Published in: European Radiology 8/2009

01-08-2009 | Interventional

Radiation exposure of medical staff from interventional x-ray procedures: a multicentre study

Authors: Uwe Häusler, Renate Czarwinski, Gunnar Brix

Published in: European Radiology | Issue 8/2009

Login to get access

Abstract

The purpose of this study was to analyse the radiation exposure of medical staff from interventional x-ray procedures. Partial-body dose measurements were performed with thermoluminescent dosimeters (TLD) in 39 physicians and nine assistants conducting 73 interventional procedures of nine different types in 14 hospitals in Germany. Fluoroscopy time and the dose–area product (DAP) were recorded too. The median (maximum) equivalent body dose per procedure was 16 (2,500) µSv for an unshielded person; the partial-body dose per procedure was 2.8 (240) µSv to the eye lens, 4.1 (730) µSv to the thyroid, 44 (1,800) µSv to one of the feet and 75 (13,000) µSv to one of the hands. A weak correlation between fluoroscopy time or DAP and the mean TLD dose was observed. Generally, the doses were within an acceptable range from a radiation hygiene point of view. However, relatively high exposures were measured to the hand in some cases and could cause a partial-body dose above the annual dose limit of 500 mSv. Thus, the use of finger dosimeters is strongly recommended.
Literature
1.
go back to reference Felmlee JP, McGough PF, Morin RL, Classic KL (1991) Hand dose measurements in interventional radiology. Health Phys 60(2):265–267PubMedCrossRef Felmlee JP, McGough PF, Morin RL, Classic KL (1991) Hand dose measurements in interventional radiology. Health Phys 60(2):265–267PubMedCrossRef
2.
go back to reference Pecher G, Koenig H, Pecher S, Gosch D, Voigt P, Schulz HG (1998) Reduction of radiation exposure for patient and investigator in interventional radiography. Rofo 169(5):505–509. (in German)PubMed Pecher G, Koenig H, Pecher S, Gosch D, Voigt P, Schulz HG (1998) Reduction of radiation exposure for patient and investigator in interventional radiography. Rofo 169(5):505–509. (in German)PubMed
3.
go back to reference Vañó E, González L, Guibelalde E, Fernández JM, Ten JI (1998) Radiation exposure to medical staff in interventional and cardiac radiology. Br J Radiol 71(849):954–960PubMed Vañó E, González L, Guibelalde E, Fernández JM, Ten JI (1998) Radiation exposure to medical staff in interventional and cardiac radiology. Br J Radiol 71(849):954–960PubMed
4.
go back to reference Servomaa A, Karppinen J (2001) The dose-area product and assessment of the occupational dose in interventional radiology. Radiat Prot Dosimetry 96(1–3):235–236PubMed Servomaa A, Karppinen J (2001) The dose-area product and assessment of the occupational dose in interventional radiology. Radiat Prot Dosimetry 96(1–3):235–236PubMed
5.
go back to reference Delichas M, Psarrakos K, Molyvda-Athanassopoulou E, Giannoglou G, Sioundas A, Hatziioannou K, Papanastassiou E (2003) Radiation exposure to cardiologists performing interventional cardiology procedures. Eur J Radiol 48(3):268–273PubMedCrossRef Delichas M, Psarrakos K, Molyvda-Athanassopoulou E, Giannoglou G, Sioundas A, Hatziioannou K, Papanastassiou E (2003) Radiation exposure to cardiologists performing interventional cardiology procedures. Eur J Radiol 48(3):268–273PubMedCrossRef
6.
go back to reference Whitby M, Martin CJ (2003) Radiation doses to the legs of radiologists performing interventional procedures: are they a cause for concern? Br J Radiol 76(905):321–327PubMedCrossRef Whitby M, Martin CJ (2003) Radiation doses to the legs of radiologists performing interventional procedures: are they a cause for concern? Br J Radiol 76(905):321–327PubMedCrossRef
7.
go back to reference Duggan L, Hood C, Warren-Forward H, Haque M, Kron T (2004) Variations in dose response with x-ray energy of LiF:Mg,Cu,P thermoluminescence dosimeters: implications for clinical dosimetry. Phys Med Biol 49(17):3831–3845PubMedCrossRef Duggan L, Hood C, Warren-Forward H, Haque M, Kron T (2004) Variations in dose response with x-ray energy of LiF:Mg,Cu,P thermoluminescence dosimeters: implications for clinical dosimetry. Phys Med Biol 49(17):3831–3845PubMedCrossRef
8.
go back to reference Whitby M, Martin CJ (2005) A study of the distribution of dose across the hands of interventional radiologists and cardiologists. Br J Radiol 78(927):219–229PubMedCrossRef Whitby M, Martin CJ (2005) A study of the distribution of dose across the hands of interventional radiologists and cardiologists. Br J Radiol 78(927):219–229PubMedCrossRef
9.
go back to reference Hidajat N, Wust P, Kreuschner M, Felix R, Schröder RJ (2006) Radiation risks for the radiologist performing transjugular intrahepatic portosystemic shunt (TIPS). Br J Radiol 79(942):483–486PubMedCrossRef Hidajat N, Wust P, Kreuschner M, Felix R, Schröder RJ (2006) Radiation risks for the radiologist performing transjugular intrahepatic portosystemic shunt (TIPS). Br J Radiol 79(942):483–486PubMedCrossRef
10.
go back to reference Shortt CP, Al-Hashimi H, Malone L, Lee MJ (2007) Staff radiation doses to the lower extremities in interventional radiology. Cardiovasc Intervent Radiol 30(6):1206–1209PubMedCrossRef Shortt CP, Al-Hashimi H, Malone L, Lee MJ (2007) Staff radiation doses to the lower extremities in interventional radiology. Cardiovasc Intervent Radiol 30(6):1206–1209PubMedCrossRef
11.
go back to reference Oonsiri S, Jumpangern C, Sanghangthum T, Krisanachinda A, Suriyapee S (2007) Radiation dose to medical staff in interventional radiology. J Med Assoc Thai 90(4):823–828PubMed Oonsiri S, Jumpangern C, Sanghangthum T, Krisanachinda A, Suriyapee S (2007) Radiation dose to medical staff in interventional radiology. J Med Assoc Thai 90(4):823–828PubMed
12.
go back to reference von Boetticher H, Lachmund J, Hoffmann W (2007) How conservative is routine personal dosimetry monitoring in diagnostic radiology? Rofo 179(7):728–732, (in German) von Boetticher H, Lachmund J, Hoffmann W (2007) How conservative is routine personal dosimetry monitoring in diagnostic radiology? Rofo 179(7):728–732, (in German)
13.
go back to reference Siiskonen T, Tapiovaara M, Kosunen A, Lehtinen M, Vartiainen E (2007) Monte Carlo simulations of occupational radiation doses in interventional radiology. Br J Radiol 80(954):460–468PubMedCrossRef Siiskonen T, Tapiovaara M, Kosunen A, Lehtinen M, Vartiainen E (2007) Monte Carlo simulations of occupational radiation doses in interventional radiology. Br J Radiol 80(954):460–468PubMedCrossRef
14.
go back to reference Kim KP, Miller DL, Balter S, Kleinerman RA, Linet MS, Kwon D, Simon SL (2008) Occupational radiation doses to operators performing cardiac catheterization procedures. Health Phys 94(3):211–227PubMedCrossRef Kim KP, Miller DL, Balter S, Kleinerman RA, Linet MS, Kwon D, Simon SL (2008) Occupational radiation doses to operators performing cardiac catheterization procedures. Health Phys 94(3):211–227PubMedCrossRef
15.
go back to reference Clerinx P, Buls N, Bosmans H, de Mey J (2008) Double-dosimetry algorithm for workers in interventional radiology. Radiat Prot Dosimetry 129(1–3):321–327PubMedCrossRef Clerinx P, Buls N, Bosmans H, de Mey J (2008) Double-dosimetry algorithm for workers in interventional radiology. Radiat Prot Dosimetry 129(1–3):321–327PubMedCrossRef
16.
go back to reference Padovani R, Vano E, Trianni A, Bokou C, Bosmans H, Bor D, Jankowski J, Torbica P, Kepler K, Dowling A, Milu C, Tsapaki V, Salat D, Vassileva J, Faulkner K (2008) Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry 129(1–3):104–107PubMedCrossRef Padovani R, Vano E, Trianni A, Bokou C, Bosmans H, Bor D, Jankowski J, Torbica P, Kepler K, Dowling A, Milu C, Tsapaki V, Salat D, Vassileva J, Faulkner K (2008) Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry 129(1–3):104–107PubMedCrossRef
17.
go back to reference von Boetticher H, Lachmund J, Hoffmann W (2008) Effective dose estimation in diagnostic radiology with two dosimeters: impact of the 2007 recommendations of the ICRP. Health Phys 95(3):337–340CrossRef von Boetticher H, Lachmund J, Hoffmann W (2008) Effective dose estimation in diagnostic radiology with two dosimeters: impact of the 2007 recommendations of the ICRP. Health Phys 95(3):337–340CrossRef
18.
go back to reference Vañó E, Gonzalez L, Fernández JM, Haskal ZJ (2008) Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology 248(3):945–953PubMedCrossRef Vañó E, Gonzalez L, Fernández JM, Haskal ZJ (2008) Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology 248(3):945–953PubMedCrossRef
19.
go back to reference Interventionelle Radiologie – Empfehlung der Strahlenschutzkommission (1997) Berichte der Strahlenschutzkommission des Bundesministeriums für Umwelt, Naturschutz und Reaktorsicherheit. Volume 9, chapt 2, p 2 Interventionelle Radiologie – Empfehlung der Strahlenschutzkommission (1997) Berichte der Strahlenschutzkommission des Bundesministeriums für Umwelt, Naturschutz und Reaktorsicherheit. Volume 9, chapt 2, p 2
20.
go back to reference International Commission on Radiological Protection (2001) Publication 85: avoidance of radiation injuries from medical interventional procedures. Pergamon, Oxford International Commission on Radiological Protection (2001) Publication 85: avoidance of radiation injuries from medical interventional procedures. Pergamon, Oxford
21.
go back to reference ISO 4037-1 (1997) X and gamma reference radiation for calibrating dosemeters and dose rate meters and for determining their response as a function of photon energy—part 1: radiation characteristics and production methods ISO 4037-1 (1997) X and gamma reference radiation for calibrating dosemeters and dose rate meters and for determining their response as a function of photon energy—part 1: radiation characteristics and production methods
22.
go back to reference International Commission on Radiological Protection (2008) Publication 103: recommendations of the ICRP. Ann ICRP 37:2–4 International Commission on Radiological Protection (2008) Publication 103: recommendations of the ICRP. Ann ICRP 37:2–4
23.
go back to reference Eder H, Panzer W, Schöfer H (2005) Is the lead-equivalent suited for rating protection properties of lead-free radiation protective clothing? Rofo 177(3):399–404. (in German)PubMed Eder H, Panzer W, Schöfer H (2005) Is the lead-equivalent suited for rating protection properties of lead-free radiation protective clothing? Rofo 177(3):399–404. (in German)PubMed
Metadata
Title
Radiation exposure of medical staff from interventional x-ray procedures: a multicentre study
Authors
Uwe Häusler
Renate Czarwinski
Gunnar Brix
Publication date
01-08-2009
Publisher
Springer-Verlag
Published in
European Radiology / Issue 8/2009
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-009-1388-4

Other articles of this Issue 8/2009

European Radiology 8/2009 Go to the issue