01-07-2016 | Original Article

Segmental quantitative myocardial perfusion with PET for the detection of significant coronary artery disease in patients with stable angina

Authors: Valentina Berti, Roberto Sciagrà, Danilo Neglia, Mikko Pietilä, Arthur J. Scholte, Stephan Nekolla, François Rouzet, Alberto Pupi, Juhani Knuuti

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 8/2016

Login to get access

Abstract

Purpose

The goal of this study is to determine the technical accuracy of segmental perfusion parameters assessed with quantitative cardiac PET imaging in the evaluation of coronary artery disease (CAD) in patients with stable angina.

Methods

A cohort of patients who participated in the EVINCI protocol underwent an evaluation of coronary anatomy by invasive coronary angiography (ICA) and/or coronary computed tomography angiography (CCTA) and PET myocardial perfusion imaging with H2 15O, 13NH3 or 82Rb. PET studies were analyzed by two independent observers blinded to clinical and instrumental data, and classified as positive or negative for significant CAD using only segmental perfusion measurements and cut-off values from literature.

Results

On a per-patient basis, the overall inter-observer agreement on PET results was 90 % (kappa = 0.79), indicating substantial agreement. On a per-vessel basis, the inter-observer agreement on PET results was 88 % (kappa = 0.74) in the RCA territory, 94 % (kappa = 0.84) in the LAD territory and 94 % (kappa = 0.85) in the LCX territory.
Segmental PET measurements correctly identified 85 % of the patients, resulting in a global sensitivity of 86 %, a specificity of 84 %, a positive predictive value (PPV) of 69 % and a negative predictive value (NPV) of 93 %.
In vessel-based analyses, quantitative perfusion parameters had a sensitivity, specificity, PPV and NPV of 92 %, 82 %, 42 % and 99 %, respectively, for the detection of significant coronary stenoses in all major coronary arteries.

Conclusions

The assessment of absolute myocardial perfusion parameters measured at a segment level lead to reliable and accurate identification of patients with significant coronary stenosis at ICA and/or CCTA.
Literature
1.
go back to reference deKemp RA, Ruddy TD, Hewitt T, Dalipaj MM, Beanlands R. Detection of serial changes in absolute myocardial perfusion with Rb-82 PET. J Nucl Med. 2000;41:1426–35.PubMed deKemp RA, Ruddy TD, Hewitt T, Dalipaj MM, Beanlands R. Detection of serial changes in absolute myocardial perfusion with Rb-82 PET. J Nucl Med. 2000;41:1426–35.PubMed
2.
go back to reference Kuhle WG, Porenta G, Huang SC, Buxton D, Gambhir SS, Hansen H, et al. Quantification of regional myocardial blood-flow using N-13 ammonia and reoriented dynamic positron emission tomographic imaging. Circulation. 1992;86:1004–17.CrossRefPubMed Kuhle WG, Porenta G, Huang SC, Buxton D, Gambhir SS, Hansen H, et al. Quantification of regional myocardial blood-flow using N-13 ammonia and reoriented dynamic positron emission tomographic imaging. Circulation. 1992;86:1004–17.CrossRefPubMed
3.
go back to reference DiCarli M, Czernin J, Hoh CK, Gerbaudo VH, Brunken RC, Huang SC, et al. Relation among stenosis severity, myocardial blood-flow, and flow reserve in patients with coronary-artery disease. Circulation. 1995;91:1944–51.CrossRef DiCarli M, Czernin J, Hoh CK, Gerbaudo VH, Brunken RC, Huang SC, et al. Relation among stenosis severity, myocardial blood-flow, and flow reserve in patients with coronary-artery disease. Circulation. 1995;91:1944–51.CrossRef
4.
go back to reference Ahn JY, Lee DS, Lee JS, Kim SK, Cheon GJ, Yeo JS, et al. Quantification of regional myocardial blood flow using dynamic H2(15)O PET and factor analysis. J Nucl Med. 2001;42:782–7.PubMed Ahn JY, Lee DS, Lee JS, Kim SK, Cheon GJ, Yeo JS, et al. Quantification of regional myocardial blood flow using dynamic H2(15)O PET and factor analysis. J Nucl Med. 2001;42:782–7.PubMed
5.
go back to reference Fiechter M, Ghadri JR, Gebhard C, Fuchs TA, Pazhenkottil AP, Nkoulou RN, et al. Diagnostic value of 13N-ammonia myocardial perfusion PET: added value of myocardial flow reserve. J Nucl Med. 2012;53:1230–4.CrossRefPubMed Fiechter M, Ghadri JR, Gebhard C, Fuchs TA, Pazhenkottil AP, Nkoulou RN, et al. Diagnostic value of 13N-ammonia myocardial perfusion PET: added value of myocardial flow reserve. J Nucl Med. 2012;53:1230–4.CrossRefPubMed
6.
go back to reference Schindler TH, Schelbert HR, Quercioli A, Dilsizian V. Cardiac PET imaging for the detection and monitoring of coronary artery disease and microvascular health. JACC Cardiovasc Imaging. 2010;3:623–40.CrossRefPubMed Schindler TH, Schelbert HR, Quercioli A, Dilsizian V. Cardiac PET imaging for the detection and monitoring of coronary artery disease and microvascular health. JACC Cardiovasc Imaging. 2010;3:623–40.CrossRefPubMed
7.
go back to reference Schindler TH, Quercioli A, Valenta I, Ambrosio G, Wahl RL, Dilsizian V. Quantitative assessment of myocardial blood flow—clinical and research applications. Semin Nucl Med. 2014;44:274–93.CrossRefPubMed Schindler TH, Quercioli A, Valenta I, Ambrosio G, Wahl RL, Dilsizian V. Quantitative assessment of myocardial blood flow—clinical and research applications. Semin Nucl Med. 2014;44:274–93.CrossRefPubMed
8.
go back to reference Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography: added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150–6.CrossRefPubMed Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography: added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150–6.CrossRefPubMed
10.
go back to reference Neglia D, Michelassi C, Trivieri MG, Sambuceti G, Giorgetti A, Pratali L, et al. Prognostic role of myocardial blood flow impairment in idiopathic left ventricular dysfunction. Circulation. 2002;105:186–93.CrossRefPubMed Neglia D, Michelassi C, Trivieri MG, Sambuceti G, Giorgetti A, Pratali L, et al. Prognostic role of myocardial blood flow impairment in idiopathic left ventricular dysfunction. Circulation. 2002;105:186–93.CrossRefPubMed
11.
go back to reference Saraste A, Kajander S, Han C, Nesterov SV, Knuuti J. PET: is myocardial flow quantification a clinical reality? J Nucl Cardiol. 2012;19:1044–59.CrossRefPubMed Saraste A, Kajander S, Han C, Nesterov SV, Knuuti J. PET: is myocardial flow quantification a clinical reality? J Nucl Cardiol. 2012;19:1044–59.CrossRefPubMed
12.
go back to reference Nesterov SV, Deshayes E, Sciagrà R, Settimo L, Declerck JM, Pan X-B, et al. Quantification of myocardial blood flow in absolute terms using 82Rb PET imaging - results of RUBY-10 study. JACC Cardiovasc Imaging. 2014;7:1119–27.CrossRefPubMedPubMedCentral Nesterov SV, Deshayes E, Sciagrà R, Settimo L, Declerck JM, Pan X-B, et al. Quantification of myocardial blood flow in absolute terms using 82Rb PET imaging - results of RUBY-10 study. JACC Cardiovasc Imaging. 2014;7:1119–27.CrossRefPubMedPubMedCentral
13.
go back to reference Yoshinaga K, Katoh C, Manabe O, Klein R, Naya M, Sakakibara M, et al. Incremental diagnostic value of regional myocardial blood flow quantification over relative perfusion imaging with generator-produced rubidium-82 PET. Circ J. 2011;75:2628–34.CrossRefPubMed Yoshinaga K, Katoh C, Manabe O, Klein R, Naya M, Sakakibara M, et al. Incremental diagnostic value of regional myocardial blood flow quantification over relative perfusion imaging with generator-produced rubidium-82 PET. Circ J. 2011;75:2628–34.CrossRefPubMed
14.
go back to reference Kajander SA, Joutsiniemi E, Saraste M, Pietila M, Ukkonen H, Saraste A, et al. Clinical value of absolute quantification of myocardial perfusion with 15O-water in coronary artery disease. Circ Cardiovasc Imaging. 2011;4:678–84.CrossRefPubMed Kajander SA, Joutsiniemi E, Saraste M, Pietila M, Ukkonen H, Saraste A, et al. Clinical value of absolute quantification of myocardial perfusion with 15O-water in coronary artery disease. Circ Cardiovasc Imaging. 2011;4:678–84.CrossRefPubMed
15.
go back to reference Liga R, Marini C, Coceani M, Filidei E, Schlueter M, Bianchi M, et al. Structural abnormalities of the coronary arterial wall--in addition to luminal narrowing--affect myocardial blood flow reserve. J Nucl Med. 2011;52:1704–12.CrossRefPubMed Liga R, Marini C, Coceani M, Filidei E, Schlueter M, Bianchi M, et al. Structural abnormalities of the coronary arterial wall--in addition to luminal narrowing--affect myocardial blood flow reserve. J Nucl Med. 2011;52:1704–12.CrossRefPubMed
16.
go back to reference Parkash R, deKemp RA, Ruddy TD, Kitsikis A, Beauschene L, Williams K, et al. Potential utility of rubidium 82 PET quantification in patients with 3-vessel coronary artery disease. J Nucl Cardiol. 2004;11:440–9.CrossRefPubMed Parkash R, deKemp RA, Ruddy TD, Kitsikis A, Beauschene L, Williams K, et al. Potential utility of rubidium 82 PET quantification in patients with 3-vessel coronary artery disease. J Nucl Cardiol. 2004;11:440–9.CrossRefPubMed
17.
go back to reference Hajjiri MM, Leavitt MB, Zheng H, Spooner AE, Fischman AJ, Gewirtz H. Comparison of positron emission tomography measurement of adenosine-stimulated absolute myocardial blood flow versus relative myocardial tracer content for physiological assessment of coronary artery stenosis severity and location. JACC Cardiovasc Imaging. 2009;2:751–8.CrossRefPubMed Hajjiri MM, Leavitt MB, Zheng H, Spooner AE, Fischman AJ, Gewirtz H. Comparison of positron emission tomography measurement of adenosine-stimulated absolute myocardial blood flow versus relative myocardial tracer content for physiological assessment of coronary artery stenosis severity and location. JACC Cardiovasc Imaging. 2009;2:751–8.CrossRefPubMed
18.
go back to reference Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG. Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med. 1999;40:1848–56.PubMed Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG. Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med. 1999;40:1848–56.PubMed
19.
go back to reference El Fakhri G, Kardan A, Sitek A, Dorbala S, Abi-Hatem N, Lahoud Y, et al. Reproducibility and accuracy of quantitative myocardial blood flow assessment with 82Rb PET: comparison with 13N-ammonia PET. J Nucl Med. 2009;50:1062–71.CrossRefPubMedPubMedCentral El Fakhri G, Kardan A, Sitek A, Dorbala S, Abi-Hatem N, Lahoud Y, et al. Reproducibility and accuracy of quantitative myocardial blood flow assessment with 82Rb PET: comparison with 13N-ammonia PET. J Nucl Med. 2009;50:1062–71.CrossRefPubMedPubMedCentral
20.
go back to reference Schindler TH, Zhang X-L, Prior JO, Cadenas J, Dahlbom M, Sayre J, et al. Assessment of intra- and interobserver reproducibility of rest and cold pressor test-stimulated myocardial blood flow with 13N-ammonia and PET. Eur J Nucl Med Mol Imaging. 2007;34:1178–88.CrossRefPubMed Schindler TH, Zhang X-L, Prior JO, Cadenas J, Dahlbom M, Sayre J, et al. Assessment of intra- and interobserver reproducibility of rest and cold pressor test-stimulated myocardial blood flow with 13N-ammonia and PET. Eur J Nucl Med Mol Imaging. 2007;34:1178–88.CrossRefPubMed
21.
go back to reference Harms HJ, Nesterov SV, Han C, Danad I, Leonora R, Raijmakers PG, et al. Comparison of clinical non-commercial tools for automated quantification of myocardial blood flow using oxygen-15-labelled water PET/CT. Eur Heart J Cardiovasc Imaging. 2014;15:431–41.CrossRefPubMed Harms HJ, Nesterov SV, Han C, Danad I, Leonora R, Raijmakers PG, et al. Comparison of clinical non-commercial tools for automated quantification of myocardial blood flow using oxygen-15-labelled water PET/CT. Eur Heart J Cardiovasc Imaging. 2014;15:431–41.CrossRefPubMed
22.
go back to reference Slomka PJ, Alexanderson E, Jacome R, Jimenez M, Romero E, Meave A, et al. Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171–81.CrossRefPubMed Slomka PJ, Alexanderson E, Jacome R, Jimenez M, Romero E, Meave A, et al. Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171–81.CrossRefPubMed
23.
go back to reference Nitzsche EU, Choi Y, Czernin J, Hoh CK, Huang SC, Schelbert HR. Noninvasive quantification of myocardial blood flow in humans. A direct comparison of the [13N]ammonia and the [15O]water techniques. Circulation. 1996;93:2000–6.CrossRefPubMed Nitzsche EU, Choi Y, Czernin J, Hoh CK, Huang SC, Schelbert HR. Noninvasive quantification of myocardial blood flow in humans. A direct comparison of the [13N]ammonia and the [15O]water techniques. Circulation. 1996;93:2000–6.CrossRefPubMed
24.
go back to reference Prior JO, Allenbach G, Valenta I, Kosinski M, Burger C, Verdun FR, et al. Quantification of myocardial blood flow with 82Rb positron emission tomography: clinical validation with 15O-water. Eur J Nucl Med Mol Imaging. 2012;39:1037–47.CrossRefPubMedPubMedCentral Prior JO, Allenbach G, Valenta I, Kosinski M, Burger C, Verdun FR, et al. Quantification of myocardial blood flow with 82Rb positron emission tomography: clinical validation with 15O-water. Eur J Nucl Med Mol Imaging. 2012;39:1037–47.CrossRefPubMedPubMedCentral
25.
go back to reference Kajander S, Joutsiniemi E, Saraste M, Pietila M, Ukkonen H, Saraste A, et al. Cardiac positron emission tomography/computed tomography imaging accurately detects anatomically and functionally significant coronary artery disease. Circulation. 2010;122:603–13.CrossRefPubMed Kajander S, Joutsiniemi E, Saraste M, Pietila M, Ukkonen H, Saraste A, et al. Cardiac positron emission tomography/computed tomography imaging accurately detects anatomically and functionally significant coronary artery disease. Circulation. 2010;122:603–13.CrossRefPubMed
26.
go back to reference Danad I, Uusitalo V, Kero T, Saraste A, Raijmakers PG, Lammertsma AA, et al. Quantitative assessment ofmyocardial perfusion in the detection of significant coronary artery disease. J Am Coll Cardiol. 2014;64:1464–75.CrossRefPubMed Danad I, Uusitalo V, Kero T, Saraste A, Raijmakers PG, Lammertsma AA, et al. Quantitative assessment ofmyocardial perfusion in the detection of significant coronary artery disease. J Am Coll Cardiol. 2014;64:1464–75.CrossRefPubMed
27.
go back to reference Neglia D, Rovai D, Caselli C, Pietilä M, Teresinska A, Aguadé-Bruix S, et al. Detection of significant coronary artery disease by noninvasive anatomical and functional imaging. Circ Cardiovasc Imaging. 2015. doi:10.1161/CIRCIMAGING.114.002179.PubMed Neglia D, Rovai D, Caselli C, Pietilä M, Teresinska A, Aguadé-Bruix S, et al. Detection of significant coronary artery disease by noninvasive anatomical and functional imaging. Circ Cardiovasc Imaging. 2015. doi:10.​1161/​CIRCIMAGING.​114.​002179.PubMed
28.
go back to reference Hesse B, Tagil K, Cuocolo A, Anagnostopoulos C, Bardies M, Bax J, et al. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005;32:855–97.CrossRefPubMed Hesse B, Tagil K, Cuocolo A, Anagnostopoulos C, Bardies M, Bax J, et al. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005;32:855–97.CrossRefPubMed
29.
go back to reference Hermansen F, Rosen SD, Fath-Ordoubadi F, Kooner JS, Clark JC, Camici PG, et al. Measurement of myocardial blood flow with oxygen-15 labelled water: comparison of different administration protocols. Eur J Nucl Med Mol Imaging. 1998;25:751–9.CrossRef Hermansen F, Rosen SD, Fath-Ordoubadi F, Kooner JS, Clark JC, Camici PG, et al. Measurement of myocardial blood flow with oxygen-15 labelled water: comparison of different administration protocols. Eur J Nucl Med Mol Imaging. 1998;25:751–9.CrossRef
30.
go back to reference DeGrado T, Hanson M, Turkington T, DeLong D, Brezinski D, Vallee J, et al. Estimation of myocardial blood flow for longitudinal studies with 13N-labeled ammonia and positron emission tomography1. J Nucl Cardiol. 1996;3:494–507.CrossRefPubMed DeGrado T, Hanson M, Turkington T, DeLong D, Brezinski D, Vallee J, et al. Estimation of myocardial blood flow for longitudinal studies with 13N-labeled ammonia and positron emission tomography1. J Nucl Cardiol. 1996;3:494–507.CrossRefPubMed
31.
go back to reference Lortie M, Beanlands RSB, Yoshinaga K, Klein R, DaSilva JN, deKemp RA. Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging. 2007;34:1765–74.CrossRefPubMed Lortie M, Beanlands RSB, Yoshinaga K, Klein R, DaSilva JN, deKemp RA. Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging. 2007;34:1765–74.CrossRefPubMed
32.
go back to reference Cerqueira MD. Standardized Myocardial Segmentation and Nomenclature for Tomographic Imaging of the Heart: A Statement for Healthcare Professionals From the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.CrossRefPubMed Cerqueira MD. Standardized Myocardial Segmentation and Nomenclature for Tomographic Imaging of the Heart: A Statement for Healthcare Professionals From the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.CrossRefPubMed
33.
go back to reference Anagnostopoulos C, Almonacid A, El Fakhri G, Curillova Z, Sitek A, Roughton M, et al. Quantitative relationship between coronary vasodilator reserve assessed by 82Rb PET imaging and coronary artery stenosis severity. Eur J Nucl Med Mol Imaging. 2008;35:1593–601.CrossRefPubMedPubMedCentral Anagnostopoulos C, Almonacid A, El Fakhri G, Curillova Z, Sitek A, Roughton M, et al. Quantitative relationship between coronary vasodilator reserve assessed by 82Rb PET imaging and coronary artery stenosis severity. Eur J Nucl Med Mol Imaging. 2008;35:1593–601.CrossRefPubMedPubMedCentral
34.
go back to reference Johnson NP, Gould KL. Physiological basis for Angina and ST-segment change. JACC Cardiovasc Imaging. 2011;4:990–8.CrossRefPubMed Johnson NP, Gould KL. Physiological basis for Angina and ST-segment change. JACC Cardiovasc Imaging. 2011;4:990–8.CrossRefPubMed
35.
go back to reference Bluemke DA, Achenbach S, Budoff M, Gerber TC, Gersh B, Hillis LD, et al. Noninvasive coronary artery Imaging - Magnetic resonance angiography and multidetector computed tomography angiography - A scientific statement from the American Heart Association Committee on cardiovascular imaging and intervention of the council on cardiovascular radiology and intervention, and the councils on clinical cardiology and cardiovascular disease in the young. Circulation. 2008;118:586–606.CrossRefPubMed Bluemke DA, Achenbach S, Budoff M, Gerber TC, Gersh B, Hillis LD, et al. Noninvasive coronary artery Imaging - Magnetic resonance angiography and multidetector computed tomography angiography - A scientific statement from the American Heart Association Committee on cardiovascular imaging and intervention of the council on cardiovascular radiology and intervention, and the councils on clinical cardiology and cardiovascular disease in the young. Circulation. 2008;118:586–606.CrossRefPubMed
36.
go back to reference Chow BJW, Abraham A, Wells GA, Chen L, Ruddy TD, Yam Y, et al. Diagnostic accuracy and impact of computed tomographic coronary angiography on utilization of invasive coronary angiography. Circ Cardiovasc Imaging. 2009;2:16–23.CrossRefPubMed Chow BJW, Abraham A, Wells GA, Chen L, Ruddy TD, Yam Y, et al. Diagnostic accuracy and impact of computed tomographic coronary angiography on utilization of invasive coronary angiography. Circ Cardiovasc Imaging. 2009;2:16–23.CrossRefPubMed
37.
go back to reference Sciagrà R. Quantitative cardiac positron emission tomography: the time is coming! Scientifica. 2012;2012:1–16.CrossRef Sciagrà R. Quantitative cardiac positron emission tomography: the time is coming! Scientifica. 2012;2012:1–16.CrossRef
38.
go back to reference Danad I, Raijmakers PG, Appelman YE, Harms HJ, de Haan S, van den Oever MLP, et al. Hybrid imaging using quantitative H215O PET and CT-based coronary angiography for the detection of coronary artery disease. J Nucl Med. 2012;54:55–63.CrossRefPubMed Danad I, Raijmakers PG, Appelman YE, Harms HJ, de Haan S, van den Oever MLP, et al. Hybrid imaging using quantitative H215O PET and CT-based coronary angiography for the detection of coronary artery disease. J Nucl Med. 2012;54:55–63.CrossRefPubMed
39.
go back to reference Neglia D, L'Abbate A. Myocardial perfusion reserve in ischemic heart disease. J Nucl Med. 2009;50:175–7.CrossRefPubMed Neglia D, L'Abbate A. Myocardial perfusion reserve in ischemic heart disease. J Nucl Med. 2009;50:175–7.CrossRefPubMed
40.
go back to reference Gould KL, Johnson NP, Bateman TM, Beanlands RS, Bengel FM, Bober R, et al. Anatomic versus physiologic assessment of coronary artery disease. J Am Coll Cardiol. 2013;62:1639–53.CrossRefPubMed Gould KL, Johnson NP, Bateman TM, Beanlands RS, Bengel FM, Bober R, et al. Anatomic versus physiologic assessment of coronary artery disease. J Am Coll Cardiol. 2013;62:1639–53.CrossRefPubMed
41.
go back to reference Johnson NP, Kirkeeide RL, Gould KL. Is discordance of coronary flow reserve and fractional flow reserve due to methodology or clinically relevant coronary pathophysiology? JACC Cardiovasc Imaging. 2012;5:193–202.CrossRefPubMed Johnson NP, Kirkeeide RL, Gould KL. Is discordance of coronary flow reserve and fractional flow reserve due to methodology or clinically relevant coronary pathophysiology? JACC Cardiovasc Imaging. 2012;5:193–202.CrossRefPubMed
42.
go back to reference Reddy KG, Nair RN, Sheehan HM, Hodgson JM. Evidence that selective endothelial dysfunction may occur in the absence of angiographic or ultrasound atherosclerosis in patients with risk factors for atherosclerosis. J Am Coll Cardiol. 1994;23:833–43.CrossRefPubMed Reddy KG, Nair RN, Sheehan HM, Hodgson JM. Evidence that selective endothelial dysfunction may occur in the absence of angiographic or ultrasound atherosclerosis in patients with risk factors for atherosclerosis. J Am Coll Cardiol. 1994;23:833–43.CrossRefPubMed
43.
go back to reference Czernin J, Muller P, Chan S, Brunken RC, Porenta G, Krivokapich J, et al. Influence of age and hemodynamics on myocardial blood-flow and flow reserve. Circulation. 1993;88:62–9.CrossRefPubMed Czernin J, Muller P, Chan S, Brunken RC, Porenta G, Krivokapich J, et al. Influence of age and hemodynamics on myocardial blood-flow and flow reserve. Circulation. 1993;88:62–9.CrossRefPubMed
44.
go back to reference Javadi MS, Lautamäki R, Merrill J, Voicu C, Epley W, McBride G, et al. Definition of vascular territories on myocardial perfusion images by integration with true coronary anatomy: a hybrid PET/CT analysis. J Nucl Med. 2010;51:198–203.CrossRefPubMed Javadi MS, Lautamäki R, Merrill J, Voicu C, Epley W, McBride G, et al. Definition of vascular territories on myocardial perfusion images by integration with true coronary anatomy: a hybrid PET/CT analysis. J Nucl Med. 2010;51:198–203.CrossRefPubMed
Metadata
Title
Segmental quantitative myocardial perfusion with PET for the detection of significant coronary artery disease in patients with stable angina
Authors
Valentina Berti
Roberto Sciagrà
Danilo Neglia
Mikko Pietilä
Arthur J. Scholte
Stephan Nekolla
François Rouzet
Alberto Pupi
Juhani Knuuti
Publication date
01-07-2016
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 8/2016
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-016-3362-0

Other articles of this Issue 8/2016

European Journal of Nuclear Medicine and Molecular Imaging 8/2016 Go to the issue