Published in: Neuroradiology 4/2013

01-04-2013 | Diagnostic Neuroradiology

Axial 3D gradient-echo imaging for improved multiple sclerosis lesion detection in the cervical spinal cord at 3T

Authors: Arzu Ozturk, Nafi Aygun, Seth A. Smith, Brian Caffo, Peter A. Calabresi, Daniel S. Reich

Published in: Neuroradiology | Issue 4/2013

Login to get access

Abstract

Introduction

In multiple sclerosis (MS), spinal cord imaging can help in diagnosis and follow-up evaluation. However, spinal cord magnetic resonance imaging (MRI) is technically challenging, and image quality, particularly in the axial plane, is typically poor compared to brain MRI. Because gradient-recalled echo (GRE) images might offer improved contrast resolution within the spinal cord at high magnetic field strength, both without and with a magnetization transfer prepulse, we compared them to T2-weighted fast-spin-echo (T2-FSE) images for the detection of MS lesions in the cervical cord at 3T.

Methods

On a clinical 3T MRI scanner, we studied 62 MS cases and 19 healthy volunteers. Axial 3D GRE sequences were performed without and with off-resonance radiofrequency irradiation. To mimic clinical practice, all images were evaluated in conjunction with linked images from a sagittal short tau inversion recovery scan, which is considered the gold standard for lesion detection in MS. Two experienced observers recorded image quality, location and size of focal lesions, atrophy, swelling, and diffuse signal abnormality independently at first and then in consensus.

Results

The number and volume of lesions detected with high confidence was more than three times as high on both GRE sequences compared to T2-FSE (p < 0.0001). Approximately 5 % of GRE scans were affected by artifacts that interfered with image interpretation, not significantly different from T2W-FSE.

Conclusions

Axial 3D GRE sequences are useful for MS lesion detection when compared to 2D T2-FSE sequences in the cervical spinal cord at 3T and should be considered when examining intramedullary spinal cord lesions.
Literature
1.
go back to reference Polman CH, Reingold SC, Banwell B, Clanet M, Cohen JA, Filippi M, Fujihara K, Havrdova E, Hutchinson M, Kappos L, Lublin FD, Montalban X, O'Connor P, Sandberg-Wollheim M, Thompson AJ, Waubant E, Weinshenker B, Wolinsky JS (2011) Diagnostic criteria for multiple sclerosis: 2010 revisions to the McDonald criteria. Ann Neurol 69(2):292–302. doi:10.1002/ana.22366 PubMedCrossRef Polman CH, Reingold SC, Banwell B, Clanet M, Cohen JA, Filippi M, Fujihara K, Havrdova E, Hutchinson M, Kappos L, Lublin FD, Montalban X, O'Connor P, Sandberg-Wollheim M, Thompson AJ, Waubant E, Weinshenker B, Wolinsky JS (2011) Diagnostic criteria for multiple sclerosis: 2010 revisions to the McDonald criteria. Ann Neurol 69(2):292–302. doi:10.​1002/​ana.​22366 PubMedCrossRef
2.
go back to reference Barkhof F (2002) The clinico-radiological paradox in multiple sclerosis revisited. Curr Opin Neurol 15(3):239–245PubMedCrossRef Barkhof F (2002) The clinico-radiological paradox in multiple sclerosis revisited. Curr Opin Neurol 15(3):239–245PubMedCrossRef
3.
go back to reference Poonawalla AH, Hou P, Nelson FA, Wolinsky JS, Narayana PA (2008) Cervical spinal cord lesions in multiple sclerosis: T1-weighted inversion-recovery MR imaging with phase-sensitive reconstruction. Radiology 246(1):258–264. doi:10.1148/radiol.2463061900 PubMedCrossRef Poonawalla AH, Hou P, Nelson FA, Wolinsky JS, Narayana PA (2008) Cervical spinal cord lesions in multiple sclerosis: T1-weighted inversion-recovery MR imaging with phase-sensitive reconstruction. Radiology 246(1):258–264. doi:10.​1148/​radiol.​2463061900 PubMedCrossRef
4.
go back to reference Vaithianathar L, Tench CR, Morgan PS, Constantinescu CS (2003) Magnetic resonance imaging of the cervical spinal cord in multiple sclerosis—a quantitative T1 relaxation time mapping approach. J Neurol 250(3):307–315. doi:10.1007/s00415-003-1001-8 PubMedCrossRef Vaithianathar L, Tench CR, Morgan PS, Constantinescu CS (2003) Magnetic resonance imaging of the cervical spinal cord in multiple sclerosis—a quantitative T1 relaxation time mapping approach. J Neurol 250(3):307–315. doi:10.​1007/​s00415-003-1001-8 PubMedCrossRef
5.
go back to reference Bot JC, Barkhof F, Lycklama a Nijeholt G, van Schaardenburg D, Voskuyl AE, Ader HJ, Pijnenburg JA, Polman CH, Uitdehaag BM, Vermeulen EG, Castelijns JA (2002) Differentiation of multiple sclerosis from other inflammatory disorders and cerebrovascular disease: value of spinal MR imaging. Radiology 223(1):46–56PubMedCrossRef Bot JC, Barkhof F, Lycklama a Nijeholt G, van Schaardenburg D, Voskuyl AE, Ader HJ, Pijnenburg JA, Polman CH, Uitdehaag BM, Vermeulen EG, Castelijns JA (2002) Differentiation of multiple sclerosis from other inflammatory disorders and cerebrovascular disease: value of spinal MR imaging. Radiology 223(1):46–56PubMedCrossRef
6.
go back to reference Lycklama G, Thompson A, Filippi M, Miller D, Polman C, Fazekas F, Barkhof F (2003) Spinal-cord MRI in multiple sclerosis. Lancet Neurol 2(9):555–562PubMedCrossRef Lycklama G, Thompson A, Filippi M, Miller D, Polman C, Fazekas F, Barkhof F (2003) Spinal-cord MRI in multiple sclerosis. Lancet Neurol 2(9):555–562PubMedCrossRef
7.
go back to reference Weier K, Mazraeh J, Naegelin Y, Thoeni A, Hirsch JG, Fabbro T, Bruni N, Duyar H, Bendfeldt K, Radue EW, Kappos L, Gass A (2012) Biplanar MRI for the assessment of the spinal cord in multiple sclerosis. Mult Scler. doi:10.1177/1352458512442754 Weier K, Mazraeh J, Naegelin Y, Thoeni A, Hirsch JG, Fabbro T, Bruni N, Duyar H, Bendfeldt K, Radue EW, Kappos L, Gass A (2012) Biplanar MRI for the assessment of the spinal cord in multiple sclerosis. Mult Scler. doi:10.​1177/​1352458512442754​
8.
go back to reference Thorpe JW, Kidd D, Moseley IF, Thompson AJ, MacManus DG, Compston DA, McDonald WI, Miller DH (1996) Spinal MRI in patients with suspected multiple sclerosis and negative brain MRI. Brain 119(Pt 3):709–714PubMedCrossRef Thorpe JW, Kidd D, Moseley IF, Thompson AJ, MacManus DG, Compston DA, McDonald WI, Miller DH (1996) Spinal MRI in patients with suspected multiple sclerosis and negative brain MRI. Brain 119(Pt 3):709–714PubMedCrossRef
9.
go back to reference Lycklama a Nijeholt GJ, Barkhof F, Scheltens P, Castelijns JA, Ader H, van Waesberghe JH, Polman C, Jongen SJ, Valk J (1997) MR of the spinal cord in multiple sclerosis: relation to clinical subtype and disability. AJNR Am J Neuroradiol 18(6):1041–1048PubMed Lycklama a Nijeholt GJ, Barkhof F, Scheltens P, Castelijns JA, Ader H, van Waesberghe JH, Polman C, Jongen SJ, Valk J (1997) MR of the spinal cord in multiple sclerosis: relation to clinical subtype and disability. AJNR Am J Neuroradiol 18(6):1041–1048PubMed
10.
go back to reference Kidd D, Thorpe JW, Kendall BE, Barker GJ, Miller DH, McDonald WI, Thompson AJ (1996) MRI dynamics of brain and spinal cord in progressive multiple sclerosis. J Neurol Neurosurg Psychiatry 60(1):15–19PubMedCrossRef Kidd D, Thorpe JW, Kendall BE, Barker GJ, Miller DH, McDonald WI, Thompson AJ (1996) MRI dynamics of brain and spinal cord in progressive multiple sclerosis. J Neurol Neurosurg Psychiatry 60(1):15–19PubMedCrossRef
11.
go back to reference Grossman RI, McGowan JC (1998) Perspectives on multiple sclerosis. AJNR Am J Neuroradiol 19(7):1251–1265PubMed Grossman RI, McGowan JC (1998) Perspectives on multiple sclerosis. AJNR Am J Neuroradiol 19(7):1251–1265PubMed
12.
go back to reference Brex PA, Miszkiel KA, O'Riordan JI, Plant GT, Moseley IF, Thompson AJ, Miller DH (2001) Assessing the risk of early multiple sclerosis in patients with clinically isolated syndromes: the role of a follow up MRI. J Neurol Neurosurg Psychiatry 70(3):390–393PubMedCrossRef Brex PA, Miszkiel KA, O'Riordan JI, Plant GT, Moseley IF, Thompson AJ, Miller DH (2001) Assessing the risk of early multiple sclerosis in patients with clinically isolated syndromes: the role of a follow up MRI. J Neurol Neurosurg Psychiatry 70(3):390–393PubMedCrossRef
13.
go back to reference Nijeholt GJ, Bergers E, Kamphorst W, Bot J, Nicolay K, Castelijns JA, van Waesberghe JH, Ravid R, Polman CH, Barkhof F (2001) Post-mortem high-resolution MRI of the spinal cord in multiple sclerosis: a correlative study with conventional MRI, histopathology and clinical phenotype. Brain 124(Pt 1):154–166PubMedCrossRef Nijeholt GJ, Bergers E, Kamphorst W, Bot J, Nicolay K, Castelijns JA, van Waesberghe JH, Ravid R, Polman CH, Barkhof F (2001) Post-mortem high-resolution MRI of the spinal cord in multiple sclerosis: a correlative study with conventional MRI, histopathology and clinical phenotype. Brain 124(Pt 1):154–166PubMedCrossRef
14.
go back to reference Lycklama a Nijeholt GJ, Castelijns JA, Weerts J, Ader H, van Waesberghe JH, Polman C, Barkhof F (1998) Sagittal MR of multiple sclerosis in the spinal cord: fast versus conventional spin-echo imaging. AJNR Am J Neuroradiol 19(2):355–360PubMed Lycklama a Nijeholt GJ, Castelijns JA, Weerts J, Ader H, van Waesberghe JH, Polman C, Barkhof F (1998) Sagittal MR of multiple sclerosis in the spinal cord: fast versus conventional spin-echo imaging. AJNR Am J Neuroradiol 19(2):355–360PubMed
15.
go back to reference Bot JC, Blezer EL, Kamphorst W, Lycklama ANGJ, Ader HJ, Castelijns JA, Ig KN, Bergers E, Ravid R, Polman C, Barkhof F (2004) The spinal cord in multiple sclerosis: relationship of high-spatial-resolution quantitative MR imaging findings to histopathologic results. Radiology 233(2):531–540. doi:10.1148/radiol.2332031572 PubMedCrossRef Bot JC, Blezer EL, Kamphorst W, Lycklama ANGJ, Ader HJ, Castelijns JA, Ig KN, Bergers E, Ravid R, Polman C, Barkhof F (2004) The spinal cord in multiple sclerosis: relationship of high-spatial-resolution quantitative MR imaging findings to histopathologic results. Radiology 233(2):531–540. doi:10.​1148/​radiol.​2332031572 PubMedCrossRef
16.
go back to reference Tartaglino LM, Friedman DP, Flanders AE, Lublin FD, Knobler RL, Liem M (1995) Multiple sclerosis in the spinal cord: MR appearance and correlation with clinical parameters. Radiology 195(3):725–732PubMed Tartaglino LM, Friedman DP, Flanders AE, Lublin FD, Knobler RL, Liem M (1995) Multiple sclerosis in the spinal cord: MR appearance and correlation with clinical parameters. Radiology 195(3):725–732PubMed
17.
go back to reference Simon JH, Li D, Traboulsee A, Coyle PK, Arnold DL, Barkhof F, Frank JA, Grossman R, Paty DW, Radue EW, Wolinsky JS (2006) Standardized MR imaging protocol for multiple sclerosis: consortium of MS centers consensus guidelines. AJNR Am J Neuroradiol 27(2):455–461PubMed Simon JH, Li D, Traboulsee A, Coyle PK, Arnold DL, Barkhof F, Frank JA, Grossman R, Paty DW, Radue EW, Wolinsky JS (2006) Standardized MR imaging protocol for multiple sclerosis: consortium of MS centers consensus guidelines. AJNR Am J Neuroradiol 27(2):455–461PubMed
18.
go back to reference Bronskill MJ, McVeigh ER, Kucharczyk W, Henkelman RM (1988) Syrinx-like artifacts on MR images of the spinal cord. Radiology 166(2):485–488PubMed Bronskill MJ, McVeigh ER, Kucharczyk W, Henkelman RM (1988) Syrinx-like artifacts on MR images of the spinal cord. Radiology 166(2):485–488PubMed
19.
go back to reference Curtin AJ, Chakeres DW, Bulas R, Boesel CP, Finneran M, Flint E (1989) MR imaging artifacts of the axial internal anatomy of the cervical spinal cord. AJR Am J Roentgenol 152(4):835–842PubMedCrossRef Curtin AJ, Chakeres DW, Bulas R, Boesel CP, Finneran M, Flint E (1989) MR imaging artifacts of the axial internal anatomy of the cervical spinal cord. AJR Am J Roentgenol 152(4):835–842PubMedCrossRef
20.
go back to reference Czervionke LF, Daniels DL, Ho PS, Yu SW, Pech P, Strandt JA, Williams AL, Haughton VM (1988) The MR appearance of gray and white matter in the cervical spinal cord. AJNR Am J Neuroradiol 9(3):557–562PubMed Czervionke LF, Daniels DL, Ho PS, Yu SW, Pech P, Strandt JA, Williams AL, Haughton VM (1988) The MR appearance of gray and white matter in the cervical spinal cord. AJNR Am J Neuroradiol 9(3):557–562PubMed
21.
go back to reference Hinks RS, Quencer RM (1988) Motion artifacts in brain and spine MR. Radiol Clin N Am 26(4):737–753PubMed Hinks RS, Quencer RM (1988) Motion artifacts in brain and spine MR. Radiol Clin N Am 26(4):737–753PubMed
22.
go back to reference Levy LM, Di Chiro G, Brooks RA, Dwyer AJ, Wener L, Frank J (1988) Spinal cord artifacts from truncation errors during MR imaging. Radiology 166(2):479–483PubMed Levy LM, Di Chiro G, Brooks RA, Dwyer AJ, Wener L, Frank J (1988) Spinal cord artifacts from truncation errors during MR imaging. Radiology 166(2):479–483PubMed
23.
go back to reference Mikulis DJ, Wood ML, Zerdoner OA, Poncelet BP (1994) Oscillatory motion of the normal cervical spinal cord. Radiology 192(1):117–121PubMed Mikulis DJ, Wood ML, Zerdoner OA, Poncelet BP (1994) Oscillatory motion of the normal cervical spinal cord. Radiology 192(1):117–121PubMed
24.
go back to reference Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM (1990) The NMR phased array. Magn Reson Med 16(2):192–225PubMedCrossRef Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM (1990) The NMR phased array. Magn Reson Med 16(2):192–225PubMedCrossRef
26.
go back to reference Smith SA, Jones CK, Gifford A, Belegu V, Chodkowski B, Farrell JA, Landman BA, Reich DS, Calabresi PA, McDonald JW, van Zijl PC (2010) Reproducibility of tract-specific magnetization transfer and diffusion tensor imaging in the cervical spinal cord at 3 tesla. NMR Biomed 23(2):207–217. doi:10.1002/nbm.1447 PubMed Smith SA, Jones CK, Gifford A, Belegu V, Chodkowski B, Farrell JA, Landman BA, Reich DS, Calabresi PA, McDonald JW, van Zijl PC (2010) Reproducibility of tract-specific magnetization transfer and diffusion tensor imaging in the cervical spinal cord at 3 tesla. NMR Biomed 23(2):207–217. doi:10.​1002/​nbm.​1447 PubMed
27.
go back to reference Kurtzke JF (1983) Rating neurologic impairment in multiple sclerosis: an expanded disability status scale (EDSS). Neurology 33(11):1444–1452PubMedCrossRef Kurtzke JF (1983) Rating neurologic impairment in multiple sclerosis: an expanded disability status scale (EDSS). Neurology 33(11):1444–1452PubMedCrossRef
28.
go back to reference Martin N, Malfair D, Zhao Y, Li D, Traboulsee A, Lang D, Vertinsky AT (2012) Comparison of MERGE and axial T2-weighted fast spin-echo sequences for detection of multiple sclerosis lesions in the cervical spinal cord. AJR Am J Roentgenol 199(1):157–162. doi:10.2214/AJR.11.7039 PubMedCrossRef Martin N, Malfair D, Zhao Y, Li D, Traboulsee A, Lang D, Vertinsky AT (2012) Comparison of MERGE and axial T2-weighted fast spin-echo sequences for detection of multiple sclerosis lesions in the cervical spinal cord. AJR Am J Roentgenol 199(1):157–162. doi:10.​2214/​AJR.​11.​7039 PubMedCrossRef
29.
go back to reference White ML, Zhang Y, Healey K (2011) Cervical spinal cord multiple sclerosis: evaluation with 2D multi-echo recombined gradient echo MR imaging. J Spinal Cord Med 34(1):93–98PubMedCrossRef White ML, Zhang Y, Healey K (2011) Cervical spinal cord multiple sclerosis: evaluation with 2D multi-echo recombined gradient echo MR imaging. J Spinal Cord Med 34(1):93–98PubMedCrossRef
30.
go back to reference Hittmair K, Mallek R, Prayer D, Schindler EG, Kollegger H (1996) Spinal cord lesions in patients with multiple sclerosis: comparison of MR pulse sequences. AJNR Am J Neuroradiol 17(8):1555–1565PubMed Hittmair K, Mallek R, Prayer D, Schindler EG, Kollegger H (1996) Spinal cord lesions in patients with multiple sclerosis: comparison of MR pulse sequences. AJNR Am J Neuroradiol 17(8):1555–1565PubMed
31.
go back to reference Hackney DB (1999) MR studies of the spinal cord in patients with multiple sclerosis: what should we do? AJNR Am J Neuroradiol 20(9):1581–1583PubMed Hackney DB (1999) MR studies of the spinal cord in patients with multiple sclerosis: what should we do? AJNR Am J Neuroradiol 20(9):1581–1583PubMed
32.
go back to reference Zackowski KM, Smith SA, Reich DS, Gordon-Lipkin E, Chodkowski BA, Sambandan DR, Shteyman M, Bastian AJ, van Zijl PC, Calabresi PA (2009) Sensorimotor dysfunction in multiple sclerosis and column-specific magnetization transfer-imaging abnormalities in the spinal cord. Brain 132(Pt 5):1200–1209. doi:10.1093/brain/awp032 PubMedCrossRef Zackowski KM, Smith SA, Reich DS, Gordon-Lipkin E, Chodkowski BA, Sambandan DR, Shteyman M, Bastian AJ, van Zijl PC, Calabresi PA (2009) Sensorimotor dysfunction in multiple sclerosis and column-specific magnetization transfer-imaging abnormalities in the spinal cord. Brain 132(Pt 5):1200–1209. doi:10.​1093/​brain/​awp032 PubMedCrossRef
33.
go back to reference Oh J, Chen M, Zackowski KM, Newsome SD, Smith SA, Prince JL, Calabresi PA, Reich DS (2012) Diffusion tensor and magnetization transfer imaging correlates of motor dysfunction in the spinal cord in multiple sclerosis. Paper presented at the American Academy of Neurology, New Orleans Oh J, Chen M, Zackowski KM, Newsome SD, Smith SA, Prince JL, Calabresi PA, Reich DS (2012) Diffusion tensor and magnetization transfer imaging correlates of motor dysfunction in the spinal cord in multiple sclerosis. Paper presented at the American Academy of Neurology, New Orleans
34.
go back to reference Keiper MD, Grossman RI, Hirsch JA, Bolinger L, Ott IL, Mannon LJ, Langlotz CP, Kolson DL (1998) MR identification of white matter abnormalities in multiple sclerosis: a comparison between 1.5T and 4T. AJNR Am J Neuroradiol 19(8):1489–1493PubMed Keiper MD, Grossman RI, Hirsch JA, Bolinger L, Ott IL, Mannon LJ, Langlotz CP, Kolson DL (1998) MR identification of white matter abnormalities in multiple sclerosis: a comparison between 1.5T and 4T. AJNR Am J Neuroradiol 19(8):1489–1493PubMed
35.
go back to reference Gauthier SA, Glanz BI, Mandel M, Tsagkaropoulos A, Neema M, Stankiewicz J, Arora A, Duan Y, Liptak Z, Egorova S, Buckle GJ, Bakshi R, Guttmann CR, Khoury SJ, Weiner HL (2009) Incidence and factors associated with treatment failure in the CLIMB multiple sclerosis cohort study. J Neurol Sci 284(1–2):116–119. doi:10.1016/j.jns.2009.04.020 PubMedCrossRef Gauthier SA, Glanz BI, Mandel M, Tsagkaropoulos A, Neema M, Stankiewicz J, Arora A, Duan Y, Liptak Z, Egorova S, Buckle GJ, Bakshi R, Guttmann CR, Khoury SJ, Weiner HL (2009) Incidence and factors associated with treatment failure in the CLIMB multiple sclerosis cohort study. J Neurol Sci 284(1–2):116–119. doi:10.​1016/​j.​jns.​2009.​04.​020 PubMedCrossRef
36.
go back to reference Stankiewicz JM, Neema M, Alsop DC, Healy BC, Arora A, Buckle GJ, Chitnis T, Guttmann CR, Hackney D, Bakshi R (2009) Spinal cord lesions and clinical status in multiple sclerosis: A 1.5T and 3T MRI study. J Neurol Sci 279(1–2):99–105. doi:10.1016/j.jns.2008.11.009 PubMedCrossRef Stankiewicz JM, Neema M, Alsop DC, Healy BC, Arora A, Buckle GJ, Chitnis T, Guttmann CR, Hackney D, Bakshi R (2009) Spinal cord lesions and clinical status in multiple sclerosis: A 1.5T and 3T MRI study. J Neurol Sci 279(1–2):99–105. doi:10.​1016/​j.​jns.​2008.​11.​009 PubMedCrossRef
37.
go back to reference Ge Y (2006) Multiple sclerosis: the role of MR imaging. AJNR Am J Neuroradiol 27(6):1165–1176PubMed Ge Y (2006) Multiple sclerosis: the role of MR imaging. AJNR Am J Neuroradiol 27(6):1165–1176PubMed
38.
go back to reference Kidd D, Thorpe JW, Thompson AJ, Kendall BE, Moseley IF, MacManus DG, McDonald WI, Miller DH (1993) Spinal cord MRI using multi-array coils and fast spin echo II. Findings in multiple sclerosis. Neurology 43(12):2632–2637 Kidd D, Thorpe JW, Thompson AJ, Kendall BE, Moseley IF, MacManus DG, McDonald WI, Miller DH (1993) Spinal cord MRI using multi-array coils and fast spin echo II. Findings in multiple sclerosis. Neurology 43(12):2632–2637
39.
go back to reference Lycklama a Nijeholt GJ, Barkhof F, Castelijns JA, van Waesberghe JH, Valk J, Jongen PJ, Hommes OR (1996) Comparison of two MR sequences for the detection of multiple sclerosis lesions in the spinal cord. AJNR Am J Neuroradiol 17(8):1533–1538PubMed Lycklama a Nijeholt GJ, Barkhof F, Castelijns JA, van Waesberghe JH, Valk J, Jongen PJ, Hommes OR (1996) Comparison of two MR sequences for the detection of multiple sclerosis lesions in the spinal cord. AJNR Am J Neuroradiol 17(8):1533–1538PubMed
41.
go back to reference Sze G, Merriam M, Oshio K, Jolesz FA (1992) Fast spin-echo imaging in the evaluation of intradural disease of the spine. AJNR Am J Neuroradiol 13(5):1383–1392PubMed Sze G, Merriam M, Oshio K, Jolesz FA (1992) Fast spin-echo imaging in the evaluation of intradural disease of the spine. AJNR Am J Neuroradiol 13(5):1383–1392PubMed
42.
go back to reference Keiper MD, Grossman RI, Brunson JC, Schnall MD (1997) The low sensitivity of fluid-attenuated inversion-recovery MR in the detection of multiple sclerosis of the spinal cord. AJNR Am J Neuroradiol 18(6):1035–1039PubMed Keiper MD, Grossman RI, Brunson JC, Schnall MD (1997) The low sensitivity of fluid-attenuated inversion-recovery MR in the detection of multiple sclerosis of the spinal cord. AJNR Am J Neuroradiol 18(6):1035–1039PubMed
43.
go back to reference Stevenson VL, Gawne-Cain ML, Barker GJ, Thompson AJ, Miller DH (1997) Imaging of the spinal cord and brain in multiple sclerosis: a comparative study between fast FLAIR and fast spin echo. J Neurol 244(2):119–124PubMedCrossRef Stevenson VL, Gawne-Cain ML, Barker GJ, Thompson AJ, Miller DH (1997) Imaging of the spinal cord and brain in multiple sclerosis: a comparative study between fast FLAIR and fast spin echo. J Neurol 244(2):119–124PubMedCrossRef
44.
go back to reference Thorpe JW, Kidd D, Kendall BE, Tofts PS, Barker GJ, Thompson AJ, MacManus DG, McDonald WI, Miller DH (1993) Spinal cord MRI using multi-array coils and fast spin echo I. Technical aspects and findings in healthy adults. Neurology 43(12):2625–2631PubMedCrossRef Thorpe JW, Kidd D, Kendall BE, Tofts PS, Barker GJ, Thompson AJ, MacManus DG, McDonald WI, Miller DH (1993) Spinal cord MRI using multi-array coils and fast spin echo I. Technical aspects and findings in healthy adults. Neurology 43(12):2625–2631PubMedCrossRef
45.
go back to reference Sze G, Kawamura Y, Negishi C, Constable RT, Merriam M, Oshio K, Jolesz F (1993) Fast spin-echo MR imaging of the cervical spine: influence of echo train length and echo spacing on image contrast and quality. AJNR Am J Neuroradiol 14(5):1203–1213PubMed Sze G, Kawamura Y, Negishi C, Constable RT, Merriam M, Oshio K, Jolesz F (1993) Fast spin-echo MR imaging of the cervical spine: influence of echo train length and echo spacing on image contrast and quality. AJNR Am J Neuroradiol 14(5):1203–1213PubMed
46.
go back to reference Enzmann DR, Rubin JB (1988) Cervical spine: MR imaging with a partial flip angle, gradient-refocused pulse sequence Part II. Spinal cord disease. Radiology 166(2):473–478PubMed Enzmann DR, Rubin JB (1988) Cervical spine: MR imaging with a partial flip angle, gradient-refocused pulse sequence Part II. Spinal cord disease. Radiology 166(2):473–478PubMed
47.
go back to reference Dale RC, de Sousa C, Chong WK, Cox TC, Harding B, Neville BG (2000) Acute disseminated encephalomyelitis, multiphasic disseminated encephalomyelitis and multiple sclerosis in children. Brain 123(Pt 12):2407–2422PubMedCrossRef Dale RC, de Sousa C, Chong WK, Cox TC, Harding B, Neville BG (2000) Acute disseminated encephalomyelitis, multiphasic disseminated encephalomyelitis and multiple sclerosis in children. Brain 123(Pt 12):2407–2422PubMedCrossRef
48.
go back to reference Finelli DA, Hurst GC, Karaman BA, Simon JE, Duerk JL, Bellon EM (1994) Use of magnetization transfer for improved contrast on gradient-echo MR images of the cervical spine. Radiology 193(1):165–171PubMed Finelli DA, Hurst GC, Karaman BA, Simon JE, Duerk JL, Bellon EM (1994) Use of magnetization transfer for improved contrast on gradient-echo MR images of the cervical spine. Radiology 193(1):165–171PubMed
Metadata
Title
Axial 3D gradient-echo imaging for improved multiple sclerosis lesion detection in the cervical spinal cord at 3T
Authors
Arzu Ozturk
Nafi Aygun
Seth A. Smith
Brian Caffo
Peter A. Calabresi
Daniel S. Reich
Publication date
01-04-2013
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 4/2013
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-012-1118-5

Other articles of this Issue 4/2013

Neuroradiology 4/2013 Go to the issue