01-03-2006 | Cardiac

Assessment of left ventricular myocardial function using 16-slice multidetector-row computed tomography: comparison with magnetic resonance imaging and echocardiography

Authors: Martin Heuschmid, Julia K. Rothfuss, Stephen Schroeder, Michael Fenchel, Norbert Stauder, Christof Burgstahler, Andreas Franow, Ronald S. Kuzo, Axel Kuettner, Stephan Miller, Claus D. Claussen, Andreas F. Kopp

Published in: European Radiology | Issue 3/2006

Login to get access

Abstract

Objective

To assess functional parameters using multidetector-row computed tomography (MDCT) and echocardiography and to compare the results with magnetic resonance imaging (MRI).

Materials and methods

End-diastolic-volume (EDV), end-systolic-volume (ESV), stroke-volume (SV), ejection-fraction (EF), and myocardial mass (MM) were calculated based on CT data sets from 52 patients. Echocardiography was performed in 24 of the 52 patients. The results from MDCT and echocardiography were compared with MRI.

Results

A strong correlation between MDCT and MRI (r=0.66–0.90) was found for all parameters. Echocardiography revealed a low or moderate correlation (0.05–0.59). Compared to MRI the average differences with MDCT were for EDV 15.1 ml, ESV 10.6 ml, SV 4.5 ml, EF 1.8%, and MM 8.2 g, for EDV determined by echocardiography 36.2 ml, ESV 6.8 ml, and EF 13.9%. Bland-Altman analysis revealed acceptable limits of agreement between MRI and MDCT.

Conclusions

MDCT enables reliable quantification of left ventricular function. Echocardiography was found to have only a moderate agreement of functional parameters with MRI.
Literature
1.
go back to reference Moise A, Bourassa MG, Theroux P, Taeymans Y, Pasternac A, Campeau L, Bois MA, Dyrda I, David PR (1985) Prognostic significance of progression of coronary artery disease. Am J Cardiol 55:941–946CrossRefPubMed Moise A, Bourassa MG, Theroux P, Taeymans Y, Pasternac A, Campeau L, Bois MA, Dyrda I, David PR (1985) Prognostic significance of progression of coronary artery disease. Am J Cardiol 55:941–946CrossRefPubMed
2.
go back to reference Emond M, Mock MB, Davis KB, Fisher LD, Holmes DR Jr, Chaitman BR, Kaiser GC, Alderman E, Killip T (1994) Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) registry. Circulation 90:2645–2657PubMed Emond M, Mock MB, Davis KB, Fisher LD, Holmes DR Jr, Chaitman BR, Kaiser GC, Alderman E, Killip T (1994) Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) registry. Circulation 90:2645–2657PubMed
4.
go back to reference Hodges MC, Rollefson WA, Sample SA, Rubal BJ, Stajduhar KC (2001) Comparison of low-volume versus standard-volume left ventriculography. Catheter Cardiovasc Interv 52:314–319CrossRefPubMed Hodges MC, Rollefson WA, Sample SA, Rubal BJ, Stajduhar KC (2001) Comparison of low-volume versus standard-volume left ventriculography. Catheter Cardiovasc Interv 52:314–319CrossRefPubMed
5.
go back to reference Peshock RM, Willett DL, Sayad DE, Hundley WG, Chwialkowski MC, Clarke GD, Parkey RW (1996) Quantitative MR imaging of the heart. Magn Reson Imaging Clin N Am 4:287–305PubMed Peshock RM, Willett DL, Sayad DE, Hundley WG, Chwialkowski MC, Clarke GD, Parkey RW (1996) Quantitative MR imaging of the heart. Magn Reson Imaging Clin N Am 4:287–305PubMed
6.
go back to reference Rathi VK, Biedermann RW (2004) Imaging of ventricular function by cardiovascular magnetic resonance. Curr Cardiol Rep 6:55–61PubMedCrossRef Rathi VK, Biedermann RW (2004) Imaging of ventricular function by cardiovascular magnetic resonance. Curr Cardiol Rep 6:55–61PubMedCrossRef
7.
go back to reference Ohnesorge B, Flohr T, Becker C, Kopp AF, Schoepf UJ, Baum U, Knez A, Klingenbeck-Regn K, Reiser MF (2000) Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience. Radiology 217:564–571PubMed Ohnesorge B, Flohr T, Becker C, Kopp AF, Schoepf UJ, Baum U, Knez A, Klingenbeck-Regn K, Reiser MF (2000) Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience. Radiology 217:564–571PubMed
8.
go back to reference Achenbach S, Giesler T, Ropers D, Ulzheimer S, Derlien H, Schulte C, Wenkel E, Moshage W, Bautz W, Daniel WG, Kalender WA, Baum U (2001) Detection of coronary artery stenoses by contrast-enhanced, retrospectively electrocardiographyically-gated, multislice spiral computed tomography. Circulation 103:2535–2538PubMed Achenbach S, Giesler T, Ropers D, Ulzheimer S, Derlien H, Schulte C, Wenkel E, Moshage W, Bautz W, Daniel WG, Kalender WA, Baum U (2001) Detection of coronary artery stenoses by contrast-enhanced, retrospectively electrocardiographyically-gated, multislice spiral computed tomography. Circulation 103:2535–2538PubMed
9.
go back to reference Kopp AF, Kuttner A, Trabold T, Heuschmid M, Schroder S, Claussen CD (2003) MDCT: cardiology indications. Eur Radiol 13(Suppl 5):M102–M115PubMed Kopp AF, Kuttner A, Trabold T, Heuschmid M, Schroder S, Claussen CD (2003) MDCT: cardiology indications. Eur Radiol 13(Suppl 5):M102–M115PubMed
10.
go back to reference Kuettner A, Kopp AF, Schroeder S, Rieger T, Brunn J, Meisner C, Heuschmid M, Trabold T, Burgstahler C, Martensen J, Schoebel W, Selbmann HK, Claussen CD (2004) Diagnostic accuracy of multidetector computed tomography coronary angiography in patients with angiographically proven coronary artery disease. J Am Coll Cardiol 43:831–839CrossRefPubMed Kuettner A, Kopp AF, Schroeder S, Rieger T, Brunn J, Meisner C, Heuschmid M, Trabold T, Burgstahler C, Martensen J, Schoebel W, Selbmann HK, Claussen CD (2004) Diagnostic accuracy of multidetector computed tomography coronary angiography in patients with angiographically proven coronary artery disease. J Am Coll Cardiol 43:831–839CrossRefPubMed
11.
go back to reference Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B (2002) New technical developments in multislice CT. II. Sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac imaging. Fortschr Roentgenstr 174:1022–1027CrossRef Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B (2002) New technical developments in multislice CT. II. Sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac imaging. Fortschr Roentgenstr 174:1022–1027CrossRef
12.
go back to reference Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054CrossRefPubMed Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054CrossRefPubMed
13.
go back to reference Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666CrossRefPubMed Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666CrossRefPubMed
14.
go back to reference Hoffmann U, Moselewski F, Cury RC, Ferencik M, Jang IK, Diaz LJ, Abbara S, Brady TJ, Achenbach S (2004) Predictive value of 16-lice multidetector spiral computed tomography to detect significant obstructive coronary artery disease in patients at high risk for coronary artery disease: patient- versus segment-based analysis. Circulation 110:2638–2643PubMedCrossRef Hoffmann U, Moselewski F, Cury RC, Ferencik M, Jang IK, Diaz LJ, Abbara S, Brady TJ, Achenbach S (2004) Predictive value of 16-lice multidetector spiral computed tomography to detect significant obstructive coronary artery disease in patients at high risk for coronary artery disease: patient- versus segment-based analysis. Circulation 110:2638–2643PubMedCrossRef
15.
go back to reference Juergens KU, Grude M, Fallenberg EM, Opitz C, Wichter T, Heindel W, Fischbach R (2002) Using ECG-gated multidetector CT to evaluate global left ventricular myocardial function in patients with coronary artery disease. Am J Roentgenol 179:1545–1550 Juergens KU, Grude M, Fallenberg EM, Opitz C, Wichter T, Heindel W, Fischbach R (2002) Using ECG-gated multidetector CT to evaluate global left ventricular myocardial function in patients with coronary artery disease. Am J Roentgenol 179:1545–1550
16.
go back to reference Kachelriess M, Kalender WA (1998) Eletrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. Med Phys 25:2417–2431PubMedCrossRef Kachelriess M, Kalender WA (1998) Eletrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. Med Phys 25:2417–2431PubMedCrossRef
17.
go back to reference Weissler AM, Harris WS, Schoenfeld CD (1968) Systolic time intervals in heart failure in man. Circulation 37:149–159PubMed Weissler AM, Harris WS, Schoenfeld CD (1968) Systolic time intervals in heart failure in man. Circulation 37:149–159PubMed
18.
go back to reference Ritchie CJ, Godwin JD, Crawford CR, Stanford W, Anno H, Kim Y (1992) Minimum scan speeds for suppression of motion artefacts in CT. Radiology 185:37–42PubMed Ritchie CJ, Godwin JD, Crawford CR, Stanford W, Anno H, Kim Y (1992) Minimum scan speeds for suppression of motion artefacts in CT. Radiology 185:37–42PubMed
19.
go back to reference Mahnken AH, Suentrup E, Wildberger JE, Heuschmid M, Niethammer M, Sinha AM, Flohr T, Buecker A, Guenther RW (2003) Quantification of cardiac function with multislice spiral CT unsing retrospective ECG-gating: comparison with MRI. Fortschr Roentgenstr 175:83–88CrossRef Mahnken AH, Suentrup E, Wildberger JE, Heuschmid M, Niethammer M, Sinha AM, Flohr T, Buecker A, Guenther RW (2003) Quantification of cardiac function with multislice spiral CT unsing retrospective ECG-gating: comparison with MRI. Fortschr Roentgenstr 175:83–88CrossRef
20.
go back to reference Schalla S, Nagel E, Lehmkuhl H, Klein C, Bornstedt A, Schnackenburg B, Schneider, Fleck E (2001) Comparison of magnetic resonance real-time imaging of left ventricular function with conventional magnetic resonance imaging and echocardiography. Am J Cardiol 87:95–99CrossRefPubMed Schalla S, Nagel E, Lehmkuhl H, Klein C, Bornstedt A, Schnackenburg B, Schneider, Fleck E (2001) Comparison of magnetic resonance real-time imaging of left ventricular function with conventional magnetic resonance imaging and echocardiography. Am J Cardiol 87:95–99CrossRefPubMed
21.
go back to reference Thompson BH, Stanford W (1994) Evaluation of cardiac function with ultrafast computed tomography. Radiol Clin North Am 32:537–551PubMed Thompson BH, Stanford W (1994) Evaluation of cardiac function with ultrafast computed tomography. Radiol Clin North Am 32:537–551PubMed
22.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310 Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
23.
go back to reference Landis J, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis J, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
24.
go back to reference Hofmann T, Meinertz T, Kasper W, Geibel A, Zehender M, Hohnloser S, Stienen U, Treese N, Just H (1988) Mode of death in idiopathic dilated cardiomyopathy: a multivariate analysis of prognostic determinants. Am Heart J 116:1455–1463PubMedCrossRef Hofmann T, Meinertz T, Kasper W, Geibel A, Zehender M, Hohnloser S, Stienen U, Treese N, Just H (1988) Mode of death in idiopathic dilated cardiomyopathy: a multivariate analysis of prognostic determinants. Am Heart J 116:1455–1463PubMedCrossRef
25.
go back to reference de Simone G, Galderisi M (2002) Quantitation of left ventricular mass and function: balancing evidence with dreams. Ital Heart J 3:562–570PubMed de Simone G, Galderisi M (2002) Quantitation of left ventricular mass and function: balancing evidence with dreams. Ital Heart J 3:562–570PubMed
26.
go back to reference Gottdiener JS, Livengood SV, Meyer PS, Chase GA (1995) Should echocardiography be performed at assess effects of antihypertensive therapy? Test-retest reliability of echocardiography for measurement of left ventricular mass and function. J Am Coll Cardiol 25:424–430PubMedCrossRef Gottdiener JS, Livengood SV, Meyer PS, Chase GA (1995) Should echocardiography be performed at assess effects of antihypertensive therapy? Test-retest reliability of echocardiography for measurement of left ventricular mass and function. J Am Coll Cardiol 25:424–430PubMedCrossRef
27.
go back to reference Rumberger JA, Behrenbeck T, Bell MR, Breen JF, Johnston DL, Holmes DR Jr, Enriquez-Sarano M (1997) Determination of ventricular ejection fraction: a comparison of available imaging methods. Mayo Clin Proc 72:860–870PubMedCrossRef Rumberger JA, Behrenbeck T, Bell MR, Breen JF, Johnston DL, Holmes DR Jr, Enriquez-Sarano M (1997) Determination of ventricular ejection fraction: a comparison of available imaging methods. Mayo Clin Proc 72:860–870PubMedCrossRef
28.
go back to reference Setser RM, Fischer SE, Lorenz CH (2000) Quantification of left ventricular function with magnetic resonance images acquired in real time. J Magn Reson Imaging 12:430–438CrossRefPubMed Setser RM, Fischer SE, Lorenz CH (2000) Quantification of left ventricular function with magnetic resonance images acquired in real time. J Magn Reson Imaging 12:430–438CrossRefPubMed
29.
go back to reference Miller S, Simonetti OP, Carr J, Kramer U, Finn JP (2002) MR Imaging of the heart with cine true fast imaging with steady-state precession: influence of spatial and temporal resolution on left ventricular functional parameters. Radiology 223:263–269PubMedCrossRef Miller S, Simonetti OP, Carr J, Kramer U, Finn JP (2002) MR Imaging of the heart with cine true fast imaging with steady-state precession: influence of spatial and temporal resolution on left ventricular functional parameters. Radiology 223:263–269PubMedCrossRef
30.
go back to reference Juergens KU, Maintz D, Grude M, Boese JM, Heimes B, Fallenberg EM, Heindel W, Fischbach R (2005) Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements? Eur Radiol 15:111–117CrossRefPubMed Juergens KU, Maintz D, Grude M, Boese JM, Heimes B, Fallenberg EM, Heindel W, Fischbach R (2005) Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements? Eur Radiol 15:111–117CrossRefPubMed
31.
go back to reference Hong C, Becker CR, Huber A, Schoepf UJ, Ohnesorge B, Knez A, Brüning R, Reiser MF (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 200:712–717CrossRef Hong C, Becker CR, Huber A, Schoepf UJ, Ohnesorge B, Knez A, Brüning R, Reiser MF (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 200:712–717CrossRef
32.
go back to reference Boehm T, Alkadhi H, Roffi M, Willmann JK, Desbiolles LM, Marincek B, Wildermuth S (2004) Time effectiveness, oberserver-dependence, and accuracy of measurements of left ventricular ejection fraction using 4-channel MDCT. Fortschr Roentgenstr 176:529–537CrossRef Boehm T, Alkadhi H, Roffi M, Willmann JK, Desbiolles LM, Marincek B, Wildermuth S (2004) Time effectiveness, oberserver-dependence, and accuracy of measurements of left ventricular ejection fraction using 4-channel MDCT. Fortschr Roentgenstr 176:529–537CrossRef
33.
go back to reference Juergens KU, Fischbach R, Grude M, Wichter T, Fallenberg EM, Opitz C, Heindel WL (2002) Evaluation of left ventricular function by retrospectively ECG-gated multislice spiral CT in comparison to CINE magnetic resonance imaging. Eur Radiol 12(Suppl):S191 Juergens KU, Fischbach R, Grude M, Wichter T, Fallenberg EM, Opitz C, Heindel WL (2002) Evaluation of left ventricular function by retrospectively ECG-gated multislice spiral CT in comparison to CINE magnetic resonance imaging. Eur Radiol 12(Suppl):S191
34.
go back to reference Wintersperger B, Hundt W, Knez A, Thilo C, Huber A, Nikolaou K, Becker CR, Reiser MF (2002) Left ventricular systolic function assessed by ECG gated multirow-detector spiral computed tomography (MDCT): comparison to ventriculography. Eur Radiol 12(Suppl):S192PubMed Wintersperger B, Hundt W, Knez A, Thilo C, Huber A, Nikolaou K, Becker CR, Reiser MF (2002) Left ventricular systolic function assessed by ECG gated multirow-detector spiral computed tomography (MDCT): comparison to ventriculography. Eur Radiol 12(Suppl):S192PubMed
35.
go back to reference Herzog C, Mehtap A, Abolmaali N, Balzer JO, Schaller S, Vogel TJ (2002) Value and reproducibility of multidetector-row CT in the assessment of cardiac function. Eur Radiol 12(Suppl):S192PubMed Herzog C, Mehtap A, Abolmaali N, Balzer JO, Schaller S, Vogel TJ (2002) Value and reproducibility of multidetector-row CT in the assessment of cardiac function. Eur Radiol 12(Suppl):S192PubMed
36.
go back to reference Heuschmid M, Kuttner A, Schroder S, Trebar B, Burgstahler C, Mahnken A, Niethammer M, Trabold T, Kopp AF, Claussen CD (2003) Left ventricular functional parameters using ECG-gated multidetector spiral CT in comparison with invasive ventriculography. Fortschr Roentgenstr 175:1349–1354CrossRef Heuschmid M, Kuttner A, Schroder S, Trebar B, Burgstahler C, Mahnken A, Niethammer M, Trabold T, Kopp AF, Claussen CD (2003) Left ventricular functional parameters using ECG-gated multidetector spiral CT in comparison with invasive ventriculography. Fortschr Roentgenstr 175:1349–1354CrossRef
37.
go back to reference Schroeder S, Kopp AF, Kuettner A, Burgstahler C, Herdeg C, Heuschmid M, Baumbach A, Claussen CD, Karsch KR, Seipel L (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography: experience in 94 patients. Clin Imaging 26:106–111CrossRefPubMed Schroeder S, Kopp AF, Kuettner A, Burgstahler C, Herdeg C, Heuschmid M, Baumbach A, Claussen CD, Karsch KR, Seipel L (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography: experience in 94 patients. Clin Imaging 26:106–111CrossRefPubMed
38.
go back to reference Ehrhard K, Oberholzer K, Gast K, Mildenberger P, Kreitner KF, Thelen M (2002) Multi-slice CT (MSCT) in cardiac function imaging: threshold-value-supported 3D volume reconstructions to determine the left ventricular ejection fraction in comparison to MRI. Fortschr Roentgenstr 174:1566–1569CrossRef Ehrhard K, Oberholzer K, Gast K, Mildenberger P, Kreitner KF, Thelen M (2002) Multi-slice CT (MSCT) in cardiac function imaging: threshold-value-supported 3D volume reconstructions to determine the left ventricular ejection fraction in comparison to MRI. Fortschr Roentgenstr 174:1566–1569CrossRef
39.
go back to reference Becker C, Schaetzl M, Feist H, Baeuml A, Schoepf UJ, Michalski G, Lechl U, Hengge M, Braeuning R, Reiser M (1999) Assessment of the effective dose for routine protocols in conventional CT, electron beam CT and coronary angiography. Fortschr Roentgenstr 170:99–104 Becker C, Schaetzl M, Feist H, Baeuml A, Schoepf UJ, Michalski G, Lechl U, Hengge M, Braeuning R, Reiser M (1999) Assessment of the effective dose for routine protocols in conventional CT, electron beam CT and coronary angiography. Fortschr Roentgenstr 170:99–104
40.
go back to reference Hunold P, Vogt FM, Schmermund A, Debatin JF, Kerkhoff G, Budde T, Erbel R, Ewen K, Barkhausen J (2003) Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT. Radiology 226:145–152PubMedCrossRef Hunold P, Vogt FM, Schmermund A, Debatin JF, Kerkhoff G, Budde T, Erbel R, Ewen K, Barkhausen J (2003) Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT. Radiology 226:145–152PubMedCrossRef
41.
go back to reference Cohnen M, Poll L, Puttmann C, Ewen K, Modder U (2001) Radiation exposure in multi-slice CT of the heart. Fortschr Roentgenstr 173:295–299CrossRef Cohnen M, Poll L, Puttmann C, Ewen K, Modder U (2001) Radiation exposure in multi-slice CT of the heart. Fortschr Roentgenstr 173:295–299CrossRef
42.
go back to reference Trabold T, Buchgeister M, Kuettner A, Heuschmid M, Kopp AF, Schroeder S, Claussen CD (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Fortschr Roentgenstr 175:1051–1055CrossRef Trabold T, Buchgeister M, Kuettner A, Heuschmid M, Kopp AF, Schroeder S, Claussen CD (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Fortschr Roentgenstr 175:1051–1055CrossRef
43.
go back to reference Mahnken AH, Koss R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kuhl HP, Gunther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720CrossRefPubMed Mahnken AH, Koss R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kuhl HP, Gunther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720CrossRefPubMed
44.
go back to reference Coche E, Vlassenbroek A, Roelants V, D'hoore W, Verschuren F, Goncette L, Maldague B (2005) Evaluation of biventricular ejection fraction with ECG-gated 16-slice CT. preliminary findings in acute pulmonary embolism in comparison with radionuclide ventriculography. Eur Radiol 15:1432–1440CrossRefPubMed Coche E, Vlassenbroek A, Roelants V, D'hoore W, Verschuren F, Goncette L, Maldague B (2005) Evaluation of biventricular ejection fraction with ECG-gated 16-slice CT. preliminary findings in acute pulmonary embolism in comparison with radionuclide ventriculography. Eur Radiol 15:1432–1440CrossRefPubMed
45.
go back to reference Koch K, Oellig F, Oberholzer K, Bender P, Kunz P, Mildenberger P, Hake U, Kreinter KF, Thelen M (2005) Assessment of right ventricular function by 16-detector-row CT: comparison with magnetic resonance imaging. Eur Radiol 15:312–318CrossRefPubMed Koch K, Oellig F, Oberholzer K, Bender P, Kunz P, Mildenberger P, Hake U, Kreinter KF, Thelen M (2005) Assessment of right ventricular function by 16-detector-row CT: comparison with magnetic resonance imaging. Eur Radiol 15:312–318CrossRefPubMed
Metadata
Title
Assessment of left ventricular myocardial function using 16-slice multidetector-row computed tomography: comparison with magnetic resonance imaging and echocardiography
Authors
Martin Heuschmid
Julia K. Rothfuss
Stephen Schroeder
Michael Fenchel
Norbert Stauder
Christof Burgstahler
Andreas Franow
Ronald S. Kuzo
Axel Kuettner
Stephan Miller
Claus D. Claussen
Andreas F. Kopp
Publication date
01-03-2006
Publisher
Springer-Verlag
Published in
European Radiology / Issue 3/2006
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-0015-2

Other articles of this Issue 3/2006

European Radiology 3/2006 Go to the issue