01-08-2007 | Original Paper

Effects of screw eccentricity on the initial stability of the acetabular cup

Authors: Jui-Ting Hsu, Chih-Han Chang, Kai-Nan An, Mark E. Zobitz, Rapin Phimolsarnti, Ronald R. Hugate, Kuo-An Lai

Published in: International Orthopaedics | Issue 4/2007

Login to get access

Abstract

One of the major failure modes of cementless acetabular components is the loosening of the acetabular cup, which is mostly attributable to insufficient initial stability. A hemispherical cup with a porous coating which is inserted with press-fit fixation and secured with several screws is one of the most widely used approaches. Many studies have found that bone screws are very helpful aids for cup fixation, but the optimal surgical technique for inserting screws has not been clearly reported. In this study, hemispherical cups were fixed into blocks of foam bone with zero to three screws. The effects of three types of screw eccentricity (a 1-mm offset and angular eccentricities of 15° and 25°) on the initial stability of the acetabular cup were evaluated. The experimental results indicate that increasing the number of screws enhances the cup stability in the case of ideal screwing (i.e., with no eccentricity). An angular eccentricity of 15° did not affect the cup stability for fixation with one or two screws. However, the presence of 25° of angular eccentricity significantly reduced the stability of the cup, while 1 mm of offset eccentricity produced an even greater impact.
Literature
1.
go back to reference Baleani M, Fognani R, Toni A (2001) Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs 25(8):664–669PubMedCrossRef Baleani M, Fognani R, Toni A (2001) Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs 25(8):664–669PubMedCrossRef
2.
go back to reference Bi Y, Van De Motter RR, Ragab AA, Goldberg VM, Anderson JM, Greenfield EM (2001) Titanium particles stimulate bone resorption by inducing differentiation of murine osteoclasts. J Bone Joint Surg Am 83-A(4):501–508PubMed Bi Y, Van De Motter RR, Ragab AA, Goldberg VM, Anderson JM, Greenfield EM (2001) Titanium particles stimulate bone resorption by inducing differentiation of murine osteoclasts. J Bone Joint Surg Am 83-A(4):501–508PubMed
3.
go back to reference Clohisy JC, Harris WH (1999) The Harris-Galante porous-coated acetabular component with screw fixation. An average ten-year follow-up study. J Bone Joint Surg Am 81(1):66–73PubMed Clohisy JC, Harris WH (1999) The Harris-Galante porous-coated acetabular component with screw fixation. An average ten-year follow-up study. J Bone Joint Surg Am 81(1):66–73PubMed
4.
go back to reference Hadjari MH, Hollis JM, Hofmann OE, Flahiff CM, Nelson CL (1994) Initial stability of porous coated acetabular implants. The effect of screw placement, screw tightness, defect type, and oversize implants. Clin Orthop Relat Res 307:117–123PubMed Hadjari MH, Hollis JM, Hofmann OE, Flahiff CM, Nelson CL (1994) Initial stability of porous coated acetabular implants. The effect of screw placement, screw tightness, defect type, and oversize implants. Clin Orthop Relat Res 307:117–123PubMed
5.
go back to reference Huiskes R (1993) Failed innovation in total hip replacement. Diagnosis and proposals for a cure. Acta Orthop Scand 64(6):699–716PubMedCrossRef Huiskes R (1993) Failed innovation in total hip replacement. Diagnosis and proposals for a cure. Acta Orthop Scand 64(6):699–716PubMedCrossRef
6.
go back to reference Huo MH, Betts F, Bogumill GP, Kenmore PI, Hayek RJ, Martinelli TJ (1993) Metallic wear debris in acetabular osteolysis in a mechanically stable cementless total hip replacement: report of a case. Orthopedics 16(11):1277–1281PubMed Huo MH, Betts F, Bogumill GP, Kenmore PI, Hayek RJ, Martinelli TJ (1993) Metallic wear debris in acetabular osteolysis in a mechanically stable cementless total hip replacement: report of a case. Orthopedics 16(11):1277–1281PubMed
7.
go back to reference Karrholm J, Snorrason F (1992) Migration of porous coated acetabular prostheses fixed with screws: roentgen stereophotogrammetric analysis. J Orthop Res 10(6):826–835PubMedCrossRef Karrholm J, Snorrason F (1992) Migration of porous coated acetabular prostheses fixed with screws: roentgen stereophotogrammetric analysis. J Orthop Res 10(6):826–835PubMedCrossRef
8.
go back to reference Kwong LM, O’Connor DO, Sedlacek RC, Krushell RJ, Maloney WJ, Harris WH (1994) A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component. J Arthroplasty 9(2):163–170PubMedCrossRef Kwong LM, O’Connor DO, Sedlacek RC, Krushell RJ, Maloney WJ, Harris WH (1994) A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component. J Arthroplasty 9(2):163–170PubMedCrossRef
9.
go back to reference Lachiewicz PF, Suh PB, Gilbert JA (1989) In vitro initial fixation of porous-coated acetabular total hip components. A biomechanical comparative study. J Arthroplasty 4(3):201–205PubMedCrossRef Lachiewicz PF, Suh PB, Gilbert JA (1989) In vitro initial fixation of porous-coated acetabular total hip components. A biomechanical comparative study. J Arthroplasty 4(3):201–205PubMedCrossRef
10.
go back to reference Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg Br 84(5):641–646PubMedCrossRef Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg Br 84(5):641–646PubMedCrossRef
11.
go back to reference Litsky AS, Pophal SG (1994) Initial mechanical stability of acetabular prostheses. Orthopedics 17(1):53–57PubMed Litsky AS, Pophal SG (1994) Initial mechanical stability of acetabular prostheses. Orthopedics 17(1):53–57PubMed
12.
go back to reference Markel DC, Hora N, Grimm M (2002) Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems. Int Orthop 26(2):72–75PubMedCrossRef Markel DC, Hora N, Grimm M (2002) Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems. Int Orthop 26(2):72–75PubMedCrossRef
13.
go back to reference Perona PG, Lawrence J, Paprosky WG, Patwardhan AG, Sartori M (1992) Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty. An in vitro comparison of different methods of initial acetabular component fixation. J Arthroplasty 7(4):537–547PubMedCrossRef Perona PG, Lawrence J, Paprosky WG, Patwardhan AG, Sartori M (1992) Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty. An in vitro comparison of different methods of initial acetabular component fixation. J Arthroplasty 7(4):537–547PubMedCrossRef
14.
go back to reference Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113PubMed Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113PubMed
15.
go back to reference Ramamurti BS, Orr TE, Bragdon CR, Lowenstein JD, Jasty M, Harris WH (1997) Factors influencing stability at the interface between a porous surface and cancellous bone: a finite element analysis of a canine in vivo micromotion experiment. J Biomed Mater Res 36(2):274–280PubMedCrossRef Ramamurti BS, Orr TE, Bragdon CR, Lowenstein JD, Jasty M, Harris WH (1997) Factors influencing stability at the interface between a porous surface and cancellous bone: a finite element analysis of a canine in vivo micromotion experiment. J Biomed Mater Res 36(2):274–280PubMedCrossRef
16.
go back to reference Schmalzried TP, Harris WH (1992) The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J Bone Joint Surg Am 74(8):1130–1139PubMed Schmalzried TP, Harris WH (1992) The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J Bone Joint Surg Am 74(8):1130–1139PubMed
17.
go back to reference Soto MO, Rodriguez JA, Ranawat CS (2000) Clinical and radiographic evaluation of the Harris-Galante cup: incidence of wear and osteolysis at 7 to 9 years follow-up. J Arthroplasty 15(2):139–145PubMedCrossRef Soto MO, Rodriguez JA, Ranawat CS (2000) Clinical and radiographic evaluation of the Harris-Galante cup: incidence of wear and osteolysis at 7 to 9 years follow-up. J Arthroplasty 15(2):139–145PubMedCrossRef
18.
go back to reference Spears IR, Pfleiderer M, Schneider E, Hille E, Morlock MM (2001) The effect of interfacial parameters on cup-bone relative micromotions. A finite element investigation. J Biomech 34(1):113–120PubMedCrossRef Spears IR, Pfleiderer M, Schneider E, Hille E, Morlock MM (2001) The effect of interfacial parameters on cup-bone relative micromotions. A finite element investigation. J Biomech 34(1):113–120PubMedCrossRef
19.
go back to reference Stranne SK, Callaghan JJ, Elder SH, Glisson RR, Seaber AV (1991) Screw-augmented fixation of acetabular components. A mechanical model to determine optimal screw placement. J Arthroplasty 6(4):301–305PubMed Stranne SK, Callaghan JJ, Elder SH, Glisson RR, Seaber AV (1991) Screw-augmented fixation of acetabular components. A mechanical model to determine optimal screw placement. J Arthroplasty 6(4):301–305PubMed
20.
go back to reference Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH (1998) Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature. J Biomed Mater Res 43(2):192–203PubMedCrossRef Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH (1998) Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature. J Biomed Mater Res 43(2):192–203PubMedCrossRef
21.
go back to reference Wilson-MacDonald J, Morscher E, Masar Z (1990) Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results. J Bone Joint Surg Br 72(3):423–430PubMed Wilson-MacDonald J, Morscher E, Masar Z (1990) Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results. J Bone Joint Surg Br 72(3):423–430PubMed
22.
go back to reference Won CH, Hearn TC, Tile M (1995) Micromotion of cementless hemispherical acetabular components. Does press-fit need adjunctive screw fixation? J Bone Joint Surg Br 77(3):484–489PubMed Won CH, Hearn TC, Tile M (1995) Micromotion of cementless hemispherical acetabular components. Does press-fit need adjunctive screw fixation? J Bone Joint Surg Br 77(3):484–489PubMed
Metadata
Title
Effects of screw eccentricity on the initial stability of the acetabular cup
Authors
Jui-Ting Hsu
Chih-Han Chang
Kai-Nan An
Mark E. Zobitz
Rapin Phimolsarnti
Ronald R. Hugate
Kuo-An Lai
Publication date
01-08-2007
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 4/2007
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-006-0226-4

Other articles of this Issue 4/2007

International Orthopaedics 4/2007 Go to the issue