01-02-2006 | Technical Note

Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients

Authors: Pablo Rodriguez, François Lellouche, Jerome Aboab, Christian Brun Buisson, Laurent Brochard

Published in: Intensive Care Medicine | Issue 2/2006

Login to get access

Abstract

Objective

To evaluate the accuracy of transcutaneous PCO2 (PtcCO2) as a surrogate for arterial PCO2 (PaCO2) in a cohort of adult critically ill patients in a medical intensive care unit (ICU).

Design

Prospective observational study comparing paired measures of transcutaneous and arterial PCO2.

Setting

A 26-bed medical ICU.

Patients

Fifty ICU patients monitored with a SenTec Digital Monitor placed at the ear lobe over prolonged periods.

Results

A total of 189 paired PCO2 measures were obtained. Twenty-one were excluded from analysis, because profound skin vasoconstriction was present (PCO2 bias = −10.8 ± 21.8  mmHg). Finally, 168 were analysed, including 137 obtained during mechanical ventilation and 82 under catecholamine treatment. Body temperature was below 36°C for 27 measurements. Mean duration of monitoring was 17 ± 17 h. The mean difference between PaCO2 and PtcCO2 was −0.2 ± 4.6  mmHg with a tight correlation (R 2 = 0.92, p > 0.01). PCO2 bias did not significantly change among three successive measurements. Changes in PaCO2 and in PtcCO2 between two blood samples were well correlated (R 2 = 0.78, p > 0.01). Variations of more than 8 mmHg in PtcCO2 had 86% sensitivity and 80% specificity to correctly predict similar changes in PaCO2 in the same direction. Catecholamine dose or respiratory support did not affect PtcCO2 accuracy. Hypothermia has only a small effect on accuracy. No complication related to a prolonged use of the sensor was observed

Conclusion

Transcutaneous PCO2 provides a safe and reliable trend-monitoring tool, provided there is no major vasoconstriction.
Appendix
Available only for authorised users
Literature
1.
go back to reference Hess DR (1998) Capnometry. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 377–400 Hess DR (1998) Capnometry. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 377–400
2.
go back to reference Simon M, Gottschall R, Gugel M, Fritz H, Mohr S, Klein U (2003) Comparison of transcutaneous and endtidal CO2-monitoring for rigid bronchoscopy during high-frequency jet ventilation. Acta Anaesthesiol Scand 47:861–867PubMedCrossRef Simon M, Gottschall R, Gugel M, Fritz H, Mohr S, Klein U (2003) Comparison of transcutaneous and endtidal CO2-monitoring for rigid bronchoscopy during high-frequency jet ventilation. Acta Anaesthesiol Scand 47:861–867PubMedCrossRef
3.
go back to reference Saura P, Blanch L, Lucangelo U, Fernandez R, Mestre J, Artigas A (1996) Use of capnography to detect hypercapnic episodes during weaning from mechanical ventilation. Intensive Care Med 22:374–381PubMedCrossRef Saura P, Blanch L, Lucangelo U, Fernandez R, Mestre J, Artigas A (1996) Use of capnography to detect hypercapnic episodes during weaning from mechanical ventilation. Intensive Care Med 22:374–381PubMedCrossRef
5.
go back to reference Meade MO, Cook DJ (1998) A framework for decision making. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 141–147 Meade MO, Cook DJ (1998) A framework for decision making. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 141–147
6.
go back to reference Palmisano BW, Severinghaus JW (1990) Transcutaneous PCO2 and PO2: a multicenter study of accuracy. J Clin Monit 6:189–195PubMedCrossRef Palmisano BW, Severinghaus JW (1990) Transcutaneous PCO2 and PO2: a multicenter study of accuracy. J Clin Monit 6:189–195PubMedCrossRef
7.
go back to reference Severinghaus JW (1981) A combined transcutaneous PO2-PCO2 electrode with electrochemical HCO3 −-stabilization. J Appl Physiol 51:1027–1032PubMed Severinghaus JW (1981) A combined transcutaneous PO2-PCO2 electrode with electrochemical HCO3 -stabilization. J Appl Physiol 51:1027–1032PubMed
8.
go back to reference Rodriguez PO, Lellouche F, Aboab J, Brun-Buisson C, Brochard L (2004) Transcutaneous arterial carbon dioxide partial pressure monitoring in ICU patients. Intensive Care Med 30:S168CrossRef Rodriguez PO, Lellouche F, Aboab J, Brun-Buisson C, Brochard L (2004) Transcutaneous arterial carbon dioxide partial pressure monitoring in ICU patients. Intensive Care Med 30:S168CrossRef
9.
go back to reference Tremper KK, Shoemaker WC, Shippy CR, Nolan LS (1981) Transcutaneous PCO2 monitoring on adult patients in the ICU and the operating room. Crit Care Med 9:752–755PubMedCrossRef Tremper KK, Shoemaker WC, Shippy CR, Nolan LS (1981) Transcutaneous PCO2 monitoring on adult patients in the ICU and the operating room. Crit Care Med 9:752–755PubMedCrossRef
10.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
11.
go back to reference Weinstein S, Obuchowski NA, Lieber ML (2005) Clinical evaluation of diagnostic tests. AJR Am J Roentgenol 184:14–19PubMed Weinstein S, Obuchowski NA, Lieber ML (2005) Clinical evaluation of diagnostic tests. AJR Am J Roentgenol 184:14–19PubMed
12.
go back to reference Obuchowski NA (2005) ROC analysis. AJR Am J Roentgenol 184:364–372PubMed Obuchowski NA (2005) ROC analysis. AJR Am J Roentgenol 184:364–372PubMed
13.
go back to reference Rohling R, Biro P (1999) Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anaesthesia. J Clin Monit 15:23–27CrossRef Rohling R, Biro P (1999) Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anaesthesia. J Clin Monit 15:23–27CrossRef
14.
go back to reference Dullenkopf A, Di Bernardo S, Berger F, Fasnacht M, Gerber A, Weiss M (2003) Evaluation of a new combined SpO2/PtcCO2 sensor in anaesthetized paediatric patients. Paed Anaesth 13:777–784CrossRef Dullenkopf A, Di Bernardo S, Berger F, Fasnacht M, Gerber A, Weiss M (2003) Evaluation of a new combined SpO2/PtcCO2 sensor in anaesthetized paediatric patients. Paed Anaesth 13:777–784CrossRef
15.
go back to reference Eberhard P, Gisiger PA, Gardaz JP, Spahn DR (2002) Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg 94:76–80 Eberhard P, Gisiger PA, Gardaz JP, Spahn DR (2002) Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg 94:76–80
16.
go back to reference Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T (1998) Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults. Chest 113:768–773PubMedCrossRef Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T (1998) Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults. Chest 113:768–773PubMedCrossRef
17.
go back to reference SenTec (2002) SenTec Digital Monitor – Mode d'emploi. Therwil, Suisse SenTec (2002) SenTec Digital Monitor – Mode d'emploi. Therwil, Suisse
18.
go back to reference Severinghaus JW (1982) Transcutaneous blood gas analysis. Resp Care 27:152–159 Severinghaus JW (1982) Transcutaneous blood gas analysis. Resp Care 27:152–159
Metadata
Title
Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients
Authors
Pablo Rodriguez
François Lellouche
Jerome Aboab
Christian Brun Buisson
Laurent Brochard
Publication date
01-02-2006
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 2/2006
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-005-0006-4

Other articles of this Issue 2/2006

Intensive Care Medicine 2/2006 Go to the issue