01-08-2014 | Computed Tomography

Improvement of image quality and radiation dose of CT perfusion of the brain by means of low-tube voltage (70 KV)

Authors: Zhen-lin Li, Hang Li, Kai Zhang, Wang-jiang Li, Xian Chen, Bin Wu, Bin Song

Published in: European Radiology | Issue 8/2014

Login to get access

Abstract

Objectives

To investigate the feasibility of 70 kV cerebral CT perfusion by comparing image quality and radiation exposure to 80 kV.

Methods

Thirty patients with suspected cerebral ischemia who underwent dual-source CT perfusion were divided into group A (80 kV, 150 mAs) and group B (70 kV, 150 mAs). Quantitative comparisons were used for maximum enhancement, signal-to-noise index (SNI), and values of cerebral blood flow (CBF), cerebral blood flow (CBV), mean transit time (MTT) on CBF, CBV, and MTT images, and radiation dose from these two groups. Qualitative perfusion images were assessed by two readers.

Results

Maximum enhancement for group B was higher than group A (P < 0.05). There were no significant differences between the two groups for SNI on CBF and CBV maps (P = 0.06 – 0.576), but significant differences for MTT when SNI was measured on frontal white matter and temporo-occipital white matter (P < 0.05). There were no differences among values of CBF, CBV, and MTT for both groups (P = 0.251–0.917). Mean image quality score in group B was higher than group A for CBF (P < 0.05), but no differences for CBV (P = 0.542) and MTT (P = 0.962). Radiation dose for group B decreased compared with group A.

Conclusions

70 kV cerebral CT perfusion reduces radiation dose without compromising image quality.

Key Points

Radiation dose is a key concern with the increased using cerebral CT perfusion. Cerebral CT perfusion of 70 kV reduces radiation dose without compromising image quality.
A 70-kV protocol could be used for cerebral CT perfusion.
Literature
1.
go back to reference Juluru K, Shih JC, Raj A et al (2013) Effects of increased image noise on image quality and quantitative interpretation in brain CT perfusion. AJNR Am J Neuroradiol 34:1506–1512PubMedCrossRef Juluru K, Shih JC, Raj A et al (2013) Effects of increased image noise on image quality and quantitative interpretation in brain CT perfusion. AJNR Am J Neuroradiol 34:1506–1512PubMedCrossRef
2.
go back to reference Koenig M, Kraus M, Theek C, Klotz E, Gehlen W, Heuser L (2001) Quantitative assessment of the ischemic brain by means of perfusion related parameters derived from perfusion CT. Stroke 32:431–437PubMedCrossRef Koenig M, Kraus M, Theek C, Klotz E, Gehlen W, Heuser L (2001) Quantitative assessment of the ischemic brain by means of perfusion related parameters derived from perfusion CT. Stroke 32:431–437PubMedCrossRef
3.
go back to reference Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurements by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432PubMedCrossRef Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurements by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432PubMedCrossRef
5.
go back to reference Eastwood JD, Lev MH, Azhari T et al (2002) CT perfusion scanning with deconvolution analysis: pilot study in patients with acute middle cerebral artery stroke. Radiology 222:227–236PubMedCrossRef Eastwood JD, Lev MH, Azhari T et al (2002) CT perfusion scanning with deconvolution analysis: pilot study in patients with acute middle cerebral artery stroke. Radiology 222:227–236PubMedCrossRef
6.
go back to reference Koenig M, Klotz E, Luka B, Venderink DJ, Spittler JF, Heuser L (1998) Perfusion CT of the brain: diagnostic approach for early detection of ischemic stroke. Radiology 209:85–93PubMed Koenig M, Klotz E, Luka B, Venderink DJ, Spittler JF, Heuser L (1998) Perfusion CT of the brain: diagnostic approach for early detection of ischemic stroke. Radiology 209:85–93PubMed
7.
go back to reference Gillard JH, Antoun NM, Burnet NG, Pickard JD (2001) Reproducibility of quantitative CT perfusion imaging. Br J Radiol 74:552–555PubMedCrossRef Gillard JH, Antoun NM, Burnet NG, Pickard JD (2001) Reproducibility of quantitative CT perfusion imaging. Br J Radiol 74:552–555PubMedCrossRef
8.
go back to reference Hoang JK, Wang C, Frush DP et al (2013) Estimation of radiation exposure for brain perfusion CT: standard protocol compared with deviations in protocol. AJR Am J Roentgenol 201:W730–W734PubMedCrossRef Hoang JK, Wang C, Frush DP et al (2013) Estimation of radiation exposure for brain perfusion CT: standard protocol compared with deviations in protocol. AJR Am J Roentgenol 201:W730–W734PubMedCrossRef
10.
go back to reference Brenner DJ, Hall EJ (2007) Computed tomography–an increasing source of radiation exposure. N Engl J Med 357:2277–2284PubMedCrossRef Brenner DJ, Hall EJ (2007) Computed tomography–an increasing source of radiation exposure. N Engl J Med 357:2277–2284PubMedCrossRef
11.
go back to reference Niesten JM, van der Schaaf IC, Riordan AJ et al (2014) Radiation dose reduction in cerebral CT perfusion imaging using iterative reconstruction. Eur Radiol 24:484–493PubMedCrossRef Niesten JM, van der Schaaf IC, Riordan AJ et al (2014) Radiation dose reduction in cerebral CT perfusion imaging using iterative reconstruction. Eur Radiol 24:484–493PubMedCrossRef
13.
go back to reference König M, Bültmann E, Bode-Schnurbus L, Koenen D, Mielke E, Heuser L (2007) Image quality in CT perfusion imaging of the brain. The role of iodine concentration. Eur Radiol 17:39–47PubMedCrossRef König M, Bültmann E, Bode-Schnurbus L, Koenen D, Mielke E, Heuser L (2007) Image quality in CT perfusion imaging of the brain. The role of iodine concentration. Eur Radiol 17:39–47PubMedCrossRef
14.
go back to reference Bodelle B, Klement D, Kerl JM et al (2013) 70 kV computed tomography of the thorax: valence for computer-assisted nodule evaluation and radiation dose - first clinical results. Acta Radiol 22 Bodelle B, Klement D, Kerl JM et al (2013) 70 kV computed tomography of the thorax: valence for computer-assisted nodule evaluation and radiation dose - first clinical results. Acta Radiol 22
15.
go back to reference Wintermark M, Maeder P, Thiran JP et al (2001) Quantitative assessment of regional cerebral blood flows by perfusion CT studies at low injection rates: a critical review of the underlying theoretical models. Eur Radiol 11:1220–1230PubMedCrossRef Wintermark M, Maeder P, Thiran JP et al (2001) Quantitative assessment of regional cerebral blood flows by perfusion CT studies at low injection rates: a critical review of the underlying theoretical models. Eur Radiol 11:1220–1230PubMedCrossRef
16.
go back to reference Kang KH, Kim HS, Kim SY (2008) Quantitative cerebrovascular reserve measured by acetazolamide-challenged dynamic ct perfusion in ischemic adult Moyamoya disease: initial experience with angiographic correlation. AJNR Am J Neuroradiol 29:1487–1493PubMedCrossRef Kang KH, Kim HS, Kim SY (2008) Quantitative cerebrovascular reserve measured by acetazolamide-challenged dynamic ct perfusion in ischemic adult Moyamoya disease: initial experience with angiographic correlation. AJNR Am J Neuroradiol 29:1487–1493PubMedCrossRef
17.
go back to reference Axel L (1983) Tissue mean transit time from dynamic computed tomography by a simple deconvolution technique. Investig Radiol 8:94–99CrossRef Axel L (1983) Tissue mean transit time from dynamic computed tomography by a simple deconvolution technique. Investig Radiol 8:94–99CrossRef
18.
go back to reference Tawfik AM, Kerl JM, Razek AA et al (2011) Image quality and radiation dose of dual-energy CT of the head and neck compared with a standard 120-kV acquisition. AJNR Am J Neuroradiol 32:1994–1999PubMedCrossRef Tawfik AM, Kerl JM, Razek AA et al (2011) Image quality and radiation dose of dual-energy CT of the head and neck compared with a standard 120-kV acquisition. AJNR Am J Neuroradiol 32:1994–1999PubMedCrossRef
19.
go back to reference Wintermark M, Maeder P, Verdun FR et al (2000) Using 80 kV versus 120 kV in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21:1881–1884PubMed Wintermark M, Maeder P, Verdun FR et al (2000) Using 80 kV versus 120 kV in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21:1881–1884PubMed
20.
go back to reference Tracers. In: Miles K, Dawson P, Blomley M (eds) Functional computed tomography. ISIS Medical Media, Oxford pp 29–45 Tracers. In: Miles K, Dawson P, Blomley M (eds) Functional computed tomography. ISIS Medical Media, Oxford pp 29–45
21.
go back to reference Koenig M, Bueltmann E, Hering von Diepenbroick V, Koenen D, Heuser L (2001) Increase of contrast signal by means of low-tube voltage in CT perfusion imaging of the brain. Radiology 221:346 Koenig M, Bueltmann E, Hering von Diepenbroick V, Koenen D, Heuser L (2001) Increase of contrast signal by means of low-tube voltage in CT perfusion imaging of the brain. Radiology 221:346
22.
go back to reference Zhang J, Wang J, Geng D, Li Y, Song D, Gu Y (2013) Whole-brain CT perfusion and CT angiography assessment of Moyamoya disease before and after surgical revascularization: preliminary study with 256-Slice CT. PLoS ONE 8:e57595PubMedCentralPubMedCrossRef Zhang J, Wang J, Geng D, Li Y, Song D, Gu Y (2013) Whole-brain CT perfusion and CT angiography assessment of Moyamoya disease before and after surgical revascularization: preliminary study with 256-Slice CT. PLoS ONE 8:e57595PubMedCentralPubMedCrossRef
23.
go back to reference Cohnen M, Wittsack HJ, Assadi S et al (2006) Radiation exposure of patients in comprehensive computed tomography of the head in acute stroke. AJNR Am J Neuroradiol 27:1741–1745PubMed Cohnen M, Wittsack HJ, Assadi S et al (2006) Radiation exposure of patients in comprehensive computed tomography of the head in acute stroke. AJNR Am J Neuroradiol 27:1741–1745PubMed
24.
go back to reference Sakamoto S, Ohba S, Shibukawa M, Kiura Y, Arita K, Kurisu K (2006) CT perfusion imaging for childhood moyamoya disease before and after surgical revascularization. Acta Neurochir (Wien) 148:77–81CrossRef Sakamoto S, Ohba S, Shibukawa M, Kiura Y, Arita K, Kurisu K (2006) CT perfusion imaging for childhood moyamoya disease before and after surgical revascularization. Acta Neurochir (Wien) 148:77–81CrossRef
25.
go back to reference Hoeffner EG, Case I, Jain R et al (2004) Cerebral perfusion CT: technique and clinical applications. Radiology 231:632–644PubMedCrossRef Hoeffner EG, Case I, Jain R et al (2004) Cerebral perfusion CT: technique and clinical applications. Radiology 231:632–644PubMedCrossRef
26.
go back to reference Frey GD, Rumboldt Z (2005) Radiation effects from perfusion CT. Radiology 234:638, author reply 638PubMedCrossRef Frey GD, Rumboldt Z (2005) Radiation effects from perfusion CT. Radiology 234:638, author reply 638PubMedCrossRef
27.
go back to reference Krissak R, Mistretta CA, Henzler T et al (2011) Noise reduction and image quality improvement of low dose and ultra low dose brain perfusion CT by HYPR-LR processing. PLoS ONE 6:e17098PubMedCentralPubMedCrossRef Krissak R, Mistretta CA, Henzler T et al (2011) Noise reduction and image quality improvement of low dose and ultra low dose brain perfusion CT by HYPR-LR processing. PLoS ONE 6:e17098PubMedCentralPubMedCrossRef
Metadata
Title
Improvement of image quality and radiation dose of CT perfusion of the brain by means of low-tube voltage (70 KV)
Authors
Zhen-lin Li
Hang Li
Kai Zhang
Wang-jiang Li
Xian Chen
Bin Wu
Bin Song
Publication date
01-08-2014
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 8/2014
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3247-1

Other articles of this Issue 8/2014

European Radiology 8/2014 Go to the issue