Published in: European Journal of Nuclear Medicine and Molecular Imaging 2/2012

01-02-2012 | Original Article

Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging

Authors: Philip Aschoff, Christian Plathow, Thomas Beyer, Matthias P. Lichy, Gunter Erb, Mehmet Ö. Öksüz, Claus D. Claussen, Christina Pfannenberg

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 2/2012

Login to get access

Abstract

Purpose

State-of-the-art positron emission tomography/computed tomography (PET/CT) systems incorporate multislice CT technology, thus facilitating the acquisition of multiphase, contrast-enhanced CT data as part of integrated PET/CT imaging protocols. We assess the influence of a highly concentrated iodinated contrast medium (CM) on quantification and image quality following CT-based attenuation correction (CT-AC) in PET/CT.

Methods

Twenty-eight patients with suspected malignant liver lesions were enrolled prospectively. PET/CT was performed 60 min after injection of 400 MBq of 18F-fluorodeoxyglucose (FDG) and following the biphasic administration of an intravenous CM (400 mg iodine/ml, Iomeron 400). PET images were reconstructed with CT-AC using any of four acquired CT image sets: non-enhanced, pre-contrast (n-PET), arterial phase (art-PET), portal venous phase (pv-PET) and late phase (late-PET). Normal tissue activity and liver lesions were assessed visually and quantitatively on each PET/CT image set.

Results

Visual assessment of PET following CT-AC revealed no noticeable difference in image appearance or quality when using any of the four CT data sets for CT-AC. A total of 44 PET-positive liver lesions was identified in 21 of 28 patients. There were no false-negative or false-positive lesions on PET. Mean standardized uptake values (SUV) in 36 evaluable lesions were: 5.5 (n-PET), 5.8 (art-PET), 5.8 (pv-PET) and 5.8 (late-PET), with the highest mean increase in mean SUV of 6%. Mean SUV changes in liver background increased by up to 10% from n-PET to pv-PET.

Conclusion

Multiphase CT data acquired with the use of highly concentrated CM can be used for qualitative assessment of liver lesions in torso FDG PET/CT. The influence on quantification of FDG uptake is small and negligible for most clinical applications.
Literature
1.
go back to reference Czernin J, Allen-Auerbach M, Schelbert HR. Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. J Nucl Med 2007;48 Suppl 1:78S–88S.PubMed Czernin J, Allen-Auerbach M, Schelbert HR. Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. J Nucl Med 2007;48 Suppl 1:78S–88S.PubMed
2.
go back to reference Antoch G, Freudenberg LS, Beyer T, Bockisch A, Debatin JF. To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT. J Nucl Med 2004;45 Suppl 1:56S–65S.PubMed Antoch G, Freudenberg LS, Beyer T, Bockisch A, Debatin JF. To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT. J Nucl Med 2004;45 Suppl 1:56S–65S.PubMed
3.
go back to reference Wong T, Paulson EK, Nelson RC, Patz Jr EF, Coleman RE. Practical approach to diagnostic CT combined with PET. AJR Am J Roentgenol 2007;188:622–9.PubMedCrossRef Wong T, Paulson EK, Nelson RC, Patz Jr EF, Coleman RE. Practical approach to diagnostic CT combined with PET. AJR Am J Roentgenol 2007;188:622–9.PubMedCrossRef
4.
go back to reference Kuehl H, Antoch G. How much CT do we need for PET/CT? A radiologist’s perspective. Nuklearmedizin 2005;44 Suppl 1:S24–31.PubMed Kuehl H, Antoch G. How much CT do we need for PET/CT? A radiologist’s perspective. Nuklearmedizin 2005;44 Suppl 1:S24–31.PubMed
5.
go back to reference Pfannenberg A, Aschoff P, Brechtel K, Müller M, Klein M, Bares R, et al. Value of contrast-enhanced multiphase CT in combined PET/CT protocols for oncological imaging. Br J Radiol 2007;80:437–45.PubMedCrossRef Pfannenberg A, Aschoff P, Brechtel K, Müller M, Klein M, Bares R, et al. Value of contrast-enhanced multiphase CT in combined PET/CT protocols for oncological imaging. Br J Radiol 2007;80:437–45.PubMedCrossRef
6.
go back to reference Morimoto T, Tateishi U, Maeda T, Arai Y, Nakajima Y, Edmund Kim E. Nodal status of malignant lymphoma in pelvic and retroperitoneal lymphatic pathways: comparison of integrated PET/CT with or without contrast enhancement. Eur J Radiol 2008;67:508–13.PubMedCrossRef Morimoto T, Tateishi U, Maeda T, Arai Y, Nakajima Y, Edmund Kim E. Nodal status of malignant lymphoma in pelvic and retroperitoneal lymphatic pathways: comparison of integrated PET/CT with or without contrast enhancement. Eur J Radiol 2008;67:508–13.PubMedCrossRef
7.
go back to reference Strobel K, Bode B, Dummer R, Veit-Haibach P, Fischer DR, Imhof L, et al. Limited value of 18F-FDG PET/CT and S-100B tumour marker in the detection of liver metastases from uveal melanoma compared to liver metastases from cutaneous melanoma. Eur J Nucl Med Mol Imaging 2009;36:1774–82.PubMedCrossRef Strobel K, Bode B, Dummer R, Veit-Haibach P, Fischer DR, Imhof L, et al. Limited value of 18F-FDG PET/CT and S-100B tumour marker in the detection of liver metastases from uveal melanoma compared to liver metastases from cutaneous melanoma. Eur J Nucl Med Mol Imaging 2009;36:1774–82.PubMedCrossRef
8.
go back to reference Soyka JD, Veit-Haibach P, Strobel K, Breitenstein S, Tschopp A, Mende KA, et al. Staging pathways in recurrent colorectal carcinoma: is contrast-enhanced 18F-FDG PET/CT the diagnostic tool of choice? J Nucl Med 2008;49:354–61.PubMedCrossRef Soyka JD, Veit-Haibach P, Strobel K, Breitenstein S, Tschopp A, Mende KA, et al. Staging pathways in recurrent colorectal carcinoma: is contrast-enhanced 18F-FDG PET/CT the diagnostic tool of choice? J Nucl Med 2008;49:354–61.PubMedCrossRef
9.
go back to reference Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25:2046–53.PubMedCrossRef Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25:2046–53.PubMedCrossRef
10.
go back to reference Hany TF, Steinert HC, Goerres GW, Buck A, von Schulthess GK. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology 2002;225:575–81.PubMedCrossRef Hany TF, Steinert HC, Goerres GW, Buck A, von Schulthess GK. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology 2002;225:575–81.PubMedCrossRef
11.
go back to reference Antoch G, Freudenberg LS, Egelhof T, Stattaus J, Jentzen W, Debatin JF, et al. Focal tracer uptake: a potential artifact in contrast-enhanced dual-modality PET/CT scans. J Nucl Med 2002;43:1339–42.PubMed Antoch G, Freudenberg LS, Egelhof T, Stattaus J, Jentzen W, Debatin JF, et al. Focal tracer uptake: a potential artifact in contrast-enhanced dual-modality PET/CT scans. J Nucl Med 2002;43:1339–42.PubMed
12.
go back to reference Beyer T, Antoch G, Bockisch A, Stattaus J. Optimized intravenous contrast administration for diagnostic whole-body 18F-FDG PET/CT. J Nucl Med 2005;46:429–35.PubMed Beyer T, Antoch G, Bockisch A, Stattaus J. Optimized intravenous contrast administration for diagnostic whole-body 18F-FDG PET/CT. J Nucl Med 2005;46:429–35.PubMed
13.
go back to reference Mawlawi O, Erasmus JJ, Munden RF, Pan T, Knight AE, Macapinlac HA, et al. Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation. AJR Am J Roentgenol 2006;186:308–9.PubMedCrossRef Mawlawi O, Erasmus JJ, Munden RF, Pan T, Knight AE, Macapinlac HA, et al. Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation. AJR Am J Roentgenol 2006;186:308–9.PubMedCrossRef
14.
go back to reference von Schulthess GK. Cost considerations regarding an integrated CT-PET system. Eur Radiol 2000;10:S377–80.CrossRef von Schulthess GK. Cost considerations regarding an integrated CT-PET system. Eur Radiol 2000;10:S377–80.CrossRef
15.
go back to reference Kinahan P, Hasegawa B, Beyer T. X-ray based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166–79.PubMedCrossRef Kinahan P, Hasegawa B, Beyer T. X-ray based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166–79.PubMedCrossRef
16.
go back to reference Nakamoto Y, Osman M, Cohade C, Marshall LT, Links JM, Kohlmyer S, et al. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuation-corrected images. J Nucl Med 2002;43:1137–43.PubMed Nakamoto Y, Osman M, Cohade C, Marshall LT, Links JM, Kohlmyer S, et al. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuation-corrected images. J Nucl Med 2002;43:1137–43.PubMed
17.
go back to reference Yau YY, Chan WS, Tam YM, Vernon P, Wong S, Coel M, et al. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error? J Nucl Med 2005;46:283–91.PubMed Yau YY, Chan WS, Tam YM, Vernon P, Wong S, Coel M, et al. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error? J Nucl Med 2005;46:283–91.PubMed
18.
go back to reference Heusner T, Kuehl H, Veit-Haibach P, Hahn S, Boy C, Forsting M, et al. Highly iodinated intravenous contrast material for PET/CT - a feasibility study. Rofo 2008;180:740–5.PubMedCrossRef Heusner T, Kuehl H, Veit-Haibach P, Hahn S, Boy C, Forsting M, et al. Highly iodinated intravenous contrast material for PET/CT - a feasibility study. Rofo 2008;180:740–5.PubMedCrossRef
19.
go back to reference Berthelsen AK, Holm S, Loft A, Klausen TL, Andersen F, Højgaard L. PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients. Eur J Nucl Med Mol Imaging 2005;32:1167–75.PubMedCrossRef Berthelsen AK, Holm S, Loft A, Klausen TL, Andersen F, Højgaard L. PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients. Eur J Nucl Med Mol Imaging 2005;32:1167–75.PubMedCrossRef
20.
21.
go back to reference Awai K, Takada K, Onishi H, Hori S. Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT. Radiology 2002;224:757–63.PubMedCrossRef Awai K, Takada K, Onishi H, Hori S. Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT. Radiology 2002;224:757–63.PubMedCrossRef
22.
go back to reference Marchianò A, Spreafico C, Lanocita R, Frigerio L, Di Tolla G, Patelli G, et al. Does iodine concentration affect the diagnostic efficacy of biphasic spiral CT in patients with hepatocellular carcinoma? Abdom Imaging 2005;30:274–80.PubMedCrossRef Marchianò A, Spreafico C, Lanocita R, Frigerio L, Di Tolla G, Patelli G, et al. Does iodine concentration affect the diagnostic efficacy of biphasic spiral CT in patients with hepatocellular carcinoma? Abdom Imaging 2005;30:274–80.PubMedCrossRef
23.
go back to reference Brechtel K, Klein M, Vogel M, Mueller M, Aschoff P, Beyer T, et al. Optimized contrast-enhanced CT protocols for diagnostic whole-body 18F-FDG PET/CT: technical aspects of single-phase versus multiphase CT imaging. J Nucl Med 2006;47:470–6.PubMed Brechtel K, Klein M, Vogel M, Mueller M, Aschoff P, Beyer T, et al. Optimized contrast-enhanced CT protocols for diagnostic whole-body 18F-FDG PET/CT: technical aspects of single-phase versus multiphase CT imaging. J Nucl Med 2006;47:470–6.PubMed
24.
go back to reference Miles KA. PET-CT in oncology: making the most of CT. Cancer Imaging 2008;4(8 Spec No A):S87–93.CrossRef Miles KA. PET-CT in oncology: making the most of CT. Cancer Imaging 2008;4(8 Spec No A):S87–93.CrossRef
25.
go back to reference Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med 2009;50:11S–20S.PubMedCrossRef Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med 2009;50:11S–20S.PubMedCrossRef
Metadata
Title
Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging
Authors
Philip Aschoff
Christian Plathow
Thomas Beyer
Matthias P. Lichy
Gunter Erb
Mehmet Ö. Öksüz
Claus D. Claussen
Christina Pfannenberg
Publication date
01-02-2012
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 2/2012
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-011-1919-5

Other articles of this Issue 2/2012

European Journal of Nuclear Medicine and Molecular Imaging 2/2012 Go to the issue