AngioCT in the management of neurointerventional patients: a prospective, consecutive series with associated dosimetry and resolution data

Authors: Philip M. White, John N. Gilmour, Nicholas W. Weir, Brian Innes, Robin J. Sellar

Published in: Neuroradiology | Issue 4/2008

Login to get access

Abstract

Introduction

Endovascular coiling of intracranial aneurysms carries a risk of complications. Early detection and management of complications can improve clinical outcomes. AngioCT is a new imaging technology enabling CT-like images to be generated on a flat-panel digital subtraction angiography system, which can provide immediate “on angio table” identification and thorough assessment of such complications. We prospectively audited its utility during aneurysm coiling in patients following subarachnoid haemorrhage (SAH).

Methods

A prospective series of 44 patients with SAH undergoing endovascular coiling with AngioCT was audited for image quality and the influence of the AngioCT on patient management. In a parallel experimental study, radiation doses were measured and image quality parameters on standard phantoms were established.

Results

In all patients, AngioCT provided adequate diagnostic information. In 40.9% of patients, AngioCT was a substantial or major factor in determining the management immediately after coiling. Using a 10-s high-dose acquisition technique, acceptable image quality could be obtained rapidly with a radiation dose just over half that for a conventional CT scan of the head (35 mGy versus approximately 60 mGy). No patient in this series required conventional CT to clarify the AngioCT appearance.

Conclusion

AngioCT has many applications in the neurointerventional setting. In particular during coiling, AngioCT provides a rapid way to clarify concerns or identify complications and in some cases was the major factor influencing further patient management immediately after coiling. AngioCT images were judged of adequate quality to be clinically useful in all patients in this series.
Literature
1.
go back to reference Molyneux AJ, Kerr RS, Yu LM et al (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef Molyneux AJ, Kerr RS, Yu LM et al (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef
2.
go back to reference van Rooij WJ, Sluzewski M, Beute GN et al (2006) Procedural complications of coiling of ruptured intracranial aneurysms: incidence and risk factors in a consecutive series of 681 patients. AJNR Am J Neuroradiol 27(7):1498–1501PubMed van Rooij WJ, Sluzewski M, Beute GN et al (2006) Procedural complications of coiling of ruptured intracranial aneurysms: incidence and risk factors in a consecutive series of 681 patients. AJNR Am J Neuroradiol 27(7):1498–1501PubMed
3.
go back to reference Brisman JL, Niimi Y, Song JK et al (2005) Aneurysmal rupture during coiling: low incidence and good outcomes at a single large volume center. Neurosurgery 57(6):1103–1109PubMedCrossRef Brisman JL, Niimi Y, Song JK et al (2005) Aneurysmal rupture during coiling: low incidence and good outcomes at a single large volume center. Neurosurgery 57(6):1103–1109PubMedCrossRef
4.
go back to reference Horowitz MB, Crammond D, Balzer J et al (2003) Aneurysm rupture during endovascular coiling: effects on cerebral transit time and neurophysiologic monitoring and the benefits of early ventriculostomy: case report. Minim Invasive Neurosurg 46(5):300–305PubMedCrossRef Horowitz MB, Crammond D, Balzer J et al (2003) Aneurysm rupture during endovascular coiling: effects on cerebral transit time and neurophysiologic monitoring and the benefits of early ventriculostomy: case report. Minim Invasive Neurosurg 46(5):300–305PubMedCrossRef
5.
go back to reference Cloft HJ, Kallmes DF (2002) Cerebral aneurysm perforations complicating therapy with Guglielmi detachable coils: a meta-analysis. AJNR Am J Neuroradiol 23(10):1706–1709PubMed Cloft HJ, Kallmes DF (2002) Cerebral aneurysm perforations complicating therapy with Guglielmi detachable coils: a meta-analysis. AJNR Am J Neuroradiol 23(10):1706–1709PubMed
6.
go back to reference Levy E, Koebbe CJ, Horowitz MB et al (2001) Rupture of intracranial aneurysms during endovascular coiling: management and outcomes. Neurosurgery 49(4):807–811PubMedCrossRef Levy E, Koebbe CJ, Horowitz MB et al (2001) Rupture of intracranial aneurysms during endovascular coiling: management and outcomes. Neurosurgery 49(4):807–811PubMedCrossRef
7.
go back to reference Heran NS, Song JK, Namba K et al (2006) The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol 27(2):330–332PubMed Heran NS, Song JK, Namba K et al (2006) The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol 27(2):330–332PubMed
8.
go back to reference Benndorf G, Strother CM, Claus B et al (2005) Angiographic CT in cerebrovascular stenting. AJNR Am J Neuroradiol 26:1813–1818PubMed Benndorf G, Strother CM, Claus B et al (2005) Angiographic CT in cerebrovascular stenting. AJNR Am J Neuroradiol 26:1813–1818PubMed
9.
go back to reference Fineberg HV, Bauman R, Sosman M (1977) Computerized cranial tomography. Effect on diagnostic and therapeutic plans. JAMA 238(3):224–227PubMedCrossRef Fineberg HV, Bauman R, Sosman M (1977) Computerized cranial tomography. Effect on diagnostic and therapeutic plans. JAMA 238(3):224–227PubMedCrossRef
11.
go back to reference Hart D, Jones DG, Wall BF (1994) Estimation of effective dose in diagnostic radiology from entrance surface dose and dose-area product measurements. NRPB-R262. Health Protection Agency, Chilton Hart D, Jones DG, Wall BF (1994) Estimation of effective dose in diagnostic radiology from entrance surface dose and dose-area product measurements. NRPB-R262. Health Protection Agency, Chilton
12.
go back to reference Edyvean S, Lewis MA, Keat N et al (2003) Measurement of the performance characteristics of diagnostic imaging systems in medicine, Part III - Computed tomography X-ray scanners (report no. 32). Institute of Physics and Engineering in Medicine, York, UK Edyvean S, Lewis MA, Keat N et al (2003) Measurement of the performance characteristics of diagnostic imaging systems in medicine, Part III - Computed tomography X-ray scanners (report no. 32). Institute of Physics and Engineering in Medicine, York, UK
13.
go back to reference Shrimpton PC, Hillier MC, Lewis MA et al (2006) National survey of doses from CT in the UK: 2003. Br J Radiol 79:968–980PubMedCrossRef Shrimpton PC, Hillier MC, Lewis MA et al (2006) National survey of doses from CT in the UK: 2003. Br J Radiol 79:968–980PubMedCrossRef
Metadata
Title
AngioCT in the management of neurointerventional patients: a prospective, consecutive series with associated dosimetry and resolution data
Authors
Philip M. White
John N. Gilmour
Nicholas W. Weir
Brian Innes
Robin J. Sellar
Publication date
01-04-2008
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 4/2008
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-007-0339-5

Other articles of this Issue 4/2008

Neuroradiology 4/2008 Go to the issue