01-10-2017 | Ultrasound

Bacterial contamination of ultrasound probes in different radiological institutions before and after specific hygiene training: do we have a general hygienical problem?

Authors: Thomas Sartoretti, Elisabeth Sartoretti, Candid Bucher, Aleksis Doert, Christoph Binkert, Klaus Hergan, Matthias Meissnitzer, Johannes Froehlich, Orpheus Kolokythas, Simon Matoori, Christina Orasch, Sebastian Kos, Sabine Sartoretti-Schefer, Andreas Gutzeit

Published in: European Radiology | Issue 10/2017

Login to get access

Abstract

Objectives

Aim was to investigate hygienic conditions of ultrasound probes before and after hygiene training in radiology institutions in comparison to bacterial contamination in public places.

Methods

In three radiology departments, bacterial contamination was evaluated using baseline agar plates for cultures taken from 36 ultrasound probes. Afterwards teams were trained by a hygiene service centre and 36 ultrasound probes were routinely disinfected with regular disinfecting wipes and then evaluated. In comparison, bacterial contamination in public places (bus poles, n = 11; toilet seats, n = 10) were analysed. Plates were routinely incubated and the number of colony forming units (CFU) analysed.

Results

Cultures taken from the probes showed a median of 53 CFU before and 0 CFU after training (p < 0.001). Cultures taken from public places showed a median of 4 CFU from toilets and 28 from bus poles and had lower bacterial load in comparison to ultrasound probes before training (p = 0.055, toilets; p = 0.772, bus poles), without statistical significance.

Conclusions

Bacterial contamination of ultrasound probes prior to hygiene training proved to be high and showed higher bacterial load than toilets seats or bus poles. Radiologists should be aware that the lack of hygiene in the field of ultrasound diagnostics puts patients at risk of healthcare-associated infections.

Key points

Hospital-associated infections are a problem for patient care.
Hygiene training of staff prevents bacterial contamination of ultrasound probes.
Disinfection of ultrasound probes is an easy method to protect patients.
Literature
1.
go back to reference Murphy RA, Chua AC (2016) Prevention of common healthcare-associated infections in humanitarian hospitals. Curr Opin Infect Dis 29:381–387CrossRefPubMed Murphy RA, Chua AC (2016) Prevention of common healthcare-associated infections in humanitarian hospitals. Curr Opin Infect Dis 29:381–387CrossRefPubMed
2.
go back to reference Klevens RM, Edwards JR, Richards CL et al (2007) Estimating health care-associated infections and deaths in US Hospitals, 2002. Public Health Rep 122:160–166CrossRefPubMedPubMedCentral Klevens RM, Edwards JR, Richards CL et al (2007) Estimating health care-associated infections and deaths in US Hospitals, 2002. Public Health Rep 122:160–166CrossRefPubMedPubMedCentral
3.
go back to reference Stone PW (2010) Economic burden of healthcare-associated infections: an American perspective. Expert Rev Pharmacoecon Outcomes Res 9:417–422CrossRef Stone PW (2010) Economic burden of healthcare-associated infections: an American perspective. Expert Rev Pharmacoecon Outcomes Res 9:417–422CrossRef
4.
go back to reference Paul R, Das NK, Dutta R, Bandyopadhyay R, Banerjee AK (2011) Bacterial contamination of the hands of doctors: a study in the medicine and dermatology wards. Indian J Dermatol Venereol Leprol 77:307–313CrossRefPubMed Paul R, Das NK, Dutta R, Bandyopadhyay R, Banerjee AK (2011) Bacterial contamination of the hands of doctors: a study in the medicine and dermatology wards. Indian J Dermatol Venereol Leprol 77:307–313CrossRefPubMed
5.
go back to reference Ejtehadi F, Ejtehadi F, Cohen J, Arasteh MM (2014) A safe and practical decontamination method to reduce the risk of bacterial colonizsation of ultrasound transducers. J Clin Ultrasound 42:395–398CrossRefPubMed Ejtehadi F, Ejtehadi F, Cohen J, Arasteh MM (2014) A safe and practical decontamination method to reduce the risk of bacterial colonizsation of ultrasound transducers. J Clin Ultrasound 42:395–398CrossRefPubMed
6.
go back to reference Karadeniz YM, Kilic D, Altan K, Altinok D, Güney S (2001) Evaluation of the role of ultrasound machines as a source of nosocomial and cross-infection. Invest Radiol 36:554–558CrossRef Karadeniz YM, Kilic D, Altan K, Altinok D, Güney S (2001) Evaluation of the role of ultrasound machines as a source of nosocomial and cross-infection. Invest Radiol 36:554–558CrossRef
7.
go back to reference Mullaney PJ, Munthali P, Vlachou P, Jenkins D, Rathod A, Entwisle J (2007) How clean is your probe? Microbiological assessment of ultrasound transducers in routine clinical use, and cost-effective ways to reduce contamination. Clin Radiol 62:694–698CrossRefPubMed Mullaney PJ, Munthali P, Vlachou P, Jenkins D, Rathod A, Entwisle J (2007) How clean is your probe? Microbiological assessment of ultrasound transducers in routine clinical use, and cost-effective ways to reduce contamination. Clin Radiol 62:694–698CrossRefPubMed
8.
go back to reference Frazee B, Fahimi J, Lambert L, Nagdev A (2011) Emergency department ultrasonographic probe contamination and experimental model of probe disinfection. Ann Emerg Med 58:56–63CrossRefPubMed Frazee B, Fahimi J, Lambert L, Nagdev A (2011) Emergency department ultrasonographic probe contamination and experimental model of probe disinfection. Ann Emerg Med 58:56–63CrossRefPubMed
9.
go back to reference Chu K, Obaid H, Babyn P, Blondeau J (2014) Bacterial contamination of ultrasound probes at a tertiary referral university medical center. AJR 203:928–932CrossRefPubMed Chu K, Obaid H, Babyn P, Blondeau J (2014) Bacterial contamination of ultrasound probes at a tertiary referral university medical center. AJR 203:928–932CrossRefPubMed
10.
go back to reference Davis CP (1996) Normal flora. In: Baron S (ed) Medical microbiology, 4th edn. University of Texas Medical Branch, Galveston Davis CP (1996) Normal flora. In: Baron S (ed) Medical microbiology, 4th edn. University of Texas Medical Branch, Galveston
11.
go back to reference Lawrence MW, Blanks J, Ayala R et al (2014) Hospital-wide survey of bacterial contamination of point-of-care ultrasound probes and coupling gel. J Ultrasound Med 33:457–462CrossRefPubMed Lawrence MW, Blanks J, Ayala R et al (2014) Hospital-wide survey of bacterial contamination of point-of-care ultrasound probes and coupling gel. J Ultrasound Med 33:457–462CrossRefPubMed
12.
go back to reference Koibuchi H, Kotani K, Taniguchi N (2013) Ultrasound probes as a possible vector of bacterial transmission. Med Ultrason 15:41–44CrossRefPubMed Koibuchi H, Kotani K, Taniguchi N (2013) Ultrasound probes as a possible vector of bacterial transmission. Med Ultrason 15:41–44CrossRefPubMed
13.
go back to reference Hayashi S, Koibuchi H, Taniguchi N, Hirai Y (2012) Evaluation of procedures for decontaminating ultrasound probes. J Med Ultrasonics 39:11–14CrossRef Hayashi S, Koibuchi H, Taniguchi N, Hirai Y (2012) Evaluation of procedures for decontaminating ultrasound probes. J Med Ultrasonics 39:11–14CrossRef
14.
go back to reference Gerlich MG, Piegsa J, Schäfer C et al (2015) Improving hospital hygiene to reduce the impact of multidrug-resistant organisms in health care - a prospective controlled multicenter study. BMC Infect Dis 15:441CrossRefPubMedPubMedCentral Gerlich MG, Piegsa J, Schäfer C et al (2015) Improving hospital hygiene to reduce the impact of multidrug-resistant organisms in health care - a prospective controlled multicenter study. BMC Infect Dis 15:441CrossRefPubMedPubMedCentral
15.
go back to reference [No authors listed] (2015) Health care-associated infections in the USA. Lancet 385:304 [No authors listed] (2015) Health care-associated infections in the USA. Lancet 385:304
16.
go back to reference Buerke B, Mellmann A, Kipp F, Heindel W, Wessling J (2012) Hyienic aspects in radiology: what the radiologist should know. Röfo 184:1099–1109PubMed Buerke B, Mellmann A, Kipp F, Heindel W, Wessling J (2012) Hyienic aspects in radiology: what the radiologist should know. Röfo 184:1099–1109PubMed
17.
go back to reference Muradali D, Gold WL, Phillips A, Wilson S (1995) Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study. AJR Am J Roentgenol 164:1521–1524CrossRefPubMed Muradali D, Gold WL, Phillips A, Wilson S (1995) Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study. AJR Am J Roentgenol 164:1521–1524CrossRefPubMed
18.
go back to reference Rutala WA, Weber DJ (2010) Guideline for disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol 31:2107–2117 Rutala WA, Weber DJ (2010) Guideline for disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol 31:2107–2117
Metadata
Title
Bacterial contamination of ultrasound probes in different radiological institutions before and after specific hygiene training: do we have a general hygienical problem?
Authors
Thomas Sartoretti
Elisabeth Sartoretti
Candid Bucher
Aleksis Doert
Christoph Binkert
Klaus Hergan
Matthias Meissnitzer
Johannes Froehlich
Orpheus Kolokythas
Simon Matoori
Christina Orasch
Sebastian Kos
Sabine Sartoretti-Schefer
Andreas Gutzeit
Publication date
01-10-2017
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 10/2017
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4812-1

Other articles of this Issue 10/2017

European Radiology 10/2017 Go to the issue