Abstract

Objectives

To compare contrast-to-noise ratios (CNRs) and iodine discrimination thresholds on iodine maps derived from dual energy CT (DECT) and subtraction CT (SCT).

Methods

A contrast-detail phantom experiment was performed with 2 to 15 mm diameter tubes containing water or iodinated contrast concentrations ranging from 0.5 mg/mL to 20 mg/mL. DECT scans were acquired at 100 kVp and at 140 kVp+Sn filtration. SCT scans were acquired at 100 kVp. Iodine maps were created by material decomposition (DECT) or by subtraction of water scans from iodine scans (SCT). Matched exposure levels varied from 8 to 15 mGy. Iodine discrimination thresholds (Cr) and response times were determined by eight observers.

Results

The adjusted mean CNR was 1.9 times higher for SCT than for DECT. Exposure level had no effect on CNR. All observers discriminated all details ≥10 mm at 12 and 15 mGy. For sub-centimetre details, the lowest calculated Cr was ≤ 0.50 mg/mL for SCT and 0.64 mg/mL for DECT. The smallest detail was discriminated at ≥4.4 mg/mL with SCT and at ≥7.4 mg/mL with DECT. Response times were lower for SCT than DECT.

Conclusions

SCT results in higher CNR and reduced iodine discrimination thresholds compared to DECT for sub-centimetre details.

Key Points

• Subtraction CT iodine maps exhibit higher CNR than dual-energy iodine maps
• Lower iodine concentrations can be discriminated for sub-cm details with SCT
• Response times are lower using SCT compared to dual-energy CT
Appendix
Available only for authorised users
Literature
8.
go back to reference Mileto A, Marin D, Alfaro-cordoba M et al (2014) Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study. Radiology 273:813–820CrossRef Mileto A, Marin D, Alfaro-cordoba M et al (2014) Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study. Radiology 273:813–820CrossRef
18.
go back to reference Mohr B, Brink M, Oostveen LJ, et al (2016) Lung iodine mapping by subtraction with image registration allowing for tissue sliding. In: Styner MA, Angelini ED (eds) Proc. SPIE, Med. Imaging, Image Process. p 978442 Mohr B, Brink M, Oostveen LJ, et al (2016) Lung iodine mapping by subtraction with image registration allowing for tissue sliding. In: Styner MA, Angelini ED (eds) Proc. SPIE, Med. Imaging, Image Process. p 978442
23.
go back to reference Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leeuwen MS (2007) Circle of Willis at CT angiography: dose reduction and image quality—reduction tube voltage and increasing tube current settings. Radiology 242(3):832–839CrossRef Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leeuwen MS (2007) Circle of Willis at CT angiography: dose reduction and image quality—reduction tube voltage and increasing tube current settings. Radiology 242(3):832–839CrossRef
25.
go back to reference Yang CL, O’Neill TR, Kramer GA (2002) Examining item difficulty and response time on perceptual ability test items. J Appl Meas 3:282–299PubMed Yang CL, O’Neill TR, Kramer GA (2002) Examining item difficulty and response time on perceptual ability test items. J Appl Meas 3:282–299PubMed
29.
go back to reference Xu Z, Panjwani SA, Lee CP, et al (2016) Evaluation of body-wise and organ-wise registrations for abdominal organs. Proc SPIE Int Soc Opt Eng. p 97841O Xu Z, Panjwani SA, Lee CP, et al (2016) Evaluation of body-wise and organ-wise registrations for abdominal organs. Proc SPIE Int Soc Opt Eng. p 97841O
31.
go back to reference Hussain FA, Mail N, Shamy AM et al (2016) A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction. J Appl Clin Med Phys 17:5903CrossRef Hussain FA, Mail N, Shamy AM et al (2016) A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction. J Appl Clin Med Phys 17:5903CrossRef
Metadata
Title
Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details?
Authors
Evelinda Baerends
Luuk J. Oostveen
Casper T. Smit
Marco Das
Ioannis Sechopoulos
Monique Brink
Frank de Lange
Mathias Prokop
Publication date
01-12-2018
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 12/2018
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5496-x

Other articles of this Issue 12/2018

European Radiology 12/2018 Go to the issue