No difference in accuracy between pinless and conventional computer-assisted surgery in total knee arthroplasty

Authors: C. Baier, G. Maderbacher, H. R. Springorum, F. Zeman, W. Fitz, J. Schaumburger, J. Grifka, J. Beckmann

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 8/2014

Login to get access

Abstract

Purpose

Many studies have demonstrated higher precision and better radiological results in Total knee arthroplasty (TKA) with computer-assisted surgery (CAS). On the other hand, studies revealed a lengthening of operation time up to 20 min for this technique and demonstrated rare additional complications as fractures and neurovascular injuries caused by the array pins and any intraoperative array dislocation leads to abortion of CAS. To combine the advantages and eliminate the disadvantages of standard CAS, we evaluated the accuracy of a so-called pinless CT-free version of knee navigation (pinless CAS) abandoning the reference pins and reducing the necessary workflow to a minimum.

Method

The present study compares the accuracy of the reference methods of two different CT-free knee navigation software versions (Brainlab Knee 2.1 and Brainlab Knee Express 2.5). Thirty patients received TKA assisted by standard CAS. Intraoperatively, the proposed bony resections of standard CAS were matched with the new pinless CAS. Postoperatively, the results were checked by evaluating the radiographs concerning leg axis, femoral flexion and tibial slope.

Results

All results concerning precise cuts (femoral as well as tibial coronal/varus–valgus alignment, femoral flexion alignment and tibial slope, resection height) were comparable between both groups (n.s.). In femoral, we found a mean deviation of coronal alignment of 0.3° (SD 0.7) and flexion of 0.2° (SD 0.8). In tibial, we found a mean deviation of coronal alignment of 0.2° (SD 0.5) and slope of 0.2° (SD 0.6). The mean additional operation time for the pinless CAS was below 2 min. The postoperative mechanical leg axis was within the threshold of 3° in all patients, tibial slope and femoral flexion matched with CAS values.

Conclusion

In clinical routine, pinless CAS can comprise the advantages of CAS leaving the disadvantages aside. It reduces surgical time and avoids complications associated with the tracking pins of conventional CAS.

Level of evidence

I.
Literature
1.
go back to reference Aunan E, Kibsgård T, Clarke-Jenssen J, Röhrl SM (2012) A new method to measure ligament balancing in total knee arthroplasty: laxity measurements in 100 knees. Arch Orthop Trauma Surg 132:1173–1181PubMedCentralPubMedCrossRef Aunan E, Kibsgård T, Clarke-Jenssen J, Röhrl SM (2012) A new method to measure ligament balancing in total knee arthroplasty: laxity measurements in 100 knees. Arch Orthop Trauma Surg 132:1173–1181PubMedCentralPubMedCrossRef
2.
go back to reference Bauwens K, Matthes G, Wich M, Gebhard F, Hanson B, Ekkernkamp A, Stengel D (2007) Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am 89:261–269PubMedCrossRef Bauwens K, Matthes G, Wich M, Gebhard F, Hanson B, Ekkernkamp A, Stengel D (2007) Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am 89:261–269PubMedCrossRef
3.
go back to reference Beldame J, Boisrenoult P, Beaufils P (2010) Pin track induced fractures around computer-assisted TKA. Orthop Traumatol Surg Res 96:249–255PubMedCrossRef Beldame J, Boisrenoult P, Beaufils P (2010) Pin track induced fractures around computer-assisted TKA. Orthop Traumatol Surg Res 96:249–255PubMedCrossRef
4.
go back to reference Bonutti P, Dethmers D, Stiehl JB (2008) Case report: femoral shaft fracture resulting from femoral tracker placement in navigated TKA. Clin Orthop Relat Res 466:1499–1502PubMedCentralPubMedCrossRef Bonutti P, Dethmers D, Stiehl JB (2008) Case report: femoral shaft fracture resulting from femoral tracker placement in navigated TKA. Clin Orthop Relat Res 466:1499–1502PubMedCentralPubMedCrossRef
5.
go back to reference Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef
6.
go back to reference Czurda T, Fennema P, Baumgartner M, Ritschl P (2010) The association between component malalignment and post-operative pain following navigation-assisted total knee arthroplasty: results of a cohort/nested case-control study. Knee Surg Sports Traumatol Arthrosc 18:863–869PubMedCrossRef Czurda T, Fennema P, Baumgartner M, Ritschl P (2010) The association between component malalignment and post-operative pain following navigation-assisted total knee arthroplasty: results of a cohort/nested case-control study. Knee Surg Sports Traumatol Arthrosc 18:863–869PubMedCrossRef
7.
go back to reference Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43PubMedCrossRef Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43PubMedCrossRef
8.
go back to reference Fitz W, Sodha S, Reichmann W, Minas T (2012) Does a modified gap-balancing technique result in medial-pivot knee kinematics in cruciate-retaining total knee arthroplasty? A pilot study. Clin Orthop Relat Res 470:91–98PubMedCentralPubMedCrossRef Fitz W, Sodha S, Reichmann W, Minas T (2012) Does a modified gap-balancing technique result in medial-pivot knee kinematics in cruciate-retaining total knee arthroplasty? A pilot study. Clin Orthop Relat Res 470:91–98PubMedCentralPubMedCrossRef
9.
go back to reference Frantz DD, Wiles AD, Leis SE, Kirsch SR (2003) Accuracy assessment protocols for electromagnetic tracking systems. Phys Med Biol 48:2241–2251PubMedCrossRef Frantz DD, Wiles AD, Leis SE, Kirsch SR (2003) Accuracy assessment protocols for electromagnetic tracking systems. Phys Med Biol 48:2241–2251PubMedCrossRef
10.
go back to reference Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89:780–785PubMedCrossRef Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89:780–785PubMedCrossRef
11.
go back to reference Lee D-H, Park J-H, Song D-I, Padhy D, Jeong W-K, Han S-B (2010) Accuracy of soft tissue balancing in TKA: comparison between navigation-assisted gap balancing and conventional measured resection. Knee Surg Sports Traumatol Arthrosc 18:381–387PubMedCrossRef Lee D-H, Park J-H, Song D-I, Padhy D, Jeong W-K, Han S-B (2010) Accuracy of soft tissue balancing in TKA: comparison between navigation-assisted gap balancing and conventional measured resection. Knee Surg Sports Traumatol Arthrosc 18:381–387PubMedCrossRef
12.
go back to reference Lopomo N, Sun L, Zaffagnini S, Giordano G, Safran MR (2010) Evaluation of formal methods in hip joint center assessment: an in vitro analysis. Clin Biomech (Bristol, Avon) 25:206–212 Lopomo N, Sun L, Zaffagnini S, Giordano G, Safran MR (2010) Evaluation of formal methods in hip joint center assessment: an in vitro analysis. Clin Biomech (Bristol, Avon) 25:206–212
13.
go back to reference Lützner J, Krummenauer F, Wolf C, Günther K-P, Kirschner S (2008) Computer-assisted and conventional total knee replacement: a comparative, prospective, randomised study with radiological and CT evaluation. J Bone Joint Surg Br 90:1039–1044PubMedCrossRef Lützner J, Krummenauer F, Wolf C, Günther K-P, Kirschner S (2008) Computer-assisted and conventional total knee replacement: a comparative, prospective, randomised study with radiological and CT evaluation. J Bone Joint Surg Br 90:1039–1044PubMedCrossRef
14.
go back to reference Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605PubMedCrossRef Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605PubMedCrossRef
15.
go back to reference Ranawat CS, Rose HA, Rich DS (1984) Total condylar knee arthroplasty for valgus and combined valgus-flexion deformity of the knee. Instr Course Lect 33:412–416PubMed Ranawat CS, Rose HA, Rich DS (1984) Total condylar knee arthroplasty for valgus and combined valgus-flexion deformity of the knee. Instr Course Lect 33:412–416PubMed
16.
go back to reference Shrout P, Fleiss J (1979) Intraclass correlations: uses in assessing rater reliability Shrout P, Fleiss J (1979) Intraclass correlations: uses in assessing rater reliability
17.
go back to reference Xie C, Liu K, Xiao L, Tang R (2012) Clinical outcomes after computer-assisted versus conventional total knee arthroplasty. Orthopedics 35:e647–e653PubMedCrossRef Xie C, Liu K, Xiao L, Tang R (2012) Clinical outcomes after computer-assisted versus conventional total knee arthroplasty. Orthopedics 35:e647–e653PubMedCrossRef
18.
go back to reference Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty. J Arthroplasty 22:1150–1161PubMedCrossRef Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty. J Arthroplasty 22:1150–1161PubMedCrossRef
19.
go back to reference Zhang G, Chen J, Chai W, Liu M, Wang Y (2011) Comparison between computer-assisted-navigation and conventional total knee arthroplasties in patients undergoing simultaneous bilateral procedures: a randomized clinical trial. J Bone Joint Surg Am 93:1190–1196PubMedCrossRef Zhang G, Chen J, Chai W, Liu M, Wang Y (2011) Comparison between computer-assisted-navigation and conventional total knee arthroplasties in patients undergoing simultaneous bilateral procedures: a randomized clinical trial. J Bone Joint Surg Am 93:1190–1196PubMedCrossRef
Metadata
Title
No difference in accuracy between pinless and conventional computer-assisted surgery in total knee arthroplasty
Authors
C. Baier
G. Maderbacher
H. R. Springorum
F. Zeman
W. Fitz
J. Schaumburger
J. Grifka
J. Beckmann
Publication date
01-08-2014
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 8/2014
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2430-2

Other articles of this Issue 8/2014

Knee Surgery, Sports Traumatology, Arthroscopy 8/2014 Go to the issue