Skip to main content
Top
Published in: European Radiology 12/2016

01-12-2016 | Cardiac

Real-time SPARSE-SENSE cardiac cine MR imaging: optimization of image reconstruction and sequence validation

Authors: Juliane Goebel, Felix Nensa, Bettina Bomas, Haemi P. Schemuth, Stefan Maderwald, Marcel Gratz, Harald H. Quick, Thomas Schlosser, Kai Nassenstein

Published in: European Radiology | Issue 12/2016

Login to get access

Abstract

Objectives

Improved real-time cardiac magnetic resonance (CMR) sequences have currently been introduced, but so far only limited practical experience exists. This study aimed at image reconstruction optimization and clinical validation of a new highly accelerated real-time cine SPARSE-SENSE sequence.

Methods

Left ventricular (LV) short-axis stacks of a real-time free-breathing SPARSE-SENSE sequence with high spatiotemporal resolution and of a standard segmented cine SSFP sequence were acquired at 1.5 T in 11 volunteers and 15 patients. To determine the optimal iterations, all volunteers’ SPARSE-SENSE images were reconstructed using 10-200 iterations, and contrast ratios, image entropies, and reconstruction times were assessed. Subsequently, the patients’ SPARSE-SENSE images were reconstructed with the clinically optimal iterations. LV volumetric values were evaluated and compared between both sequences.

Results

Sufficient image quality and acceptable reconstruction times were achieved when using 80 iterations. Bland-Altman plots and Passing-Bablok regression showed good agreement for all volumetric parameters.

Conclusions

80 iterations are recommended for iterative SPARSE-SENSE image reconstruction in clinical routine. Real-time cine SPARSE-SENSE yielded comparable volumetric results as the current standard SSFP sequence. Due to its intrinsic low image acquisition times, real-time cine SPARSE-SENSE imaging with iterative image reconstruction seems to be an attractive alternative for LV function analysis.

Key Points

A highly accelerated real-time CMR sequence using SPARSE-SENSE was evaluated.
SPARSE-SENSE allows free breathing in real-time cardiac cine imaging.
For clinically optimal SPARSE-SENSE image reconstruction, 80 iterations are recommended.
Real-time SPARSE-SENSE imaging yielded comparable volumetric results as the reference SSFP sequence.
The fast SPARSE-SENSE sequence is an attractive alternative to standard SSFP sequences.
Appendix
Available only for authorised users
Literature
1.
go back to reference Finn JP, Nael K, Deshpande V, Ratib O, Laub G (2006) Cardiac MR imaging: state of the technology. Radiology 241:338–354CrossRefPubMed Finn JP, Nael K, Deshpande V, Ratib O, Laub G (2006) Cardiac MR imaging: state of the technology. Radiology 241:338–354CrossRefPubMed
2.
go back to reference Feng L, Srichai MB, Lim RP et al (2013) Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE. Magn Reson Med 70:64–74CrossRefPubMed Feng L, Srichai MB, Lim RP et al (2013) Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE. Magn Reson Med 70:64–74CrossRefPubMed
3.
go back to reference Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRefPubMed Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRefPubMed
4.
go back to reference Vincenti G, Monney P, Chaptinel J et al (2014) Compressed sensing single-breath-hold CMR for fast quantification of LV function, volumes, and mass. J Am Coll Cardiol Img 7:882–892CrossRef Vincenti G, Monney P, Chaptinel J et al (2014) Compressed sensing single-breath-hold CMR for fast quantification of LV function, volumes, and mass. J Am Coll Cardiol Img 7:882–892CrossRef
5.
go back to reference Stalder AF, Schmidt M, Quick HH et al (2014) Highly undersampled contrast-enhanced MRA with iterative reconstruction: Integration in a clinical setting. Magn Reson Med. doi:10.1002/mrm.25565 Stalder AF, Schmidt M, Quick HH et al (2014) Highly undersampled contrast-enhanced MRA with iterative reconstruction: Integration in a clinical setting. Magn Reson Med. doi:10.​1002/​mrm.​25565
6.
go back to reference Gonzalez RC, Woods RE, Eddins SL (2003) In: Anonymous Digital Image Processing Using MATLAB, pp Chapter 11 Gonzalez RC, Woods RE, Eddins SL (2003) In: Anonymous Digital Image Processing Using MATLAB, pp Chapter 11
7.
go back to reference Nassenstein K, Eberle H, Maderwald S et al (2010) Single breath-hold magnetic resonance cine imaging for fast assessment of global and regional left ventricular function in clinical routine. Eur Radiol 20:2341–2347CrossRefPubMed Nassenstein K, Eberle H, Maderwald S et al (2010) Single breath-hold magnetic resonance cine imaging for fast assessment of global and regional left ventricular function in clinical routine. Eur Radiol 20:2341–2347CrossRefPubMed
8.
go back to reference Mehta CR, Patel NR (1986) Algorithm 643. FEXACT: a fortran subroutine for fisher's exact test on unordered r*c contingency tables. ACM Trans Math Softw 12:154–161CrossRef Mehta CR, Patel NR (1986) Algorithm 643. FEXACT: a fortran subroutine for fisher's exact test on unordered r*c contingency tables. ACM Trans Math Softw 12:154–161CrossRef
9.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed
10.
go back to reference Passing H, Bablok (1983) A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, part I. J Clin Chem Clin Biochem 21:709–720PubMed Passing H, Bablok (1983) A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, part I. J Clin Chem Clin Biochem 21:709–720PubMed
11.
go back to reference Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU (2003) Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging 17:323–329CrossRefPubMed Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU (2003) Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging 17:323–329CrossRefPubMed
12.
go back to reference Buda AJ, Pinsky MR, Ingels NB Jr, Daughters GT 2nd, Stinson EB, Alderman EL (1979) Effect of intrathoracic pressure on left ventricular performance. N Engl J Med 301:453–459CrossRefPubMed Buda AJ, Pinsky MR, Ingels NB Jr, Daughters GT 2nd, Stinson EB, Alderman EL (1979) Effect of intrathoracic pressure on left ventricular performance. N Engl J Med 301:453–459CrossRefPubMed
13.
go back to reference Voit D, Zhang S, Unterberg-Buchwald C, Sohns JM, Lotz J, Frahm J (2013) Real-time cardiovascular magnetic resonance at 1.5 T using balanced SSFP and 40 ms resolution. J Cardiovasc Magn Reson 15:79CrossRefPubMedPubMedCentral Voit D, Zhang S, Unterberg-Buchwald C, Sohns JM, Lotz J, Frahm J (2013) Real-time cardiovascular magnetic resonance at 1.5 T using balanced SSFP and 40 ms resolution. J Cardiovasc Magn Reson 15:79CrossRefPubMedPubMedCentral
14.
go back to reference Jahnke C, Nagel E, Gebker R et al (2007) Four-dimensional single breathhold magnetic resonance imaging using kt-BLAST enables reliable assessment of left- and right-ventricular volumes and mass. J Magn Reson Imaging 25:737–742CrossRefPubMed Jahnke C, Nagel E, Gebker R et al (2007) Four-dimensional single breathhold magnetic resonance imaging using kt-BLAST enables reliable assessment of left- and right-ventricular volumes and mass. J Magn Reson Imaging 25:737–742CrossRefPubMed
15.
go back to reference Eberle HC, Nassenstein K, Jensen CJ et al (2010) Rapid MR assessment of left ventricular systolic function after acute myocardial infarction using single breath-hold cine imaging with the temporal parallel acquisition technique (TPAT) and 4D guide-point modelling analysis of left ventricular function. Eur Radiol 20:73–80CrossRefPubMed Eberle HC, Nassenstein K, Jensen CJ et al (2010) Rapid MR assessment of left ventricular systolic function after acute myocardial infarction using single breath-hold cine imaging with the temporal parallel acquisition technique (TPAT) and 4D guide-point modelling analysis of left ventricular function. Eur Radiol 20:73–80CrossRefPubMed
16.
go back to reference Young AA, Cowan BR, Schoenberg SO, Wintersperger BJ (2008) Feasibility of single breath-hold left ventricular function with 3 tesla TSENSE acquisition and 3D modeling analysis. J Cardiovasc Magn Reson 10:24CrossRefPubMedPubMedCentral Young AA, Cowan BR, Schoenberg SO, Wintersperger BJ (2008) Feasibility of single breath-hold left ventricular function with 3 tesla TSENSE acquisition and 3D modeling analysis. J Cardiovasc Magn Reson 10:24CrossRefPubMedPubMedCentral
17.
Metadata
Title
Real-time SPARSE-SENSE cardiac cine MR imaging: optimization of image reconstruction and sequence validation
Authors
Juliane Goebel
Felix Nensa
Bettina Bomas
Haemi P. Schemuth
Stefan Maderwald
Marcel Gratz
Harald H. Quick
Thomas Schlosser
Kai Nassenstein
Publication date
01-12-2016
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 12/2016
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4301-y

Other articles of this Issue 12/2016

European Radiology 12/2016 Go to the issue