Long-term follow-up (11 years plus) results of JRI (Furlong) total hip arthroplasty in young patients: cause for concern regarding acetabular cup?

Authors: D. K. Sharma, Stuart Brooks

Published in: International Orthopaedics | Issue 5/2006

Login to get access

Abstract

We describe the clinical and radiological long-term outcomes of 77 primary total hip replacements in 69 patients using the fully hydroxyapatite-coated JRI (Furlong) total hip replacement. The total cases followed up were 77 hips, performed at a mean duration of 11 years and 2 months. Twelve hips could not be followed up for various reasons, which are discussed in the results section. The mean Harris hip score was 89. Seventeen acetabular cups were revised for aseptic loosening. Only one femoral stem was revised, for fracture. By Engh’s criteria there were a further two unstable cups with no symptoms, and all femoral stems were stable. Kaplan–Meier survivorship analysis revealed a survival of 98.8% for the femoral stem, 78.7% for the acetabular cup, and a combined survival of 77.8% for both components. Our findings suggest that the JRI (Furlong) hip gives a durable femoral stem implant fixation, whereas the prosthesis–bone interface achieved with the acetabular component is questionable.
Literature
1.
go back to reference Bloebaum RD, Bachus KN, Rubman MH, Dorr LD (1993) Post-mortem comparative analysis of titanium and hydroxyapatite porous coated femoral implants retrieved from the same patient. J Arthroplasty 8:203–211PubMed Bloebaum RD, Bachus KN, Rubman MH, Dorr LD (1993) Post-mortem comparative analysis of titanium and hydroxyapatite porous coated femoral implants retrieved from the same patient. J Arthroplasty 8:203–211PubMed
2.
go back to reference Capello WN, D’Antonio JA, Manley MT, Feiberg JR (1998) Hydroxy-apatite in total hip arthroplasty: clinical results and critical issues. Clin Orthop 355:200–211PubMedCrossRef Capello WN, D’Antonio JA, Manley MT, Feiberg JR (1998) Hydroxy-apatite in total hip arthroplasty: clinical results and critical issues. Clin Orthop 355:200–211PubMedCrossRef
3.
go back to reference David A, Eitenmuller J, Muher G et al (1995) Mechanical and histological evaluation of hydroxyapatite-coated, titanium-coated and grit-blasted surfaces under weight-bearing surfaces. Arch Orthop Trauma Surg 114:112–118PubMedCrossRef David A, Eitenmuller J, Muher G et al (1995) Mechanical and histological evaluation of hydroxyapatite-coated, titanium-coated and grit-blasted surfaces under weight-bearing surfaces. Arch Orthop Trauma Surg 114:112–118PubMedCrossRef
4.
go back to reference Donnelly WJ, Kobayashi A, Freeman MA et al (1997) Radiological and survival comparison of four methods of fixation of proximal femoral stem. J Bone Joint Surg [Br] 79:351–360CrossRef Donnelly WJ, Kobayashi A, Freeman MA et al (1997) Radiological and survival comparison of four methods of fixation of proximal femoral stem. J Bone Joint Surg [Br] 79:351–360CrossRef
5.
go back to reference Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement; the factors governing bone ingrowth, stress shielding and clinical results. J Bone Joint Surg [Br] 79:45–55 Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement; the factors governing bone ingrowth, stress shielding and clinical results. J Bone Joint Surg [Br] 79:45–55
6.
go back to reference Jaffe WL, Scott DF (1996) Current concept review: total hip arthroplasty with hydroxyapatite–coated prosthesis. J Bone Joint Surg 78:1918–1934PubMed Jaffe WL, Scott DF (1996) Current concept review: total hip arthroplasty with hydroxyapatite–coated prosthesis. J Bone Joint Surg 78:1918–1934PubMed
7.
go back to reference Furlong RJ, Osborn JF (1991) Fixation of hip prostheses by hydroxyapatite ceramic coating. J Bone Joint Surg 73:741–745 Furlong RJ, Osborn JF (1991) Fixation of hip prostheses by hydroxyapatite ceramic coating. J Bone Joint Surg 73:741–745
8.
go back to reference Havelin LI, Espehaug B, Engesaeter LB (2002) The performance of two hydroxyapatite-coated acetabular cups compared with Charnley cup. J Bone Joint Surg [Br] 84:839–844CrossRef Havelin LI, Espehaug B, Engesaeter LB (2002) The performance of two hydroxyapatite-coated acetabular cups compared with Charnley cup. J Bone Joint Surg [Br] 84:839–844CrossRef
9.
go back to reference Hofmann AA, Bachus KN, Bloebaum (1993) Comparative study of human cancellous bone remodelling to titanium and hydroxyapatite coated implants. J Arthroplasty 8:157–161PubMedCrossRef Hofmann AA, Bachus KN, Bloebaum (1993) Comparative study of human cancellous bone remodelling to titanium and hydroxyapatite coated implants. J Arthroplasty 8:157–161PubMedCrossRef
10.
go back to reference Kuo-An Lai, Wun-Jer Shen, Chyun-Hsiang Chen, Chyun-Yu Yang et al (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll. J Bone Joint Surg [Br] 84:641–646CrossRef Kuo-An Lai, Wun-Jer Shen, Chyun-Hsiang Chen, Chyun-Yu Yang et al (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll. J Bone Joint Surg [Br] 84:641–646CrossRef
11.
go back to reference McNally SA, Shepperd JAN, Mann CV, Walczak JP (2000) The results at nine to twelve years of the use of a hydroxyapatite-coated femoral stem. J Bone Joint Surg [Br] 82:378–382CrossRef McNally SA, Shepperd JAN, Mann CV, Walczak JP (2000) The results at nine to twelve years of the use of a hydroxyapatite-coated femoral stem. J Bone Joint Surg [Br] 82:378–382CrossRef
12.
go back to reference Rahbek O, Overgaard S, Soballe S, Bunger C (1996) Hydroxyapatite coating might prevent peri-implant particle migration: a pilot study in dogs. Acta Ortho Scand 67 (Suppl 267):58–59 Rahbek O, Overgaard S, Soballe S, Bunger C (1996) Hydroxyapatite coating might prevent peri-implant particle migration: a pilot study in dogs. Acta Ortho Scand 67 (Suppl 267):58–59
13.
go back to reference Rokkum M, Brandt M, Bye K et al (1999) Polyethylene wear, osteolysis and acetabular loosening with an HA-coated prosthesis; a follow up of 94 consecutive arthroplasties. J Bone Joint Surg [Br] 81:582–589 Rokkum M, Brandt M, Bye K et al (1999) Polyethylene wear, osteolysis and acetabular loosening with an HA-coated prosthesis; a follow up of 94 consecutive arthroplasties. J Bone Joint Surg [Br] 81:582–589
14.
go back to reference Sharma DK, Brooks S, Creasy S, Lewis B (2004) Fracture of fully hydroxyapatite-coated titanium femoral stem of a total hip replacement—a report of 3 cases. Acta Ortho Scand 75(6):768–771 Sharma DK, Brooks S, Creasy S, Lewis B (2004) Fracture of fully hydroxyapatite-coated titanium femoral stem of a total hip replacement—a report of 3 cases. Acta Ortho Scand 75(6):768–771
15.
go back to reference Shen WJ, Chung KC, Wang GJ, McLaughlin RE (1992) Mechanical failure of hydroxyapatite and polysulfone-coated titanium rods in a weight bearing canine model. J Arthroplasty 7:43–49PubMedCrossRef Shen WJ, Chung KC, Wang GJ, McLaughlin RE (1992) Mechanical failure of hydroxyapatite and polysulfone-coated titanium rods in a weight bearing canine model. J Arthroplasty 7:43–49PubMedCrossRef
16.
go back to reference Soballe K, Hansen ES, Brockstedt-Rasmussen H et al (1991) Bone graft incorporation around titanium alloy and hydroxyapatite-coated implants in dogs. Clin Orthop 272:255–258PubMed Soballe K, Hansen ES, Brockstedt-Rasmussen H et al (1991) Bone graft incorporation around titanium alloy and hydroxyapatite-coated implants in dogs. Clin Orthop 272:255–258PubMed
Metadata
Title
Long-term follow-up (11 years plus) results of JRI (Furlong) total hip arthroplasty in young patients: cause for concern regarding acetabular cup?
Authors
D. K. Sharma
Stuart Brooks
Publication date
01-10-2006
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 5/2006
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-006-0083-1

Other articles of this Issue 5/2006

International Orthopaedics 5/2006 Go to the issue