Published in: Intensive Care Medicine 2/2011

01-02-2011 | Physiological and Technological Notes

Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure

Authors: P.-E. Gancel, E. Roupie, L. Guittet, S. Laplume, N. Terzi

Published in: Intensive Care Medicine | Issue 2/2011

Login to get access

Abstract

Purpose

Transcutaneous CO2 monitors are widely used in neonatal ICUs. Until recently, these devices performed poorly in adults. Recent technical modifications have produced transcutaneous CO2 monitors that have performed well in adults with chronic illnesses. We evaluated the accuracy of one of these devices, the TOSCA® 500, in adults admitted to an emergency department for acute respiratory failure.

Methods

We prospectively collected 29 pairs of simultaneous transcutaneous arterial CO2 (PtcCO2) and arterial CO2 (PaCO2) values in 21 consecutive adults with acute respiratory failure (acute heart failure, n = 6; COPD exacerbation, n = 8; acute pneumonia, n = 6; and pulmonary embolism, n = 1). Agreement between PaCO2 and PtcCO2 was evaluated using the Bland-Altman method.

Results

Mean arterial oxygen saturation was 90%, arterial oxygen tension ranged from 32 to 215 mmHg, and PaCO2 ranged from 23 to 84 mmHg. The mean difference between PaCO2 and PtcCO2 was 0.1 mmHg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from −6 to 6.2 mmHg. None of the patients experienced adverse effects from heating of the device clipped to the earlobe.

Conclusion

PtcCO2 showed good agreement with PaCO2 in adults with acute respiratory failure.
Literature
1.
go back to reference Crawford A (2004) An audit of the patient’s experience of arterial blood gas testing. Br J Nurs 13:529–532PubMed Crawford A (2004) An audit of the patient’s experience of arterial blood gas testing. Br J Nurs 13:529–532PubMed
2.
go back to reference Zaloga GP (1990) Evaluation of bedside testing options for the critical care unit. Chest 97:185S–190SPubMed Zaloga GP (1990) Evaluation of bedside testing options for the critical care unit. Chest 97:185S–190SPubMed
3.
go back to reference Thorson SH, Marini JJ, Pierson DJ, Hudson LD (1983) Variability of arterial blood gas values in stable patients in the icu. Chest 84:14–18CrossRefPubMed Thorson SH, Marini JJ, Pierson DJ, Hudson LD (1983) Variability of arterial blood gas values in stable patients in the icu. Chest 84:14–18CrossRefPubMed
4.
go back to reference Eberhard P (2007) The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg 105:S48–S52CrossRefPubMed Eberhard P (2007) The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg 105:S48–S52CrossRefPubMed
5.
go back to reference Rosner V, Hannhart B, Chabot F, Polu JM (1999) Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. Eur Respir J 13:1044–1047CrossRefPubMed Rosner V, Hannhart B, Chabot F, Polu JM (1999) Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. Eur Respir J 13:1044–1047CrossRefPubMed
6.
go back to reference Sanders MH, Kern NB, Costantino JP, Stiller RA, Strollo PJ Jr, Studnicki KA, Coates JA, Richards TJ (1994) Accuracy of end-tidal and transcutaneous pco2 monitoring during sleep. Chest 106:472–483CrossRefPubMed Sanders MH, Kern NB, Costantino JP, Stiller RA, Strollo PJ Jr, Studnicki KA, Coates JA, Richards TJ (1994) Accuracy of end-tidal and transcutaneous pco2 monitoring during sleep. Chest 106:472–483CrossRefPubMed
7.
go back to reference Cuvelier A, Grigoriu B, Molano LC, Muir JF (2005) Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation. Chest 127:1744–1748CrossRefPubMed Cuvelier A, Grigoriu B, Molano LC, Muir JF (2005) Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation. Chest 127:1744–1748CrossRefPubMed
8.
go back to reference Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M (2008) Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med 34:1340–1344CrossRefPubMed Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M (2008) Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med 34:1340–1344CrossRefPubMed
9.
go back to reference McVicar J, Eager R (2009) Validation study of a transcutaneous carbon dioxide monitor in patients in the emergency department. Emerg Med J 26:344–346CrossRefPubMed McVicar J, Eager R (2009) Validation study of a transcutaneous carbon dioxide monitor in patients in the emergency department. Emerg Med J 26:344–346CrossRefPubMed
10.
go back to reference Parker SM, Gibson GJ (2007) Evaluation of a transcutaneous carbon dioxide monitor (“Tosca”) in adult patients in routine respiratory practice. Respir Med 101:261–264CrossRefPubMed Parker SM, Gibson GJ (2007) Evaluation of a transcutaneous carbon dioxide monitor (“Tosca”) in adult patients in routine respiratory practice. Respir Med 101:261–264CrossRefPubMed
11.
go back to reference Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA (2005) Transcutaneous pco2 monitoring in critically ill adults: clinical evaluation of a new sensor. Crit Care Med 33:2203–2206CrossRefPubMed Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA (2005) Transcutaneous pco2 monitoring in critically ill adults: clinical evaluation of a new sensor. Crit Care Med 33:2203–2206CrossRefPubMed
12.
go back to reference Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW (2004) Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees c. J Clin Monit Comput 18:343–345CrossRefPubMed Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW (2004) Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees c. J Clin Monit Comput 18:343–345CrossRefPubMed
Metadata
Title
Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure
Authors
P.-E. Gancel
E. Roupie
L. Guittet
S. Laplume
N. Terzi
Publication date
01-02-2011
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 2/2011
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-010-2076-1

Other articles of this Issue 2/2011

Intensive Care Medicine 2/2011 Go to the issue