01-06-2016 | Technical Note

Transjugular Intrahepatic Portosystemic Shunt Flow Reduction with Adjustable Polytetrafluoroethylene-Covered Balloon-Expandable Stents Using the “Sheath Control” Technique

Authors: Robert C. Blue, Grace C. Lo, Edward Kim, Rahul S. Patel, F. Scott Nowakowski, Robert A. Lookstein, Aaron M. Fischman

Published in: CardioVascular and Interventional Radiology | Issue 6/2016

Login to get access

Abstract

Purpose

A complication of transjugular intrahepatic portosystemic shunts (TIPS) placement is refractory portosystemic encephalopathy (PSE) often requiring TIPS reduction. We report the results of a “sheath control technique” utilizing constraining sheaths during deployment of polytetrafluoroethylene (PTFE)-covered balloon-expandable stents, minimizing stent migration, and providing additional procedural control.

Methods

TIPS reduction was performed in 10 consecutive patients for PSE using Atrium iCast covered stents (Atrium Maquet Getinge Group, Germany). Within the indwelling TIPS stent, a 9 mm × 59 mm iCast stent was deployed with 2 cm exposed from the sheath’s distal end and the majority of the stent within the sheath to create the distal hourglass shape. During balloon retraction, the stent was buttressed by the sheath. The proximal portion of the stent was angioplastied to complete the hourglass configuration, and the central portion of the stent was dilated to 5 mm. Demographics, pre- and post-procedure laboratory values, and outcomes were recorded.

Results

Ten patients underwent TIPS reduction with 100 % technical success. There was no stent migration during stent deployment. All patients experienced initial improvement of encephalopathy. One patient ultimately required complete TIPS occlusion for refractory PSE, and another developed TIPS occlusion 36 days post-procedure. There was no significant trend toward change in patients’ MELD scores immediately post-procedure or at 30 days (p = 0.46, p = 0.47, respectively).

Conclusion

TIPS reduction using Atrium iCast PTFE balloon-expandable stents using the “sheath control technique” is safe and effective, and minimizes the risk of stent migration.
Literature
1.
go back to reference Casadaban LC, Parvinian A, Minocha J, et al. Clearing the confusion over hepatic encephalopathy after TIPS creation: incidence, prognostic factors, and clinical outcomes. Dig Dis Sci. 2015;60:1059–66.CrossRefPubMed Casadaban LC, Parvinian A, Minocha J, et al. Clearing the confusion over hepatic encephalopathy after TIPS creation: incidence, prognostic factors, and clinical outcomes. Dig Dis Sci. 2015;60:1059–66.CrossRefPubMed
2.
go back to reference LaBerge JM, Somberg KA, Lake JR, et al. Two-year outcome following transjugular intrahepatic portosystemic shunt for variceal bleeding: results in 90 patients. Gastroenterology. 1995;108:1143–51.CrossRefPubMed LaBerge JM, Somberg KA, Lake JR, et al. Two-year outcome following transjugular intrahepatic portosystemic shunt for variceal bleeding: results in 90 patients. Gastroenterology. 1995;108:1143–51.CrossRefPubMed
3.
go back to reference Maleux G, Heye S, Verslype C, et al. Management of transjugular intrahepatic portosystemic shunt induced refractory hepatic encephalopathy with the parallel technique: results of a clinical follow-up study. J Vasc Interv Radiol. 2007;18:986–92 quiz 993.CrossRefPubMed Maleux G, Heye S, Verslype C, et al. Management of transjugular intrahepatic portosystemic shunt induced refractory hepatic encephalopathy with the parallel technique: results of a clinical follow-up study. J Vasc Interv Radiol. 2007;18:986–92 quiz 993.CrossRefPubMed
4.
go back to reference Vorwerk D, Gunther RW, Marschall HU, et al. Adaptable hourglass stent to reduce shunt flow after TIPS insertion. Rofo. 1995;162:436–8.CrossRefPubMed Vorwerk D, Gunther RW, Marschall HU, et al. Adaptable hourglass stent to reduce shunt flow after TIPS insertion. Rofo. 1995;162:436–8.CrossRefPubMed
5.
go back to reference Fanelli F, Salvatori FM, Rabuffi P, et al. Management of refractory hepatic encephalopathy after insertion of TIPS: long-term results of shunt reduction with hourglass-shaped balloon-expandable stent-graft. AJR Am J Roentgenol. 2009;193:1696–702.CrossRefPubMed Fanelli F, Salvatori FM, Rabuffi P, et al. Management of refractory hepatic encephalopathy after insertion of TIPS: long-term results of shunt reduction with hourglass-shaped balloon-expandable stent-graft. AJR Am J Roentgenol. 2009;193:1696–702.CrossRefPubMed
6.
go back to reference Saket RR, Sze DY, Razavi MK, et al. TIPS reduction with use of stents or stent-grafts. J Vasc Interv Radiol. 2004;15:745–51.CrossRefPubMed Saket RR, Sze DY, Razavi MK, et al. TIPS reduction with use of stents or stent-grafts. J Vasc Interv Radiol. 2004;15:745–51.CrossRefPubMed
7.
go back to reference Jacquier A, Vidal V, Monnet O, et al. A modified procedure for transjugular intrahepatic portosystemic shunt flow reduction. J Vasc Interv Radiol. 2006;17:1359–63.CrossRefPubMed Jacquier A, Vidal V, Monnet O, et al. A modified procedure for transjugular intrahepatic portosystemic shunt flow reduction. J Vasc Interv Radiol. 2006;17:1359–63.CrossRefPubMed
8.
go back to reference Kroma G, Lopera J, Cura M, et al. Transjugular intrahepatic portosystemic shunt flow reduction with adjustable polytetrafluoroethylene-covered balloon-expandable stents. J Vasc Interv Radiol. 2009;20:981–6.CrossRefPubMed Kroma G, Lopera J, Cura M, et al. Transjugular intrahepatic portosystemic shunt flow reduction with adjustable polytetrafluoroethylene-covered balloon-expandable stents. J Vasc Interv Radiol. 2009;20:981–6.CrossRefPubMed
9.
go back to reference Madoff DC, Wallace MJ. Reduced stents and stent-grafts for the management of hepatic encephalopathy after transjugular intrahepatic portosystemic shunt creation. Semin Intervent Radiol. 2005;22:316–28.CrossRefPubMedPubMedCentral Madoff DC, Wallace MJ. Reduced stents and stent-grafts for the management of hepatic encephalopathy after transjugular intrahepatic portosystemic shunt creation. Semin Intervent Radiol. 2005;22:316–28.CrossRefPubMedPubMedCentral
10.
go back to reference Quaretti P, Michieletti E, Rossi S. Successful treatment of TIPS-induced hepatic failure with an hourglass stent-graft: a simple new technique for reducing shunt flow. J Vasc Interv Radiol. 2001;12:887–90.CrossRefPubMed Quaretti P, Michieletti E, Rossi S. Successful treatment of TIPS-induced hepatic failure with an hourglass stent-graft: a simple new technique for reducing shunt flow. J Vasc Interv Radiol. 2001;12:887–90.CrossRefPubMed
11.
go back to reference Barrio J, Ripoll C, Banares R, et al. Comparison of transjugular intrahepatic portosystemic shunt dysfunction in PTFE-covered stent-grafts versus bare stents. Eur J Radiol. 2005;55:120–4.CrossRefPubMed Barrio J, Ripoll C, Banares R, et al. Comparison of transjugular intrahepatic portosystemic shunt dysfunction in PTFE-covered stent-grafts versus bare stents. Eur J Radiol. 2005;55:120–4.CrossRefPubMed
Metadata
Title
Transjugular Intrahepatic Portosystemic Shunt Flow Reduction with Adjustable Polytetrafluoroethylene-Covered Balloon-Expandable Stents Using the “Sheath Control” Technique
Authors
Robert C. Blue
Grace C. Lo
Edward Kim
Rahul S. Patel
F. Scott Nowakowski
Robert A. Lookstein
Aaron M. Fischman
Publication date
01-06-2016
Publisher
Springer US
Published in
CardioVascular and Interventional Radiology / Issue 6/2016
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-015-1249-2

Other articles of this Issue 6/2016

CardioVascular and Interventional Radiology 6/2016 Go to the issue