Comparison of fixed-flexion positioning with fluoroscopic semi-flexed positioning for quantifying radiographic joint-space width in the knee: test-retest reproducibility

Authors: C. Peterfy, J. Li, S. Zaim, J. Duryea, J. Lynch, Y. Miaux, W. Yu, H. K. Genant

Published in: Skeletal Radiology | Issue 3/2003

Login to get access

Abstract

Objective

To compare fixed-flexion radiography of the knee with fluoroscopic semi-flexed radiography in terms of the reproducibility of measurements of minimum joint-space width (JSW) in the medial femorotibial joint.

Methods

Posteroanterior radiographs of the right knees of 18 normal volunteers were acquired with the patients standing on an upright fluoroscopy table, the feet externally rotated 10° and the toes touching the vertical table. Knees were positioned and radiographed with two different techniques: (1) semi-flexed positioning under fluoroscopic guidance using a horizontal X-ray beam; and (2) fixed-flexion positioning, with the knees and thighs touching the vertical table, using 10° caudal beam angulation without fluoroscopy. Foot maps were drawn in each case. Subjects were repositioned and radiographed twice using each technique. The posteroanterior beam angle that optimally projected the medial tibia plateau with the patient in the fixed-flexion position was also determined for each subject in a separate examination using fluoroscopy. Ten patients with osteoarthritis were also examined with the fixed-flexion technique using a conventional radiographic unit. Minimum medial joint-space width (JSW) in the medial femorotibial joint was measured manually with a graduated lens and also with a semi-automated computer algorithm.

Results

Reproducibility errors (root-mean-square SD) for manual and automated JSW measurement were 0.2 mm and 0.1 mm, respectively, for fluoroscopic semi-flexed positioning in volunteers; 0.3 mm and 0.1 mm, respectively, for fixed-flexion positioning in volunteers; and 0.2 mm and 0.1 mm, respectively, for fixed-flexion positioning in osteoarthritic patients. The optimal beam angle for visualizing the joint space was 9.0°±3.6°.

Conclusion

Fixed-flexion, non-fluoroscopic radiography of the knee can provide reproducible JSW measurement using widely available X-ray equipment. This technique is more feasible for multicenter clinical studies and routine clinical use than are methods that rely on fluoroscopic alignment of the tibial plateau.
Literature
1.
go back to reference Peterfy C. Imaging techniques. In: Klippel J, Dieppe P, eds. Rheumatology, 2nd edn, vol 1. Philadelphia: Mosby, 1997:14.1–14.18. Peterfy C. Imaging techniques. In: Klippel J, Dieppe P, eds. Rheumatology, 2nd edn, vol 1. Philadelphia: Mosby, 1997:14.1–14.18.
2.
go back to reference Buckland-Wright JC, Macfarlane DG, Jasani MK, Lynch JA. Quantitative microfocal radiographic assessment of osteoarthritis of the knee from weight bearing tunnel and semiflexed standing views. J Rheum 1994; 21:1734–1741.PubMed Buckland-Wright JC, Macfarlane DG, Jasani MK, Lynch JA. Quantitative microfocal radiographic assessment of osteoarthritis of the knee from weight bearing tunnel and semiflexed standing views. J Rheum 1994; 21:1734–1741.PubMed
3.
go back to reference Mazzuca S, Brandt K, Katz B. Is conventional radiography suitable for evaluation of a disease-modifying drug in patients with knee osteoarthritis? Osteoarthritis Cart 1997; 5:217–226. Mazzuca S, Brandt K, Katz B. Is conventional radiography suitable for evaluation of a disease-modifying drug in patients with knee osteoarthritis? Osteoarthritis Cart 1997; 5:217–226.
4.
go back to reference Mazzuca S, Brandt K, Buckland-Wright J, et al. Field test of the reproducibility of automated measurements of medial tibiofemoral joint space width derived from standardized knee radiographs. J Rheumatol 1999; 26:1359–1365.PubMed Mazzuca S, Brandt K, Buckland-Wright J, et al. Field test of the reproducibility of automated measurements of medial tibiofemoral joint space width derived from standardized knee radiographs. J Rheumatol 1999; 26:1359–1365.PubMed
5.
go back to reference Duryea J, Li J, Peterfy C, Gordon C, Genant H. Trainable rule-based algorithm for the measurement of joint space width in digital radiographic images of the knee. Med Phys 2000; 27:580–591.CrossRefPubMed Duryea J, Li J, Peterfy C, Gordon C, Genant H. Trainable rule-based algorithm for the measurement of joint space width in digital radiographic images of the knee. Med Phys 2000; 27:580–591.CrossRefPubMed
6.
go back to reference Mazzuca S, Brandt K, Buckwalter K. Knee pain reduces joint space width in conventional standing anteroposterior radiographs of osteoarthritic knees. Arthritis Rheum (in press). Mazzuca S, Brandt K, Buckwalter K. Knee pain reduces joint space width in conventional standing anteroposterior radiographs of osteoarthritic knees. Arthritis Rheum (in press).
7.
go back to reference Buckland-Wright JC, Clarke GS, Chikanza IC, Grahame R. Quantitative microfocal radiography detects changes in erosion area in patients with early rheumatoid arthritis treated with myocrisine. J Rheumatol 1993; 20:243–247.PubMed Buckland-Wright JC, Clarke GS, Chikanza IC, Grahame R. Quantitative microfocal radiography detects changes in erosion area in patients with early rheumatoid arthritis treated with myocrisine. J Rheumatol 1993; 20:243–247.PubMed
8.
go back to reference Buckland-Wright J, Wolfe F, Ward R, Flowers N, Hayne C. Substantial superiority of semiflexed (MTP) views in knee osteoarthritis: a comparative radiographic study, without fluoroscopy, of standing extended, semiflexed (MTP), and schuss views. J Rheumatol 1999; 26:2664–2674.PubMed Buckland-Wright J, Wolfe F, Ward R, Flowers N, Hayne C. Substantial superiority of semiflexed (MTP) views in knee osteoarthritis: a comparative radiographic study, without fluoroscopy, of standing extended, semiflexed (MTP), and schuss views. J Rheumatol 1999; 26:2664–2674.PubMed
Metadata
Title
Comparison of fixed-flexion positioning with fluoroscopic semi-flexed positioning for quantifying radiographic joint-space width in the knee: test-retest reproducibility
Authors
C. Peterfy
J. Li
S. Zaim
J. Duryea
J. Lynch
Y. Miaux
W. Yu
H. K. Genant
Publication date
01-03-2003
Publisher
Springer-Verlag
Published in
Skeletal Radiology / Issue 3/2003
Print ISSN: 0364-2348
Electronic ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-002-0603-z

Other articles of this Issue 3/2003

Skeletal Radiology 3/2003 Go to the issue