Linearity of patient positioning detection

A phantom study of skin markers, cone beam computed tomography, and 3D ultrasound

Authors: Hendrik Ballhausen, Sheila Hieber, Minglun Li, Katia Parodi, Claus Belka, Michael Reiner

Published in: Strahlentherapie und Onkologie | Issue 5/2015

Login to get access

Abstract

Background

Three-dimensional ultrasound (3D-US) is a modality complementary to kilovoltage cone beam computed tomography (kV-CBCT) and skin markers for patient positioning detection. This study compares the linearity of evaluations based on measurements using a modern 3D-US system (Elekta Clarity®; Elekta, Stockholm, Sweden), a kV-CBCT system (Elekta iView®), and skin markers.

Materials and methods

An investigator deliberately displaced a multimodal phantom by up to ± 30 mm along different axes. The following data points were acquired: 27 along the lateral axis, 29 along the longitudinal axis, 27 along the vertical axis, and 27 along the space diagonal. At each of these 110 positions, the displacements according to skin’ markers were recorded and scans were performed using both 3D-US and kV-CBCT. Shifts were detected by matching bony anatomy or soft tissue density to a reference planning CT in the case of kV-CBCT and for 3D-US, by matching ultrasound volume data to a reference planning volume. A consensus value was calculated from the average of the four modalities. With respect to this consensus value, the linearity (offset and regression coefficient, i.e., slope), average offset, systematic error, and random error of all four modalities were calculated for each axis.

Results

Linearity was similar for all four modalities, with regression coefficients between 0.994 and 1.012, and all offsets below 1 mm. The systematic errors of skin markers and 3D-US were higher than for kV-CBCT, but random errors were similar. In particular, 3D-US demonstrated an average offset of 0.36 mm to the right, 0.08 mm inferiorly, and 0.15 mm anteriorly; the systematic error was 0.36 mm laterally, 0.35 mm longitudinally, and 0.22 mm vertically; the random error was 0.15 mm laterally, 0.30 mm longitudinally, and 0.12 mm vertically. A total of 109 out of 110 (99 %) 3D-US measurements were within 1 mm of the consensus value on either axis.

Conclusion

The linearity of 3D-US was no worse than that of skin markers or kV-CBCT. Average offsets, systematic errors, and random errors were all below 1 mm. Optimal margins in the order of 1 mm could be achieved in the controlled laboratory setting of this phantom study.
Literature
1.
go back to reference Bohrer et al (2008) Reduced rectal toxicity with ultrasound-based image guided radiotherapy using BAT (B-mode acquisition and targeting system) for prostate cancer. Strahlenther Onkol 184(12):674–678 Bohrer et al (2008) Reduced rectal toxicity with ultrasound-based image guided radiotherapy using BAT (B-mode acquisition and targeting system) for prostate cancer. Strahlenther Onkol 184(12):674–678
2.
go back to reference Pinkawa et al (2008) Image-guided radiotherapy for prostate cancer. Implementation of ultrasound-based prostate localization for the analysis of inter- and intrafraction organ motion. Strahlenther Onkol 184(12):679–685 Pinkawa et al (2008) Image-guided radiotherapy for prostate cancer. Implementation of ultrasound-based prostate localization for the analysis of inter- and intrafraction organ motion. Strahlenther Onkol 184(12):679–685
3.
go back to reference Chadha M, Young A, Geraghty C, Masino R, Harrison L (2011) Image guidance using 3D-ultrasound (3D-US) for daily positioning of lumpectomy cavity for boost irradiation. Radiat Oncol 6:45CrossRefPubMedCentralPubMed Chadha M, Young A, Geraghty C, Masino R, Harrison L (2011) Image guidance using 3D-ultrasound (3D-US) for daily positioning of lumpectomy cavity for boost irradiation. Radiat Oncol 6:45CrossRefPubMedCentralPubMed
4.
go back to reference Ballhausen et al (2013) The random walk model of intrafraction movement. Physics in Medicine and Biology 58:2413–2427 Ballhausen et al (2013) The random walk model of intrafraction movement. Physics in Medicine and Biology 58:2413–2427
5.
go back to reference Ballhausen et al (2015) Intra-fraction motion of the prostate is a random walk. Physics in Medicine and Biology 60:549–563 Ballhausen et al (2015) Intra-fraction motion of the prostate is a random walk. Physics in Medicine and Biology 60:549–563
6.
go back to reference Wong P, Muanza T, Reynard E, Robert K, Barker J, Sultanem K (2011) Use of three-dimensional ultrasound in the detection of breast tumor bed displacement during radiotherapy. Int J Radiat Oncol Biol Phys 79:39–45CrossRefPubMed Wong P, Muanza T, Reynard E, Robert K, Barker J, Sultanem K (2011) Use of three-dimensional ultrasound in the detection of breast tumor bed displacement during radiotherapy. Int J Radiat Oncol Biol Phys 79:39–45CrossRefPubMed
7.
go back to reference Robinson D, Liu D, Steciw S, Field C, Daly H, Saibishkumar EP, Fallone G, Parliament M, Amanie J (2012) An evaluation of the Clarity 3D ultrasound system for prostate localization. J Appl Clin Med Phys 13:3753PubMed Robinson D, Liu D, Steciw S, Field C, Daly H, Saibishkumar EP, Fallone G, Parliament M, Amanie J (2012) An evaluation of the Clarity 3D ultrasound system for prostate localization. J Appl Clin Med Phys 13:3753PubMed
8.
go back to reference van der Meer S, Bloemen-van Gurp E, Hermans J, Voncken R, Heuvelmans D, Gubbels C, Fontanarosa D, Visser P, Lutgens L, van Gils F, Verhaegen F (2013) Critical assessment of intramodality 3D ultrasound imaging for prostate IGRT compared to fiducial markers. Med Phys 40:071707CrossRefPubMed van der Meer S, Bloemen-van Gurp E, Hermans J, Voncken R, Heuvelmans D, Gubbels C, Fontanarosa D, Visser P, Lutgens L, van Gils F, Verhaegen F (2013) Critical assessment of intramodality 3D ultrasound imaging for prostate IGRT compared to fiducial markers. Med Phys 40:071707CrossRefPubMed
9.
go back to reference Ballhausen H, Hieber S, Li M, Belka C, Reiner M (2014) Millimeter precision in ultrasound based patient positioning: experimental quantification of inherent technical limitations. Med Phys 41:081718CrossRefPubMed Ballhausen H, Hieber S, Li M, Belka C, Reiner M (2014) Millimeter precision in ultrasound based patient positioning: experimental quantification of inherent technical limitations. Med Phys 41:081718CrossRefPubMed
10.
go back to reference van Herk M, Remeijer P, Lebesque JV (2002) Inclusion of geometric uncertainties in treatment plan evaluation. Int J Radiat Oncol Biol Phys 52 1407–1422CrossRefPubMed van Herk M, Remeijer P, Lebesque JV (2002) Inclusion of geometric uncertainties in treatment plan evaluation. Int J Radiat Oncol Biol Phys 52 1407–1422CrossRefPubMed
Metadata
Title
Linearity of patient positioning detection
A phantom study of skin markers, cone beam computed tomography, and 3D ultrasound
Authors
Hendrik Ballhausen
Sheila Hieber
Minglun Li
Katia Parodi
Claus Belka
Michael Reiner
Publication date
01-05-2015
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 5/2015
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-015-0811-9

Other articles of this Issue 5/2015

Strahlentherapie und Onkologie 5/2015 Go to the issue