Published in: European Journal of Plastic Surgery 5/2015

Open Access 01-10-2015 | Original Paper

A blinking periorbital prosthesis using surface electromyographic signals of the orbicularis oculi muscle

Authors: Tadashi Akamatsu, Norimichi Kawashima, Takeshi Tsunekuni, Kotaro Imagawa, Muneo Miyasaka

Published in: European Journal of Plastic Surgery | Issue 5/2015

Login to get access

Abstract

Background

Recent advances in human–machine interface technology have enabled the development of multifunctional, primarily orthopedic myoelectric prostheses. We developed a noninvasive blinking periorbital prosthesis that can synchronize with blinking of the intact eyelid by using surface electromyographic signals of the orbicularis oculi muscle.

Methods

Myoelectric potentials of the orbicularis oculi muscle while blinking were measured with surface electrodes on the eyelid in four healthy adults. Possible cross talk introduced via the electrodes was also measured and assessed to determine whether cross talk would affect surface electromyographic measurements while blinking.

Results

The amplitude of the surface myoelectric potential of the orbicularis oculi muscle was sufficiently high for the practical use of blinking prostheses. Our blinking model was successfully synchronized with blinks of the subjects’ eyelids under experimental conditions without cross talk between the orbicularis oculi muscle and other muscles.

Conclusions

Although our study revealed several problems, the use of surface electromyographic signals could be a promising and useful technique for synchronizing blinking of the prosthetic eyelid with blinking of the intact eyelid.
Level of Evidence: Level V, therapeutic study.
Appendix
Available only for authorised users
Literature
1.
go back to reference Zecca M, Micera S, Carrozza MC, Dario P (2002) Control of multifunctional prosthetic hands by processing the electromyographic signal. Crit Rev Biomed Eng 30:459–485CrossRefPubMed Zecca M, Micera S, Carrozza MC, Dario P (2002) Control of multifunctional prosthetic hands by processing the electromyographic signal. Crit Rev Biomed Eng 30:459–485CrossRefPubMed
2.
go back to reference Graupe D, Salahi J, Kohn KH (1982) Multifunctional prosthesis and orthosis control via microcomputer identification of temporal pattern differences in single-site myoelectric signals. J Biomed Eng 4:17–22CrossRefPubMed Graupe D, Salahi J, Kohn KH (1982) Multifunctional prosthesis and orthosis control via microcomputer identification of temporal pattern differences in single-site myoelectric signals. J Biomed Eng 4:17–22CrossRefPubMed
3.
go back to reference Parker PA, Scott RN (1986) Myoelectric control of prostheses. Crit Rev Biomed Eng 13:283–310PubMed Parker PA, Scott RN (1986) Myoelectric control of prostheses. Crit Rev Biomed Eng 13:283–310PubMed
4.
go back to reference Williams TW 3rd (1990) Practical methods for controlling powered upper-extremity prostheses. Assist Technol 2:3–18CrossRefPubMed Williams TW 3rd (1990) Practical methods for controlling powered upper-extremity prostheses. Assist Technol 2:3–18CrossRefPubMed
5.
go back to reference Selcuk CT, Sahin U, Celebioglu S, Erbas O, Aydin C, Yuce S (2011) Complex craniofacial reconstruction with prostheses as an alternative method to autogenous reconstruction. J Craniofac Surg 22:2090–2093CrossRefPubMed Selcuk CT, Sahin U, Celebioglu S, Erbas O, Aydin C, Yuce S (2011) Complex craniofacial reconstruction with prostheses as an alternative method to autogenous reconstruction. J Craniofac Surg 22:2090–2093CrossRefPubMed
6.
go back to reference Tanner PB, Mobley SR (2006) External auricular and facial prosthetics: a collaborative effort of the reconstructive surgeon and anaplastologist. Facial Plast Surg Clin North Am 14:137–145CrossRefPubMed Tanner PB, Mobley SR (2006) External auricular and facial prosthetics: a collaborative effort of the reconstructive surgeon and anaplastologist. Facial Plast Surg Clin North Am 14:137–145CrossRefPubMed
7.
go back to reference French DJ, Jackson IT, Tolman DE (1988) A system of osseointegrated implants and its application to dental and facial rehabilitation. Eur J Plast Surg 11:14–21CrossRef French DJ, Jackson IT, Tolman DE (1988) A system of osseointegrated implants and its application to dental and facial rehabilitation. Eur J Plast Surg 11:14–21CrossRef
8.
go back to reference Honda M, Niimi A, Ueda M (1999) New orbital prosthesis with a blinking eyelid: report of a case. J Oral Maxillofac Surg 57:730–733CrossRefPubMed Honda M, Niimi A, Ueda M (1999) New orbital prosthesis with a blinking eyelid: report of a case. J Oral Maxillofac Surg 57:730–733CrossRefPubMed
9.
go back to reference Klein M, Menneking H, Hasenpusch M, Schmitz H, Locke HG, Bier J (1999) A new orbital prosthesis with a mobile upper eyelid. J Facial Somat Prosthetics 5:11–17 Klein M, Menneking H, Hasenpusch M, Schmitz H, Locke HG, Bier J (1999) A new orbital prosthesis with a mobile upper eyelid. J Facial Somat Prosthetics 5:11–17
10.
go back to reference Yanagisawa T, Hirata M, Saitoh Y, Kato A, Shibuya D, Kamitani Y, Yoshimine T (2009) Neural decoding using gyral and intrasulcal electrocorticograms. Neuroimage 45:1099–1106CrossRefPubMed Yanagisawa T, Hirata M, Saitoh Y, Kato A, Shibuya D, Kamitani Y, Yoshimine T (2009) Neural decoding using gyral and intrasulcal electrocorticograms. Neuroimage 45:1099–1106CrossRefPubMed
11.
go back to reference Yanagisawa T, Hirata M, Saitoh Y, Kishima H, Matsushita K, Goto T, Fukuma R, Yokoi H, Kamitani Y, Yoshimine T (2012) Electrocorticographic control of a prosthetic arm in paralyzed patients. Ann Neurol 71:353–361CrossRefPubMed Yanagisawa T, Hirata M, Saitoh Y, Kishima H, Matsushita K, Goto T, Fukuma R, Yokoi H, Kamitani Y, Yoshimine T (2012) Electrocorticographic control of a prosthetic arm in paralyzed patients. Ann Neurol 71:353–361CrossRefPubMed
12.
go back to reference Deeny S, Chicoine C, Hargrove L, Parrish T, Jayaraman A (2014) A simple ERP method for quantitative analysis of cognitive workload in myoelectric prosthesis control and human-machine interaction. PLoS One 9:e112091PubMedCentralCrossRefPubMed Deeny S, Chicoine C, Hargrove L, Parrish T, Jayaraman A (2014) A simple ERP method for quantitative analysis of cognitive workload in myoelectric prosthesis control and human-machine interaction. PLoS One 9:e112091PubMedCentralCrossRefPubMed
13.
go back to reference Evinger C, Manning KA, Sibony PA (1991) Eyelid movements. Mechanisms and normal data. Invest Ophthalmol Vis Sci 32:387–400PubMed Evinger C, Manning KA, Sibony PA (1991) Eyelid movements. Mechanisms and normal data. Invest Ophthalmol Vis Sci 32:387–400PubMed
14.
go back to reference Wu Z, Begley CG, Situ P, Simpson T, Liu H (2014) The effects of mild ocular surface stimulation and concentration on spontaneous blink parameters. Curr Eye Res 39:9–20PubMedCentralCrossRefPubMed Wu Z, Begley CG, Situ P, Simpson T, Liu H (2014) The effects of mild ocular surface stimulation and concentration on spontaneous blink parameters. Curr Eye Res 39:9–20PubMedCentralCrossRefPubMed
15.
go back to reference Wurth SM, Hargrove LJ (2014) A real-time comparison between direct control, sequential pattern recognition control and simultaneous pattern recognition control using a Fitts’ law style assessment procedure. J Neuroeng Rehabil 11:91PubMedCentralCrossRefPubMed Wurth SM, Hargrove LJ (2014) A real-time comparison between direct control, sequential pattern recognition control and simultaneous pattern recognition control using a Fitts’ law style assessment procedure. J Neuroeng Rehabil 11:91PubMedCentralCrossRefPubMed
16.
go back to reference Zhang Y, Li KY, Jin C, Wang YT, Geng L, Sun YJ, Tian HC, Liu JQ, Jin XJ (2014) Comparative studies on the implantation of nano platinum black and pure platinum electrodes in the rabbit orbicularis oculi muscle. J Laryngol Otol 128:679–689CrossRefPubMed Zhang Y, Li KY, Jin C, Wang YT, Geng L, Sun YJ, Tian HC, Liu JQ, Jin XJ (2014) Comparative studies on the implantation of nano platinum black and pure platinum electrodes in the rabbit orbicularis oculi muscle. J Laryngol Otol 128:679–689CrossRefPubMed
17.
go back to reference Salerno GM, McClellan GA, Bleicher JN, Stromberg BV, Cheng SC (1991) Electrical stimulation treatment of dog’s denervated orbicularis oculi muscle. Ann Plast Surg 26:431–440CrossRefPubMed Salerno GM, McClellan GA, Bleicher JN, Stromberg BV, Cheng SC (1991) Electrical stimulation treatment of dog’s denervated orbicularis oculi muscle. Ann Plast Surg 26:431–440CrossRefPubMed
18.
go back to reference Salerno GM, Bleicher JN, Stromberg BV, Cheng SC (1990) Electrophysiological study of the denervated orbicularis oculi muscle in dogs. Ann Plast Surg 24:24–31CrossRefPubMed Salerno GM, Bleicher JN, Stromberg BV, Cheng SC (1990) Electrophysiological study of the denervated orbicularis oculi muscle in dogs. Ann Plast Surg 24:24–31CrossRefPubMed
Metadata
Title
A blinking periorbital prosthesis using surface electromyographic signals of the orbicularis oculi muscle
Authors
Tadashi Akamatsu
Norimichi Kawashima
Takeshi Tsunekuni
Kotaro Imagawa
Muneo Miyasaka
Publication date
01-10-2015
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Plastic Surgery / Issue 5/2015
Print ISSN: 0930-343X
Electronic ISSN: 1435-0130
DOI
https://doi.org/10.1007/s00238-015-1111-4

Other articles of this Issue 5/2015

European Journal of Plastic Surgery 5/2015 Go to the issue