01-12-2016 | Magnetic Resonance
Nonenhanced magnetic resonance angiography (MRA) of the calf arteries at 3 Tesla: intraindividual comparison of 3D flow-dependent subtractive MRA and 2D flow-independent non-subtractive MRA
Published in: European Radiology | Issue 12/2016
Login to get accessAbstract
Purpose
To prospectively compare 3D flow-dependent subtractive MRA vs. 2D flow-independent non-subtractive MRA for assessment of the calf arteries at 3 Tesla.
Methods
Forty-two patients with peripheral arterial occlusive disease underwent nonenhanced MRA of calf arteries at 3 Tesla with 3D flow-dependent subtractive MRA (fast spin echo sequence; 3D-FSE-MRA) and 2D flow-independent non-subtractive MRA (balanced steady-state-free-precession sequence; 2D-bSSFP-MRA). Moreover, all patients underwent contrast-enhanced MRA (CE-MRA) as standard-of-reference. Two readers performed a per-segment evaluation for image quality (4 = excellent to 0 = non-diagnostic) and severity of stenosis.
Results
Image quality scores of 2D-bSSFP-MRA were significantly higher compared to 3D-FSE-MRA (medians across readers: 4 vs. 3; p < 0.0001) with lower rates of non-diagnostic vessel segments on 2D-bSSFP-MRA (reader 1: <1 % vs. 15 %; reader 2: 1 % vs. 29 %; p < 0.05). Diagnostic performance of 2D-bSSFP-MRA and 3D-FSE-MRA across readers showed sensitivities of 89 % (214/240) vs. 70 % (168/240), p = 0.0153; specificities: 91 % (840/926) vs. 63 % (585/926), p < 0.0001; and diagnostic accuracies of 90 % (1054/1166) vs. 65 % (753/1166), p < 0.0001.
Conclusion
2D flow-independent non-subtractive MRA (2D-bSSFP-MRA) is a robust nonenhanced MRA technique for assessment of the calf arteries at 3 Tesla with significantly higher image quality and diagnostic accuracy compared to 3D flow-dependent subtractive MRA (3D-FSE-MRA).
Key Points
• 2D flow-independent non-subtractive MRA (2D-bSSFP-MRA) is a robust NE-MRA technique at 3T
• 2D-bSSFP-MRA outperforms 3D flow-dependent subtractive MRA (3D-FSE-MRA) as NE-MRA of calf arteries
• 2D-bSSFP-MRA is a promising alternative to CE-MRA for calf PAOD evaluation