Published in: Knee Surgery, Sports Traumatology, Arthroscopy 7/2012

01-07-2012 | Elbow

Accuracy and inter-observer reliability of visual estimation compared to clinical goniometry of the elbow

Authors: Davide Blonna, Peter C. Zarkadas, James S. Fitzsimmons, Shawn W. O’Driscoll

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 7/2012

Login to get access

Abstract

Purpose

To test the hypothesis that visual estimation by a trained observer is as accurate and reliable as clinical goniometry for measuring elbow range of motion.

Methods

Instrument validity and inter-observer reliability of visual estimation was evaluated on a consecutive series of 50 elbow contractures. Four observers with different levels of elbow experience first estimated extension and flexion of the contracted elbows and then measured them with a blinded goniometer.

Results

Instrument validity for visually-based goniometry was extremely high. ICC scores were 0.97 for both extension and flexion estimations. Systematic error was negligible (1°) with upper limits of agreement being 9° (95% CI: 7°–11°) and 8° (95% CI: 6°–10°), respectively, for extension and flexion. For the expert surgeon, 92% of the visual estimates were within 5° of the value obtained by clinical goniometry. Between experienced observers (elbow surgeon and physician assistant), the ICC’s were very high—0.96 for extension and 0.93 for flexion. The systematic errors were low, from −1° to 1° with upper limit of agreement being 11° (95% CI: 8°–14°). However, agreement was poor between an inexperienced study coordinator and the others (ICC’s: 0.51–0.38, systematic errors: 8°–18°, upper limit of agreement: 32°–40°). The accuracy of the visual estimations made by the experienced elbow surgeon was as good as the measurements taken with a goniometer by the physician assistant or the clinical fellow and better than those taken by an inexperienced study coordinator.

Conclusions

The trained human eye is highly capable of accurately estimating the range of motion of the elbow, compared to conventional clinical goniometry, depending on the experience of the observer.

Level of evidence

Diagnostic study, Level II.
Footnotes
1
The data on goniometric measurements are the same as data used in an arm of this investigation in which photographs were used to measure range of motion and compared to measurements taken by standard clinical goniometry [5].
 
Literature
1.
go back to reference Armstrong AD, MacDermid JC, Chinchalkar S, Stevens RS, King GJ (1998) Reliability of range-of-motion measurement in the elbow and forearm. J Shoulder Elbow Surg 7:573–580PubMedCrossRef Armstrong AD, MacDermid JC, Chinchalkar S, Stevens RS, King GJ (1998) Reliability of range-of-motion measurement in the elbow and forearm. J Shoulder Elbow Surg 7:573–580PubMedCrossRef
2.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef
3.
go back to reference Bland JM, Altman DG (1995) Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346:1085–1087PubMedCrossRef Bland JM, Altman DG (1995) Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346:1085–1087PubMedCrossRef
4.
go back to reference Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef
5.
6.
go back to reference Boone DC, Azen SP, Lin CM, Spence C, Baron C, Lee L (1978) Reliability of goniometric measurements. Phys Ther 58:1355–1390PubMed Boone DC, Azen SP, Lin CM, Spence C, Baron C, Lee L (1978) Reliability of goniometric measurements. Phys Ther 58:1355–1390PubMed
7.
go back to reference Bovens AMPM, van Baak MA, Vrencken JGPM, Wijnen JAG, Verstappen FTJ (1990) Variability and reliability of joint measurements. Am J Sports Med 18:58–63PubMedCrossRef Bovens AMPM, van Baak MA, Vrencken JGPM, Wijnen JAG, Verstappen FTJ (1990) Variability and reliability of joint measurements. Am J Sports Med 18:58–63PubMedCrossRef
8.
go back to reference Brosseau L, Balmer S, Tousignant M, O’Sullivan JP, Goudreault C, Goudreault M, Gringras S (2001) Intra- and intertester reliability and criterion validity of the parallelogram and universal goniometers for measuring maximum active knee flexion and extension of patients with knee restrictions. Arch Phys Med Rehabil 82:396–402PubMedCrossRef Brosseau L, Balmer S, Tousignant M, O’Sullivan JP, Goudreault C, Goudreault M, Gringras S (2001) Intra- and intertester reliability and criterion validity of the parallelogram and universal goniometers for measuring maximum active knee flexion and extension of patients with knee restrictions. Arch Phys Med Rehabil 82:396–402PubMedCrossRef
9.
go back to reference Cleffken B, van Breukelen G, van Mameren H, Brink P, Olde Damink S (2007) Test-retest reproducibility of elbow goniometric measurements in a rigid double-blinded protocol: intervals for distinguishing between measurement error and clinical change. J Shoulder Elbow Surg 16:788–794PubMedCrossRef Cleffken B, van Breukelen G, van Mameren H, Brink P, Olde Damink S (2007) Test-retest reproducibility of elbow goniometric measurements in a rigid double-blinded protocol: intervals for distinguishing between measurement error and clinical change. J Shoulder Elbow Surg 16:788–794PubMedCrossRef
10.
11.
go back to reference Fish DR, Wingate L (1985) Sources of goniometric error at the elbow. Phys Ther 65:1666–1670PubMed Fish DR, Wingate L (1985) Sources of goniometric error at the elbow. Phys Ther 65:1666–1670PubMed
12.
go back to reference Gajdosik RL, Bohannon RW (1987) Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Phys Ther 67:1867–1872PubMed Gajdosik RL, Bohannon RW (1987) Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Phys Ther 67:1867–1872PubMed
13.
go back to reference Haas M (1991) The reliability of reliability. J Manipulative Physiol Ther 14:199–208PubMed Haas M (1991) The reliability of reliability. J Manipulative Physiol Ther 14:199–208PubMed
14.
go back to reference Haas M (1991) Statistical methodology for reliability studies. J Manipulative Physiol Ther 14:119–132PubMed Haas M (1991) Statistical methodology for reliability studies. J Manipulative Physiol Ther 14:119–132PubMed
15.
go back to reference Haight HJ, Dahm DL, Smith J, Krause DA (2005) Measuring standing hindfoot alignment: reliability of goniometric and visual measurements. Arch Phys Med Rehabil 86:571–575PubMedCrossRef Haight HJ, Dahm DL, Smith J, Krause DA (2005) Measuring standing hindfoot alignment: reliability of goniometric and visual measurements. Arch Phys Med Rehabil 86:571–575PubMedCrossRef
16.
go back to reference Hellebrandt FA, Duvall EN, Moore ML (1949) The measurement of joint motion: part III. Reliability of goniometry. Phys Ther Rev 29:302–307 Hellebrandt FA, Duvall EN, Moore ML (1949) The measurement of joint motion: part III. Reliability of goniometry. Phys Ther Rev 29:302–307
17.
go back to reference Jakobsen TL, Christensen M, Christensen SS, Olsen M, Bandholm T (2010) Reliability of knee joint range of motion and circumference measurements after total knee arthroplasty: does tester experience matter? Physiother Res Int 15:126–134PubMedCrossRef Jakobsen TL, Christensen M, Christensen SS, Olsen M, Bandholm T (2010) Reliability of knee joint range of motion and circumference measurements after total knee arthroplasty: does tester experience matter? Physiother Res Int 15:126–134PubMedCrossRef
18.
go back to reference Low JL (1976) The reliability of joint measurement. Physiotherapy 62:227–229PubMed Low JL (1976) The reliability of joint measurement. Physiotherapy 62:227–229PubMed
19.
go back to reference Piriyaprasarth P, Morris ME (2007) Psychometric properties of measurement tools for quantifying knee joint position and movement: a systematic review. Knee 14:2–8PubMedCrossRef Piriyaprasarth P, Morris ME (2007) Psychometric properties of measurement tools for quantifying knee joint position and movement: a systematic review. Knee 14:2–8PubMedCrossRef
20.
go back to reference Rothstein JM, Miller PJ, Roettger RF (1983) Goniometric reliability in a clinical setting. Elbow and knee measurements. Phys Ther 63:1611–1615PubMed Rothstein JM, Miller PJ, Roettger RF (1983) Goniometric reliability in a clinical setting. Elbow and knee measurements. Phys Ther 63:1611–1615PubMed
21.
go back to reference Shultz SJ, Nguyen AD, Windley TC, Kulas AS, Botic TL, Beynnon BD (2006) Intratester and intertester reliability of clinical measures of lower extremity anatomic characteristics: implications for multicenter studies. Clin J Sport Med 16:155–161PubMedCrossRef Shultz SJ, Nguyen AD, Windley TC, Kulas AS, Botic TL, Beynnon BD (2006) Intratester and intertester reliability of clinical measures of lower extremity anatomic characteristics: implications for multicenter studies. Clin J Sport Med 16:155–161PubMedCrossRef
22.
go back to reference Watkins MA, Riddle DL, Lamb RL, Personius WJ (1991) Reliability of goniometric measurements and visual estimates of knee range of motion obtained in a clinical setting. Phys Ther 71:90–96 Discussion 96–97PubMed Watkins MA, Riddle DL, Lamb RL, Personius WJ (1991) Reliability of goniometric measurements and visual estimates of knee range of motion obtained in a clinical setting. Phys Ther 71:90–96 Discussion 96–97PubMed
23.
go back to reference Youdas JW, Bogard CL, Suman VJ (1993) Reliability of goniometric measurements and visual estimates of ankle joint active range of motion obtained in a clinical setting. Arch Phys Med Rehabil 74:1113–1118PubMedCrossRef Youdas JW, Bogard CL, Suman VJ (1993) Reliability of goniometric measurements and visual estimates of ankle joint active range of motion obtained in a clinical setting. Arch Phys Med Rehabil 74:1113–1118PubMedCrossRef
Metadata
Title
Accuracy and inter-observer reliability of visual estimation compared to clinical goniometry of the elbow
Authors
Davide Blonna
Peter C. Zarkadas
James S. Fitzsimmons
Shawn W. O’Driscoll
Publication date
01-07-2012
Publisher
Springer-Verlag
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 7/2012
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-011-1720-9

Other articles of this Issue 7/2012

Knee Surgery, Sports Traumatology, Arthroscopy 7/2012 Go to the issue