Axial and torsional stability of supracondylar femur osteotomies: biomechanical comparison of the stability of five different plate and osteotomy configurations

Authors: J.-M. Brinkman, C. Hurschler, J. D. Agneskirchner, D. Freiling, R. J. van Heerwaarden

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 4/2011

Login to get access

Abstract

Purpose

Little is known regarding the biomechanical stability and stiffness of implants and techniques used in supracondylar femur osteotomies (SCO). Therefore, fixation stability and stiffness of implants to bone was investigated under simulated physiological loading conditions using a composite femur model and a 3D motion-analysis system.

Methods

Five osteotomy configurations were investigated: (1) oblique medial closing-wedge fixated with an angle-stable implant; (2) oblique and (3) perpendicular medial closing-wedge, both fixated with an angled blade plate; and lateral opening-wedge fixated with (4) a spacer plate and (5) an angle-stable lateral implant. The motion measured at the osteotomy was used to calculate the stiffness and stability of the constructs.

Results

The least amount of motion and highest stiffness was measured in the medial oblique closing-wedge osteotomy fixated with the angled blade plate. The lateral opening-wedge techniques were less stable and had a lower stiffness compared with the medial; the oblique saw cuts were more stable and had a higher stiffness than the perpendicular.

Conclusion

This experimental study presents baseline data on the differences in the primary stability of bone–implant constructs used in SCO. The data in this study can be used as reference for future testing of SCO techniques. Furthermore, it is recommended that based on the differences found, the early postoperative rehabilitation protocol is tailored to the stability and stiffness of the fixation method used.
Literature
1.
go back to reference Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14:291–300PubMedCrossRef Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14:291–300PubMedCrossRef
2.
go back to reference Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90:1548–1557PubMedCrossRef Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90:1548–1557PubMedCrossRef
3.
go back to reference Cusick RP, Lucas GL, McQueen DA, Graber CD (2000) Construct stiffness of different fixation methods for supracondylar femoral fractures above total knee prostheses. Am J Orthop 29:695–699PubMed Cusick RP, Lucas GL, McQueen DA, Graber CD (2000) Construct stiffness of different fixation methods for supracondylar femoral fractures above total knee prostheses. Am J Orthop 29:695–699PubMed
4.
go back to reference Dvir Z, Prushansky T (2000) Reproducibility and instrument validity of a new ultrasonography-based system for measuring cervical spine kinematics. Clin Biomech 15:658–664CrossRef Dvir Z, Prushansky T (2000) Reproducibility and instrument validity of a new ultrasonography-based system for measuring cervical spine kinematics. Clin Biomech 15:658–664CrossRef
5.
go back to reference Firoozbakhsh K, Behzadi K, DeCoster TA, Moneim MS, Naraghi FF (1995) Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma 9:152–157PubMedCrossRef Firoozbakhsh K, Behzadi K, DeCoster TA, Moneim MS, Naraghi FF (1995) Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma 9:152–157PubMedCrossRef
6.
go back to reference Franco V, Cipolla M, Gerullo G, Gianni E, Puddu G (2004) Open wedge osteotomy of the distal femur in the valgus knee. Orthopade 33:185–192PubMedCrossRef Franco V, Cipolla M, Gerullo G, Gianni E, Puddu G (2004) Open wedge osteotomy of the distal femur in the valgus knee. Orthopade 33:185–192PubMedCrossRef
7.
go back to reference Frigg R (2003) Development of the locking compression plate. Injury 34(Supplement 2):6–10CrossRef Frigg R (2003) Development of the locking compression plate. Injury 34(Supplement 2):6–10CrossRef
8.
go back to reference van Heerwaarden RJ, Wymenga AB, Freiling D, Lobenhoffer P (2007) Distal medial closed wedge varus femur osteotomy stabilized with Tomofix plate fixator. Op Tech Orthopaedics 17:12–21CrossRef van Heerwaarden RJ, Wymenga AB, Freiling D, Lobenhoffer P (2007) Distal medial closed wedge varus femur osteotomy stabilized with Tomofix plate fixator. Op Tech Orthopaedics 17:12–21CrossRef
9.
go back to reference Heiner AD, Brown TD (2001) Structural properties of a new design of composite replicate femurs and tibias. J Biomech 34:773–781PubMedCrossRef Heiner AD, Brown TD (2001) Structural properties of a new design of composite replicate femurs and tibias. J Biomech 34:773–781PubMedCrossRef
10.
go back to reference Hsu RW, Himeno S, Coventry MB, Chao EY (1990) Normal axial alignment of the lower extremity and load-bearing distribution at the knee. Clin Orthop 90:215–227 Hsu RW, Himeno S, Coventry MB, Chao EY (1990) Normal axial alignment of the lower extremity and load-bearing distribution at the knee. Clin Orthop 90:215–227
11.
go back to reference Learmonth ID (1990) A simple technique for varus supracondylar osteotomy in genu valgum. J Bone Joint Surg Br 72:235–237PubMed Learmonth ID (1990) A simple technique for varus supracondylar osteotomy in genu valgum. J Bone Joint Surg Br 72:235–237PubMed
13.
go back to reference Marti RK, Schröder J, Witteveen A (2000) The closed wedge varus supracondylar osteotomy. Oper Tech Sports Med 8:48–55CrossRef Marti RK, Schröder J, Witteveen A (2000) The closed wedge varus supracondylar osteotomy. Oper Tech Sports Med 8:48–55CrossRef
14.
go back to reference McDermott AG, Finklestein JA, Farine I, Boynton EL, MacIntosh DL, Gross A (1988) Distal femoral varus osteotomy for valgus deformity of the knee. J Bone Joint Surg Am 70:110–116PubMed McDermott AG, Finklestein JA, Farine I, Boynton EL, MacIntosh DL, Gross A (1988) Distal femoral varus osteotomy for valgus deformity of the knee. J Bone Joint Surg Am 70:110–116PubMed
15.
go back to reference Miniaci A, Grossmann SP, Jakob RP (1990) Supracondylar femoral varus osteotomy in the treatment of valgus knee deformity. Am J Knee Surg 3:65–73 Miniaci A, Grossmann SP, Jakob RP (1990) Supracondylar femoral varus osteotomy in the treatment of valgus knee deformity. Am J Knee Surg 3:65–73
16.
go back to reference Miniaci A, Watson LW (2002) Distal femoral osteotomy. In: Fu FF (ed) Knee surgery, 1994. Williams & Wilkins, New York, pp 1173–1180 Miniaci A, Watson LW (2002) Distal femoral osteotomy. In: Fu FF (ed) Knee surgery, 1994. Williams & Wilkins, New York, pp 1173–1180
17.
go back to reference Stahelin T, Hardegger F, Ward JC (2000) Supracondylar osteotomy of the femur with use of compression Osteosynthesis with a malleable implant. J Bone Joint Surg Am 82:712–722PubMed Stahelin T, Hardegger F, Ward JC (2000) Supracondylar osteotomy of the femur with use of compression Osteosynthesis with a malleable implant. J Bone Joint Surg Am 82:712–722PubMed
18.
go back to reference Stahelin T, Hardegger F (2004) Incomplete, supracondylar femur osteotomy. A minimally invasive compression osteosynthesis with soft implant. Orthopade 33:178–184PubMedCrossRef Stahelin T, Hardegger F (2004) Incomplete, supracondylar femur osteotomy. A minimally invasive compression osteosynthesis with soft implant. Orthopade 33:178–184PubMedCrossRef
19.
go back to reference Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):55–62CrossRef Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):55–62CrossRef
20.
go back to reference Stoffel K, Dieter U, Stachowiak G, Gachter A, Kuster MS (2003) Biomechanical testing of the LCP—how can stability in locked internal fixators be controlled? Injury 34(Suppl 2):11–19CrossRef Stoffel K, Dieter U, Stachowiak G, Gachter A, Kuster MS (2003) Biomechanical testing of the LCP—how can stability in locked internal fixators be controlled? Injury 34(Suppl 2):11–19CrossRef
21.
go back to reference Stoffel K, Stachowiak G, Kuster M (2004) Open wedge high tibial osteotomy: biomechanical investigation of the modified Arthrex Osteotomy Plate (Puddu Plate) and the TomoFix Plate. Clin Biomech 19:944–950CrossRef Stoffel K, Stachowiak G, Kuster M (2004) Open wedge high tibial osteotomy: biomechanical investigation of the modified Arthrex Osteotomy Plate (Puddu Plate) and the TomoFix Plate. Clin Biomech 19:944–950CrossRef
22.
go back to reference Szivek JA, Weng M, Karpman R (1990) Variability in the torsional and bending reponse of a commercially available composite “femur”. J Appl Biomater 1:183–186CrossRef Szivek JA, Weng M, Karpman R (1990) Variability in the torsional and bending reponse of a commercially available composite “femur”. J Appl Biomater 1:183–186CrossRef
23.
go back to reference Wagner M (2003) General principles for the clinical use of the LCP. Injury 34(Suppl 2):31–42CrossRef Wagner M (2003) General principles for the clinical use of the LCP. Injury 34(Suppl 2):31–42CrossRef
24.
go back to reference Walsh JC, Quinlan JF, Stapleton R, FitzPatrick DP, McCormack D (2007) Three-dimensional motion analysis of the lumbar spine during “free squat” weight lift training. Am J Sports Med 35:927–932PubMedCrossRef Walsh JC, Quinlan JF, Stapleton R, FitzPatrick DP, McCormack D (2007) Three-dimensional motion analysis of the lumbar spine during “free squat” weight lift training. Am J Sports Med 35:927–932PubMedCrossRef
25.
go back to reference Zlowodzki M, Williamson S, Cole PA, Zardiackas LD, Kregor PJ (2004) Biomechanical evaluation of the less invasive stabilization system, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures. J Orthop Trauma 8:494–502CrossRef Zlowodzki M, Williamson S, Cole PA, Zardiackas LD, Kregor PJ (2004) Biomechanical evaluation of the less invasive stabilization system, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures. J Orthop Trauma 8:494–502CrossRef
Metadata
Title
Axial and torsional stability of supracondylar femur osteotomies: biomechanical comparison of the stability of five different plate and osteotomy configurations
Authors
J.-M. Brinkman
C. Hurschler
J. D. Agneskirchner
D. Freiling
R. J. van Heerwaarden
Publication date
01-04-2011
Publisher
Springer-Verlag
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 4/2011
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-010-1281-3

Other articles of this Issue 4/2011

Knee Surgery, Sports Traumatology, Arthroscopy 4/2011 Go to the issue