Published in: World Journal of Surgery 3/2013

01-03-2013

Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?

Authors: Bellal Joseph, Pantelis Hadjizacharia, Hassan Aziz, Kara Snyder, Julie Wynne, Narong Kulvatunyou, Andrew Tang, Terence O’Keeffe, Rifat Latifi, Randall Friese, Peter Rhee

Published in: World Journal of Surgery | Issue 3/2013

Login to get access

Abstract

Background

Advances in technology have allowed for continuous noninvasive hemoglobin monitoring (SpHb), which may enable earlier detection of hemorrhage and more efficient surgical and/or blood transfusion management. The use of SpHb has not been described in the trauma population. The purpose of the present study was to evaluate the accuracy of a SpHb measurement device in severely injured trauma patients.

Methods

We performed a prospective cohort analysis of severely injured trauma patients admitted to the intensive care unit (ICU) at our level I trauma center over a 6 month period. Serial IHb (invasive hemoglobin) levels and SpHb for the first 72 h were measured. Each SpHb measurement was matched with a corresponding IHb measurement. We defined normal Hgb as >8 mg/dL and low Hgb as <8 mg/dL. Data were then grouped based on Hgb level. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and Spearman correlation coefficient plot were calculated.

Results

A total of 23 trauma patients with 89 data pairs were reviewed. Eighty-six percent of the patients were male with a mean age of 32 years and a mean injury severity score (ISS) of 21.1 ± 14. Invasive hemoglobin had a range of 7.2–16.9 and SpHb had a range of 3.3–15.2. The average mean and difference between IHb and SpHb were 10.7 and 1, respectively. Continuous noninvasive hemoglobin measurement did not record data points 13.5 % of the time. The Spearman correlation plot revealed a correlation of R = 0.670 (p < 0.001). After dichotomization with Hgb > 8, SpHb was found to have a sensitivity of 91 %, PPV 96 %, specificity 40 %, NPV 20 %, and an accuracy of 88 %.

Conclusions

The continuous noninvasive hemoglobin monitor does not appear to represent serum hemoglobin levels accurately in severely injured trauma patients. However, we were able to identify utility for this noninvasive tool when Hgb was dichotomized into normal or low levels.
Literature
1.
go back to reference Kauvar DS, Lefering R, Wade CE (2006) Impact of hemorrhage on trauma outcome: an overview of epidemiology, clinical presentations, and therapeutic considerations. J Trauma 60:S3–S11PubMedCrossRef Kauvar DS, Lefering R, Wade CE (2006) Impact of hemorrhage on trauma outcome: an overview of epidemiology, clinical presentations, and therapeutic considerations. J Trauma 60:S3–S11PubMedCrossRef
2.
go back to reference Acosta JA, Yang JC, Winchell RJ et al (1998) Lethal injuries and time to death in a level I trauma center. J Am Coll Surg 186:528–533PubMedCrossRef Acosta JA, Yang JC, Winchell RJ et al (1998) Lethal injuries and time to death in a level I trauma center. J Am Coll Surg 186:528–533PubMedCrossRef
3.
go back to reference Shapiro NI, Kociszewski C, Harrison T et al (2003) Isolated prehospital hypotension after traumatic injuries: a predictor of mortality? J Emerg Med 25:175–179PubMedCrossRef Shapiro NI, Kociszewski C, Harrison T et al (2003) Isolated prehospital hypotension after traumatic injuries: a predictor of mortality? J Emerg Med 25:175–179PubMedCrossRef
4.
go back to reference Bruns B, Lindsey M, Rowe K et al (2007) Hemoglobin drops within minutes of injuries and predicts need for an intervention to stop hemorrhage. J Trauma 63:312–315PubMedCrossRef Bruns B, Lindsey M, Rowe K et al (2007) Hemoglobin drops within minutes of injuries and predicts need for an intervention to stop hemorrhage. J Trauma 63:312–315PubMedCrossRef
5.
go back to reference Shoemaker WC, Peitzman AB, Bellamy R et al (1996) Resuscitation from severe hemorrhage. Crit Care Med 24:S12–S23PubMedCrossRef Shoemaker WC, Peitzman AB, Bellamy R et al (1996) Resuscitation from severe hemorrhage. Crit Care Med 24:S12–S23PubMedCrossRef
6.
go back to reference Gehring H, Hornberger C, Dibbelt L et al (2002) Accuracy of point of care testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand 46:980–986PubMedCrossRef Gehring H, Hornberger C, Dibbelt L et al (2002) Accuracy of point of care testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand 46:980–986PubMedCrossRef
7.
go back to reference Mokken FC, van der Waart FJ, Henny CP et al (1996) Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol 73:135–137PubMedCrossRef Mokken FC, van der Waart FJ, Henny CP et al (1996) Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol 73:135–137PubMedCrossRef
8.
go back to reference Yang ZW, Yang SH, Chen L et al (2001) Comparison of blood counts in venous, fingertip, and arterial blood and their measurement variation. Clin Lab Haem 23:155–159CrossRef Yang ZW, Yang SH, Chen L et al (2001) Comparison of blood counts in venous, fingertip, and arterial blood and their measurement variation. Clin Lab Haem 23:155–159CrossRef
9.
go back to reference Macknet M, Norton S, Kimball-Jones P et al (2007) Continuous noninvasive measurement of hemoglobin via pulse CO-oximetry. Anesth Analg 105:S–108 Macknet M, Norton S, Kimball-Jones P et al (2007) Continuous noninvasive measurement of hemoglobin via pulse CO-oximetry. Anesth Analg 105:S–108
10.
go back to reference Macknet MR, Allard M, Applegate RL II et al (2010) The Accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg 111:1424–1426PubMedCrossRef Macknet MR, Allard M, Applegate RL II et al (2010) The Accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg 111:1424–1426PubMedCrossRef
11.
go back to reference Clarke JR, Trooskin SZ, Doshi PJ et al (2002) Time to laparotomy for intra-abdominal bleeding from trauma does affect survival for delays up to 90 min. J Trauma 52:420–425PubMedCrossRef Clarke JR, Trooskin SZ, Doshi PJ et al (2002) Time to laparotomy for intra-abdominal bleeding from trauma does affect survival for delays up to 90 min. J Trauma 52:420–425PubMedCrossRef
12.
go back to reference Gayat E, Bodin A, Sportiello C et al (2011) Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med 57:330–333PubMedCrossRef Gayat E, Bodin A, Sportiello C et al (2011) Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med 57:330–333PubMedCrossRef
13.
go back to reference Causey MW, Miller S, Foster A et al (2011) Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg 201:590–596CrossRef Causey MW, Miller S, Foster A et al (2011) Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg 201:590–596CrossRef
14.
go back to reference Miller RD, Ward TA, Shiboski SC et al (2011) Comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg 112:858–863PubMedCrossRef Miller RD, Ward TA, Shiboski SC et al (2011) Comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg 112:858–863PubMedCrossRef
15.
go back to reference Berkow L, Rotolo S, Mirski E (2011) Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg 113:1396–1402PubMedCrossRef Berkow L, Rotolo S, Mirski E (2011) Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg 113:1396–1402PubMedCrossRef
16.
go back to reference Butwick A, Hilton G, Carvalho B (2012) Non-invasive haemoglobin measurement in patients undergoing elective caesarean section. Br J Anaesth 108:271–277PubMedCrossRef Butwick A, Hilton G, Carvalho B (2012) Non-invasive haemoglobin measurement in patients undergoing elective caesarean section. Br J Anaesth 108:271–277PubMedCrossRef
Metadata
Title
Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?
Authors
Bellal Joseph
Pantelis Hadjizacharia
Hassan Aziz
Kara Snyder
Julie Wynne
Narong Kulvatunyou
Andrew Tang
Terence O’Keeffe
Rifat Latifi
Randall Friese
Peter Rhee
Publication date
01-03-2013
Publisher
Springer-Verlag
Published in
World Journal of Surgery / Issue 3/2013
Print ISSN: 0364-2313
Electronic ISSN: 1432-2323
DOI
https://doi.org/10.1007/s00268-012-1871-y

Other articles of this Issue 3/2013

World Journal of Surgery 3/2013 Go to the issue