Introduction

The Guglielmi detachable coil (GDC) 360°, a new complex shaped bare platinum coil, became available in Europe for aneurysm treatment in September 2005. The purpose of this study was to assess the feasibility and safety of selective embolization of intracranial aneurysms with the GDC 360° in 52 consecutive patients.

Methods

All patients included in this study were registered in a prospectively maintained database. We assessed the patient clinical history, aneurysm shape and dimensions, technical details and complications of the procedures, degree of aneurysm occlusion, and clinical findings upon discharge. In all patients, the first coil deployed was a GDC 360°.

Results

Over a 6-month period, we intended to treat 52 aneurysms with the GDC 360° in 52 patients. Of these 52 patients, 42 (81%) were treated in the context of subarachnoid haemorrhage. In 51 of 52 patients, the underlying aneurysm was successfully treated by coil embolization. Six procedures (11.5%) were complicated by the formation of thrombus in the parent artery during the intervention. One patient suffered a stroke related to the procedure. Angiograms obtained immediately after the procedure showed complete occlusion of the aneurysmal sac in 38 of 51 procedures (74.5%), a neck remnant in 11 (21.6%), and a residual aneurysm in 2 (3.9%). In 43 of 51 patients (84.3%), clinical assessment demonstrated independent clinical status, whereas 7 patients (13.7%) required assistance in the activities of daily living upon hospital discharge. One patient (2.0%) died after development of a severe vasospasm 10 days after the endovascular procedure.

Conclusion

The GDC 360° can be safely used for the endovascular occlusion of intracranial aneurysms.
Literature
1.
go back to reference Molyneux A, Kerr R, Stratton I, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group et al (2002) International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet 360:1267–1274PubMedCrossRef Molyneux A, Kerr R, Stratton I, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group et al (2002) International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet 360:1267–1274PubMedCrossRef
2.
go back to reference Molyneux A, Kerr R, Yu LM, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group et al (2005) International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef Molyneux A, Kerr R, Yu LM, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group et al (2005) International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef
3.
go back to reference Moret J, Cognard C, Weill A, Castaings L, Rey A (1997) Reconstruction technique in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases. J Neuroradiol 24:30–44PubMed Moret J, Cognard C, Weill A, Castaings L, Rey A (1997) Reconstruction technique in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases. J Neuroradiol 24:30–44PubMed
4.
go back to reference Fiorella D, Albuquerque FC, Han P, McDougall CG (2004) Preliminary experience using the Neuroform stent for the treatment of cerebral aneurysms. Neurosurgery 54:6–16PubMedCrossRef Fiorella D, Albuquerque FC, Han P, McDougall CG (2004) Preliminary experience using the Neuroform stent for the treatment of cerebral aneurysms. Neurosurgery 54:6–16PubMedCrossRef
5.
go back to reference Raymond J, Guilbert F, Weill A et al (2004) Safety, science, and sales: a request for valid clinical trials to assess new devices for endovascular treatment of intracranial aneurysms. AJNR Am J Neuroradiol 25:1128–1130PubMed Raymond J, Guilbert F, Weill A et al (2004) Safety, science, and sales: a request for valid clinical trials to assess new devices for endovascular treatment of intracranial aneurysms. AJNR Am J Neuroradiol 25:1128–1130PubMed
6.
go back to reference Murayama Y, Nien YL, Duckwiler G et al (2003) Guglielmi detachable coil embolisation of cerebral aneurysms: 11 years’ experience. J Neurosurg 98:959–966PubMedCrossRef Murayama Y, Nien YL, Duckwiler G et al (2003) Guglielmi detachable coil embolisation of cerebral aneurysms: 11 years’ experience. J Neurosurg 98:959–966PubMedCrossRef
7.
go back to reference Guglielmi G, Vinuela F, Dion J, Duckwiler G (1991) Electrothrombosis of saccular aneurysms via endovascular approach – Part 2: Preliminary clinical experience. J Neurosurg 75:8–14PubMed Guglielmi G, Vinuela F, Dion J, Duckwiler G (1991) Electrothrombosis of saccular aneurysms via endovascular approach – Part 2: Preliminary clinical experience. J Neurosurg 75:8–14PubMed
8.
go back to reference Jennett B, Bond M (1975) Assessment of outcome after severe brain damage: a practical scale. Lancet 1:480–484PubMedCrossRef Jennett B, Bond M (1975) Assessment of outcome after severe brain damage: a practical scale. Lancet 1:480–484PubMedCrossRef
9.
go back to reference Roy D, Milot G, Raymond J (2001) Endovascular treatment of unruptured aneurysms. Stroke 32:1998–2004PubMed Roy D, Milot G, Raymond J (2001) Endovascular treatment of unruptured aneurysms. Stroke 32:1998–2004PubMed
10.
go back to reference Brilstra EH, Rinkel GJ, van der Graaf Y, van Rooij WJ, Algra A (1999) Treatment of intracranial aneurysms by embolisation with coils: a systematic review. Stroke 30:470–476PubMed Brilstra EH, Rinkel GJ, van der Graaf Y, van Rooij WJ, Algra A (1999) Treatment of intracranial aneurysms by embolisation with coils: a systematic review. Stroke 30:470–476PubMed
11.
go back to reference Ng P, Khangure MS, Phatouros CC, Bynevelt M, ApSimon H, McAuliffe W (2002) Endovascular treatment of intracranial aneurysms with Guglielmi detachable coils: analysis of midterm angiographic and clinical outcomes. Stroke 33:210–217PubMedCrossRef Ng P, Khangure MS, Phatouros CC, Bynevelt M, ApSimon H, McAuliffe W (2002) Endovascular treatment of intracranial aneurysms with Guglielmi detachable coils: analysis of midterm angiographic and clinical outcomes. Stroke 33:210–217PubMedCrossRef
12.
go back to reference Friedman JA, Nichols DA, Meyer FB et al (2003) Guglielmi detachable coil treatment of ruptured saccular cerebral aneurysms: retrospective review of a 10-year single-center experience. AJNR Am J Neuroradiol 24:526–533PubMed Friedman JA, Nichols DA, Meyer FB et al (2003) Guglielmi detachable coil treatment of ruptured saccular cerebral aneurysms: retrospective review of a 10-year single-center experience. AJNR Am J Neuroradiol 24:526–533PubMed
13.
go back to reference Sluzewski M, van Rooij WJ, Rinkel GJ, Wijnalda D (2003) Endovascular treatment of ruptured intracranial aneurysms with detachable coils: long-term clinical and serial angiographic results. Radiology 227:720–724PubMedCrossRef Sluzewski M, van Rooij WJ, Rinkel GJ, Wijnalda D (2003) Endovascular treatment of ruptured intracranial aneurysms with detachable coils: long-term clinical and serial angiographic results. Radiology 227:720–724PubMedCrossRef
14.
go back to reference Cognard C, Weill A, Spelle L, Piotin M, Castaings L, Rey A, Moret J (1999) Long-term angiographic follow-up of 169 intracranial berry aneurysms occluded with detachable coils. Radiology 212:348–356PubMed Cognard C, Weill A, Spelle L, Piotin M, Castaings L, Rey A, Moret J (1999) Long-term angiographic follow-up of 169 intracranial berry aneurysms occluded with detachable coils. Radiology 212:348–356PubMed
15.
go back to reference Slob MJ, Sluzewski M, van Rooij WJ, Roks G, Rinkel GJ (2004) Additional coiling of previously coiled cerebral aneurysms: clinical and angiographic results. AJNR Am J Neuroradiol 25:1373–1376PubMed Slob MJ, Sluzewski M, van Rooij WJ, Roks G, Rinkel GJ (2004) Additional coiling of previously coiled cerebral aneurysms: clinical and angiographic results. AJNR Am J Neuroradiol 25:1373–1376PubMed
16.
go back to reference Murayama Y, Vinuela F, Tateshima S et al (2002) Cellular responses of bioabsorbable polymeric material and Guglielmi detachable coil in experimental aneurysms. Stroke 33:1120–1128PubMedCrossRef Murayama Y, Vinuela F, Tateshima S et al (2002) Cellular responses of bioabsorbable polymeric material and Guglielmi detachable coil in experimental aneurysms. Stroke 33:1120–1128PubMedCrossRef
17.
go back to reference Murayama Y, Tateshima S, Gonzalez NR, Vinuela F (2003) Matrix and bioabsorbable polymeric coils accelerate healing of intracranial aneurysms: long-term experimental study. Stroke 34:2031–2037PubMedCrossRef Murayama Y, Tateshima S, Gonzalez NR, Vinuela F (2003) Matrix and bioabsorbable polymeric coils accelerate healing of intracranial aneurysms: long-term experimental study. Stroke 34:2031–2037PubMedCrossRef
18.
go back to reference Raymond J, Roy D, Leblanc P et al (2003) Endovascular treatment of intracranial aneurysms with radioactive coils: initial clinical experience. Stroke 34:2801–2806PubMedCrossRef Raymond J, Roy D, Leblanc P et al (2003) Endovascular treatment of intracranial aneurysms with radioactive coils: initial clinical experience. Stroke 34:2801–2806PubMedCrossRef
19.
go back to reference Cloft HJ, Kallmes DF (2004) Aneurysm packing with HydroCoil Embolic System versus platinum coils: initial clinical experience. AJNR Am J Neuroradiol 25:60–62PubMed Cloft HJ, Kallmes DF (2004) Aneurysm packing with HydroCoil Embolic System versus platinum coils: initial clinical experience. AJNR Am J Neuroradiol 25:60–62PubMed
20.
go back to reference Taschner CA, Leclerc X, Rachdi H, Barros AM, Pruvo JP (2005) Matrix detachable coils for the endovascular treatment of intracranial aneurysms: analysis of early angiographic and clinical outcomes. Stroke 36:2176–2180PubMedCrossRef Taschner CA, Leclerc X, Rachdi H, Barros AM, Pruvo JP (2005) Matrix detachable coils for the endovascular treatment of intracranial aneurysms: analysis of early angiographic and clinical outcomes. Stroke 36:2176–2180PubMedCrossRef
21.
go back to reference Niimi Y, Song J, Madrid M, Berenstein A (2006) Endosaccular treatment of intracranial aneurysms using matrix coils: early experience and midterm follow-up. Stroke 37:1028–1032PubMedCrossRef Niimi Y, Song J, Madrid M, Berenstein A (2006) Endosaccular treatment of intracranial aneurysms using matrix coils: early experience and midterm follow-up. Stroke 37:1028–1032PubMedCrossRef
22.
go back to reference Fiorella D, Albuquerque FC, McDougall CG (2006) Durability of aneurysm embolisation with matrix detachable coils. Neurosurgery 58:51–59PubMedCrossRef Fiorella D, Albuquerque FC, McDougall CG (2006) Durability of aneurysm embolisation with matrix detachable coils. Neurosurgery 58:51–59PubMedCrossRef
23.
go back to reference Tamatani S, Ito Y, Abe H, Koike T, Takeuchi S, Tanaka R (2002) Evaluation of the stability of aneurysms after embolisation using detachable coils: correlation between stability of aneurysms and embolised volume of aneurysms. AJNR Am J Neuroradiol 23:762–767PubMed Tamatani S, Ito Y, Abe H, Koike T, Takeuchi S, Tanaka R (2002) Evaluation of the stability of aneurysms after embolisation using detachable coils: correlation between stability of aneurysms and embolised volume of aneurysms. AJNR Am J Neuroradiol 23:762–767PubMed
Metadata
Title
Safety of endovascular treatment of intracranial aneurysms with a new, complex shaped Guglielmi detachable coil
Authors
Christian A. Taschner
Xavier Leclerc
Jean-Yves Gauvrit
Anis Kerkeni
Mohamed El-Mahdy
Jean-Paul Lejeune
Jean-Pierre Pruvo
Publication date
01-09-2007
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 9/2007
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-007-0247-8

Other articles of this Issue 9/2007

Neuroradiology 9/2007 Go to the issue