01-07-2007 | Breast

Silicone-selective multishot echo-planar imaging for rapid MRI survey of breast implants

Authors: Yasuo Amano, Ritsu Aoki, Shinichiro Kumita, Tatsuo Kumazaki

Published in: European Radiology | Issue 7/2007

Login to get access

Abstract

The purpose of this study was to assess the usefulness of silicone-selective multishot echo-planar imaging (EPI) for the rapid magnetic resonance imaging (MRI) survey of breast implants. Twenty patients with bilateral breast implants underwent MRI. The use of inversion recovery and magnetization transfer pulses led to silicone-selective images. The rapid MRI survey required only 43 s, which accurately distinguished between silicone and non-silicone materials in the implants, and therefore induced the final MRI sequences appropriate for the detailed characterization of the implants. In 5 of the 20 patients, the rapid MRI survey showed implanted materials that were not indicated from clinical information. This silicone-selective multishot EPI allows a rapid survey of breast implants, which is useful to avoid unnecessary sequences in these patients.
Literature
1.
go back to reference Robinson OG Jr, Bradley EL, Wilson DS (1995) Analysis of explanted silicone implants: a report of 300 patients. Ann Plast Surg 34:1–7PubMedCrossRef Robinson OG Jr, Bradley EL, Wilson DS (1995) Analysis of explanted silicone implants: a report of 300 patients. Ann Plast Surg 34:1–7PubMedCrossRef
2.
go back to reference Gabriel SE, O’Fallon WM, Kurland LT, Beard CM, Woods JE, Melton LJ III (1994) Risk of connective-tissue diseases and other disorders after breast implantation. N Engl J Med 330:1697–1702PubMedCrossRef Gabriel SE, O’Fallon WM, Kurland LT, Beard CM, Woods JE, Melton LJ III (1994) Risk of connective-tissue diseases and other disorders after breast implantation. N Engl J Med 330:1697–1702PubMedCrossRef
3.
go back to reference Huch RA, Kunzi W, Debatin JF, Wiesner W, Krestin GP (1998) MR imaging of the augmented breast. Eur Radiol 8:371–376PubMedCrossRef Huch RA, Kunzi W, Debatin JF, Wiesner W, Krestin GP (1998) MR imaging of the augmented breast. Eur Radiol 8:371–376PubMedCrossRef
4.
go back to reference Brown SL, Silverman BG, Berg WA (1997) Rupture of silicone-gel breast implants: causes, sequelae, and diagnosis. Lancet 350:1531–1537PubMedCrossRef Brown SL, Silverman BG, Berg WA (1997) Rupture of silicone-gel breast implants: causes, sequelae, and diagnosis. Lancet 350:1531–1537PubMedCrossRef
5.
go back to reference Brown SL, Pennello G, Berg WA, Soo MS, Middleton MS (2001) Silicone gel breast implant rupture, extracapsular silicone, and health status in a population of women. J Rheumatol 28:996–1003PubMed Brown SL, Pennello G, Berg WA, Soo MS, Middleton MS (2001) Silicone gel breast implant rupture, extracapsular silicone, and health status in a population of women. J Rheumatol 28:996–1003PubMed
6.
go back to reference Everson LI, Parantainen H, Detlie T, Stillman AE, Olson PN, Landis G, Foshager MC, Cunningham B, Griffiths HJ (1994) Diagnosis of breast implant rupture: imaging findings and relative efficacies of imaging techniques. AJR Am J Roentgenol 163:57–60PubMed Everson LI, Parantainen H, Detlie T, Stillman AE, Olson PN, Landis G, Foshager MC, Cunningham B, Griffiths HJ (1994) Diagnosis of breast implant rupture: imaging findings and relative efficacies of imaging techniques. AJR Am J Roentgenol 163:57–60PubMed
7.
go back to reference Ikeda DM, Borofsky HB, Herfkens RJ, Sawyer-Glover AM, Birdwell RL, Glover GH (1999) Silicone breast implant rupture: pitfalls of magnetic resonance imaging and relative efficacies of magnetic resonance, mammography, and ultrasound. Plast Reconstr Surg 104:2054–2061PubMedCrossRef Ikeda DM, Borofsky HB, Herfkens RJ, Sawyer-Glover AM, Birdwell RL, Glover GH (1999) Silicone breast implant rupture: pitfalls of magnetic resonance imaging and relative efficacies of magnetic resonance, mammography, and ultrasound. Plast Reconstr Surg 104:2054–2061PubMedCrossRef
8.
go back to reference Monticciolo DL, Nelson RC, Dixon WT, Bostwick J III, Mukundan S, Hester TR (1994) MR detection of leakage from silicone breast implants: value of a silicone-selective pulse sequence. AJR Am J Roentgenol 163:51–56PubMed Monticciolo DL, Nelson RC, Dixon WT, Bostwick J III, Mukundan S, Hester TR (1994) MR detection of leakage from silicone breast implants: value of a silicone-selective pulse sequence. AJR Am J Roentgenol 163:51–56PubMed
9.
go back to reference Soo MS, Kornguth PJ, Walsh R, Elenberger CD, Georgiade GS (1996) Complex radial folds versus subtle signs of intracapsular rupture of breast implants: MR findings with surgical correlation. AJR Am J Roentgenol 166:1421–1427PubMed Soo MS, Kornguth PJ, Walsh R, Elenberger CD, Georgiade GS (1996) Complex radial folds versus subtle signs of intracapsular rupture of breast implants: MR findings with surgical correlation. AJR Am J Roentgenol 166:1421–1427PubMed
10.
go back to reference McKinnon GC (1993) Ultrafast interleaved gradient-echo-planar imaging on a standard scanner. Magn Reson Med 30:609–616PubMedCrossRef McKinnon GC (1993) Ultrafast interleaved gradient-echo-planar imaging on a standard scanner. Magn Reson Med 30:609–616PubMedCrossRef
Metadata
Title
Silicone-selective multishot echo-planar imaging for rapid MRI survey of breast implants
Authors
Yasuo Amano
Ritsu Aoki
Shinichiro Kumita
Tatsuo Kumazaki
Publication date
01-07-2007
Publisher
Springer-Verlag
Published in
European Radiology / Issue 7/2007
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0533-6

Other articles of this Issue 7/2007

European Radiology 7/2007 Go to the issue