Published in: International Orthopaedics 10/2013

01-10-2013 | Original Paper

Metal hypersensitivity and metal ion levels in patients with coated or uncoated total knee arthroplasty: a randomised controlled study

Authors: Jörg Lützner, Albrecht Hartmann, Gerd Dinnebier, Petra Spornraft-Ragaller, Christine Hamann, Stephan Kirschner

Published in: International Orthopaedics | Issue 10/2013

Login to get access

Abstract

Purpose

Metal ion release by orthopaedic implants may cause local and systemic effects and induce hypersensitivity reactions. Coated implants have been developed to prevent or reduce these effects. This study was initiated to investigate the safety of a novel coating for total knee arthroplasty (TKA) implants.

Methods

A total of 120 patients undergoing primary TKA with no history of hypersensitivity and no other metal implant were randomised to receive either a coated or uncoated implant. Chromium (Cr), cobalt (Co), molybdenum (Mb) and nickel (Ni) hypersensitivity patch testing and plasma ion concentrations were evaluated pre-operatively and one year post-operatively.

Results

At the one year follow-up both groups demonstrated significant improvement in knee function and quality of life. One new weakly positive reaction to Co in the TKA group with coated implant and two doubtful skin reactions to Ni (one in each group) were noted. Even with sensitisation to implant materials no skin reactions were observed. Plasma metal ion concentrations did not increase and were not elevated at the one year follow-up in either group.

Conclusions

Sensitisation after TKA was rare and had no influence on clinical results. TKA with coated implant and standard TKA demonstrated no plasma metal ion elevation.
Literature
1.
go back to reference Basko-Plluska JL, Thyssen JP, Schalock PC (2011) Cutaneous and systemic hypersensitivity reactions to metallic implants. Dermatitis 22(2):65–79PubMed Basko-Plluska JL, Thyssen JP, Schalock PC (2011) Cutaneous and systemic hypersensitivity reactions to metallic implants. Dermatitis 22(2):65–79PubMed
2.
go back to reference Bergschmidt P, Bader R, Ganzer D et al (2012) Ceramic femoral components in total knee arthroplasty - two year follow-up results of an international prospective multi-centre study. Open Orthop J 6:172–178PubMedCrossRef Bergschmidt P, Bader R, Ganzer D et al (2012) Ceramic femoral components in total knee arthroplasty - two year follow-up results of an international prospective multi-centre study. Open Orthop J 6:172–178PubMedCrossRef
3.
go back to reference Cadosch D, Chan E, Gautschi OP, Filgueira L (2009) Metal is not inert: role of metal ions released by biocorrosion in aseptic loosening–current concepts. J Biomed Mater Res A 91(4):1252–1262PubMed Cadosch D, Chan E, Gautschi OP, Filgueira L (2009) Metal is not inert: role of metal ions released by biocorrosion in aseptic loosening–current concepts. J Biomed Mater Res A 91(4):1252–1262PubMed
4.
go back to reference Eben R, Walk R, Summer B et al (2009) Implant allergy register—a first report. Orthopade 38(6):557–562PubMedCrossRef Eben R, Walk R, Summer B et al (2009) Implant allergy register—a first report. Orthopade 38(6):557–562PubMedCrossRef
5.
go back to reference Ezzet KA, Hermida JC, Colwell CW Jr, D’Lima DD (2004) Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator. Clin Orthop Relat Res 428:120–124PubMedCrossRef Ezzet KA, Hermida JC, Colwell CW Jr, D’Lima DD (2004) Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator. Clin Orthop Relat Res 428:120–124PubMedCrossRef
6.
go back to reference Friesenbichler J, Maurer-Ertl W, Sadoghi P et al (2012) Serum metal ion levels after rotating-hinge knee arthroplasty: comparison between a standard device and a megaprosthesis. Int Orthop 36(3):539–544PubMedCrossRef Friesenbichler J, Maurer-Ertl W, Sadoghi P et al (2012) Serum metal ion levels after rotating-hinge knee arthroplasty: comparison between a standard device and a megaprosthesis. Int Orthop 36(3):539–544PubMedCrossRef
7.
go back to reference Frigerio E, Pigatto PD, Guzzi G, Altomare G (2011) Metal sensitivity in patients with orthopaedic implants: a prospective study. Contact Dermatitis 64(5):273–279PubMedCrossRef Frigerio E, Pigatto PD, Guzzi G, Altomare G (2011) Metal sensitivity in patients with orthopaedic implants: a prospective study. Contact Dermatitis 64(5):273–279PubMedCrossRef
8.
go back to reference Galetz MC, Fleischmann EW, Konrad CH, Schuetz A, Glatzel U (2010) Abrasion resistance of oxidized zirconium in comparison with CoCrMo and titanium nitride coatings for artificial knee joints. J Biomed Mater Res B Appl Biomater 93(1):244–251PubMed Galetz MC, Fleischmann EW, Konrad CH, Schuetz A, Glatzel U (2010) Abrasion resistance of oxidized zirconium in comparison with CoCrMo and titanium nitride coatings for artificial knee joints. J Biomed Mater Res B Appl Biomater 93(1):244–251PubMed
9.
go back to reference Göbel F, Ulbricht S, Hein W, Bernstein A (2008) Radiological mid-term results of total knee arthroplasty with femoral components of different materials. Z Orthop Unfall 146(2):194–199PubMedCrossRef Göbel F, Ulbricht S, Hein W, Bernstein A (2008) Radiological mid-term results of total knee arthroplasty with femoral components of different materials. Z Orthop Unfall 146(2):194–199PubMedCrossRef
10.
go back to reference Goldhahn S, Sawaguchi T, Audigé L et al (2009) Complication reporting in orthopaedic trials. A systematic review of randomized controlled trials. J Bone Joint Surg Am 91(8):1847–1853PubMedCrossRef Goldhahn S, Sawaguchi T, Audigé L et al (2009) Complication reporting in orthopaedic trials. A systematic review of randomized controlled trials. J Bone Joint Surg Am 91(8):1847–1853PubMedCrossRef
11.
go back to reference Granchi D, Cenni E, Giunti A, Baldini N (2012) Metal hypersensitivity testing in patients undergoing joint replacement: a systematic review. J Bone Joint Surg Br 94(8):1126–1134PubMed Granchi D, Cenni E, Giunti A, Baldini N (2012) Metal hypersensitivity testing in patients undergoing joint replacement: a systematic review. J Bone Joint Surg Br 94(8):1126–1134PubMed
12.
go back to reference Granchi D, Cenni E, Tigani D et al (2008) Sensitivity to implant materials in patients with total knee arthroplasties. Biomaterials 29(10):1494–1500PubMedCrossRef Granchi D, Cenni E, Tigani D et al (2008) Sensitivity to implant materials in patients with total knee arthroplasties. Biomaterials 29(10):1494–1500PubMedCrossRef
14.
go back to reference Hallab NJ, Jacobs JJ (2009) Biologic effects of implant debris. Bull NYU Hosp Jt Dis 67(2):182–188PubMed Hallab NJ, Jacobs JJ (2009) Biologic effects of implant debris. Bull NYU Hosp Jt Dis 67(2):182–188PubMed
15.
go back to reference Jacobs JJ, Gilbert JL, Urban RM (1998) Corrosion of metal orthopaedic implants. J Bone Joint Surg Am 80(2):268–282PubMed Jacobs JJ, Gilbert JL, Urban RM (1998) Corrosion of metal orthopaedic implants. J Bone Joint Surg Am 80(2):268–282PubMed
16.
go back to reference Kirschner S, Lützner J, Günther KP, Eberlein-Gonska M, Krummenauer F (2010) Adverse events in total knee arthroplasty: results of a physician independent survey in 260 patients. Patient Saf Surg 4(1):12PubMedCrossRef Kirschner S, Lützner J, Günther KP, Eberlein-Gonska M, Krummenauer F (2010) Adverse events in total knee arthroplasty: results of a physician independent survey in 260 patients. Patient Saf Surg 4(1):12PubMedCrossRef
17.
go back to reference Kluess D, Mittelmeier W, Bader R (2010) Intraoperative impaction of total knee replacements: an explicit finite-element-analysis of principal stresses in ceramic vs. cobalt-chromium femoral components. Clin Biomech (Bristol, Avon) 25(10):1018–1024CrossRef Kluess D, Mittelmeier W, Bader R (2010) Intraoperative impaction of total knee replacements: an explicit finite-element-analysis of principal stresses in ceramic vs. cobalt-chromium femoral components. Clin Biomech (Bristol, Avon) 25(10):1018–1024CrossRef
18.
go back to reference Langton DJ, Jameson SS, Joyce TJ et al (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46PubMed Langton DJ, Jameson SS, Joyce TJ et al (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46PubMed
19.
go back to reference Luetzner J, Krummenauer F, Lengel AM, Ziegler J, Witzleb WC (2007) Serum metal ion exposure after total knee arthroplasty. Clin Orthop Relat Res 461:136–142PubMed Luetzner J, Krummenauer F, Lengel AM, Ziegler J, Witzleb WC (2007) Serum metal ion exposure after total knee arthroplasty. Clin Orthop Relat Res 461:136–142PubMed
20.
go back to reference Niki Y, Matsumoto H, Otani T et al (2005) Screening for symptomatic metal sensitivity: a prospective study of 92 patients undergoing total knee arthroplasty. Biomaterials 26(9):1019–1026PubMedCrossRef Niki Y, Matsumoto H, Otani T et al (2005) Screening for symptomatic metal sensitivity: a prospective study of 92 patients undergoing total knee arthroplasty. Biomaterials 26(9):1019–1026PubMedCrossRef
21.
go back to reference Reich J, Hovy L, Lindenmaier HL et al (2010) Preclinical evaluation of coated knee implants for allergic patients. Orthopade 39(5):495–502PubMedCrossRef Reich J, Hovy L, Lindenmaier HL et al (2010) Preclinical evaluation of coated knee implants for allergic patients. Orthopade 39(5):495–502PubMedCrossRef
22.
go back to reference Reich J, Thomas P, Hovy L et al (2007) Allergy solutions for orthopaedic knee-implants. Biomaterialien 8(2):109–110 Reich J, Thomas P, Hovy L et al (2007) Allergy solutions for orthopaedic knee-implants. Biomaterialien 8(2):109–110
23.
go back to reference Schäfer T, Böhler E, Ruhdorfer S et al (2001) Epidemiology of contact allergy in adults. Allergy 56(12):1192–1196PubMedCrossRef Schäfer T, Böhler E, Ruhdorfer S et al (2001) Epidemiology of contact allergy in adults. Allergy 56(12):1192–1196PubMedCrossRef
24.
go back to reference Schnuch A, Aberer W, Agathos M et al (2008) Performing patch testing with contact allergens. J Dtsch Dermatol Ges 6(9):770–775PubMed Schnuch A, Aberer W, Agathos M et al (2008) Performing patch testing with contact allergens. J Dtsch Dermatol Ges 6(9):770–775PubMed
25.
go back to reference Smith AJ, Dieppe P, Vernon K, Porter M, Blom AW et al (2012) Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet 379(9822):1199–1204PubMedCrossRef Smith AJ, Dieppe P, Vernon K, Porter M, Blom AW et al (2012) Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet 379(9822):1199–1204PubMedCrossRef
26.
go back to reference Spector BM, Ries MD, Bourne RB et al (2001) Wear performance of ultra-high molecular weight polyethylene on oxidized zirconium total knee femoral components. J Bone Joint Surg Am 83-A(Suppl 2 Pt 2):80–86PubMed Spector BM, Ries MD, Bourne RB et al (2001) Wear performance of ultra-high molecular weight polyethylene on oxidized zirconium total knee femoral components. J Bone Joint Surg Am 83-A(Suppl 2 Pt 2):80–86PubMed
27.
go back to reference Tharani R, Dorey FJ, Schmalzried TP (2001) The risk of cancer following total hip or knee arthroplasty. J Bone Joint Surg Am 83-A(5):774–780PubMed Tharani R, Dorey FJ, Schmalzried TP (2001) The risk of cancer following total hip or knee arthroplasty. J Bone Joint Surg Am 83-A(5):774–780PubMed
28.
go back to reference Thomas P, Schuh A, Ring J, Thomsen M (2008) Orthopedic surgical implants and allergies: joint statement by the implant allergy working group (AK 20) of the DGOOC (German association of orthopedics and orthopedic surgery), DKG (German contact dermatitis research group) and dgaki (German society for allergology and clinical immunology)]. Orthopade 37(1):75–88PubMedCrossRef Thomas P, Schuh A, Ring J, Thomsen M (2008) Orthopedic surgical implants and allergies: joint statement by the implant allergy working group (AK 20) of the DGOOC (German association of orthopedics and orthopedic surgery), DKG (German contact dermatitis research group) and dgaki (German society for allergology and clinical immunology)]. Orthopade 37(1):75–88PubMedCrossRef
29.
go back to reference Tsukamoto R, Chen S, Asano T et al (2006) Improved wear performance with crosslinked UHMWPE and zirconia implants in knee simulation. Acta Orthop 77(3):505–511PubMedCrossRef Tsukamoto R, Chen S, Asano T et al (2006) Improved wear performance with crosslinked UHMWPE and zirconia implants in knee simulation. Acta Orthop 77(3):505–511PubMedCrossRef
30.
go back to reference Vassiliou K, Elfick AP, Scholes SC, Unsworth A (2006) The effect of ‘running-in’ on the tribology and surface morphology of metal-on-metal Birmingham hip resurfacing device in simulator studies. Proc Inst Mech Eng H 220(2):269–277PubMedCrossRef Vassiliou K, Elfick AP, Scholes SC, Unsworth A (2006) The effect of ‘running-in’ on the tribology and surface morphology of metal-on-metal Birmingham hip resurfacing device in simulator studies. Proc Inst Mech Eng H 220(2):269–277PubMedCrossRef
31.
go back to reference Webley M, Kates A, Snaith ML (1978) Metal sensitivity in patients with a hinge arthroplasty of the knee. Ann Rheum Dis 37(4):373–375PubMedCrossRef Webley M, Kates A, Snaith ML (1978) Metal sensitivity in patients with a hinge arthroplasty of the knee. Ann Rheum Dis 37(4):373–375PubMedCrossRef
32.
go back to reference Whittingham-Jones PM, Dunstan E, Altaf H et al (2008) Immune responses in patients with metal-on-metal hip articulations: a long-term follow-up. J Arthroplasty 23(8):1212–1218PubMedCrossRef Whittingham-Jones PM, Dunstan E, Altaf H et al (2008) Immune responses in patients with metal-on-metal hip articulations: a long-term follow-up. J Arthroplasty 23(8):1212–1218PubMedCrossRef
Metadata
Title
Metal hypersensitivity and metal ion levels in patients with coated or uncoated total knee arthroplasty: a randomised controlled study
Authors
Jörg Lützner
Albrecht Hartmann
Gerd Dinnebier
Petra Spornraft-Ragaller
Christine Hamann
Stephan Kirschner
Publication date
01-10-2013
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 10/2013
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-013-2010-6

Other articles of this Issue 10/2013

International Orthopaedics 10/2013 Go to the issue