Skip to main content
Top
Published in: International Orthopaedics 12/2015

01-12-2015 | Original Paper

Using the anatomical tibial axis for total knee arthroplasty alignment may lead to an internal rotation error

Authors: Csaba Forster-Horvath, Valerie Kremo, Magdalena Müller-Gerbl, Andrej Maria Nowakowski

Published in: International Orthopaedics | Issue 12/2015

Login to get access

Abstract

Introduction

Despite intensive research, current total knee arthroplasty (TKA) designs do not always provide the correct kinematics for the native joint and thus further optimisation is necessary. Several studies support the importance of malrotation of the tibial components in the failure of TKA. We hypothesise that using the anatomical tibial axis (ATA) to align tibial component rotation on the resected tibial surface may lead to an internal rotation error due to relative anterior shift of the lateral articular surface centre compared to the medial one. The aim of this study was to compare the anatomical tibial axis of the physiological tibial joint surface to the resected one.

Method

Twenty formalin-fixed cadaveric knees were obtained for study. After computed tomography scanning the data of each specimen were entered into a standardised coordinate system and virtual bone cuts were performed with 6, 8 and 10 mm resection depths. The positions of the articular surface centres were determined at each resection depth.

Results

The lateral articular surface centre had moved anteriorly after the resection by a mean 1.475 mm, while the medial one had not changed significantly. Resecting the tibia at a 6-mm cut and using the transverse tibial axis to align the prosthetic tibial plateau will result in a mean 4.0° (95 % confidence interval, 2.5-5.5°) of internal rotation compared to the uncut tibia.

Discussion

The ATA lies in 6 degrees of external rotation compared to the perpendicular to the posterior tibial condylar axis (PTCA). Graw et al. suggest aligning the tibial component in 10 degrees of external rotation to the latter. Thus, if we accept the above suggestion, the ATA is 4 degrees internally rotated compared to the same line on the resected proximal tibia. These prior studies appear to be in accordance with our findings.

Conclusions

We conclude that using the ATA on the resected tibial surface may contribute to an internal rotation error.
Literature
1.
go back to reference Ethgen O, Bruyere O, Richy F, Dardennes C, Reginster JY (2004) Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature. J Bone Joint Surg Am 86-A(5):963–974PubMed Ethgen O, Bruyere O, Richy F, Dardennes C, Reginster JY (2004) Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature. J Bone Joint Surg Am 86-A(5):963–974PubMed
2.
go back to reference van Essen GJ, Chipchase LS, O’Connor D, Krishnan J (1998) Primary total knee replacement: short-term outcomes in an Australian population. J Qual Clin Pract 18(2):135–142PubMed van Essen GJ, Chipchase LS, O’Connor D, Krishnan J (1998) Primary total knee replacement: short-term outcomes in an Australian population. J Qual Clin Pract 18(2):135–142PubMed
4.
go back to reference Sesen H, Demirkale I, Karaduman M, Vural CA, Okkaoglu MC, Altay M (2014) Why two-thirds of patients accepted the second session in staged bilateral total knee arthroplasty: a prospective analysis of 111 patients. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-014-3251-7 PubMed Sesen H, Demirkale I, Karaduman M, Vural CA, Okkaoglu MC, Altay M (2014) Why two-thirds of patients accepted the second session in staged bilateral total knee arthroplasty: a prospective analysis of 111 patients. Knee Surg Sports Traumatol Arthrosc. doi:10.​1007/​s00167-014-3251-7 PubMed
5.
7.
9.
go back to reference Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM (2002) Insall Award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res 404:7–13CrossRefPubMed Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM (2002) Insall Award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res 404:7–13CrossRefPubMed
13.
go back to reference Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res 392:46–55CrossRefPubMed Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res 392:46–55CrossRefPubMed
15.
go back to reference Luyckx T, Peeters T, Vandenneucker H, Victor J, Bellemans J (2012) Is adapted measured resection superior to gap-balancing in determining femoral component rotation in total knee replacement? J Bone Joint Surg (Br) 94(9):1271–1276. doi:10.1302/0301-620X.94B9.28670 CrossRef Luyckx T, Peeters T, Vandenneucker H, Victor J, Bellemans J (2012) Is adapted measured resection superior to gap-balancing in determining femoral component rotation in total knee replacement? J Bone Joint Surg (Br) 94(9):1271–1276. doi:10.​1302/​0301-620X.​94B9.​28670 CrossRef
16.
go back to reference Fehring TK (2000) Rotational malalignment of the femoral component in total knee arthroplasty. Clin Orthop Relat Res 380:72–79CrossRefPubMed Fehring TK (2000) Rotational malalignment of the femoral component in total knee arthroplasty. Clin Orthop Relat Res 380:72–79CrossRefPubMed
18.
19.
go back to reference Kobayashi H, Akamatsu Y, Kumagai K, Kusayama Y, Ishigatsubo R, Muramatsu S, Saito T (2014) The surgical epicondylar axis is a consistent reference of the distal femur in the coronal and axial planes. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-014-2867-y Kobayashi H, Akamatsu Y, Kumagai K, Kusayama Y, Ishigatsubo R, Muramatsu S, Saito T (2014) The surgical epicondylar axis is a consistent reference of the distal femur in the coronal and axial planes. Knee Surg Sports Traumatol Arthrosc. doi:10.​1007/​s00167-014-2867-y
20.
go back to reference Incavo SJ, Coughlin KM, Pappas C, Beynnon BD (2003) Anatomic rotational relationships of the proximal tibia, distal femur, and patella: implications for rotational alignment in total knee arthroplasty. J Arthroplasty 18(5):643–648CrossRefPubMed Incavo SJ, Coughlin KM, Pappas C, Beynnon BD (2003) Anatomic rotational relationships of the proximal tibia, distal femur, and patella: implications for rotational alignment in total knee arthroplasty. J Arthroplasty 18(5):643–648CrossRefPubMed
22.
go back to reference Dalury DF (2001) Observations of the proximal tibia in total knee arthroplasty. Clin Orthop Relat Res 389:150–155CrossRefPubMed Dalury DF (2001) Observations of the proximal tibia in total knee arthroplasty. Clin Orthop Relat Res 389:150–155CrossRefPubMed
24.
go back to reference Eckhoff DG, Metzger RG, Vandewalle MV (1995) Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res 321:28–31PubMed Eckhoff DG, Metzger RG, Vandewalle MV (1995) Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res 321:28–31PubMed
25.
go back to reference Huddleston JI, Scott RD, Wimberley DW (2005) Determination of neutral tibial rotational alignment in rotating platform TKA. Clin Orthop Relat Res 440:101–106CrossRefPubMed Huddleston JI, Scott RD, Wimberley DW (2005) Determination of neutral tibial rotational alignment in rotating platform TKA. Clin Orthop Relat Res 440:101–106CrossRefPubMed
27.
go back to reference Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T (2007) Determining the rotational alignment of the tibial component at total knee replacement: a comparison of two techniques. J Bone Joint Surg (Br) 89(1):45–49. doi:10.1302/0301-620X.89B1.17728 CrossRef Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T (2007) Determining the rotational alignment of the tibial component at total knee replacement: a comparison of two techniques. J Bone Joint Surg (Br) 89(1):45–49. doi:10.​1302/​0301-620X.​89B1.​17728 CrossRef
28.
go back to reference Akagi M, Mori S, Nishimura S, Nishimura A, Asano T, Hamanishi C (2005) Variability of extraarticular tibial rotation references for total knee arthroplasty. Clin Orthop Relat Res 436:172–176CrossRefPubMed Akagi M, Mori S, Nishimura S, Nishimura A, Asano T, Hamanishi C (2005) Variability of extraarticular tibial rotation references for total knee arthroplasty. Clin Orthop Relat Res 436:172–176CrossRefPubMed
29.
go back to reference Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219CrossRefPubMed Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219CrossRefPubMed
30.
go back to reference Kaech AMN, Müller-Gerbl M (2002) Implementierung eines standardisierten Koordinatensystems für dreidimensionale CT-Darstellungen des Kniegelenks mit VGStudio max. Dissertation, University of Basel, Basel, pp 1–33 Kaech AMN, Müller-Gerbl M (2002) Implementierung eines standardisierten Koordinatensystems für dreidimensionale CT-Darstellungen des Kniegelenks mit VGStudio max. Dissertation, University of Basel, Basel, pp 1–33
31.
32.
go back to reference Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144CrossRefPubMed Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144CrossRefPubMed
37.
40.
go back to reference Iorio R, Mazza D, Drogo P, Bolle G, Conteduca F, Redler A, Valeo L, Conteduca J, Ferretti A (2015) Clinical and radiographic outcomes of an accelerometer-based system for the tibial resection in total knee arthroplasty. Int Orthop 39(3):461–466. doi:10.1007/s00264-014-2541-5 CrossRefPubMed Iorio R, Mazza D, Drogo P, Bolle G, Conteduca F, Redler A, Valeo L, Conteduca J, Ferretti A (2015) Clinical and radiographic outcomes of an accelerometer-based system for the tibial resection in total knee arthroplasty. Int Orthop 39(3):461–466. doi:10.​1007/​s00264-014-2541-5 CrossRefPubMed
41.
go back to reference Maderbacher G, Schaumburger J, Keshmiri A, Barthel M, Springorum HR, Craiovan B, Grifka J, Baier C (2015) Pinless navigation in total knee arthroplasty: Navigation reduced by the maximum? Int Orthop 39(3):455–460. doi:10.1007/s00264-014-2529-1 CrossRefPubMed Maderbacher G, Schaumburger J, Keshmiri A, Barthel M, Springorum HR, Craiovan B, Grifka J, Baier C (2015) Pinless navigation in total knee arthroplasty: Navigation reduced by the maximum? Int Orthop 39(3):455–460. doi:10.​1007/​s00264-014-2529-1 CrossRefPubMed
Metadata
Title
Using the anatomical tibial axis for total knee arthroplasty alignment may lead to an internal rotation error
Authors
Csaba Forster-Horvath
Valerie Kremo
Magdalena Müller-Gerbl
Andrej Maria Nowakowski
Publication date
01-12-2015
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 12/2015
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-2858-8

Other articles of this Issue 12/2015