Short-term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia

Authors: K. Engelke, B. Stampa, W. Timm, B. Dardzinski, A. E. de Papp, H. K. Genant, T. Fuerst

Published in: Osteoporosis International | Issue 8/2012

Login to get access

Abstract

Summary

In vivo hr-pQCT precision was determined in 42 postmenopausal women using double baseline measurements from a multicenter trial of odanacatib. Errors, e.g., at the radius below 1.3% for BMD and below 6.3% for trabecular structure, were comparable to single-center results. Motion artifacts remain a challenge, particularly at the forearm.

Introduction

The short-term in vivo precision of BMD, trabecular bone structure, cortical thickness and porosity of the forearm and tibia was measured by hr-pQCT. Also the effect of image quality on precision was evaluated.

Methods

In 42 postmenopausal women (age 64.4 ± 6.8 years) out of 214 subjects enrolled in a multi center advanced imaging phase III study of odanacatib (DXA spine or hip T-scores between −1.5 and −3.5), double baseline hr-pQCT (XtremeCT) measurements with repositioning were performed. The standard ultradistal location and a second, more proximally located VOI were measured at the radius and tibia to better assess cortical thickness and porosity. Image analysis and quality grading (grades: perfect, slight artifacts, pronounced artifacts, unacceptable) were performed centrally.

Results

At the radius RMS%CV values varied from 0.7% to 1.3% for BMD and BV/TV and from 5.6% to 6.3% for Tb.Sp, Tb.Th, Tb.N, and cortical porosity. Numerically at the tibia, precision errors were approx. 0.5% lower for BMD and 1% to 2% lower for structural parameters although most differences were insignificant. In the radius but not in the tibia, precision errors for cortical thickness were smaller at the distal compared to the ultradistal location (1% versus 2%).

Conclusions

BMD precision errors were lower than those for trabecular architecture and cortical porosity. Motion artifacts remain a challenge, particularly at the forearm. Quality grading remains subjective, and more objective evaluation methods are needed. Precision in the context of a multicenter clinical trial, with centralized training and scan analysis, was comparable to single-center results previously reported.
Literature
1.
go back to reference Boutroy S, Bouxsein ML, Munoz F, Delmas PD (2005) In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab 90(12):6508–6515PubMedCrossRef Boutroy S, Bouxsein ML, Munoz F, Delmas PD (2005) In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab 90(12):6508–6515PubMedCrossRef
2.
go back to reference Laib A, Hauselmann HJ, Ruegsegger P (1998) In vivo high resolution 3D-QCT of the human forearm. Technol Health Care 6(5–6):329–337PubMed Laib A, Hauselmann HJ, Ruegsegger P (1998) In vivo high resolution 3D-QCT of the human forearm. Technol Health Care 6(5–6):329–337PubMed
3.
go back to reference Müller R, Hildebrand T, Häuselmann HJ, Rüegsegger P (1996) In vivo reproducibility of noninvasive bone biopsies using 3d-pQCT. J Bone Miner Res 11:1745–1750PubMedCrossRef Müller R, Hildebrand T, Häuselmann HJ, Rüegsegger P (1996) In vivo reproducibility of noninvasive bone biopsies using 3d-pQCT. J Bone Miner Res 11:1745–1750PubMedCrossRef
4.
go back to reference Burghardt AJ, Buie HR, Laib A, Majumdar S, Boyd SK (2010) Reproducibility of direct quantitative measures of cortical bone microarchitecture of the distal radius and tibia by HR-pQCT. Bone 47(3):519–528PubMedCrossRef Burghardt AJ, Buie HR, Laib A, Majumdar S, Boyd SK (2010) Reproducibility of direct quantitative measures of cortical bone microarchitecture of the distal radius and tibia by HR-pQCT. Bone 47(3):519–528PubMedCrossRef
5.
go back to reference Buie HR, Campbell GM, Klinck RJ, MacNeil JA, Boyd SK (2007) Automatic segmentation of cortical and trabecular compartments based on a dual threshold technique for in vivo micro-CT bone analysis. Bone 41(4):505–515PubMedCrossRef Buie HR, Campbell GM, Klinck RJ, MacNeil JA, Boyd SK (2007) Automatic segmentation of cortical and trabecular compartments based on a dual threshold technique for in vivo micro-CT bone analysis. Bone 41(4):505–515PubMedCrossRef
6.
go back to reference Burghardt AJ, Kazakia GJ, Majumdar S (2007) A local adaptive threshold strategy for high resolution peripheral quantitative computed tomography of trabecular bone. Ann Biomed Eng 35(10):1678–1686PubMedCrossRef Burghardt AJ, Kazakia GJ, Majumdar S (2007) A local adaptive threshold strategy for high resolution peripheral quantitative computed tomography of trabecular bone. Ann Biomed Eng 35(10):1678–1686PubMedCrossRef
7.
go back to reference MacNeil JA, Boyd SK (2007) Accuracy of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys 29(10):1096–1105PubMedCrossRef MacNeil JA, Boyd SK (2007) Accuracy of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys 29(10):1096–1105PubMedCrossRef
8.
go back to reference XtremeCT User’s Guide, Revision 5.05 (2005) SCANCO MEDICAL AG, Bassersdorf Switzerland XtremeCT User’s Guide, Revision 5.05 (2005) SCANCO MEDICAL AG, Bassersdorf Switzerland
9.
go back to reference Mueller TL, Stauber M, Kohler T, Eckstein F, Muller R, van Lenthe GH (2009) Non-invasive bone competence analysis by high-resolution pQCT: an in vitro reproducibility study on structural and mechanical properties at the human radius. Bone 44(2):364–371PubMedCrossRef Mueller TL, Stauber M, Kohler T, Eckstein F, Muller R, van Lenthe GH (2009) Non-invasive bone competence analysis by high-resolution pQCT: an in vitro reproducibility study on structural and mechanical properties at the human radius. Bone 44(2):364–371PubMedCrossRef
10.
go back to reference MacNeil JA, Boyd SK (2008) Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys 30(6):792–799PubMedCrossRef MacNeil JA, Boyd SK (2008) Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys 30(6):792–799PubMedCrossRef
11.
go back to reference Kazakia GJ, Hyun B, Burghardt AJ, Krug R, Newitt DC, de Papp AE, Link TM, Majumdar S (2008) In vivo determination of bone structure in postmenopausal women: a comparison of HR-pQCT and high-field MR imaging. J Bone Miner Res 23(4):463–474PubMedCrossRef Kazakia GJ, Hyun B, Burghardt AJ, Krug R, Newitt DC, de Papp AE, Link TM, Majumdar S (2008) In vivo determination of bone structure in postmenopausal women: a comparison of HR-pQCT and high-field MR imaging. J Bone Miner Res 23(4):463–474PubMedCrossRef
12.
go back to reference Swinford RR, Warden SJ (2010) Factors affecting short-term precision of musculoskeletal measures using peripheral quantitative computed tomography (pQCT). Osteoporos Int 21(11):1863–1870PubMedCrossRef Swinford RR, Warden SJ (2010) Factors affecting short-term precision of musculoskeletal measures using peripheral quantitative computed tomography (pQCT). Osteoporos Int 21(11):1863–1870PubMedCrossRef
13.
go back to reference Grampp S, Lang P, Jergas M, Gluer CC, Mathur A, Engelke K, Genant HK (1995) Assessment of the skeletal status by peripheral quantitative computed tomography of the forearm: short-term precision in vivo and comparison to dual X-ray absorptiometry. J Bone Miner Res 10(10):1566–1576PubMedCrossRef Grampp S, Lang P, Jergas M, Gluer CC, Mathur A, Engelke K, Genant HK (1995) Assessment of the skeletal status by peripheral quantitative computed tomography of the forearm: short-term precision in vivo and comparison to dual X-ray absorptiometry. J Bone Miner Res 10(10):1566–1576PubMedCrossRef
14.
go back to reference Koenig C, Wey H, Binkley T (2008) Precision of the XCT 3000 and comparison of densitometric measurements in distal radius scans between XCT 3000 and XCT 2000 peripheral quantitative computed tomography scanners. J Clin Densitom 11(4):575–580PubMedCrossRef Koenig C, Wey H, Binkley T (2008) Precision of the XCT 3000 and comparison of densitometric measurements in distal radius scans between XCT 3000 and XCT 2000 peripheral quantitative computed tomography scanners. J Clin Densitom 11(4):575–580PubMedCrossRef
15.
go back to reference Engelke K, Libanati C, Liu Y, Wang H, Austin M, Fuerst T, Stampa B, Timm W, Genant HK (2009) Quantitative computed tomography (QCT) of the forearm using general purpose spiral whole-body CT scanners: accuracy, precision and comparison with dual-energy X-ray absorptiometry (DXA). Bone 45(1):110–118PubMedCrossRef Engelke K, Libanati C, Liu Y, Wang H, Austin M, Fuerst T, Stampa B, Timm W, Genant HK (2009) Quantitative computed tomography (QCT) of the forearm using general purpose spiral whole-body CT scanners: accuracy, precision and comparison with dual-energy X-ray absorptiometry (DXA). Bone 45(1):110–118PubMedCrossRef
16.
go back to reference Eisa F, Museyko O, Hess A, Kalender WA, Engelke K (2010) Comparison of anatomic coordinate systems with rigid multi-resolution 3D registration for the reproducible positioning of analysis volumes of interest in QCT. Phys Med Biol 55(5):1429–1439PubMedCrossRef Eisa F, Museyko O, Hess A, Kalender WA, Engelke K (2010) Comparison of anatomic coordinate systems with rigid multi-resolution 3D registration for the reproducible positioning of analysis volumes of interest in QCT. Phys Med Biol 55(5):1429–1439PubMedCrossRef
17.
go back to reference Sode M, Burghardt AJ, Pialat JB, Link TM, Majumdar S (2011) Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia. Bone 48(6):1291–1297PubMedCrossRef Sode M, Burghardt AJ, Pialat JB, Link TM, Majumdar S (2011) Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia. Bone 48(6):1291–1297PubMedCrossRef
18.
go back to reference Fuerst T, Engelke K, Mc Cracken P, Stampa B, Timm W, Dardzinski B, Genant HK, De Papp A (2010) Short term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia. In 32th Annual Meeting of the American Society for Bone and Mineral Research. Toronto, Canada: JBMR Fuerst T, Engelke K, Mc Cracken P, Stampa B, Timm W, Dardzinski B, Genant HK, De Papp A (2010) Short term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia. In 32th Annual Meeting of the American Society for Bone and Mineral Research. Toronto, Canada: JBMR
Metadata
Title
Short-term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia
Authors
K. Engelke
B. Stampa
W. Timm
B. Dardzinski
A. E. de Papp
H. K. Genant
T. Fuerst
Publication date
01-08-2012
Publisher
Springer-Verlag
Published in
Osteoporosis International / Issue 8/2012
Print ISSN: 0937-941X
Electronic ISSN: 1433-2965
DOI
https://doi.org/10.1007/s00198-011-1829-1

Other articles of this Issue 8/2012

Osteoporosis International 8/2012 Go to the issue