01-04-2013 | Clinical Investigation

Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure

Authors: Loredana D’Ercole, Pietro Quaretti, Nicola Cionfoli, Catherine Klersy, Milena Bocchiola, Giuseppe Rodolico, Andrea Azzaretti, Francesco Lisciandro, Tommaso Cascella, Federico Zappoli Thyrion

Published in: CardioVascular and Interventional Radiology | Issue 2/2013

Login to get access

Abstract

Purpose

To measure the maximum entrance skin dose (MESD) on patients undergoing carotid artery stenting (CAS) using embolic-protection devices, to analyze the dependence of dose and exposure parameters on anatomical, clinical, and technical factors affecting the procedure complexity, to obtain some local diagnostic reference levels (DRLs), and to evaluate whether overcoming DRLs is related to procedure complexity.

Materials and Methods

MESD were evaluated with radiochromic films in 31 patients (mean age 72 ± 7 years). Five of 33 (15 %) procedures used proximal EPD, and 28 of 33 (85 %) procedures used distal EPD. Local DRLs were derived from the recorded exposure parameters in 93 patients (65 men and 28 women, mean age 73 ± 9 years) undergoing 96 CAS with proximal (33 %) or distal (67 %) EPD. Four bilateral lesions were included.

Results

MESD values (mean 0.96 ± 0.42 Gy) were <2 Gy without relevant dependence on procedure complexity. Local DRL values for kerma area product (KAP), fluoroscopy time (FT), and number of frames (NFR) were 269 Gy cm2, 28 minutes, and 251, respectively. Only simultaneous bilateral treatment was associated with KAP (odds ratio [OR] 10.14, 95 % confidence interval [CI] 1–102.7, p < 0.05) and NFR overexposures (OR 10.8, 95 % CI 1.1–109.5, p < 0.05). Type I aortic arch decreased the risk of FT overexposure (OR 0.4, 95 % CI 0.1–0.9, p = 0.042), and stenosis ≥ 90 % increased the risk of NFR overexposure (OR 2.8, 95 % CI 1.1–7.4, p = 0.040). At multivariable analysis, stenosis ≥ 90 % (OR 2.8, 95 % CI 1.1–7.4, p = 0.040) and bilateral treatment (OR 10.8, 95 % CI 1.1–109.5, p = 0.027) were associated with overexposure for two or more parameters.

Conclusion

Skin doses are not problematic in CAS with EPD because these procedures rarely lead to doses >2 Gy.
Literature
1.
go back to reference Yadav JS, Wholey MH, Kuntz RE et al (2004) Protected carotid-artery stenting versus endarterectomy in high-risk patients. N Engl J Med 351:1493–1501PubMedCrossRef Yadav JS, Wholey MH, Kuntz RE et al (2004) Protected carotid-artery stenting versus endarterectomy in high-risk patients. N Engl J Med 351:1493–1501PubMedCrossRef
2.
go back to reference Mas JL, Chatellier G, Beyssen B et al (2006) Endarterectomy versus stenting in patients with symptomatic severe carotid stenosis. N Engl J Med 355:1660–1671PubMedCrossRef Mas JL, Chatellier G, Beyssen B et al (2006) Endarterectomy versus stenting in patients with symptomatic severe carotid stenosis. N Engl J Med 355:1660–1671PubMedCrossRef
3.
go back to reference Ringleb PA, Allenberg J, Berger J et al (2006) 30 day results from the SPACE trial of stent-protected angioplasty versus carotid endarterectomy in symptomatic patients: A randomised non-inferiority trial. Lancet 368:1239–1247PubMedCrossRef Ringleb PA, Allenberg J, Berger J et al (2006) 30 day results from the SPACE trial of stent-protected angioplasty versus carotid endarterectomy in symptomatic patients: A randomised non-inferiority trial. Lancet 368:1239–1247PubMedCrossRef
4.
go back to reference Ederle J, Dobson J, Featherstone RL et al (2010) Carotid artery stenting compared with endarterectomy in patients with symptomatic carotid stenosis (International Carotid Stenting Study): An interim analysis of a randomised controlled trial. Lancet 375:985–997PubMedCrossRef Ederle J, Dobson J, Featherstone RL et al (2010) Carotid artery stenting compared with endarterectomy in patients with symptomatic carotid stenosis (International Carotid Stenting Study): An interim analysis of a randomised controlled trial. Lancet 375:985–997PubMedCrossRef
5.
go back to reference Brott TG, Hobson RW, George H et al (2010) Stenting versus endarterectomy for treatment of carotid-artery stenosis. N Engl J Med 363(1):11–23PubMedCrossRef Brott TG, Hobson RW, George H et al (2010) Stenting versus endarterectomy for treatment of carotid-artery stenosis. N Engl J Med 363(1):11–23PubMedCrossRef
6.
go back to reference Biasi GM, Froio A, Diethrich EB et al (2004) Carotid plaque echolucency increases the risk of stroke in carotid stenting: The imaging in carotid angioplasty and risk of stroke (ICAROS) study. Circulation 110:756–762PubMedCrossRef Biasi GM, Froio A, Diethrich EB et al (2004) Carotid plaque echolucency increases the risk of stroke in carotid stenting: The imaging in carotid angioplasty and risk of stroke (ICAROS) study. Circulation 110:756–762PubMedCrossRef
7.
go back to reference Faggioli GL, Ferri M, Freyrie A et al (2007) Aortic arch anomalies are associated with increased risk of neurological events in carotid stent procedures. Eur J Vasc Endovasc Surg 33(4):436–441PubMedCrossRef Faggioli GL, Ferri M, Freyrie A et al (2007) Aortic arch anomalies are associated with increased risk of neurological events in carotid stent procedures. Eur J Vasc Endovasc Surg 33(4):436–441PubMedCrossRef
8.
go back to reference Chiam PT, Roubin GS, Ivyer SS et al (2008) Carotid artery stenting in elderly patients: Importance of case selection. Catheter Cardiovasc Interv 72:318–324PubMedCrossRef Chiam PT, Roubin GS, Ivyer SS et al (2008) Carotid artery stenting in elderly patients: Importance of case selection. Catheter Cardiovasc Interv 72:318–324PubMedCrossRef
9.
10.
go back to reference Mantovani L, D’Ercole L, Lisciandro F et al (2006) Radiochromic films for improved evaluation of patient dose in liver interventions. J Vasc Interv Radiol 17(5):855–862PubMedCrossRef Mantovani L, D’Ercole L, Lisciandro F et al (2006) Radiochromic films for improved evaluation of patient dose in liver interventions. J Vasc Interv Radiol 17(5):855–862PubMedCrossRef
11.
go back to reference D’Ercole L, Mantovani L, Zappoli Thyrion F et al (2007) A study on maximum skin dose in cerebral embolization procedures with radiochromic films. AJNR Am J Neuroradiol 28:503–507PubMed D’Ercole L, Mantovani L, Zappoli Thyrion F et al (2007) A study on maximum skin dose in cerebral embolization procedures with radiochromic films. AJNR Am J Neuroradiol 28:503–507PubMed
12.
go back to reference Delle Canne S, Carosi A, Bufacchi A et al (2006) Use of GAFCHROMIC XR type R films for skin-dose measurements in interventional radiology: Validation of a dosimetric procedure on a sample of patients undergoing interventional cardiology. Physica Medica 22(3):105–110PubMedCrossRef Delle Canne S, Carosi A, Bufacchi A et al (2006) Use of GAFCHROMIC XR type R films for skin-dose measurements in interventional radiology: Validation of a dosimetric procedure on a sample of patients undergoing interventional cardiology. Physica Medica 22(3):105–110PubMedCrossRef
13.
go back to reference Faj D, Steiner R, Trifunovic D et al. (2008) Patient dosimetry in interventional cardiology at the university Hospital of Osljek. Radiat Protect Dosim 128(4):485–490CrossRef Faj D, Steiner R, Trifunovic D et al. (2008) Patient dosimetry in interventional cardiology at the university Hospital of Osljek. Radiat Protect Dosim 128(4):485–490CrossRef
14.
go back to reference Balter S, Miller DL, Vano E et al (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35(2):673–680PubMedCrossRef Balter S, Miller DL, Vano E et al (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35(2):673–680PubMedCrossRef
15.
go back to reference Giordano C, D’Ercole L, Gobbi R et al (2010) Coronary angiography and percutaneous transluminal coronary angioplasty procedures: Evaluation of patients’ maximum skin dose using Gafchromic films and a comparison of local levels with reference levels proposed in the literature. Physica Medica 26:224–232PubMedCrossRef Giordano C, D’Ercole L, Gobbi R et al (2010) Coronary angiography and percutaneous transluminal coronary angioplasty procedures: Evaluation of patients’ maximum skin dose using Gafchromic films and a comparison of local levels with reference levels proposed in the literature. Physica Medica 26:224–232PubMedCrossRef
16.
go back to reference Ying CK, Kandaiya S (2010) Patient skin dose measurements during coronary interventional procedures using Gafchromic film. J Radiol Prot 30:585–596PubMedCrossRef Ying CK, Kandaiya S (2010) Patient skin dose measurements during coronary interventional procedures using Gafchromic film. J Radiol Prot 30:585–596PubMedCrossRef
17.
go back to reference D’Ercole L, Azzaretti A, Zappoli Thyrion F et al (2010) Measurement of patient skin dose in vertebroplasty using radiochromic dosimetry film. Spine 35:1304–1306PubMed D’Ercole L, Azzaretti A, Zappoli Thyrion F et al (2010) Measurement of patient skin dose in vertebroplasty using radiochromic dosimetry film. Spine 35:1304–1306PubMed
18.
go back to reference International Commission on Radiological Protection (2000) ICRP Publication 85. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30:7–67 International Commission on Radiological Protection (2000) ICRP Publication 85. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30:7–67
19.
go back to reference European Commission (2000) Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiat Protect 109:4–25 European Commission (2000) Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiat Protect 109:4–25
20.
go back to reference Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part II: Skin dose. J Vasc Interv Radiol 14:977–990PubMedCrossRef Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part II: Skin dose. J Vasc Interv Radiol 14:977–990PubMedCrossRef
21.
go back to reference Fletcher DW, Miller DL, Balter S et al (2002) Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 13:391–397PubMedCrossRef Fletcher DW, Miller DL, Balter S et al (2002) Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 13:391–397PubMedCrossRef
22.
go back to reference Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part I: Overall measures of dose. J Vasc Interv Radiol 14:711–727PubMedCrossRef Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part I: Overall measures of dose. J Vasc Interv Radiol 14:711–727PubMedCrossRef
23.
go back to reference Bor D, Türkay T, Olgar T et al (2006) Variations of patient doses in interventional examinations at different angiographic units. Cardiovasc Interv Radiol 29:797–806CrossRef Bor D, Türkay T, Olgar T et al (2006) Variations of patient doses in interventional examinations at different angiographic units. Cardiovasc Interv Radiol 29:797–806CrossRef
24.
go back to reference European Union (1997) Council Directive 97/43 Euratom of 30 June 1997, Official Journal of the European Communities, 1997; Legislation L. 180 European Union (1997) Council Directive 97/43 Euratom of 30 June 1997, Official Journal of the European Communities, 1997; Legislation L. 180
25.
go back to reference Gianfelice D, Lepanto L, Perreault P et al (2000) Effect of the learning process on procedure times and radiation exposure for CT fluoroscopy-guided percutaneous biopsy procedures. J Vasc Interv Radiol 11:1217–1221PubMedCrossRef Gianfelice D, Lepanto L, Perreault P et al (2000) Effect of the learning process on procedure times and radiation exposure for CT fluoroscopy-guided percutaneous biopsy procedures. J Vasc Interv Radiol 11:1217–1221PubMedCrossRef
26.
go back to reference Andrews RT, Brown PH (2000) Uterine arterial embolization: Factors influencing patient radiation exposure. Radiology 217:713–722PubMed Andrews RT, Brown PH (2000) Uterine arterial embolization: Factors influencing patient radiation exposure. Radiology 217:713–722PubMed
27.
go back to reference Watson LE, Riggs MW, Bourland PD (1997) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef Watson LE, Riggs MW, Bourland PD (1997) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef
28.
go back to reference Verdun FR, Aroua A, Trueb R et al (2005) Diagnostic and interventional radiology: A strategy to introduce reference dose level taking into account the national practice. Radiat Prot Dosimetry 114(1–3):188–191PubMedCrossRef Verdun FR, Aroua A, Trueb R et al (2005) Diagnostic and interventional radiology: A strategy to introduce reference dose level taking into account the national practice. Radiat Prot Dosimetry 114(1–3):188–191PubMedCrossRef
29.
go back to reference Aroua A, Rickli H, Stauffer JC et al (2007) How to set up and apply reference levels in fluoroscopy at a national level? Eur Radiol 17:1621–1633PubMedCrossRef Aroua A, Rickli H, Stauffer JC et al (2007) How to set up and apply reference levels in fluoroscopy at a national level? Eur Radiol 17:1621–1633PubMedCrossRef
30.
go back to reference Stratis AI, Anthopoulos PL, Gavaliatsis IP et al (2009) Patient dose in cardiac radiology. Hellenic J Cardiol 50:17–25PubMed Stratis AI, Anthopoulos PL, Gavaliatsis IP et al (2009) Patient dose in cardiac radiology. Hellenic J Cardiol 50:17–25PubMed
Metadata
Title
Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure
Authors
Loredana D’Ercole
Pietro Quaretti
Nicola Cionfoli
Catherine Klersy
Milena Bocchiola
Giuseppe Rodolico
Andrea Azzaretti
Francesco Lisciandro
Tommaso Cascella
Federico Zappoli Thyrion
Publication date
01-04-2013
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 2/2013
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-012-0392-2

Other articles of this Issue 2/2013

CardioVascular and Interventional Radiology 2/2013 Go to the issue