Published in: European Radiology 7/2008

01-07-2008 | Musculoskeletal

Real-time MR fluoroscopy-navigated lumbar facet joint injections: feasibility and technical properties

Authors: J. Fritz, S. Clasen, A. Boss, C. Thomas, C. W. König, C. D. Claussen, P. L. Pereira

Published in: European Radiology | Issue 7/2008

Login to get access

Abstract

We prospectively evaluated the feasibility and technical properties of a dedicated interventional magnetic resonance (MR) imaging protocol for near-real-time MR fluoroscopy-guided bilateral lumbar facet joint injections. A total of 44 facet joint injections were performed in 22 patients using a C-shaped open 0.2-T MR imaging system (Magnetom Concerto, Siemens Medical Solutions, Erlangen, Germany). A T1/T2*-weighted fast-imaging-with-steady-precession (FISP) sequence with an end-to-end latency of 1.2 s facilitated sufficient near real-time MR imaging guidance in all cases. A T1-weighted two dimensional fast-low-angle-shot (FLASH2D) MR sequence identified final needle tip location. Different angles of the needle path had only minimal influence on the appearance of the needle artifact produced by both sequences, resulting in a symmetrical needle tip artifact. The joint cavity was successfully punctured in 79.5% (35/44) of joints, which was followed by intra-articular fluid accumulation in 75% (33/44). Inaccessible joints demonstrated a significantly (p=0.044) higher number of posterior osteophytes (66.7%, 6/9 joints) compared to accessible joints (26%, 9/35 joints). No complications occurred. Table time showed significant shortening over time with average table time of 33 (21–68.5) min. We conclude that MR fluoroscopy-navigated lumbar facet joint injections are feasible and safe.
Appendix
Available only for authorised users
Literature
1.
go back to reference Carrino JA, Morrison WB, Parker L, Schweitzer ME, Levin DC, Sunshine JH (2002) Spinal injection procedures: volume, provider distribution, and reimbursement in the US medicare population from 1993 to 1999. Radiology 225:723–729CrossRefPubMed Carrino JA, Morrison WB, Parker L, Schweitzer ME, Levin DC, Sunshine JH (2002) Spinal injection procedures: volume, provider distribution, and reimbursement in the US medicare population from 1993 to 1999. Radiology 225:723–729CrossRefPubMed
2.
go back to reference Bogduk N (1997) International Spinal Injection Society guidelines for the performance of spinal injection procedures. Part 1: Zygapophysial joint blocks. Clin J Pain 13:285–302CrossRefPubMed Bogduk N (1997) International Spinal Injection Society guidelines for the performance of spinal injection procedures. Part 1: Zygapophysial joint blocks. Clin J Pain 13:285–302CrossRefPubMed
3.
go back to reference Gorbach C, Schmid MR, Elfering A, Hodler J, Boos N (2006) Therapeutic efficacy of facet joint blocks. AJR Am J Roentgenol 186:1228–1233CrossRefPubMed Gorbach C, Schmid MR, Elfering A, Hodler J, Boos N (2006) Therapeutic efficacy of facet joint blocks. AJR Am J Roentgenol 186:1228–1233CrossRefPubMed
4.
go back to reference Murtagh FR (1988) Computed tomography and fluoroscopy guided anesthesia and steroid injection in facet syndrome. Spine 13:686–689PubMedCrossRef Murtagh FR (1988) Computed tomography and fluoroscopy guided anesthesia and steroid injection in facet syndrome. Spine 13:686–689PubMedCrossRef
5.
go back to reference Meleka S, Patra A, Minkoff E, Murphy K (2005) Value of CT fluoroscopy for lumbar facet blocks. AJNR Am J Neuroradiol 26:1001–1003PubMed Meleka S, Patra A, Minkoff E, Murphy K (2005) Value of CT fluoroscopy for lumbar facet blocks. AJNR Am J Neuroradiol 26:1001–1003PubMed
6.
go back to reference Hilfiker PR, Weishaupt D, Schmid M, Dubno B, Hodler J, Debatin JF (1999) Real-time MR-guided joint puncture and arthrography: preliminary results. Eur Radiol 9:201–204CrossRefPubMed Hilfiker PR, Weishaupt D, Schmid M, Dubno B, Hodler J, Debatin JF (1999) Real-time MR-guided joint puncture and arthrography: preliminary results. Eur Radiol 9:201–204CrossRefPubMed
7.
go back to reference König CW, Schott UG, Pereira PL et al (2002) MR-guided lumbar sympathicolysis. Eur Radiol 12:1388–1393CrossRefPubMed König CW, Schott UG, Pereira PL et al (2002) MR-guided lumbar sympathicolysis. Eur Radiol 12:1388–1393CrossRefPubMed
8.
go back to reference Schulz T, Puccini S, Schneider JP, Kahn T (2004) Interventional and intraoperative MR: review and update of techniques and clinical experience. Eur Radiol 14:2212–2227CrossRefPubMed Schulz T, Puccini S, Schneider JP, Kahn T (2004) Interventional and intraoperative MR: review and update of techniques and clinical experience. Eur Radiol 14:2212–2227CrossRefPubMed
9.
go back to reference Ojala R, Vahala E, Karppinen J et al (2000) Nerve root infiltration of the first sacral root with MRI guidance. J Magn Reson Imaging 12:556–561CrossRefPubMed Ojala R, Vahala E, Karppinen J et al (2000) Nerve root infiltration of the first sacral root with MRI guidance. J Magn Reson Imaging 12:556–561CrossRefPubMed
10.
go back to reference Pereira PL, Gunaydin I, Trubenbach J et al (2000) Interventional MR imaging for injection of sacroiliac joints in patients with sacroiliitis. AJR Am J Roentgenol 175:265–266PubMed Pereira PL, Gunaydin I, Trubenbach J et al (2000) Interventional MR imaging for injection of sacroiliac joints in patients with sacroiliitis. AJR Am J Roentgenol 175:265–266PubMed
11.
go back to reference Sequeiros RB, Ojala RO, Klemola R, Vaara TJ, Jyrkinen L, Tervonen OA (2002) MRI-guided periradicular nerve root infiltration therapy in low-field (0.23-T) MRI system using optical instrument tracking. Eur Radiol 12:1331–1337CrossRefPubMed Sequeiros RB, Ojala RO, Klemola R, Vaara TJ, Jyrkinen L, Tervonen OA (2002) MRI-guided periradicular nerve root infiltration therapy in low-field (0.23-T) MRI system using optical instrument tracking. Eur Radiol 12:1331–1337CrossRefPubMed
12.
go back to reference Sequeiros RB, Klemola R, Ojala R, Jyrkinen L, Vaara T, Tervonen O (2003) Percutaneous MR-guided discography in a low-field system using optical instrument tracking: a feasibility study. J Magn Reson Imaging 17:214–219CrossRefPubMed Sequeiros RB, Klemola R, Ojala R, Jyrkinen L, Vaara T, Tervonen O (2003) Percutaneous MR-guided discography in a low-field system using optical instrument tracking: a feasibility study. J Magn Reson Imaging 17:214–219CrossRefPubMed
13.
go back to reference Lewin JS, Petersilge CA, Hatem SF et al (1998) Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system. AJR Am J Roentgenol 170:1593–1601PubMed Lewin JS, Petersilge CA, Hatem SF et al (1998) Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system. AJR Am J Roentgenol 170:1593–1601PubMed
14.
go back to reference Zangos S, Eichler K, Engelmann K et al (2005) MR-guided transgluteal biopsies with an open low-field system in patients with clinically suspected prostate cancer: technique and preliminary results. Eur Radiol 15:174–182CrossRefPubMed Zangos S, Eichler K, Engelmann K et al (2005) MR-guided transgluteal biopsies with an open low-field system in patients with clinically suspected prostate cancer: technique and preliminary results. Eur Radiol 15:174–182CrossRefPubMed
15.
go back to reference Zangos S, Eichler K, Wetter A et al (2006) MR-guided biopsies of lesions in the retroperitoneal space: technique and results. Eur Radiol 16:307–312CrossRefPubMed Zangos S, Eichler K, Wetter A et al (2006) MR-guided biopsies of lesions in the retroperitoneal space: technique and results. Eur Radiol 16:307–312CrossRefPubMed
16.
go back to reference Duerk JL, Lewin JS, Wendt M, Petersilge C (1998) Remember true FISP? A high SNR, near 1-second imaging method for T2-like contrast in interventional MRI at.2 T. J Magn Reson Imaging 8:203–208CrossRefPubMed Duerk JL, Lewin JS, Wendt M, Petersilge C (1998) Remember true FISP? A high SNR, near 1-second imaging method for T2-like contrast in interventional MRI at.2 T. J Magn Reson Imaging 8:203–208CrossRefPubMed
17.
go back to reference D’Aprile P, Tarantino A, Jinkins JR, Brindicci D (2007) The value of fat saturation sequences and contrast medium administration in MRI of degenerative disease of the posterior/perispinal elements of the lumbosacral spine. Eur Radiol 17:523–531CrossRefPubMed D’Aprile P, Tarantino A, Jinkins JR, Brindicci D (2007) The value of fat saturation sequences and contrast medium administration in MRI of degenerative disease of the posterior/perispinal elements of the lumbosacral spine. Eur Radiol 17:523–531CrossRefPubMed
18.
19.
go back to reference Boyle GE, Ahern M, Cooke J, Sheehy NP, Meaney JF (2006) An interactive taxonomy of MR imaging sequences. Radiographics 26:e24CrossRefPubMed Boyle GE, Ahern M, Cooke J, Sheehy NP, Meaney JF (2006) An interactive taxonomy of MR imaging sequences. Radiographics 26:e24CrossRefPubMed
20.
go back to reference Lewin JS, Duerk JL, Jain VR, Petersilge CA, Chao CP, Haaga JR (1996) Needle localization in MR-guided biopsy and aspiration: effects of field strength, sequence design, and magnetic field orientation. AJR Am J Roentgenol 166:1337–1345PubMed Lewin JS, Duerk JL, Jain VR, Petersilge CA, Chao CP, Haaga JR (1996) Needle localization in MR-guided biopsy and aspiration: effects of field strength, sequence design, and magnetic field orientation. AJR Am J Roentgenol 166:1337–1345PubMed
21.
go back to reference Lynch MC, Taylor JF (1986) Facet joint injection for low back pain. A clinical study. J Bone Joint Surg Br 68:138–141PubMed Lynch MC, Taylor JF (1986) Facet joint injection for low back pain. A clinical study. J Bone Joint Surg Br 68:138–141PubMed
Metadata
Title
Real-time MR fluoroscopy-navigated lumbar facet joint injections: feasibility and technical properties
Authors
J. Fritz
S. Clasen
A. Boss
C. Thomas
C. W. König
C. D. Claussen
P. L. Pereira
Publication date
01-07-2008
Publisher
Springer-Verlag
Published in
European Radiology / Issue 7/2008
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-008-0890-4

Other articles of this Issue 7/2008

European Radiology 7/2008 Go to the issue