Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose

Authors: Eun-Suk Cho, Tae-Sub Chung, Sung Jun Ahn, KyoungHoon Chong, Jang Hun Baek, Sang Hyun Suh

Published in: European Radiology | Issue 5/2015

Login to get access

Abstract

Objectives

To determine the feasibility of using a 70-kVp protocol compared with a 120-kVp protocol for cerebral CT angiography. An additional target was to investigate a possible reduction in the volume of contrast medium (CM) using the 70-kVp protocol.

Methods

Attenuation value and CNR for iodine were determined at various tube voltage settings using a phantom. Sixty-nine volunteers were randomly assigned to one of three protocols: group A (120-kVp and CM 64 mL), group B (70-kVp and CM 64 mL), or group C (70-kVp and CM 40 mL). The attenuation value, SNR, and CNR of cerebral arteries, subjective image quality, and radiation dose were compared among the groups.

Results

The vascular attenuation, SNR, and CNR of group B were significantly higher than those of group A. Group C had a significantly higher vascular attenuation than group A. Groups B and C were significantly better than group A with respect to subjective image quality. An effective dose of 70-kVp was 10 % lower than that of 120-kVp.

Conclusions

Using 70-kVp improved arterial enhancement, SNR, and CNR, and provided better subjective image quality, using a 10 % lower effective dose. Furthermore, the 70-kVp protocol may both reduce volume of CM by 37.5 % and improve arterial enhancement.

Key Points

Cerebral CT angiography at 70-kVp substantially improved vascular enhancement
Subjective image quality was better at 70-kVp, with lower radiation dose
The volume of contrast media can be substantially reduced at 70-kVp
Literature
1.
2.
go back to reference Tamura Y, Utsunomiya D, Sakamoto T et al (2010) Reduction of contrast material volume in 3D angiography of the brain using MDCT. AJR Am J Roentgenol 195:455–458CrossRefPubMed Tamura Y, Utsunomiya D, Sakamoto T et al (2010) Reduction of contrast material volume in 3D angiography of the brain using MDCT. AJR Am J Roentgenol 195:455–458CrossRefPubMed
3.
go back to reference Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leeuwen MS (2007) Circle of Willis at CT angiography: dose reduction and image quality—reducing tube voltage and increasing tube current settings. Radiology 242:832–839CrossRefPubMed Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leeuwen MS (2007) Circle of Willis at CT angiography: dose reduction and image quality—reducing tube voltage and increasing tube current settings. Radiology 242:832–839CrossRefPubMed
4.
go back to reference Schindera ST, Graca P, Patak MA et al (2009) Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose. Invest Radiol 44:650–655CrossRefPubMed Schindera ST, Graca P, Patak MA et al (2009) Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose. Invest Radiol 44:650–655CrossRefPubMed
5.
go back to reference Heyer CM, Mohr PS, Lemburg SP, Peters SA, Nicolas V (2007) Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study. Radiology 245:577–583CrossRefPubMed Heyer CM, Mohr PS, Lemburg SP, Peters SA, Nicolas V (2007) Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study. Radiology 245:577–583CrossRefPubMed
6.
go back to reference Schueller-Weidekamm C, Schaefer-Prokop CM, Weber M, Herold CJ, Prokop M (2006) CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology 241:899–907CrossRefPubMed Schueller-Weidekamm C, Schaefer-Prokop CM, Weber M, Herold CJ, Prokop M (2006) CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology 241:899–907CrossRefPubMed
7.
go back to reference Bahner ML, Bengel A, Brix G, Zuna I, Kauczor HU, Delorme S (2005) Improved vascular opacification in cerebral computed tomography angiography with 80 kVp. Invest Radiol 40:229–234CrossRefPubMed Bahner ML, Bengel A, Brix G, Zuna I, Kauczor HU, Delorme S (2005) Improved vascular opacification in cerebral computed tomography angiography with 80 kVp. Invest Radiol 40:229–234CrossRefPubMed
8.
go back to reference Cho ES, Chung TS, Oh DK et al (2012) Cerebral computed tomography angiography using a low tube voltage (80 kVp) and a moderate concentration of iodine contrast material: a quantitative and qualitative comparison with conventional computed tomography angiography. Invest Radiol 47:142–147CrossRefPubMed Cho ES, Chung TS, Oh DK et al (2012) Cerebral computed tomography angiography using a low tube voltage (80 kVp) and a moderate concentration of iodine contrast material: a quantitative and qualitative comparison with conventional computed tomography angiography. Invest Radiol 47:142–147CrossRefPubMed
9.
go back to reference Szucs-Farkas Z, Schaller C, Bensler S, Patak MA, Vock P, Schindera ST (2009) Detection of pulmonary emboli with CT angiography at reduced radiation exposure and contrast material volume: comparison of 80 kVp and 120 kVp protocols in a matched cohort. Invest Radiol 44:793–799CrossRefPubMed Szucs-Farkas Z, Schaller C, Bensler S, Patak MA, Vock P, Schindera ST (2009) Detection of pulmonary emboli with CT angiography at reduced radiation exposure and contrast material volume: comparison of 80 kVp and 120 kVp protocols in a matched cohort. Invest Radiol 44:793–799CrossRefPubMed
10.
go back to reference Szucs-Farkas Z, Schibler F, Cullmann J et al (2011) Diagnostic accuracy of pulmonary CT angiography at low tube voltage: intraindividual comparison of a normal-dose protocol at 120 kVp and a low-dose protocol at 80 kVp using reduced amount of contrast medium in a simulation study. AJR Am J Roentgenol 197:W852–859CrossRefPubMed Szucs-Farkas Z, Schibler F, Cullmann J et al (2011) Diagnostic accuracy of pulmonary CT angiography at low tube voltage: intraindividual comparison of a normal-dose protocol at 120 kVp and a low-dose protocol at 80 kVp using reduced amount of contrast medium in a simulation study. AJR Am J Roentgenol 197:W852–859CrossRefPubMed
11.
go back to reference Sahani DV, Kalva SP, Hahn PF, Saini S (2007) 16-MDCT angiography in living kidney donors at various tube potentials: impact on image quality and radiation dose. AJR Am J Roentgenol 188:115–120CrossRefPubMed Sahani DV, Kalva SP, Hahn PF, Saini S (2007) 16-MDCT angiography in living kidney donors at various tube potentials: impact on image quality and radiation dose. AJR Am J Roentgenol 188:115–120CrossRefPubMed
12.
go back to reference Hunsaker AR, Oliva IB, Cai T et al (2010) Contrast opacification using a reduced volume of iodinated contrast material and low peak kilovoltage in pulmonary CT angiography: Objective and subjective evaluation. AJR Am J Roentgenol 195:W118–124CrossRefPubMed Hunsaker AR, Oliva IB, Cai T et al (2010) Contrast opacification using a reduced volume of iodinated contrast material and low peak kilovoltage in pulmonary CT angiography: Objective and subjective evaluation. AJR Am J Roentgenol 195:W118–124CrossRefPubMed
13.
go back to reference Gnannt R, Winklehner A, Goetti R, Schmidt B, Kollias S, Alkadhi H (2012) Low kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol. AJNR Am J Neuroradiol 33:1014–1019CrossRefPubMed Gnannt R, Winklehner A, Goetti R, Schmidt B, Kollias S, Alkadhi H (2012) Low kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol. AJNR Am J Neuroradiol 33:1014–1019CrossRefPubMed
14.
go back to reference Mehran R, Aymong ED, Nikolsky E et al (2004) A simple risk score for prediction of contrast-induced nephropathy after percutaneous coronary intervention: development and initial validation. J Am Coll Cardiol 44:1393–1399PubMed Mehran R, Aymong ED, Nikolsky E et al (2004) A simple risk score for prediction of contrast-induced nephropathy after percutaneous coronary intervention: development and initial validation. J Am Coll Cardiol 44:1393–1399PubMed
15.
go back to reference Gleeson TG, Bulugahapitiya S (2004) Contrast-induced nephropathy. AJR Am J Roentgenol 183:1673–1689CrossRefPubMed Gleeson TG, Bulugahapitiya S (2004) Contrast-induced nephropathy. AJR Am J Roentgenol 183:1673–1689CrossRefPubMed
16.
go back to reference Tepel M, Aspelin P, Lameire N (2006) Contrast-induced nephropathy: a clinical and evidence-based approach. Circulation 113:1799–1806CrossRefPubMed Tepel M, Aspelin P, Lameire N (2006) Contrast-induced nephropathy: a clinical and evidence-based approach. Circulation 113:1799–1806CrossRefPubMed
17.
go back to reference Mehran R, Nikolsky E (2006) Contrast-induced nephropathy: definition, epidemiology, and patients at risk. Kidney Int Suppl:S11–15 Mehran R, Nikolsky E (2006) Contrast-induced nephropathy: definition, epidemiology, and patients at risk. Kidney Int Suppl:S11–15
18.
go back to reference Barrett BJ, Parfrey PS (2006) Clinical practice. Preventing nephropathy induced by contrast medium. N Engl J Med 354:379–386CrossRefPubMed Barrett BJ, Parfrey PS (2006) Clinical practice. Preventing nephropathy induced by contrast medium. N Engl J Med 354:379–386CrossRefPubMed
19.
go back to reference Furtado AD, Adraktas DD, Brasic N et al (2010) The triple rule-out for acute ischemic stroke: imaging the brain, carotid arteries, aorta, and heart. AJNR Am J Neuroradiol 31:1290–1296CrossRefPubMed Furtado AD, Adraktas DD, Brasic N et al (2010) The triple rule-out for acute ischemic stroke: imaging the brain, carotid arteries, aorta, and heart. AJNR Am J Neuroradiol 31:1290–1296CrossRefPubMed
20.
go back to reference Ertl-Wagner BB, Hoffmann RT, Bruning R et al (2004) Multi-detector row CT angiography of the brain at various kilovoltage settings. Radiology 231:528–535CrossRefPubMed Ertl-Wagner BB, Hoffmann RT, Bruning R et al (2004) Multi-detector row CT angiography of the brain at various kilovoltage settings. Radiology 231:528–535CrossRefPubMed
22.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed
23.
go back to reference Curry TSDJ, Murry RC (1990) Basic interactions between X-rays and matter. In: Curry TS, Dowdey JE, Murry RC (eds) Christensen’s physics of diagnostic radiology, 4th edn. Lippincott Williams & Wilkins, Philadelphia, pp 61–69 Curry TSDJ, Murry RC (1990) Basic interactions between X-rays and matter. In: Curry TS, Dowdey JE, Murry RC (eds) Christensen’s physics of diagnostic radiology, 4th edn. Lippincott Williams & Wilkins, Philadelphia, pp 61–69
24.
go back to reference Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRefPubMed Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRefPubMed
25.
go back to reference Yoon DY, Lim KJ, Choi CS et al (2007) Sixteen-detector row CT angiography of the brain: comparison of 3 different volumes of the contrast material. J Comput Assist Tomogr 31:671–676CrossRefPubMed Yoon DY, Lim KJ, Choi CS et al (2007) Sixteen-detector row CT angiography of the brain: comparison of 3 different volumes of the contrast material. J Comput Assist Tomogr 31:671–676CrossRefPubMed
26.
go back to reference Tumlin J, Stacul F, Adam A et al (2006) Pathophysiology of contrast-induced nephropathy. Am J Cardiol 98:14K–20KCrossRefPubMed Tumlin J, Stacul F, Adam A et al (2006) Pathophysiology of contrast-induced nephropathy. Am J Cardiol 98:14K–20KCrossRefPubMed
27.
go back to reference Machida H, Takeuchi H, Tanaka I et al (2013) Improved delineation of arteries in the posterior fossa of the brain by model-based iterative reconstruction in volume-rendered 3D CT angiography. AJNR Am J Neuroradiol 34:971–975CrossRefPubMed Machida H, Takeuchi H, Tanaka I et al (2013) Improved delineation of arteries in the posterior fossa of the brain by model-based iterative reconstruction in volume-rendered 3D CT angiography. AJNR Am J Neuroradiol 34:971–975CrossRefPubMed
Metadata
Title
Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose
Authors
Eun-Suk Cho
Tae-Sub Chung
Sung Jun Ahn
KyoungHoon Chong
Jang Hun Baek
Sang Hyun Suh
Publication date
01-05-2015
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 5/2015
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3540-z

Other articles of this Issue 5/2015

European Radiology 5/2015 Go to the issue