Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis

Authors: Helmut D. Hummler, Anja Engelmann, Frank Pohlandt, Josef Högel, Axel R. Franz

Published in: Intensive Care Medicine | Issue 4/2004

Login to get access

Abstract

Objective

To test the effects of low perfusion caused by emerging sepsis on the performance of two new pulse oximetry techniques: Masimo SET in comparison with Nellcor Oxismart XL.

Design

Cohort study with random allocation of two pulse oximetry devices to two sensor sites.

Setting

University animal research facility.

Subjects

Twenty-five adult, anesthetized, ventilated rabbits.

Interventions

Pneumonia/sepsis was induced by tracheal instillation of E. coli.

Measurements and results

Oxygen saturation was measured by pulse oximetry (SpO2) and recorded continuously until death. Arterial oxygen saturation (SaO2) was measured hourly by CO oximetry and whenever a difference of >5% between the devices occurred. SpO2 sensors were positioned at both forelegs and switched hourly. There was no difference in total signal dropout time [median 3.8 min (range 0.4–66.6 min) vs 3.3 min (range 0–94.5 min), Masimo SET vs Oxismart XL]. There were fewer episodes with a false SpO2 reading [1 (range 0–7) vs 2 (range 0–17)] using the Masimo SET vs the Oxismart XL as verified by CO oximetry; p<0.05. Average bias (SpO2–SaO2) was significantly different between the two devices, and variability of bias values increased across time with both devices.

Conclusions

Both devices were capable to measure SpO2 during most of the experimental time in this model of low perfusion and therefore appear to be highly sensitive to pick up a signal; however, low perfusion caused by emerging sepsis may result in inaccurate measurements with both devices. These episodes were less common with the Masimo SET vs the Oxismart XL.
Literature
1.
go back to reference Rusch TL, Sankar R, Scharf JE (1996) Signal processing methods for pulse oximetry. Comput Biol Med 26:143–159CrossRefPubMed Rusch TL, Sankar R, Scharf JE (1996) Signal processing methods for pulse oximetry. Comput Biol Med 26:143–159CrossRefPubMed
2.
go back to reference Hay WW Jr, Rodden DJ, Collins SM, Melara DL, Hale KA, Fashaw LM (2002) Reliability of conventional and new pulse oximetry in neonatal patients. J Perinatol 22:360–366CrossRefPubMed Hay WW Jr, Rodden DJ, Collins SM, Melara DL, Hale KA, Fashaw LM (2002) Reliability of conventional and new pulse oximetry in neonatal patients. J Perinatol 22:360–366CrossRefPubMed
3.
go back to reference Bohnhorst B, Peter CS, Poets CF (2000) Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters. Crit Care Med 28:1565–1568CrossRefPubMed Bohnhorst B, Peter CS, Poets CF (2000) Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters. Crit Care Med 28:1565–1568CrossRefPubMed
4.
go back to reference Trivedi NS, Ghouri AF, Shah NK, Lai E, Barker SJ (1997) Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. J Clin Anesth 9:179–183CrossRefPubMed Trivedi NS, Ghouri AF, Shah NK, Lai E, Barker SJ (1997) Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. J Clin Anesth 9:179–183CrossRefPubMed
5.
go back to reference Langton JE, Lassey D, Hanning CD (1990) Comparison of four pulse oximeters: effects of venous occlusion and cold-induced peripheral vasoconstriction. Br J Anaesth 65:245–247PubMed Langton JE, Lassey D, Hanning CD (1990) Comparison of four pulse oximeters: effects of venous occlusion and cold-induced peripheral vasoconstriction. Br J Anaesth 65:245–247PubMed
6.
go back to reference Gehring H, Hornberger C, Matz H, Konecny E, Schmucker P (2002) The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia. Respir Care 47:48–60PubMed Gehring H, Hornberger C, Matz H, Konecny E, Schmucker P (2002) The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia. Respir Care 47:48–60PubMed
7.
go back to reference Barrington KJ, Ryan CA, Finer NN (1986) Pulse oximetry during hemorrhagic hypotension and cardiopulmonary resuscitation in the rabbit. J Crit Care 1:241–246 Barrington KJ, Ryan CA, Finer NN (1986) Pulse oximetry during hemorrhagic hypotension and cardiopulmonary resuscitation in the rabbit. J Crit Care 1:241–246
8.
go back to reference Ibanez J, Velasco J, Raurich JM (1991) The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy. Intensive Care Med 17:484–486PubMed Ibanez J, Velasco J, Raurich JM (1991) The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy. Intensive Care Med 17:484–486PubMed
9.
go back to reference Secker C, Spiers P (1997) Accuracy of pulse oximetry in patients with low systemic vascular resistance. Anaesthesia 52:127–130PubMed Secker C, Spiers P (1997) Accuracy of pulse oximetry in patients with low systemic vascular resistance. Anaesthesia 52:127–130PubMed
10.
go back to reference Brun-Buisson C, Doyon F, Carlet J, French Bacteremia-Sepsis Study Group (1996) Bacteremia and severe sepsis in adults: a multicenter prospective survey in ICUs and wards of 24 hospitals. Am J Respir Crit Care Med 154:617–624PubMed Brun-Buisson C, Doyon F, Carlet J, French Bacteremia-Sepsis Study Group (1996) Bacteremia and severe sepsis in adults: a multicenter prospective survey in ICUs and wards of 24 hospitals. Am J Respir Crit Care Med 154:617–624PubMed
11.
go back to reference Hummler HD, Pohlandt F, Franz AR (2002) Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbits. Crit Care Med 30:2501–2508CrossRefPubMed Hummler HD, Pohlandt F, Franz AR (2002) Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbits. Crit Care Med 30:2501–2508CrossRefPubMed
12.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 2:307–310 Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 2:307–310
13.
go back to reference Villanueva R, Bell C, Kain ZN, Colingo KA (1999) Effect of peripheral perfusion on accuracy of pulse oximetry in children. J Clin Anesth 11:317–322CrossRefPubMed Villanueva R, Bell C, Kain ZN, Colingo KA (1999) Effect of peripheral perfusion on accuracy of pulse oximetry in children. J Clin Anesth 11:317–322CrossRefPubMed
14.
go back to reference Vicenzi MN, Gombotz H, Krenn H, Dorn C, Rehak P (2000) Transesophageal versus surface pulse oximetry in intensive care unit patients. Crit Care Med 28:2268–2270 Vicenzi MN, Gombotz H, Krenn H, Dorn C, Rehak P (2000) Transesophageal versus surface pulse oximetry in intensive care unit patients. Crit Care Med 28:2268–2270
Metadata
Title
Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis
Authors
Helmut D. Hummler
Anja Engelmann
Frank Pohlandt
Josef Högel
Axel R. Franz
Publication date
01-04-2004
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 4/2004
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-2116-1

Other articles of this Issue 4/2004

Intensive Care Medicine 4/2004 Go to the issue