Intra- and extra-articular planes of reference for use in total hip arthroplasty: a preliminary study

Authors: Jerome Hausselle, Pierre Etienne Moreau, Loic Wessely, Emmanuel de Thomasson, Ayman Assi, Sebastien Parratte, Jerome Essig, Wafa Skalli

Published in: International Orthopaedics | Issue 8/2012

Login to get access

Abstract

Purpose

Acetabular component malalignment in total hip arthroplasty can lead to potential complications such as dislocation, component impingement and excessive wear. Computer-assisted orthopaedic surgery systems generally use the anterior pelvic plane (APP). Our aim was to investigate the reliability of anatomical landmarks accessible during surgery and to define new potential planes of reference.

Methods

Three types of palpations were performed: virtual, on dry bones and on two cadaveric specimens. Four landmarks were selected, the reproducibility of their positioning ranging from 0.9 to 2.3 mm. We then defined five planes and tested them during palpations on two cadaveric specimens.

Results

Two planes produced a mean orientation error of 5.0° [standard deviation (SD 3.3°)] and 5.6° (SD 2.7°).

Conclusions

Even if further studies are needed to test the reliability of such planes on a larger scale in vivo during surgery, these results demonstrated the feasibility of defining a new plane of reference as an alternative to the APP.
Literature
1.
go back to reference Lewinneck GE, Lewis JL, Tarr R, Compere CL, Zimmermann JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220 Lewinneck GE, Lewis JL, Tarr R, Compere CL, Zimmermann JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220
2.
go back to reference DeChenne CL, Jayarama U, Lovellb T, Dongc N, Cusick M (2005) A novel acetabular alignment guide for THR using selective anatomic landmarks on the pelvis. J Biomech 38:1902–1908PubMedCrossRef DeChenne CL, Jayarama U, Lovellb T, Dongc N, Cusick M (2005) A novel acetabular alignment guide for THR using selective anatomic landmarks on the pelvis. J Biomech 38:1902–1908PubMedCrossRef
3.
go back to reference Haaker RGA, Tiedjen K, Ottersbach A, Rubenthaler F, Stockheim M, Stiehl JB (2007) Comparison of conventional versus computer-navigated acetabular component insertion. J Arthroplasty 22:151–158PubMedCrossRef Haaker RGA, Tiedjen K, Ottersbach A, Rubenthaler F, Stockheim M, Stiehl JB (2007) Comparison of conventional versus computer-navigated acetabular component insertion. J Arthroplasty 22:151–158PubMedCrossRef
4.
go back to reference Jolles BM, Genoud P, Hoffmeyer P (2001) Accuracy of computer-assisted cup placement in total hip arthroplasty. Int Congress Series 1230:314–318CrossRef Jolles BM, Genoud P, Hoffmeyer P (2001) Accuracy of computer-assisted cup placement in total hip arthroplasty. Int Congress Series 1230:314–318CrossRef
5.
go back to reference Kalteis T, Handel M, Herold T, Perlick L, Baethis H, Grifka J (2005) Greater accuracy in positioning of the acetabular cup by using an image-free navigation system. Int Orthopaedics 29:272–276CrossRef Kalteis T, Handel M, Herold T, Perlick L, Baethis H, Grifka J (2005) Greater accuracy in positioning of the acetabular cup by using an image-free navigation system. Int Orthopaedics 29:272–276CrossRef
6.
go back to reference Minoda Y, Ohzono K, Aihara M, Umeda N, Tomita M, Hayakawa K (2010) Are acetabular component alignment guides for total hip arthroplasty accurate? J Arthroplasty 25:986–989PubMedCrossRef Minoda Y, Ohzono K, Aihara M, Umeda N, Tomita M, Hayakawa K (2010) Are acetabular component alignment guides for total hip arthroplasty accurate? J Arthroplasty 25:986–989PubMedCrossRef
7.
go back to reference Najarian BC, Kilgore JE, Markel DC (2009) Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplasty 24:15–21PubMedCrossRef Najarian BC, Kilgore JE, Markel DC (2009) Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplasty 24:15–21PubMedCrossRef
8.
go back to reference Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, Michaelis U, Kemper G, Grützner PA, Steffen R, von Knoch M, Holland-Letz T, Bernsmann K (2004) The accuracy of free-hand cup positioning - a CT based measurement of cup placement in 105 total hip arthroplasties. Int Orthopaedics 28:198–201 Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, Michaelis U, Kemper G, Grützner PA, Steffen R, von Knoch M, Holland-Letz T, Bernsmann K (2004) The accuracy of free-hand cup positioning - a CT based measurement of cup placement in 105 total hip arthroplasties. Int Orthopaedics 28:198–201
9.
go back to reference Wolf A, DiGioia AM III, Mor AB, Jaramaz B (2005) A kinematic model for calculating cup alignment error during total hip arthroplasty. J Biomech 38:2257–2265PubMedCrossRef Wolf A, DiGioia AM III, Mor AB, Jaramaz B (2005) A kinematic model for calculating cup alignment error during total hip arthroplasty. J Biomech 38:2257–2265PubMedCrossRef
10.
go back to reference Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2011) Limitations of imageless computer-assisted navigation for total hip arthroplasty. J Arthroplasty 26:596–605PubMedCrossRef Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2011) Limitations of imageless computer-assisted navigation for total hip arthroplasty. J Arthroplasty 26:596–605PubMedCrossRef
11.
go back to reference Sotereanos NG, Miller MC, Smith B, Hube R, Sewecke JJ, Wohlrab D (2006) Using intraoperative pelvic landmarks for acetabular component placement in total hip arthroplasty. J Arthroplasty 21:832–840PubMedCrossRef Sotereanos NG, Miller MC, Smith B, Hube R, Sewecke JJ, Wohlrab D (2006) Using intraoperative pelvic landmarks for acetabular component placement in total hip arthroplasty. J Arthroplasty 21:832–840PubMedCrossRef
12.
go back to reference Yoon YS, Hodgson AJ, Tonetti J, Masri BA, Duncan CP (2008) Resolving inconsistencies in defining the target orientation for the acetabular cup angles in total hip arthroplasty. Clin Biomech 23:253–259CrossRef Yoon YS, Hodgson AJ, Tonetti J, Masri BA, Duncan CP (2008) Resolving inconsistencies in defining the target orientation for the acetabular cup angles in total hip arthroplasty. Clin Biomech 23:253–259CrossRef
13.
go back to reference Ybinger T, Kumpan W, Hoffart HE, Muschalik B, Bullmann W, Zweymüller K (2007) Accuracy of navigation-assisted acetabular component positioning studied by computed tomography measurements. J Arthroplasty 22:812–817PubMedCrossRef Ybinger T, Kumpan W, Hoffart HE, Muschalik B, Bullmann W, Zweymüller K (2007) Accuracy of navigation-assisted acetabular component positioning studied by computed tomography measurements. J Arthroplasty 22:812–817PubMedCrossRef
14.
go back to reference Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA (2007) Squeaking in ceramic-on-ceramic hips. The importance of acetabular component orientation. J Arthroplasty 22:496–503PubMedCrossRef Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA (2007) Squeaking in ceramic-on-ceramic hips. The importance of acetabular component orientation. J Arthroplasty 22:496–503PubMedCrossRef
15.
go back to reference Parratte S, Argenson JN (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am 89:494–499PubMedCrossRef Parratte S, Argenson JN (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am 89:494–499PubMedCrossRef
16.
go back to reference Kruger S, Zambelli PY, Leyvraz PF, Jolles BM (2009) Computer-assisted placement technique in hip resurfacing arthroplasty: improvement in accuracy? Int Orthop 33:27–33PubMedCrossRef Kruger S, Zambelli PY, Leyvraz PF, Jolles BM (2009) Computer-assisted placement technique in hip resurfacing arthroplasty: improvement in accuracy? Int Orthop 33:27–33PubMedCrossRef
17.
go back to reference Kumar MA, Shetty MS, Kiran KG, Kini AR (2012) Validation of navigation assisted cup placement in total hip arthroplasty. Int Orthop 36:17–22PubMedCrossRef Kumar MA, Shetty MS, Kiran KG, Kini AR (2012) Validation of navigation assisted cup placement in total hip arthroplasty. Int Orthop 36:17–22PubMedCrossRef
18.
go back to reference Pitto RP, Malak S, Anderson IA (2009) Accuracy of a computer-assisted navigation system in resurfacing hip arthroplasty. Int Orthop 33:391–395PubMedCrossRef Pitto RP, Malak S, Anderson IA (2009) Accuracy of a computer-assisted navigation system in resurfacing hip arthroplasty. Int Orthop 33:391–395PubMedCrossRef
19.
go back to reference Schnurr C, Michael JWP, Eysel P, Konig DP (2009) Imageless navigation of hip resurfacing arthroplasty increases the implant accuracy. Int Orthop 33:365–372PubMedCrossRef Schnurr C, Michael JWP, Eysel P, Konig DP (2009) Imageless navigation of hip resurfacing arthroplasty increases the implant accuracy. Int Orthop 33:365–372PubMedCrossRef
20.
go back to reference Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2008) Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech 23:1004–1011CrossRef Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2008) Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech 23:1004–1011CrossRef
21.
go back to reference Pinoit Y, May O, Girard J, Laffargue P, Ala Eddine T, Migaud H (2007) Low accuracy of the anterior pelvic plane to guide the position of the cup with imageless computer assistance: variation of position in 106 patients. Rev Chir Orthop Reparatrice Appar Mot 93:455–460PubMedCrossRef Pinoit Y, May O, Girard J, Laffargue P, Ala Eddine T, Migaud H (2007) Low accuracy of the anterior pelvic plane to guide the position of the cup with imageless computer assistance: variation of position in 106 patients. Rev Chir Orthop Reparatrice Appar Mot 93:455–460PubMedCrossRef
22.
go back to reference Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK (1995) Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 5:262–270PubMedCrossRef Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK (1995) Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 5:262–270PubMedCrossRef
23.
go back to reference Rousseau MA, Lazennec JY, Boyer P, Mora N, Gorin M, Catonné Y (2009) Optimization of total hip arthroplasty implantation. J Arthroplasty 24:22–26PubMedCrossRef Rousseau MA, Lazennec JY, Boyer P, Mora N, Gorin M, Catonné Y (2009) Optimization of total hip arthroplasty implantation. J Arthroplasty 24:22–26PubMedCrossRef
Metadata
Title
Intra- and extra-articular planes of reference for use in total hip arthroplasty: a preliminary study
Authors
Jerome Hausselle
Pierre Etienne Moreau
Loic Wessely
Emmanuel de Thomasson
Ayman Assi
Sebastien Parratte
Jerome Essig
Wafa Skalli
Publication date
01-08-2012
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 8/2012
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-012-1516-7

Other articles of this Issue 8/2012

International Orthopaedics 8/2012 Go to the issue