01-07-2018 | Computed Tomography

Dual-energy CT: a phantom comparison of different platforms for abdominal imaging

Authors: Thorsten Sellerer, Peter B. Noël, Manuel Patino, Anushri Parakh, Sebastian Ehn, Sascha Zeiter, Jasmin A. Holz, Johannes Hammel, Alexander A. Fingerle, Franz Pfeiffer, David Maintz, Ernst J. Rummeny, Daniela Muenzel, Dushyant V. Sahani

Published in: European Radiology | Issue 7/2018

Login to get access

Abstract

Objectives

Evaluation of imaging performance across dual-energy CT (DECT) platforms, including dual-layer CT (DLCT), rapid-kVp-switching CT (KVSCT) and dual-source CT (DSCT).

Methods

A semi-anthropomorphic abdomen phantom was imaged on these DECT systems. Scans were repeated three times for CTDIvol levels of 10 mGy, 20 mGy, 30 mGy and different fat-simulating extension rings. Over the available range of virtual-monoenergetic images (VMI), noise as well as quantitative accuracy of hounsfield units (HU) and iodine concentrations were evaluated.

Results

For all VMI levels, HU values could be determined with high accuracy compared to theoretical values. For KVSCT and DSCT, a noise increase was observed towards lower VMI levels. A patient-size dependent increase in the uncertainty of quantitative iodine concentrations is observed for all platforms. For a medium patient size the iodine concentration root-mean-square deviation at 20 mGy is 0.17 mg/ml (DLCT), 0.30 mg/ml (KVSCT) and 0.77mg/ml (DSCT).

Conclusion

Noticeable performance differences are observed between investigated DECT systems. Iodine concentrations and VMI HUs are accurately determined across all DECT systems. KVSCT and DLCT deliver slightly more accurate iodine concentration values than DSCT for investigated scenarios. In DLCT, low-noise and high-image contrast at low VMI levels may help to increase diagnostic information in abdominal CT.

Key Points

• Current dual-energy CT platforms provide accurate, reliable quantitative information.
• Dual-energy CT cross-platform evaluation revealed noticeable performance differences between different systems.
• Dual-layer CT offers constant noise levels over the complete energy range.
Literature
1.
go back to reference Conway JRW, Carragher NO, Timpson P (2014) Developments in preclinical cancer imaging : innovating the discovery of therapeutics. Nat Publ Gr 14:314–328 Conway JRW, Carragher NO, Timpson P (2014) Developments in preclinical cancer imaging : innovating the discovery of therapeutics. Nat Publ Gr 14:314–328
2.
go back to reference Kirchhof P, Sipido KR, Cowie MR et al (2014) The continuum of personalized cardiovascular medicine: a position paper of the European Society of Cardiology. Eur Heart J:3250–3257 Kirchhof P, Sipido KR, Cowie MR et al (2014) The continuum of personalized cardiovascular medicine: a position paper of the European Society of Cardiology. Eur Heart J:3250–3257
3.
go back to reference Rudin M, Weissleder R (2003) Molecular imaging in drug discovery and development. Nat Rev Drug Discov 2:123–131CrossRefPubMed Rudin M, Weissleder R (2003) Molecular imaging in drug discovery and development. Nat Rev Drug Discov 2:123–131CrossRefPubMed
5.
go back to reference Patino M, Fuentes JM, Singh S, Hahn PF, Sahani DV (2015) Iterative Reconstruction Techniques in Abdominopelvic CT: Technical Concepts and Clinical Implementation. Am J Roentgenol 205:W19–W31CrossRef Patino M, Fuentes JM, Singh S, Hahn PF, Sahani DV (2015) Iterative Reconstruction Techniques in Abdominopelvic CT: Technical Concepts and Clinical Implementation. Am J Roentgenol 205:W19–W31CrossRef
6.
go back to reference Noël PB, Renger B, Fiebich M et al (2013) Does Iterative Reconstruction Lower CT Radiation Dose: Evaluation of 15,000 Examinations. Chen C-T, editor. PLoS One Public Library Sci 8:e81141CrossRef Noël PB, Renger B, Fiebich M et al (2013) Does Iterative Reconstruction Lower CT Radiation Dose: Evaluation of 15,000 Examinations. Chen C-T, editor. PLoS One Public Library Sci 8:e81141CrossRef
8.
go back to reference Willemink MJ, Leiner T, de Jong PA et al (2013) Iterative reconstruction techniques for computed tomography part 2: initial results in dose reduction and image quality. Eur Radiol 23:1632–1642CrossRefPubMed Willemink MJ, Leiner T, de Jong PA et al (2013) Iterative reconstruction techniques for computed tomography part 2: initial results in dose reduction and image quality. Eur Radiol 23:1632–1642CrossRefPubMed
9.
go back to reference McCollough CH, Chen GH, Kalender W et al (2012) Achieving Routine Submillisievert CT Scanning: Report from the Summit on Management of Radiation Dose in CT. Radiology 264:567–580CrossRefPubMedPubMedCentral McCollough CH, Chen GH, Kalender W et al (2012) Achieving Routine Submillisievert CT Scanning: Report from the Summit on Management of Radiation Dose in CT. Radiology 264:567–580CrossRefPubMedPubMedCentral
10.
go back to reference Muenzel D, Kabus S, Gramer B et al (2013) Dynamic CT Perfusion Imaging of the Myocardium: A Technical Note on Improvement of Image Quality. PLoS One8 Muenzel D, Kabus S, Gramer B et al (2013) Dynamic CT Perfusion Imaging of the Myocardium: A Technical Note on Improvement of Image Quality. PLoS One8
11.
go back to reference Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurement by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432CrossRefPubMed Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurement by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432CrossRefPubMed
12.
go back to reference Fischer MA, Kartalis N, Grigoriadis A et al (2015) Perfusion computed tomography for detection of hepatocellular carcinoma in patients with liver cirrhosis. Eur Radiol 25:3123–3132CrossRefPubMed Fischer MA, Kartalis N, Grigoriadis A et al (2015) Perfusion computed tomography for detection of hepatocellular carcinoma in patients with liver cirrhosis. Eur Radiol 25:3123–3132CrossRefPubMed
13.
go back to reference McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications. Radiology 276:637–653CrossRefPubMedPubMedCentral McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications. Radiology 276:637–653CrossRefPubMedPubMedCentral
14.
go back to reference Patino M, Prochowski A, Agrawal MD et al (2016) Material Separation Using Dual-Energy CT: Current and Emerging Applications. RadioGraphics Radiol Soc N Am 36:1087–1105 Patino M, Prochowski A, Agrawal MD et al (2016) Material Separation Using Dual-Energy CT: Current and Emerging Applications. RadioGraphics Radiol Soc N Am 36:1087–1105
15.
go back to reference Johnson TRC, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: Initial experience. Eur Radiol 16:1409–1415CrossRefPubMed Johnson TRC, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: Initial experience. Eur Radiol 16:1409–1415CrossRefPubMed
16.
go back to reference Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol Springer-Verlag 16:256–268CrossRef Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol Springer-Verlag 16:256–268CrossRef
17.
go back to reference Pinho DF, Kulkarni NM, Krishnaraj A, Kalva SP, Sahani DV (2013) Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: Image quality, enhancement, diagnosis and radiation dose. Eur Radiol 23:351–359CrossRefPubMed Pinho DF, Kulkarni NM, Krishnaraj A, Kalva SP, Sahani DV (2013) Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: Image quality, enhancement, diagnosis and radiation dose. Eur Radiol 23:351–359CrossRefPubMed
18.
go back to reference Karcaaltincaba M, Aykut A (2010) Dual-energy CT revisited by multidetector ct: review of principles and clinical applications. Diagn Interv Radiol 17:181–194PubMed Karcaaltincaba M, Aykut A (2010) Dual-energy CT revisited by multidetector ct: review of principles and clinical applications. Diagn Interv Radiol 17:181–194PubMed
19.
go back to reference Vlassenbroek A. Dual Layer CT (2011) In: Johnson T, Fink C, Schönberg SO, Reiser MF, editors. Dual Energy CT in Clinical Practice. Berlin, Heidelberg: Springer Berlin Heidelberg; 21–34 Vlassenbroek A. Dual Layer CT (2011) In: Johnson T, Fink C, Schönberg SO, Reiser MF, editors. Dual Energy CT in Clinical Practice. Berlin, Heidelberg: Springer Berlin Heidelberg; 21–34
20.
go back to reference Hickethier DM, Byrtus T, Wybranski J et al (2017) Image quality evaluation of dual-layer spectral detector CT of the chest and comparison with conventional CT imaging. Eur J Radiol 52 Hickethier DM, Byrtus T, Wybranski J et al (2017) Image quality evaluation of dual-layer spectral detector CT of the chest and comparison with conventional CT imaging. Eur J Radiol 52
21.
go back to reference Altman A, Carmi R (2009) TU-E-210A-03: A Double-Layer Detector, Dual-Energy CT — Principles, Advantages and Applications. Med Phys Am Assoc Phys Med 36:2750 Altman A, Carmi R (2009) TU-E-210A-03: A Double-Layer Detector, Dual-Energy CT — Principles, Advantages and Applications. Med Phys Am Assoc Phys Med 36:2750
22.
go back to reference Agrawal MD, Pinho DF, Kulkarni NM, Hahn PF, Guimaraes AR, Sahani DV (2014) Oncologic Applications of Dual-Energy CT in the Abdomen. RadioGraphics 34:589–612CrossRefPubMed Agrawal MD, Pinho DF, Kulkarni NM, Hahn PF, Guimaraes AR, Sahani DV (2014) Oncologic Applications of Dual-Energy CT in the Abdomen. RadioGraphics 34:589–612CrossRefPubMed
23.
go back to reference Marin D, Boll DT, Mileto A, Nelson RC (2014) State of the art: dual-energy CT of the abdomen. Radiology 271:327–342CrossRefPubMed Marin D, Boll DT, Mileto A, Nelson RC (2014) State of the art: dual-energy CT of the abdomen. Radiology 271:327–342CrossRefPubMed
24.
go back to reference Flors L, Leiva-Salinas C, Norton PT, Patrie JT, Hagspiel KD (2013) Endoleak Detection After Endovascular Repair of Thoracic Aortic Aneurysm Using Dual-Source Dual-Energy CT: Suitable Scanning Protocols and Potential Radiation Dose Reduction. Am J Roentgenol 200:451–460CrossRef Flors L, Leiva-Salinas C, Norton PT, Patrie JT, Hagspiel KD (2013) Endoleak Detection After Endovascular Repair of Thoracic Aortic Aneurysm Using Dual-Source Dual-Energy CT: Suitable Scanning Protocols and Potential Radiation Dose Reduction. Am J Roentgenol 200:451–460CrossRef
25.
go back to reference Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: Phantom study and preliminary experience with renal masses. Am J Roentgenol 196:693–700CrossRef Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: Phantom study and preliminary experience with renal masses. Am J Roentgenol 196:693–700CrossRef
26.
go back to reference Almeida IP, Schyns LEJR, Ollers MC (2017) Dual-energy CT quantitative imaging: a comparison study between twin-beam and dual-source CT scanners 171–9 Almeida IP, Schyns LEJR, Ollers MC (2017) Dual-energy CT quantitative imaging: a comparison study between twin-beam and dual-source CT scanners 171–9
27.
go back to reference Pelgrim GJ, van Hamersvelt RW, Willemink MJ et al. (2017) Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT. Eur Radiol Pelgrim GJ, van Hamersvelt RW, Willemink MJ et al. (2017) Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT. Eur Radiol
28.
go back to reference Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol Springer Berlin Heidelberg 24:512–518 Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol Springer Berlin Heidelberg 24:512–518
30.
go back to reference Kalender W, Perman W, Vetter J, Klotz E (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13:334–339CrossRefPubMed Kalender W, Perman W, Vetter J, Klotz E (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13:334–339CrossRefPubMed
31.
go back to reference Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Investig Radiol 47:656–661CrossRef Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Investig Radiol 47:656–661CrossRef
32.
go back to reference Li Y, Shi G, Wang S, Wang S, Wu R (2013) Iodine quantification with dual-energy CT: Phantom study and preliminary experience with VX2 residual tumour in rabbits after radiofrequency ablation. Br J Radiol 86 Li Y, Shi G, Wang S, Wang S, Wu R (2013) Iodine quantification with dual-energy CT: Phantom study and preliminary experience with VX2 residual tumour in rabbits after radiofrequency ablation. Br J Radiol 86
33.
go back to reference Euler A, Parakh A, Falkowski AL et al (2016) Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter. Invest Radiol 1 Euler A, Parakh A, Falkowski AL et al (2016) Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter. Invest Radiol 1
34.
go back to reference Scheffel H, Alkadhi H, Plass A et al (2006) Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747CrossRefPubMedPubMedCentral Scheffel H, Alkadhi H, Plass A et al (2006) Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747CrossRefPubMedPubMedCentral
Metadata
Title
Dual-energy CT: a phantom comparison of different platforms for abdominal imaging
Authors
Thorsten Sellerer
Peter B. Noël
Manuel Patino
Anushri Parakh
Sebastian Ehn
Sascha Zeiter
Jasmin A. Holz
Johannes Hammel
Alexander A. Fingerle
Franz Pfeiffer
David Maintz
Ernst J. Rummeny
Daniela Muenzel
Dushyant V. Sahani
Publication date
01-07-2018
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 7/2018
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-5238-5

Other articles of this Issue 7/2018

European Radiology 7/2018 Go to the issue