Abstract

Backgrounds

Porous tantalum osteonecrosis implants have been used in femoral head necrosis for several years, while the clinical outcomes were mixed. As a joint-preserving surgery, early necrosis deterioration and conversion to total hip arthroplasy failed our expectation. We hence investigate an observational study with retrieval analysis to find out the underlying reasons.

Methods

Thirteen patients were treated with core decompression and implantation of a tantalum rod. The cases were evaluated both functionally and radiologically. We retrieved and analyzed the micro-structural changes and the histopathologic features of four early failed femoral heads with scanning electron microscopy, histopathologic examination, and micro-CT scaning.

Results

All implants were placed in proper positions. One-year survival rate was 64.29 % with a HSS score of 81.11 ± 15.62. Four patients converted to arthroplasty in a mean time of 305 days (0.84 years), with a HSS score of 43.75 ± 7.5 at the last follow-up. A liquid layer surrounded the tantalum implant was noted on MRI in all four failed cases. Volume render CT remodeling revealed interspace between the metal and bone. Scanning electron microscopy and histopathologic examination indicated sparse and isolated bone ingrowth into the implants. The remodeled trabecular bone and the increased density around the peri-implant area were illustrated with micro-CT scaning.

Conclusions

The deterioration of early failed tantalum implant exceeds the nature of osteonecrosis progression. Rather than insufficient mechanical support resulting in improper position and invalid bone ingrowth, nullification of core decompression and consequential intra-osseous pressurization probably led to early failure of porous tantalum osteonecrosis implants.
Appendix
Available only for authorised users
Literature
1.
go back to reference Zhang Y, Li L, Shi Z et al (2013) Porous tantalum rod implant is an effective and safe choice for early-stage femoral head necrosis: a meta-analysis of clinical trials. Eur J Orthop Surg Traumatol Orthop Traumatol 23(2):211–217. doi:10.1007/s00590-012-0962-7 CrossRef Zhang Y, Li L, Shi Z et al (2013) Porous tantalum rod implant is an effective and safe choice for early-stage femoral head necrosis: a meta-analysis of clinical trials. Eur J Orthop Surg Traumatol Orthop Traumatol 23(2):211–217. doi:10.​1007/​s00590-012-0962-7 CrossRef
3.
go back to reference Bobyn JD, Poggie RA, Krygier JJ et al (2004) Clinical validation of a structural porous tantalum biomaterial for adult reconstruction. J Bone Joint Surg Am 86-A(2):123–129PubMed Bobyn JD, Poggie RA, Krygier JJ et al (2004) Clinical validation of a structural porous tantalum biomaterial for adult reconstruction. J Bone Joint Surg Am 86-A(2):123–129PubMed
4.
go back to reference Floerkemeier T, Thorey F, Daentzer D et al (2011) Clinical and radiological outcome of the treatment of osteonecrosis of the femoral head using the osteonecrosis intervention implant. Int Orthop 35(4):489–495. doi:10.1007/s00264-009-0940-9 CrossRefPubMed Floerkemeier T, Thorey F, Daentzer D et al (2011) Clinical and radiological outcome of the treatment of osteonecrosis of the femoral head using the osteonecrosis intervention implant. Int Orthop 35(4):489–495. doi:10.​1007/​s00264-009-0940-9 CrossRefPubMed
5.
go back to reference Veillette CJH, Mehdian H, Schemitsch EH et al (2006) Survivorship analysis and radiographic outcome following tantalum rod insertion for osteonecrosis of the femoral head. J Bone Joint Surg Am 88A3(11):48–55. doi:10.2106/JBJS.F.00538 CrossRef Veillette CJH, Mehdian H, Schemitsch EH et al (2006) Survivorship analysis and radiographic outcome following tantalum rod insertion for osteonecrosis of the femoral head. J Bone Joint Surg Am 88A3(11):48–55. doi:10.​2106/​JBJS.​F.​00538 CrossRef
7.
go back to reference Tanzer M, Bobyn JD, Krygier JJ et al (2008) Histopathologic retrieval analysis of clinically failed porous tantalum osteonecrosis implants. J Bone Joint Surg Am 90A(6):1282–1289. doi:10.2106/JBJS.F.00847 CrossRef Tanzer M, Bobyn JD, Krygier JJ et al (2008) Histopathologic retrieval analysis of clinically failed porous tantalum osteonecrosis implants. J Bone Joint Surg Am 90A(6):1282–1289. doi:10.​2106/​JBJS.​F.​00847 CrossRef
8.
go back to reference Osman WS, Bassiony AA, Asal MK et al (2015) Failure of a properly positioned tantalum rod for treatment of early femoral head necrosis and conversion to total hip arthroplasty. Eur Orthop Traumatol. doi:10.1007/s12570-015-0317-z Osman WS, Bassiony AA, Asal MK et al (2015) Failure of a properly positioned tantalum rod for treatment of early femoral head necrosis and conversion to total hip arthroplasty. Eur Orthop Traumatol. doi:10.​1007/​s12570-015-0317-z
9.
go back to reference Floerkemeier T, Lutz A, Nackenhorst U et al (2011) Core decompression and osteonecrosis intervention rod in osteonecrosis of the femoral head: clinical outcome and finite element analysis. Int Orthop 35(10):1461–1466. doi:10.1007/s00264-010-1138-x CrossRefPubMed Floerkemeier T, Lutz A, Nackenhorst U et al (2011) Core decompression and osteonecrosis intervention rod in osteonecrosis of the femoral head: clinical outcome and finite element analysis. Int Orthop 35(10):1461–1466. doi:10.​1007/​s00264-010-1138-x CrossRefPubMed
11.
go back to reference Gonzalez Del Pino J, Knapp K, Gomez Castresana F et al (1990) Revascularization of femoral head ischemic necrosis with vascularized bone graft: a CT scan experimental study. Skelet Radiol 19(3):197–202. doi:10.1007/BF00204095 CrossRef Gonzalez Del Pino J, Knapp K, Gomez Castresana F et al (1990) Revascularization of femoral head ischemic necrosis with vascularized bone graft: a CT scan experimental study. Skelet Radiol 19(3):197–202. doi:10.​1007/​BF00204095 CrossRef
13.
go back to reference Hong Y, Zhong H, Lin T et al (2015) Comparison of core decompression and conservative treatment for avascular necrosis of femoral head at early stage: a meta-analysis. Int J Clin Exp Med 8(4):5207–5216PubMedPubMedCentral Hong Y, Zhong H, Lin T et al (2015) Comparison of core decompression and conservative treatment for avascular necrosis of femoral head at early stage: a meta-analysis. Int J Clin Exp Med 8(4):5207–5216PubMedPubMedCentral
14.
go back to reference Castro FJ, Barrack RL (2000) Core decompression and conservative treatment for avascular necrosis of the femoral head: a meta-analysis. Am J Orthop (Belle Mead NJ) 29(3):187–194 Castro FJ, Barrack RL (2000) Core decompression and conservative treatment for avascular necrosis of the femoral head: a meta-analysis. Am J Orthop (Belle Mead NJ) 29(3):187–194
15.
go back to reference Tsao AK, Roberson JR, Christie MJ et al (2005) Biomechanical and clinical evaluations of a porous tantalum implant for the treatment of early-stage osteonecrosis. J Bone Joint Surg Am 87A2(1):22–27. doi:10.2106/JBJS.E.00490 CrossRef Tsao AK, Roberson JR, Christie MJ et al (2005) Biomechanical and clinical evaluations of a porous tantalum implant for the treatment of early-stage osteonecrosis. J Bone Joint Surg Am 87A2(1):22–27. doi:10.​2106/​JBJS.​E.​00490 CrossRef
16.
go back to reference Miao H, Ye D, Liang W et al (2015) Effect of osteonecrosis intervention rod versus core decompression using multiple small drill holes on early stages of necrosis of the femoral head: a prospective study on a series of 60 patients with a minimum 1-year-follow-up. Open Orthop J 9(1):179–184. doi:10.2174/1874325001509010179 CrossRefPubMedPubMedCentral Miao H, Ye D, Liang W et al (2015) Effect of osteonecrosis intervention rod versus core decompression using multiple small drill holes on early stages of necrosis of the femoral head: a prospective study on a series of 60 patients with a minimum 1-year-follow-up. Open Orthop J 9(1):179–184. doi:10.​2174/​1874325001509010​179 CrossRefPubMedPubMedCentral
19.
go back to reference Liu Y, Yan L, Zhou S et al (2015) Tantalum rod implantation for femoral head osteonecrosis: survivorship analysis and determination of prognostic factors for total hip arthroplasty. Int Orthop. doi:10.1007/s00264-015-2897-1 Liu Y, Yan L, Zhou S et al (2015) Tantalum rod implantation for femoral head osteonecrosis: survivorship analysis and determination of prognostic factors for total hip arthroplasty. Int Orthop. doi:10.​1007/​s00264-015-2897-1
20.
Metadata
Title
Early failures of porous tantalum osteonecrosis implants: a case series with retrieval analysis
Authors
Xianzuo Zhang
Jian Wang
Jun Xiao
Zhanjun Shi
Publication date
01-09-2016
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 9/2016
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-3087-x

Other articles of this Issue 9/2016

International Orthopaedics 9/2016 Go to the issue