01-10-2012 | Physiological and Technical Notes

Comparison of volumetric capnography and mixed expired gas methods to calculate physiological dead space in mechanically ventilated ICU patients

Authors: Pratik Sinha, Neil Soni

Published in: Intensive Care Medicine | Issue 10/2012

Login to get access

Abstract

Introduction

Physiological dead space should be a routine measurement in ventilated patients but measuring dead space using the Douglas bag (DB) method is cumbersome and requires corrections for compressed ventilator gas. These factors make this method impractical in the critical care setting. Volumetric capnography (VCAP) offers a relatively simple solution to calculating dead space. Few studies have been conducted to directly compare dead space measured by VCAP and the DB method in critically unwell adults.

Method

Prospective observational study of 48 mechanically ventilated adults ICU patients. Dead space was calculated simultaneously using VCAP (CO2SMO) and the Bohr–Enghoff equation. In total, 168 paired readings were taken. Single-breath CO2 waveform areas under the curve were computed automatically by software to calculate physiological dead space. The calculated value of \( P_{{\bar{E}_{{{\text{CO}}_{2} }} }} \) was also recorded from the CO2SMO device. Exhaust ventilator gas was collected in a 10-l mixing chamber. \( P_{{\bar{E}_{{{\text{CO}}_{2} }} }} \) was measured in the chamber following correction for compressed gas.

Results

The study demonstrated good agreement between physiological V D/V T calculated by VCAP and corrected (mean bias 0.03), and uncorrected (mean bias 0.02) Bohr–Enghoff method. There was good correlation between the two methods of measurement (VCAP vs corrected r 2 = 0.90 P < 0.001, VCAP vs uncorrected r 2 = 0.90, P < 0.001). There was good correlation between \( P_{{\bar{E}_{{{\text{CO}}_{2} }} }} \) calculated by the CO2SMO and in the exhaust collected gas (mean bias 0.08).

Conclusions

VCAP shows good agreement with Douglas Bag method in measuring physiological V D/V T over a wide range of dead space fractions.
Literature
1.
go back to reference Cepkova M, Kapur V, Ren X, Quinn T, Zhuo H, Foster E, Liu KD, Matthay MA (2007) Pulmonary dead space fraction and pulmonary artery systolic pressure as early predictors of clinical outcome in acute lung injury. Chest 132:836–842PubMedCrossRef Cepkova M, Kapur V, Ren X, Quinn T, Zhuo H, Foster E, Liu KD, Matthay MA (2007) Pulmonary dead space fraction and pulmonary artery systolic pressure as early predictors of clinical outcome in acute lung injury. Chest 132:836–842PubMedCrossRef
2.
go back to reference Kallet RH, Alonso JA, Pittet JF, Matthay MA (2004) Prognostic value of the pulmonary dead-space fraction during the first 6 days of acute respiratory distress syndrome. Respir Care 49:1008–1014PubMed Kallet RH, Alonso JA, Pittet JF, Matthay MA (2004) Prognostic value of the pulmonary dead-space fraction during the first 6 days of acute respiratory distress syndrome. Respir Care 49:1008–1014PubMed
3.
go back to reference Nuckton TJ, Alonso JA, Kallet RH, Daniel BM, Pittet JF, Eisner MD, Matthay MA (2002) Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome. N Engl J Med 346:1281–1286PubMedCrossRef Nuckton TJ, Alonso JA, Kallet RH, Daniel BM, Pittet JF, Eisner MD, Matthay MA (2002) Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome. N Engl J Med 346:1281–1286PubMedCrossRef
4.
go back to reference Raurich JM, Vilar M, Colomar A, Ibanez J, Ayestaran I, Perez-Barcena J, Llompart-Pou JA (2010) Prognostic value of the pulmonary dead-space fraction during the early and intermediate phases of acute respiratory distress syndrome. Respir Care 55:282–287PubMed Raurich JM, Vilar M, Colomar A, Ibanez J, Ayestaran I, Perez-Barcena J, Llompart-Pou JA (2010) Prognostic value of the pulmonary dead-space fraction during the early and intermediate phases of acute respiratory distress syndrome. Respir Care 55:282–287PubMed
5.
go back to reference Siddiki H, Kojicic M, Li G, Yilmaz M, Thompson TB, Hubmayr RD, Gajic O (2010) Bedside quantification of dead-space fraction using routine clinical data in patients with acute lung injury: secondary analysis of two prospective trials. Crit Care 14:R141PubMedCrossRef Siddiki H, Kojicic M, Li G, Yilmaz M, Thompson TB, Hubmayr RD, Gajic O (2010) Bedside quantification of dead-space fraction using routine clinical data in patients with acute lung injury: secondary analysis of two prospective trials. Crit Care 14:R141PubMedCrossRef
6.
go back to reference Crossman PF, Bushnell LS, Hedley-Whyte J (1970) Dead space during artificial ventilation: gas compression and mechanical dead space. J Appl Physiol 28:94–97PubMed Crossman PF, Bushnell LS, Hedley-Whyte J (1970) Dead space during artificial ventilation: gas compression and mechanical dead space. J Appl Physiol 28:94–97PubMed
7.
go back to reference Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493PubMedCrossRef Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493PubMedCrossRef
8.
go back to reference Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493PubMedCrossRef Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493PubMedCrossRef
9.
go back to reference Fletcher R, Jonson B, Cumming G, Brew J (1981) The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 53:77–88PubMedCrossRef Fletcher R, Jonson B, Cumming G, Brew J (1981) The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 53:77–88PubMedCrossRef
10.
go back to reference Fletcher R, Jonson B, Cumming G, Brew J (1981) The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 53:77–88PubMedCrossRef Fletcher R, Jonson B, Cumming G, Brew J (1981) The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 53:77–88PubMedCrossRef
11.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef
12.
go back to reference Arnold JH, Thompson JE, Arnold LW (1996) Single breath CO2 analysis: description and validation of a method. Crit Care Med 24:96–102PubMedCrossRef Arnold JH, Thompson JE, Arnold LW (1996) Single breath CO2 analysis: description and validation of a method. Crit Care Med 24:96–102PubMedCrossRef
13.
go back to reference Kallet RH, Daniel BM, Garcia O, Matthay MA (2005) Accuracy of physiologic dead space measurements in patients with acute respiratory distress syndrome using volumetric capnography: comparison with the metabolic monitor method. Respir Care 50:462–467PubMed Kallet RH, Daniel BM, Garcia O, Matthay MA (2005) Accuracy of physiologic dead space measurements in patients with acute respiratory distress syndrome using volumetric capnography: comparison with the metabolic monitor method. Respir Care 50:462–467PubMed
14.
go back to reference Lum L, Saville A, Venkataraman ST (1998) Accuracy of physiologic deadspace measurement in intubated pediatric patients using a metabolic monitor: comparison with the Douglas bag method. Crit Care Med 26:760–764PubMedCrossRef Lum L, Saville A, Venkataraman ST (1998) Accuracy of physiologic deadspace measurement in intubated pediatric patients using a metabolic monitor: comparison with the Douglas bag method. Crit Care Med 26:760–764PubMedCrossRef
15.
go back to reference Sinha P, Flower O, Soni N (2011) Deadspace ventilation: a waste of breath! Intensive Care Med 37:735–746PubMedCrossRef Sinha P, Flower O, Soni N (2011) Deadspace ventilation: a waste of breath! Intensive Care Med 37:735–746PubMedCrossRef
16.
go back to reference Fletcher R, Werner O, Nordstrom L, Jonson B (1983) Sources of error and their correction in the measurement of carbon dioxide elimination using the Siemens-Elema CO2 Analyzer. Br J Anaesth 55:177–185PubMedCrossRef Fletcher R, Werner O, Nordstrom L, Jonson B (1983) Sources of error and their correction in the measurement of carbon dioxide elimination using the Siemens-Elema CO2 Analyzer. Br J Anaesth 55:177–185PubMedCrossRef
17.
go back to reference Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J (2002) Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 30:2566–2574PubMedCrossRef Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J (2002) Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 30:2566–2574PubMedCrossRef
Metadata
Title
Comparison of volumetric capnography and mixed expired gas methods to calculate physiological dead space in mechanically ventilated ICU patients
Authors
Pratik Sinha
Neil Soni
Publication date
01-10-2012
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 10/2012
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-012-2670-5

Other articles of this Issue 10/2012

Intensive Care Medicine 10/2012 Go to the issue