Low-Dose CT of the Paranasal Sinuses: Minimizing X-Ray Exposure with Spectral Shaping

Authors: Wolfgang Wuest, Matthias May, Marc Saake, Michael Brand, Michael Uder, Michael Lell

Published in: European Radiology | Issue 11/2016

Login to get access

Abstract

Objectives

Shaping the energy spectrum of the X-ray beam has been shown to be beneficial in low-dose CT. This study’s aim was to investigate dose and image quality of tin filtration at 100 kV for pre-operative planning in low-dose paranasal CT imaging in a large patient cohort.

Methods

In a prospective trial, 129 patients were included. 64 patients were randomly assigned to the study protocol (100 kV with additional tin filtration, 150mAs, 192x0.6-mm slice collimation) and 65 patients to the standard low-dose protocol (100 kV, 50mAs, 128 × 0.6-mm slice collimation). To assess the image quality, subjective parameters were evaluated using a five-point scale. This scale was applied on overall image quality and contour delineation of critical anatomical structures.

Results

All scans were of diagnostic image quality. Bony structures were of good diagnostic image quality in both groups, soft tissues were of sufficient diagnostic image quality in the study group because of a high level of noise. Radiation exposure was very low in both groups, but significantly lower in the study group (CTDIvol 1.2 mGy vs. 4.4 mGy, p < 0.001).

Conclusions

Spectral optimization (tin filtration at 100 kV) allows for visualization of the paranasal sinus with sufficient image quality at a very low radiation exposure.

Key Points

Spectral optimization (tin filtration) is beneficial to low-dose parasinus CT
Tin filtration at 100 kV yields sufficient image quality for pre-operative planning
Diagnostic parasinus CT can be performed with an effective dose <0.05 mSv
Literature
1.
go back to reference Hopper RA, Salemy S, Sze RW (2006) Diagnosis of midface fractures with CT: what the surgeon needs to know. Radiographics 26:783–793CrossRefPubMed Hopper RA, Salemy S, Sze RW (2006) Diagnosis of midface fractures with CT: what the surgeon needs to know. Radiographics 26:783–793CrossRefPubMed
2.
go back to reference Duvoisin B, Landry M, Chapuis L, Krayenbuhl M, Schnyder P (1991) Low-dose CT and inflammatory disease of the paranasal sinuses. Neuroradiology 33:4036CrossRef Duvoisin B, Landry M, Chapuis L, Krayenbuhl M, Schnyder P (1991) Low-dose CT and inflammatory disease of the paranasal sinuses. Neuroradiology 33:4036CrossRef
3.
go back to reference Tack D, Widelec J, De Maertelaer V, Bailly JM, Delcour C, Gevenois PA (2003) Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis. AJR Am J Roentgenol 181:939–944CrossRefPubMed Tack D, Widelec J, De Maertelaer V, Bailly JM, Delcour C, Gevenois PA (2003) Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis. AJR Am J Roentgenol 181:939–944CrossRefPubMed
4.
go back to reference Kaur J, Chopra R (2010) Three dimensional CT reconstruction for the evaluation and surgical planning of mid face fractures: a 100 case study. J Maxillofac Oral Surg 9:323–328CrossRefPubMed Kaur J, Chopra R (2010) Three dimensional CT reconstruction for the evaluation and surgical planning of mid face fractures: a 100 case study. J Maxillofac Oral Surg 9:323–328CrossRefPubMed
5.
go back to reference Abul-Kasim K, Strombeck A, Sahlstrand-Johnson P (2009) Low-dose computed tomography of the paranasal sinuses: radiation doses and reliability analysis. Am J Otolaryngol 32:47–51CrossRefPubMed Abul-Kasim K, Strombeck A, Sahlstrand-Johnson P (2009) Low-dose computed tomography of the paranasal sinuses: radiation doses and reliability analysis. Am J Otolaryngol 32:47–51CrossRefPubMed
6.
go back to reference Hoang JK, Eastwood JD, Tebbit CL, Glastonbury CM (2010) Multiplanar sinus CT: a systematic approach to imaging before functional endoscopic sinus surgery. Am J Roentgenol 194:W527–W536CrossRef Hoang JK, Eastwood JD, Tebbit CL, Glastonbury CM (2010) Multiplanar sinus CT: a systematic approach to imaging before functional endoscopic sinus surgery. Am J Roentgenol 194:W527–W536CrossRef
7.
go back to reference Perisinakis K, Raissaki M, Theocharopoulos N, Damilakis J, Gourtsoyiannis N (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32:1024–1030CrossRefPubMed Perisinakis K, Raissaki M, Theocharopoulos N, Damilakis J, Gourtsoyiannis N (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32:1024–1030CrossRefPubMed
8.
go back to reference Hein E, Rogalla P, Klingebiel R, Hamm B (2002) Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol 12:1693–1696CrossRefPubMed Hein E, Rogalla P, Klingebiel R, Hamm B (2002) Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol 12:1693–1696CrossRefPubMed
9.
go back to reference Brem MH, Zamani AA, Riva R et al (2007) Multidetector CT of the paranasal sinus: potential for radiation dose reduction. Radiology 243:847–852CrossRefPubMed Brem MH, Zamani AA, Riva R et al (2007) Multidetector CT of the paranasal sinus: potential for radiation dose reduction. Radiology 243:847–852CrossRefPubMed
10.
go back to reference Hagtvedt T, Aalokken TM, Notthellen J, Kolbenstvedt A (2003) A new low-dose CT examination compared with standard-dose CT in the diagnosis of acute sinusitis. Eur Radiol 13:976–980PubMed Hagtvedt T, Aalokken TM, Notthellen J, Kolbenstvedt A (2003) A new low-dose CT examination compared with standard-dose CT in the diagnosis of acute sinusitis. Eur Radiol 13:976–980PubMed
11.
go back to reference Hojreh A, Czerny C, Kainberger F (2005) Dose classification scheme for computed tomography of the paranasal sinuses. Eur J Radiol 56:31–37CrossRefPubMed Hojreh A, Czerny C, Kainberger F (2005) Dose classification scheme for computed tomography of the paranasal sinuses. Eur J Radiol 56:31–37CrossRefPubMed
12.
go back to reference Wuest W, May MS, Brand M, Bayerl N, Krauss A, Uder M, Lell M (2015) Improved Image Quality in Head and Neck CT Using a 3D Iterative Approach toReduce Metal Artifact. AJNR Am J Neuroradiol 36(10):1988--93 Wuest W, May MS, Brand M, Bayerl N, Krauss A, Uder M, Lell M (2015) Improved Image Quality in Head and Neck CT Using a 3D Iterative Approach toReduce Metal Artifact. AJNR Am J Neuroradiol 36(10):1988--93
13.
go back to reference Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166CrossRefPubMed Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166CrossRefPubMed
14.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed
15.
16.
go back to reference Chodick G, Bekiroglu N, Hauptmann M et al (2008) Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists. Am J Epidemiol 168:620–631CrossRefPubMedPubMedCentral Chodick G, Bekiroglu N, Hauptmann M et al (2008) Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists. Am J Epidemiol 168:620–631CrossRefPubMedPubMedCentral
17.
go back to reference Yuan MK, Tsai DC, Chang SC et al (2013) The risk of cataract associated with repeated head and neck CT studies: a nationwide population-based study. AJR Am J Roentgenol 201:626–630CrossRefPubMed Yuan MK, Tsai DC, Chang SC et al (2013) The risk of cataract associated with repeated head and neck CT studies: a nationwide population-based study. AJR Am J Roentgenol 201:626–630CrossRefPubMed
18.
go back to reference ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. 2007. Elsevier, Orlando ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. 2007. Elsevier, Orlando
19.
go back to reference NCRP NCoRPaM (1985) Induction of thyroid cancer by ionizing radiation. NCRP report 80. NCRP Publications, Bethesda NCRP NCoRPaM (1985) Induction of thyroid cancer by ionizing radiation. NCRP report 80. NCRP Publications, Bethesda
20.
go back to reference Cohen MD (2009) Pediatric CT radiation dose: how low can you go? Am J Roentgenol 192:1292–1303CrossRef Cohen MD (2009) Pediatric CT radiation dose: how low can you go? Am J Roentgenol 192:1292–1303CrossRef
21.
go back to reference Aalokken TM, Hagtvedt T, Dalen I, Kolbenstvedt A (2003) Conventional sinus radiography compared with CT in the diagnosis of acute sinusitis. Dentomaxillofac Radiol 32:60–62CrossRefPubMed Aalokken TM, Hagtvedt T, Dalen I, Kolbenstvedt A (2003) Conventional sinus radiography compared with CT in the diagnosis of acute sinusitis. Dentomaxillofac Radiol 32:60–62CrossRefPubMed
22.
go back to reference Hoxworth JM, Lal D, Fletcher GP et al (2014) Radiation dose reduction in paranasal sinus CT using model-based iterative reconstruction. AJNR Am J Neuroradiol 35:644–649CrossRefPubMed Hoxworth JM, Lal D, Fletcher GP et al (2014) Radiation dose reduction in paranasal sinus CT using model-based iterative reconstruction. AJNR Am J Neuroradiol 35:644–649CrossRefPubMed
Metadata
Title
Low-Dose CT of the Paranasal Sinuses: Minimizing X-Ray Exposure with Spectral Shaping
Authors
Wolfgang Wuest
Matthias May
Marc Saake
Michael Brand
Michael Uder
Michael Lell
Publication date
01-11-2016
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2016
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4263-0

Other articles of this Issue 11/2016

European Radiology 11/2016 Go to the issue