaskas BC, Schell MJ, Miglioretti DL (2004) Recall and detection rates in screening mammography. Cancer 101:2710–2711 author reply 1-2CrossRefPubMed
28.
go back to reference Schell MJ, Yankaskas BC, Ballard-Barbash R et al (2007) Evidence-based target recall rates for screening mammography. Radiology 243:681–689CrossRefPubMed Schell MJ, Yankaskas BC, Ballard-Barbash R et al (2007) Evidence-based target recall rates for screening mammography. Radiology 243:681–689CrossRefPubMed
29.
go back to reference Dang PA, Freer PE, Humphrey KL, Halpern EF, Rafferty EA (2014) Addition of tomosynthesis to conventional digital mammography: effect on image interpretation time of screening examinations. Radiology 270:49–56CrossRefPubMed Dang PA, Freer PE, Humphrey KL, Halpern EF, Rafferty EA (2014) Addition of tomosynthesis to conventional digital mammography: effect on image interpretation time of screening examinations. Radiology 270:49–56CrossRefPubMed
30.
go back to reference Tucker L, Gilbert FJ, Astley SM et al (2017) Does reader performance with digital breast tomosynthesis vary according to experience with two-dimensional mammography? Radiology 283:371–380CrossRefPubMed Tucker L, Gilbert FJ, Astley SM et al (2017) Does reader performance with digital breast tomosynthesis vary according to experience with two-dimensional mammography? Radiology 283:371–380CrossRefPubMed
31.
go back to reference Zhang J, Grimm LJ, Lo JY et al (2015) Does breast imaging experience during residency translate into improved initial performance in digital breast tomosynthesis? J Am Coll Radiol 12:728–732CrossRefPubMed Zhang J, Grimm LJ, Lo JY et al (2015) Does breast imaging experience during residency translate into improved initial performance in digital breast tomosynthesis? J Am Coll Radiol 12:728–732CrossRefPubMed
Metadata
Title
A randomized controlled trial of digital breast tomosynthesis versus digital mammography in population-based screening in Bergen: interim analysis of performance indicators from the To-Be trial
Authors
Hildegunn S. Aase
Åsne S Holen
Kristin Pedersen
Nehmat Houssami
Ingfrid S. Haldorsen
Sofie Sebuødegård
Berit Hanestad
Solveig Hofvind
Publication date
01-03-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 3/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5690-x

Other articles of this Issue 3/2019

European Radiology 3/2019 Go to the issue